CORDIS AVANTI TRANS-RADIAL CATHETER SHEATH INTRODUCER KIT

K962746 · Cordis Corp. · DYB · Oct 11, 1996 · Cardiovascular

Device Facts

Record IDK962746
Device NameCORDIS AVANTI TRANS-RADIAL CATHETER SHEATH INTRODUCER KIT
ApplicantCordis Corp.
Product CodeDYB · Cardiovascular
Decision DateOct 11, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

The Avanti™ Trans-Radial Catheter Sheath Introducer kit is intended for use in (radial) arterial procedures requiring percutaneous introduction of catheters and other intravascular devices. As with other currently marketed Cordis CSIs mentioned in this submission, these devices provide vascular access for various intravascular devices through the valve while simultaneously maintaining hemostasis. Infusion of fluids into the vasculature and withdrawal of blood from the vasculature are possible using the sheath sideport.

Device Story

Kit provides vascular access for intravascular devices via radial artery; maintains hemostasis during procedure; allows fluid infusion and blood withdrawal via sideport. Used by physicians in clinical settings for percutaneous catheterization. Benefits include simplified radial access through inclusion of vessel dilator and needle.

Clinical Evidence

Bench testing only; biocompatibility testing performed on new materials.

Technological Characteristics

Catheter sheath introducer kit; includes sheath, dilator, and needle. Materials tested for biocompatibility. Class II device per 21 CFR 870.1340.

Indications for Use

Indicated for patients undergoing radial arterial procedures requiring percutaneous introduction of catheters and other intravascular devices.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Submission Summary (Full Text)

{0} OCT 11 1996 K962746 # Summary of Safety and Effectiveness Pursuant to Section 513(I) of the Federal Food, Drug, and Cosmetic Act ## I. General Provisions Classification Name: Catheter Sheath Introducer (CSI) Kit Common or Usual Name: Sheath Introducer Kit or Trans-Radial Introducer Kit Proprietary Name: Cordis Avanti™ Trans-Radial Sheath Introducer Kit Applicant Name and Address: Cordis P.O. Box 025700 Miami, FL 33015-5700 ## II. Name of Predicate Devices Cordis Avanti™ Catheter Sheath Introducer K945616 and K932733 Cordis Plus Catheter Sheath Introducer K911794 Arrow Radial Artery Catheterization Set K810675 ## III. Classification Catheter Sheath Introducer Kits are Cardiovascular Diagnostic Devices and are classified as Class II devices according to 21 CFR 870.1340. ## IV. Performance Standards Performance standards have not been established by the FDA under Section 514 of the Food, Drug and Cosmetic Act. {1} # V. Intended Use and Device Description The Avanti™ Trans-Radial Catheter Sheath Introducer kit is intended for use in (radial) arterial procedures requiring percutaneous introduction of catheters and other intravascular devices. As with other currently marketed Cordis CSIs mentioned in this submission, these devices provide vascular access for various intravascular devices through the valve while simultaneously maintaining hemostasis. Infusion of fluids into the vasculature and withdrawal of blood from the vasculature are possible using the sheath sideport. # VI. Biocompatibility All appropriate biocompatibility tests were performed on the new material used for the Trans-Radial CSI Kit. All other materials were tested on previously concurred devices and passed. # VII. Summary of Substantial Equivalence Except for the site of entry (radial instead of femoral) and the addition of the radial vessel dilator and needle (thus making it a complete kit), the Cordis Avanti™ Trans-Radial Catheter Sheath Introducer Kit has the same intended use as other sheath introducers already on the market (i.e., Cordis Avanti™ CSI and Cordis Plus CSI). In addition, the Avanti™ Trans-Radial CSI Kit is similar in its basic design, construction, and indication for use to currently marketed catheter sheath introducer systems.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...