INTRASTAR PRE-WIRED FACEPLATE FOR IN-THE-EAR HEARING AIDS

K961593 · Segi Hearing Aid Co. · LRB · Nov 7, 1996 · Ear, Nose, Throat

Device Facts

Record IDK961593
Device NameINTRASTAR PRE-WIRED FACEPLATE FOR IN-THE-EAR HEARING AIDS
ApplicantSegi Hearing Aid Co.
Product CodeLRB · Ear, Nose, Throat
Decision DateNov 7, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3300
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The intended use of these models is to amplify sound pressure waves and transmit the signal to the external ear through the medium of air to compensate for impaired hearing.

Device Story

Intrastar pre-wired faceplates function as components for in-the-ear hearing aids; device amplifies sound pressure waves; transmits signal to external ear via air conduction; compensates for hearing impairment; intended for use by patients with hearing loss; operated by hearing healthcare professionals during fitting/assembly; output provides amplified sound to assist patient communication and auditory perception.

Clinical Evidence

No clinical data; substantial equivalence based on bench testing and comparison of performance characteristics to predicate devices.

Technological Characteristics

In-the-ear hearing aid pre-wired faceplates; analog amplification circuitry; frequency range 200-8000 Hz; maximum SSPL90 124.8-133.2 dB; full-on gain 38.3-55.4 dB; battery current 0.74-1.37 mA.

Indications for Use

Indicated for individuals with impaired hearing requiring sound amplification via in-the-ear hearing aid devices.

Regulatory Classification

Identification

An air-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing that conducts sound to the ear through the air. An air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable. The air-conduction hearing aid generic type excludes the group hearing aid or group auditory trainer, master hearing aid, and the tinnitus masker, regulated under §§ 874.3320, 874.3330, and 874.3400, respectively.

Predicate Devices

Submission Summary (Full Text)

{0} K961593 NOV - 7 1996 501(k) Summary Date: April 10, 1996 Submission contact person: Leo Weinberger, Esq. 24 Fifth Avenue, New York, N.Y. 10011 (212) 228-4404; Fax: (212) 598-0904 The following information is submitted for approval as required by Title 21, Code of Federal Regulations, Subchapter H-Medical Devices, Subpart E, Paragraph 807.81-Premarket Notification. Device names: Trade names: Intrastar pre-wired faceplate in-the-ear hearing aid models: SP-W "A1" SP-W "K1" SP-W "K4" "G1" "G3" Common name: hearing aid pre-wired faceplate Classification name: hearing aid Substantial equivalence: The functions and design of the above-named devices are substantially equivalent to existing conventional devices installed in in-the-ear hearing aids currently available through commercial distribution, such as: | 510(k) Number | Trade/Proprietary Name | Manufacturer | | --- | --- | --- | | K884871 | Resound In-The-Ear Hearing Enhancement System | Resound Corp. | | (Attachment 3) | | | | K902846 | PRIZM Hearing System | AudioScience, Inc. | | (Attachment 4) | | | Description: Descriptions of the device functions, scientific concepts, and physical and performance characteristics of the above-mentioned Intrastar models are set forth in Attachments 1 and 2. - 9 - {1} Use: Segi Hearing Aid Co. is introducing into commercial distribution the above-mentioned Intrastar models of in-the-ear hearing aid pre-wired faceplates. The intended use of these models is to amplify sound pressure waves and transmit the signal to the external ear through the medium of air to compensate for impaired hearing. Their function and design are substantially equivalent to those installed in existing conventional in-the-ear hearing aids currently available through commercial distribution. Comparison of technological characteristics: | | Intrastar All models | Resound K884871 | AudioScience K902846 | | --- | --- | --- | --- | | Maximum SSPL90 | 124.8-133.2 dB | 114 dB SPL | 122 dB SPL | | HF Avg. | 119.6-126.4 dB | 112 dB SPL | 114 dB SPL | | Full on gain | 38.3-55.4 dB | 54 dB max | 54 dB | | Reference test gain | 38.3-49.4 dB | | 54 dB | | Frequency range | 200-8000 Hz | 200-7000 Hz | 100-6800 Hz | | THD 500 Hz | 1.6-5.2% | | <5% | | THD 800 Hz | 1.8-13.4% | | <5% | | THD 1600 Hz | 0.3-7.7% | | <5% | | Equivalent input noise | 19.0-29.4 dB | 28 dB SPL | 21-26 dB | | Battery current | 0.74-1.37 mA | 1.3 mA | 2.2 mA | - 10 -
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...