MIDAS REX III MOTOR

K962483 · Midas Rex Pneumatic Tools, Inc. · HBB · Nov 19, 1996 · Neurology

Device Facts

Record IDK962483
Device NameMIDAS REX III MOTOR
ApplicantMidas Rex Pneumatic Tools, Inc.
Product CodeHBB · Neurology
Decision DateNov 19, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4370
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Midas Rex Motor is designed for skull base and other microsurgical applications, including otolaryngology, ear, nose and throat, craniofacial and maxillofacial specifications in which a small, lightweight, high speed bone dissection system is desired.

Device Story

Pneumatic cranial drill motor; lightweight, ergonomic design; variable speed operation. Input: compressed air (20-150 psi). Output: high-speed mechanical bone dissection. Used in OR by surgeons for microsurgical applications. Integrates with existing Midas Rex GS dissecting tools, tube attachments, fixed/rotating footed attachments, telescoping microsystems, and micro mini drill guides. Facilitates precise bone removal in skull base and ENT procedures.

Clinical Evidence

Bench testing only; no clinical data provided. Substantial equivalence based on design and performance similarity to predicate devices.

Technological Characteristics

Pneumatic motor; operating pressure 20-150 psi (2-10 bar), recommended 100 psi (6.9 bar). Lightweight, small diameter housing. Quick disconnect interface for GS dissecting tools and attachments. Mechanical energy source.

Indications for Use

Indicated for patients requiring skull base and microsurgical bone dissection procedures, including otolaryngology, ENT, craniofacial, and maxillofacial surgeries.

Regulatory Classification

Identification

A pneumatic cranial drill motor is a pneumatically operated power source used with removable rotating surgical cutting tools or drill bits on a patient's skull.

Predicate Devices

Submission Summary (Full Text)

{0} NOV 19 1996 K962483 # 510 (K) SUMMARY OF SAFETY AND EFFECTIVENESS Date prepared: June 17, 1996 Submitter: Midas Rex Pneumatic Tools Inc. 3001 Race Street Fort Worth, Texas 76111 (817) 831-4181 Contact Person: Michael Fowler Director Regulatory Affairs Trade (proprietary) Name: Midas Rex Motor (Midas III) Common / Classification Name: Pneumatic Cranial Drill Motor Device Classification: Class II Predicate Device: Midas Rex Motors, Midas (I, II, & Convertible) K953434 Class II Midas (I, II, & Convertible) K952300 Class I Whirl Wind TM (Pre-1976) Predicate Attachments: Midas GS and GS20 Surgical Tool System K952463 Class I Midas Drill Guide Attachment K954076 Class I Midas Telescoping System K953433 Class I Premarket Notification: Midas Rex Pneumatic Tools Inc. Billmansafety {1} Premarket Notification: Midas Rex Pneumatic Tools. Cstansafety # Description Of the Device: Ergonomically, designed lightweight motor featuring small, overall diameter for comfort and ease of handling. The Midas Rex III motor operates at variable speeds. Operating pressures range from 20-150 psi (2-10 bar), with a suggested operating pressure of 100 psi (6.9 bar). The Midas Rex III has a standard quick disconnect GS dissecting tool and tube attachment system, which includes a fixed or rotating footed attachment, the telescoping microsystems, and the micro mini drill guide. # Statement of Intended Use: The Midas Rex Motor is designed for skull base and other microsurgical applications, including otolaryngology, ear, nose and throat, craniofacial and maxillofacial specifications in which a small, lightweight, high speed bone dissection system is desired. # Technological Characteristics: The modified device has the same technological characteristics (i.e., design, material, energy source) as the predicate device. The specifications are as follows:
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...