CYTO-STAT TRICHROME CD8-FITC/CD4-RDI/CD3-PC5 MONOCLONAL ANTIBODY REAGENT WITH CYTO-STAT TRICHROME MSIGGI-FITC/MSIGGI-RD1
Applicant
Coulter Corp.
Product Code
GKZ · Hematology
Decision Date
Dec 23, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
CYTO-STAT® triCHROME™ CD8-FITC/CD4-RD1/CD3-PC5 Monoclonal Antibody Reagent is a three-color fluorescent reagent comprised of three murine monoclonal antibodies. Each antibody is labeled with a different color fluorochrome. The reagent allows simultaneous identification and enumeration of total CD3+, total CD4+, total CD8+, dual CD3+/CD4+ and dual CD3+/CD8+ lymphocytes in whole blood by flow cytometry. An isotypic control, CYTO-STAT® triCHROME™ MsIgG1-FITC/MsIgG1-RD1/MsIgG1-PC5, is used to monitor nonspecific staining. CYTO-STAT® triCHROME™ MsIgG1-FITC/MsIgG1-RD1/MsIgG1-PC5 Isotypic Control is a three-color fluorescent reagent comprised of three murine monoclonal antibodies. Each antibody is labeled with a different color fluorochrome. This product is intended for use as a quality control reagent to monitor the levels of nonspecific antibody binding in cell surface staining procedures which use CYTO-STAT® triCHROME™ CD8-FITC/CD4-RD1/CD3-PC5 Monoclonal Antibody Reagent to identify and enumerate total CD3+, total CD4+, total CD8+, dual CD3+/CD4+ and dual CD3+/CD8+ lymphocytes in whole blood by flow cytometry.
Device Story
Device consists of three-color fluorescent murine monoclonal antibody reagents (CD8-FITC/CD4-RD1/CD3-PC5) and corresponding isotypic controls. Used in clinical laboratories by trained personnel to perform immunophenotyping of whole blood via flow cytometry. Reagents bind to specific lymphocyte surface markers; fluorochromes allow simultaneous detection of multiple cell populations (CD3+, CD4+, CD8+). Flow cytometer (e.g., COULTER® EPICS® XL-MCL) measures immunofluorescence; software gates on lymphocytes to identify and enumerate cell subsets. Output provides percentage and absolute counts of T-lymphocyte populations. Results assist clinicians in assessing immune status in patients with conditions like HIV, autoimmune disease, or post-transplant status.
Clinical Evidence
Bench testing and comparative clinical study. Accuracy evaluated using normal and abnormal (HIV, transplant, autoimmune) blood specimens (n=diverse population, ages 18-85). Specimens assayed in parallel with predicate device. Metrics included regression/correlation analysis, means, SD, and 95% CIs. Results showed identical enumeration of targeted lymphocytes. Linearity confirmed via serial dilution of control cells. Precision (within-run and interlaboratory) demonstrated via CV analysis. No clinical diagnostic outcomes reported; study focused on analytical performance equivalence.
Technological Characteristics
Three-color fluorescent murine monoclonal antibody reagent. Fluorochromes: FITC, RD1, PC5. Principle: Immunofluorescence analysis via flow cytometry. Connectivity: Standalone reagent system for use with flow cytometers (e.g., COULTER® EPICS® XL-MCL). Sterilization: Not specified. Software: Rule-based gating for lymphocyte identification.
Indications for Use
Indicated for the identification and enumeration of total CD3+, total CD4+, total CD8+, dual CD3+/CD4+, and dual CD3+/CD8+ lymphocytes in whole blood samples from patients (ages 18-85, including those with HIV, organ transplants, or autoimmune disease) using flow cytometry.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
CYTO-STAT®/COULTER CLONE® CD3-ECD/T4-RD1/T8-FITC Monoclonal Antibody Reagent with CYTO-STAT®/COULTER CLONE® MsIgG2b-ECD/MsIgG1-RD1/MsIgG1-FITC Isotypic Control (K950742)
Submission Summary (Full Text)
{1}
DEC 23 1996
K964618
COULTER
COULTER CORPORATION
P.O. BOX 169015
Miami, Florida 33116-9015 USA
Customer Service: (800) 526-7694
Product Information: (800) 526-6932
(305) 380-3800
(800) 327-6531
Date: October 11, 1996
Title: Summary of Safety and Effectiveness Information for "K950742 510(k) Additional Information"
Coultor Corporation
Miami, Florida USA
Coultor Leasing Corporation
Miami, Florida USA
Coultor Electronics, Pty. Ltd.
Sydney, Australia
Coultor Electronics Ind. & Com., Ltda.
Rio de Janeiro, Brazil
Coultor Electronics of Canada, Ltd.
Burlington, Ontario, Canada
Coultor Electronics, Ltd.
Luton Bedfordshire, England
Coultor Electronics France, S.A.
Margency, France
Coultor Electronics GmbH
Krefeld, Germany
Coultor Electronics (HK), Ltd.
Hong Kong
Coultor K. K.
Tokyo, Japan
Coultor de Mexico S.A., DE C.V.
Mexico City, Mexico
Coultor Electronics, Ltd.
Mijdrecht, Netherlands
Coultor Electronics, Pty. Ltd.
Auckland, New Zealand
Coultor Electronics Sales of P.R., Inc.
San Juan, Puerto Rico
Coultor Electronics, Ltd.
Johannesburg, South Africa
Coultor Electronics, Ltd.
Istanbul, Turkey
Coultor Electronics, S.A.
Caracas, Venezuela
Product: CYTO-STAT® triCHROME™ CD8-FITC/CD4-RD1/CD3-PC5 Monoclonal Antibody Reagent with CYTO-STAT® triCHROME™ MsIgG1-FITC/MsIgG1-RD1/MsIgG1-PC5 Isotypic Control
Company: Coulter Corporation
11800 SW 147 Avenue
Miami, FL 33196-2500
Contact: Dr. Marion S. Gaide (M/C: 31-B06)
Senior Regulatory Affairs Specialist
Corporate Regulatory Affairs
Telephone: 305-380-2594
Common or Usual or Classification Name: Lymphocyte Immunophenotyping Monoclonal Antibody Reagent and Isotypic Control
Product Classification: Product Code: GKZ; C.F.R. Section: 864.5220
Classification Panel: Hematology and Pathology Devices; Device Class: II
Intended Use: CYTO-STAT® triCHROME™ CD8-FITC/CD4-RD1/CD3-PC5 Monoclonal Antibody Reagent is a three-color fluorescent reagent comprised of three murine monoclonal antibodies. Each antibody is labeled with a different color fluorochrome. The reagent allows simultaneous identification and enumeration of total CD3+, total CD4+, total CD8+, dual CD3+/CD4+ and dual CD3+/CD8+ lymphocytes in whole blood by flow cytometry. An isotypic control, CYTO-STAT® triCHROME™ MsIgG1-FITC/MsIgG1-RD1/MsIgG1-PC5, is used to monitor nonspecific staining.
CYTO-STAT® triCHROME™ MsIgG1-FITC/MsIgG1-RD1/MsIgG1-PC5 Isotypic Control is a three-color fluorescent reagent comprised of three murine monoclonal antibodies. Each antibody is labeled with a different color fluorochrome. This product is intended for use as a quality control reagent to monitor the levels of nonspecific antibody binding in cell surface staining procedures which use CYTO-STAT® triCHROME™ CD8-FITC/CD4-RD1/CD3-PC5 Monoclonal Antibody Reagent to identify and enumerate total CD3+, total CD4+, total CD8+, dual CD3+/CD4+ and dual CD3+/CD8+ lymphocytes in whole blood by flow cytometry.
Substantial Equivalence: "K950742 510(k) Premarket Notification"
CYTO-STAT®/COULTER CLONE® CD3-ECD/T4-RD1/T8-FITC Monoclonal Antibody Reagent with
CYTO-STAT®/COULTER CLONE® MsIgG2b-ECD/MsIgG1-RD1/MsIgG1-FITC Isotypic Control
Science Serving Humanity
4232216-E R 11-94
{2}
2
# Product Differences:
CD8/CD4/CD3 with MsIgG1/MsIgG1/MsIgG1 and CD3/T4/T8 with MsIgG2b/MsIgG1/MsIgG1 are essentially identical with respect to features and principles of operation. Both the revised and original product systems use the same, well-established, state-of-the-art technologies of immunophenotyping with monoclonal antibodies and flow cytometry to measure cellular components in whole blood via immunofluorescence analysis. Further, the intended use of each system is the same. The liquid monoclonal antibody reagents allow simultaneous identification and enumeration of more than one lymphocyte population (total CD3+, total CD4+, total CD8+, dual CD3+/CD4+ and dual CD3+/CD8+) in a single specimen using a single reagent. Each system also uses separate reagents to 1) identify a lymphocyte gate (i.e., CYTO-STAT®/COULTER CLONE® Mo2-RD1/KC56 (T-200)-FITC); and 2) monitor nonspecific binding (i.e., isotypic control).
The few differences (italicized) between the reagent and isotypic control formulations are as follows:
1. CD3 Clone: CD8/CD4/CD3: UCHT1
CD3/T4/T8: HIT3a
2. CD3 Isotype: CD8/CD4/CD3: MsIgG1
CD3/T4/T8: MsIgG2a
3. Isotypic Control Clone: MsIgG1/MsIgG1/MsIgG1: ————; MsIgG1=2T8-2F5
MsIgG2b/MsIgG1/MsIgG1: MsIgG2b=MPC11; MsIgG1=2T8-2F5
4. Isotypic Control Isotype: MsIgG1/MsIgG1/MsIgG1: ————; MsIgG1
MsIgG2b/MsIgG1/MsIgG1: MsIgGb; MsIgG1
5. CD3 Fluorochrome: CD8/CD4/CD3: PC5 (Phycoerythrin-Cy5)
CD3/T4/T8: ECD (Energy-Coupled Dye)
# Product Testing:
Product testing to assess the performance of the CD8/CD4/CD3 product system is described below. Studies were designed in line with instructions for use given in the Package Inserts, Product Manuals, and performance specifications. Specimens were assayed with the CD3/T4/T8 product system for comparison purposes. The results of product testing demonstrated that CD8/CD4/CD3 met all performance specifications and provided mature T (CD3+), inducer T (CD4+; dual CD3+/CD4+) and suppressor/cytotoxic T (CD8+; dual CD3+/CD8+) T lymphocyte values comparable to those of CD3/T4/T8.
## 1. Accuracy:
Normal and abnormal (e.g., Human Immunodeficiency Virus, organ transplant, autoimmune disease, low white blood cell count) whole blood specimens were collected from geographically diverse populations of males and females unselected as to race and ranging in age from 18 to 85 years. Specimens were divided, processed as lysed preparations and assayed in parallel with CD8/CD4/CD3 and CD3/T4/T8. The CD3+, CD4+, CD8+, CD3+/CD4+ and CD3+/CD8+ percentages expressed in terms of the total lymphocyte count and absolute counts (cells/μL) were determined with a COULTER® EPICS® XL-MCL flow cytometer gated on lymphocytes. White blood cell counts and 5-part differentials were obtained for all specimens using the COULTER® STKS. Absolute counts were determined using both the Standard (Indirect) Method, and the Flow-Count (Direct) Method with Flow-Count™ Fluorospheres. All values were corrected for lymphocyte purity (Lymphocyte Gate Limits: lymphocyte recovery ≥ 90%; lymphocyte purity ≥ 85%).
Results analyzed in terms of minimums, maximums, means ± 1 SD, confidence intervals with 95% limits, regression and correlation analyses, and analyses of variance demonstrated that CD8/CD4/CD3 and CD3/T4/T8 identify and enumerate essentially identical numbers of the targeted lymphocytes in whole blood specimens.
{3}
2. Linearity:
Three replicate measurements were made on a concentrated COULTER™ CYTO-TROL™ Control Cells sample serially diluted to achieve a range of CD3+, CD4+ (CD3+/CD4+) and CD8+ (CD3+/CD8+) lymphocyte concentrations. Samples were assayed with CD8/CD4/CD3 and analyzed on a COULTER® EPICS® XL-MCL flow cytometer gated on lymphocytes. Values were expressed in terms of absolute counts (cells/μL).
Results analyzed in terms of regression and correlation analyses for recovered versus expected absolute counts demonstrated linearity of the assay.
3. Within Run (Intralaboratory) Precision:
Ten replicate measurements were made for each of three levels of CD3+, CD4+ (CD3+/CD4+) and CD8+ (CD3+/CD8+) lymphocyte concentrations using a COULTER® EPICS® XL-MCL flow cytometer gated on lymphocytes. Levels were obtained by selective depletions of a normal whole blood specimen and assayed with CD8/CD4/CD3. Values were expressed in terms of % of the total lymphocyte count.
Results analyzed in terms of mean ± 1 SD and CV demonstrated Within Run (Intralaboratory) Precision of the assay.
4. Interlaboratory Precision:
Ten replicate measurements were made on the same day using different laboratories and COULTER® EPICS® XL-MCL flow cytometers. All measurements were made on a single normal whole blood specimen divided and assayed with CD8/CD4/CD3. Values were expressed in terms of % of the total lymphocyte count.
Results analyzed in terms of mean ± 1 SD and CV demonstrated Interlaboratory Precision of the assay.
Marion S. Gaide, Ph.D.
Senior Regulatory Affairs Specialist
Corporate Regulatory Affairs
October 11, 1996
Date
843se
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.