ORTHODOC PREOPERATIVE PLANNER MODEL 500

K960685 · Integrated Surgical Systems, Inc. · HWT · Jan 17, 1997 · Orthopedic

Device Facts

Record IDK960685
Device NameORTHODOC PREOPERATIVE PLANNER MODEL 500
ApplicantIntegrated Surgical Systems, Inc.
Product CodeHWT · Orthopedic
Decision DateJan 17, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.4800
Device ClassClass 1
AttributesSoftware as a Medical Device

Indications for Use

The ORTHODOC Preoperative Planner is a software device indicated for the planning of primary total hip arthroplasties on a personal computer. The software is intended to import computed tomography data, provide a constructed image of the data, and use computerized files, representing implant patterns of femoral components, to overlay the constructed image to aid surgeons in their selection and positioning of the femoral components.

Device Story

Software-only preoperative planning tool; runs on Pentium-based PC. Inputs: CT scan data; CAD-based geometric implant surface model files. Operation: imports CT data; constructs image; allows surgeon to superimpose/manipulate scaled 2D graphic images of femoral stem components over CT femur images using mouse pointer. Output: visual overlay of implant patterns on patient anatomy. Usage: clinical setting; operated by surgeons. Benefit: aids selection of optimal stem size, position, neck length, and head diameter to achieve surgical goals. Device does not operate other medical devices or act directly on patients.

Clinical Evidence

No pre-clinical or clinical performance data generated.

Technological Characteristics

Software-only; runs on Pentium-based PC. Utilizes CAD geometric data to generate implant surface model files. Interface: mouse-driven manipulation of 2D graphic overlays on CT-derived images. No hardware components; no direct patient contact.

Indications for Use

Indicated for planning primary total hip arthroplasties in patients requiring femoral component selection and positioning.

Regulatory Classification

Identification

A template for clinical use is a device that consists of a pattern or guide intended for medical purposes, such as selecting or positioning orthopedic implants or guiding the marking of tissue before cutting.

Predicate Devices

Submission Summary (Full Text)

{0} JAN 7 1997 Integrated Surgical Systems, Inc. ORTHODOC Preoperative Planning Software 510(k) Summary K960685 Submitter: Integrated Surgical Systems, Inc. 829 W. Stadium Lane Sacramento, CA 95834 (916) 646-3487 Trade name: ORTHODOCR Model 500 Preoperative Planning Software Common name: Templating System Classification name: Template for clinical use (21 CFR 888.4800) Indications for use: The ORTHODOC Preoperative Planner is a software device indicated for the planning of primary total hip arthroplasties on a personal computer. The software is intended to import computed tomography data, provide a constructed image of the data, and use computerized files, representing implant patterns of femoral components, to overlay the constructed image to aid surgeons in their selection and positioning of the femoral components. Legally marketed predicate: All femoral component templates supplied by femoral component manufacturers. For example, Depuy AML templates and Howmedica Osteolock HA templates. Device description: This software is designed to run on a PentiumR-based personal computer. The device is a software-only product and is not intended to operate any other medical device or act directly on a patient. The application stores scaled graphic images of femoral stem components and provides a means for surgeons to superimpose these patterns on a CT image of the patient's femur. The implant patterns can be moved over the femur images to select what in the surgeon's judgement is the best stem size, position, neck length, and head diameter to achieve the physician's surgical goal. The library of implant patterns, called surface model files, is developed from computer-aided-design (CAD) geometric data supplied by the implant's manufacturer. The implant surface model file used by ORTHODOC contains implant identification information as well as geometric data which defines the shape and dimensions of the implant surface. 12 {1} Premarket Notification Integrated Surgical Systems, Inc. ORTHODOC Preoperative Planning Software ## Technological characteristics: The implant patterns of the predicates are scaled computer-generated images that are plotted, photographed, and contact printed to maintain accurate scale. These are held over the x-ray and moved by hand to position them to assist the surgeon in planning the stem's placement. The ORTHODOC software contains a library of computer-generated femoral component images which can be called-up and superimposed over a CT image of the patients femur. The images are moved by hand, employing a mouse pointer, to position them to assist the surgeon in planning the stem's placement. ## Performance data: There has been no pre-clinical or clinical performance data generated upon which to make a decision of substantial equivalence.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...