CYTO-STAT TRICHROME CD45-FITC/CD4-RDI/CD3-PC5 MONOCLONAL ANITBODY REAGENT W/ISOTYPIC CONTROL
Applicant
Coulter Corp.
Product Code
GKZ · Hematology
Decision Date
Jan 6, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Intended Use
CYTO-STAT® triCHROME™ CD45-FITC/CD4-RD1/CD3-PC5 is a three-color fluorescent reagent comprised of three murine monoclonal antibodies. Each antibody is labeled with a different color fluorochrome. The reagent identifies a lymphocyte gate based on CD45 staining (vs. side scatter) and allows simultaneous identification and enumeration of total CD3+, total CD4+ and dual-stained CD3+/CD4+ lymphocytes in whole blood by flow cytometry. An isotypic control, CYTO-STAT® triCHROME™ CD45-FITC/MsIgG1-RD1/MsIgG1-PC5, is used to monitor non-specific binding.
Device Story
Reagent kit uses three murine monoclonal antibodies (CD45-FITC, CD4-RD1, CD3-PC5) to label whole blood samples; flow cytometry identifies lymphocyte populations. CD45 staining vs. side scatter defines lymphocyte gate; simultaneous enumeration of CD3+, CD4+, and CD3+/CD4+ cells occurs in single specimen. Isotypic control monitors non-specific binding. Used in clinical laboratories by trained technicians/pathologists. Output provides percentage and absolute counts of lymphocyte subsets; assists clinicians in assessing immune status in HIV, transplant, and autoimmune patients.
Clinical Evidence
Performance validated via comparative study against predicate (CD3/T4) using normal and abnormal (HIV, transplant, autoimmune) whole blood specimens (n=diverse population, ages 18-85). Metrics included accuracy (regression/correlation), linearity (recovered vs. expected), and precision (within-day/interlaboratory CVs). Results demonstrated comparable identification and enumeration of targeted lymphocytes between subject and predicate.
Technological Characteristics
Three-color fluorescent murine monoclonal antibody reagent. Conjugates: CD45-FITC (Fluorescein Isothiocyanate), CD4-RD1 (Phycoerythrin), CD3-PC5 (Phycoerythrin-Cy5). Operates via flow cytometry using lymphocyte gating. No electronic or software components integral to the reagent itself; requires external flow cytometer for analysis.
Indications for Use
Indicated for identification and enumeration of total CD3+, total CD4+, and dual-stained CD3+/CD4+ lymphocytes in whole blood via flow cytometry. Applicable to adult populations (18-85 years) including normal individuals and those with HIV, organ transplant, autoimmune disease, or low white blood cell counts.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
K962305 — CYTO-STAT TRICHROME CD45-FITC/CD8-RD1/CD3-PC5 MONOCLONAL ANTIBODY REAGENT IWTH ISOTYPIC CONTROL · Coulter Corp. · Jan 6, 1997
K982166 — CYTO-STAT TRICHROME CD45-FITC/CD56-RD1/CD3-PC5 MONOCLONAL ANTIBODY REAGENT WITH CYTO-STAT TRICHROME CD45-FITC/MSIGG1-RD1 · Coulter Corp. · Nov 5, 1998
K982167 — CYTO-STAT TRICHROME CD45-FITC/CD19-RD1/CD3-PC5 MONOCLONAL ANTIBODY REAGENT WITH CYTO-STAT TRICHROME CD45-FITC/MSIGG1-RD1 · Coulter Corp. · Nov 9, 1998
K964618 — CYTO-STAT TRICHROME CD8-FITC/CD4-RDI/CD3-PC5 MONOCLONAL ANTIBODY REAGENT WITH CYTO-STAT TRICHROME MSIGGI-FITC/MSIGGI-RD1 · Coulter Corp. · Dec 23, 1996
K963263 — TETRAONE SYSTEM FOR EPICS XL FLOW CYTOMETRY SYSTEMS AND CYTO-STAT TERTRACHROME CD45-FITC/CD4-RDI/CD8-ECD/CD3-PC5 · Coulter Corp. · Jan 22, 1997
Submission Summary (Full Text)
{0}
X962768
COULTER
Jan 6, 1997
COULTER CORPORATION
P.O. Box 169015
Miami, Florida 33116-9015 USA
(305) 380-3800
(800) 327-6531
Product Information: (800) 526-6932
Date: June 14, 1996
Title: Summary of Safety and Effectiveness Information For 510(k) Premarket Notification
Product: CYTO-STAT® triCHROME™ CD45-FITC/CD4-RD1/CD3-PC5 Monoclonal Antibody Reagent With Isotypic Control
Company: Coulter Corporation
11800 SW 147 Avenue
Miami, FL 33196-2500
Contact: Dr. Marion S. Gaide (M/C: 31-B06)
Senior Regulatory Affairs Specialist
Corporate Regulatory Affairs
Telephone: 305-380-2594
Common or Usual or Classification Name: Lymphocyte Immunophenotyping Monoclonal Antibody Reagents
Product Classification: Product Code: GKZ; C.F.R. Section: 864.5220; Classification Panel: Hematology and Pathology Devices; Device Class: II
Intended Use: CYTO-STAT® triCHROME™ CD45-FITC/CD4-RD1/CD3-PC5 is a three-color fluorescent reagent comprised of three murine monoclonal antibodies. Each antibody is labeled with a different color fluorochrome. The reagent identifies a lymphocyte gate based on CD45 staining (vs. side scatter) and allows simultaneous identification and enumeration of total CD3+, total CD4+ and dual-stained CD3+/CD4+ lymphocytes in whole blood by flow cytometry. An isotypic control, CYTO-STAT® triCHROME™ CD45-FITC/MsIgG1-RD1/MsIgG1-PC5, is used to monitor non-specific binding.
Substantial Equivalence: 510(k) Premarket Notification: K922745
CYTO-STAT®/COULTER CLONE® CD3(IgG1)-FITC/T4-RD1
Monoclonal Antibody Reagent
Product Differences: CD45/CD4/CD3 and CD3/T4 are essentially identical with respect to features and principles of operation. Each liquid reagent allows simultaneous identification and enumeration of more than one T lymphocyte population (CD3+; CD4+; CD3+/CD4+) in a single specimen using a single reagent. Each reagent also requires a separate isotypic control to monitor non-specific binding.
The one difference between the reagents is that CD45/CD4/CD3 contains CD45 to identify a lymphocyte gate for making CD3, CD4 and CD3+/CD4+ measurements. In contrast, CD3/T4 requires a separate reagent, CYTO-STAT®/COULTER CLONE® Mo2-RD1/KC56 (T-200)-FITC, for this purpose.
Mab Conjugation: CD45: FITC (Fluorescein Isothiocyanate). CD4: RD1 (Phycoerythrin). CD3: PC5 (Phycoerythrin-Cy5).
Science Serving Humanity
4232216-D R 12-93
1046
{1}
Product Testing:
Product testing to assess the performance of CD45/CD4/CD3 is described below. Studies were designed in line with instructions for use in the Product Package Insert and performance specifications. Specimens were assayed with CD3/T4 for comparison purposes. The results of product testing demonstrated that CD45/CD4/CD3 met all performance specifications and provided mature T (CD3+) and inducer T (CD4+; CD3+/CD4+) lymphocyte values comparable to those of CD3/T4.
1. Accuracy:
Normal and abnormal (e.g., Human Immunodeficiency Virus, organ transplant, autoimmune disease, low white blood cell count) whole blood specimens were collected from geographically diverse populations of males and females unselected as to race and ranging in age from 18 to 85 years. Specimens were divided, processed as lysed preparations and assayed in parallel with CD45/CD4/CD3 and CD3/T4. CD3+, CD4+ and CD3+/CD4+ percentages expressed in terms of the total lymphocyte count and absolute counts (cells/μL) were determined with COULTER® EPICS® XL/XL-MCL flow cytometers gated on lymphocytes. White blood cell counts and 3-part differentials were obtained for all specimens.
Results analyzed in terms of minimums, maximums, means ± 1 SD, regression and correlation analyses, and analyses of variance demonstrated that CD45/CD4/CD3 and CD3/T4 identify and enumerate essentially identical numbers of the targeted lymphocytes in whole blood specimens.
2. Linearity:
Three replicate measurements were made on a concentrated normal whole blood specimen serially diluted to achieve a range of CD3+ and CD4+ (CD3+/CD4+) lymphocyte concentrations. Samples were assayed with CD45/CD4/CD3 and analyzed on a COULTER EPICS XL/XL-MCL flow cytometer gated on lymphocytes. Values were expressed in terms of absolute counts (cells/μL).
Results analyzed in terms of regression and correlation analyses for recovered versus expected absolute counts demonstrated linearity of the assay.
3. Precision (Within Day/Intralaboratory):
Ten replicate measurements were made for each of three levels of CD3+ and CD4+ (CD3+/CD4+) lymphocyte concentrations on the same day using a COULTER EPICS XL/XL-MCL flow cytometer gated on lymphocytes. Levels were obtained by selective depletions of a normal whole blood specimen and assayed with CD45/CD4/CD3. Values were expressed in terms of % of the total lymphocyte count.
Results analyzed in terms of mean ± 1 SD and CV demonstrated within day/intralaboratory precision of the assay.
4. Precision (Interlaboratory):
Ten replicate measurements on were made on the same day using different laboratories and COULTER EPICS XL/XL-MCL flow cytometers. All measurements were made on a single normal whole blood specimen divided and assayed with CD45/CD4/CD3. Values were expressed in terms of % of the total lymphocyte count.
Results analyzed in terms of mean ± 1 SD and CV demonstrated interlaboratory precision of the assay.
Marion S. Gaide, Ph.D.
Senior Regulatory Affairs Specialist
Corporate Regulatory Affairs
4543se
June 14, 1996
1217
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.