K960513 · Minntech Corp. · MED · Oct 1, 1997 · General Hospital
Device Facts
Record ID
K960513
Device Name
PERACT 20 LIQUID STERILANT
Applicant
Minntech Corp.
Product Code
MED · General Hospital
Decision Date
Oct 1, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6885
Device Class
Class 2
Indications for Use
Peract™ 20 Liquid Sterilant/Disinfectant is intended for sterilization or high level disinfection of medical and surgical instruments that require submersion. Peract™ 20 Liquid Sterilant/Disinfectant should be used only with heat sensitive medical and surgical instruments that are not compatible with other sterilization or high level disinfection processes that can be biologically monitored. Peract™ 20 Liquid Sterilant/Disinfectant should be used under the following contact conditions: Sterilization: 8 hours, 68°F (20°C), 500 ppm; High Level Disinfection: 25 minutes, 68°F (20°C), 500 ppm. Peract™ 20 Indicator Test Strips are intended for verifying the minimum effective concentration of peracetic acid in Peract™ 20 Liquid Sterilant/Disinfectant during reuse.
Device Story
Peract™ 20 is a single-component liquid germicide containing peracetic acid and hydrogen peroxide; requires no mixing/activation. Used for submersion sterilization or high-level disinfection of heat-sensitive medical/surgical instruments. Mechanism involves oxidation of sulfhydryl/sulfur bonds in microbial proteins/enzymes. Accompanied by indicator test strips to verify minimum effective concentration (MEC) of 500 ppm peracetic acid. Healthcare providers (clinicians/technicians) perform manual submersion of instruments in the solution for specified contact times (25 mins for disinfection; 8 hours for sterilization) at 68°F. Output is a processed, disinfected/sterilized instrument. Benefits include effective microbial reduction (sporicidal, tuberculocidal, virucidal, fungicidal, bactericidal) for devices unable to withstand heat-based sterilization.
Clinical Evidence
Bench testing only. Efficacy demonstrated via AOAC sporicidal testing, tuberculocidal testing, and use-dilution testing against S. aureus, S. choleraesuis, and P. aeruginosa. Virucidal efficacy confirmed against Poliovirus Type 2, Influenza A2, HIV-1, and HSV-1. Fungicidal efficacy confirmed against T. mentagrophytes. Simulated-use testing performed on endoscopes. Biocompatibility testing (cytotoxicity, hemolysis, acute toxicity, mucosal irritation) and material compatibility testing confirmed safety and device integrity. Stability studies confirmed shelf life and performance at MEC.
Technological Characteristics
Single-component liquid germicide; active ingredients: peracetic acid and hydrogen peroxide. Packaged in 1-gallon polyethylene bottles. Indicator test strips utilize chemical reaction to verify 500 ppm peracetic acid MEC. Compatible with wide range of endoscope materials. Stability tested per Liquid Chemical Germicide Document section (III)(F)(3).
Indications for Use
Indicated for sterilization or high-level disinfection of heat-sensitive, immersible medical and surgical instruments incompatible with other monitorable processes.
Regulatory Classification
Identification
A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.
Special Controls
*Classification.* Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.
Predicate Devices
ProCide® NS
Cidex®
Submission Summary (Full Text)
{0}
OCT-01-97 WED 10:04 AM MINNTECH
FAX NO. 6125533387
P. 02
K960513
# PERACT™ 20 LIQUID STERILANT/DISINFECTANT AND PERACT™ 20 INDICATOR TEST STRIPS
OCT - 1 1997
510(k) Summary of Safety and Effectiveness
Minntech Corporation, 14605 28th Ave. N, Mpls, MN 55447
Telephone: 800-328-3345
Official Contact: Robert Johnson
Vice President, Regulatory Affairs and Quality Assurance
Minntech Corporation has provided the following information to the U.S. Food and Drug Administration to support that Peract™ 20 Liquid Sterilant/Disinfectant is substantially equivalent to other sterilants currently in commercial distribution in the United States.
## 1. Device Description
Peract™ 20 Liquid Sterilant/Disinfectant is a single component germicide which does not require mixing or activation. The product is used full strength, without dilution. It is packaged in one gallon polyethylene bottles and has a one year shelf life.
The active ingredient in Peract™ 20 are peracetic acid and hydrogen peroxide. As discussed in Block’s¹ article, the mechanism of microbial action is believed to be oxidizing sulfhydryl and sulfur bonds in proteins and enzymes, particularly in the cell walls.
Peract™ 20 Peracetic Acid Indicator Test Strips are provided to verify that the minimum effective concentration (MEC) (500 ppm peracetic acid) of Peract™ 20 is present. Three test strips are used with each application.
## 2. Intended Use
Peract™ 20 Liquid Sterilant/Disinfectant is intended for sterilization or high level disinfection of medical and surgical instruments that require submersion.
Peract™ 20 Liquid Sterilant/Disinfectant should be used only with heat sensitive medical and surgical instruments that are not compatible with other sterilization or high level disinfection processes that can be biologically monitored.
¹ Block, Seymour S., Disinfection, Sterilization and Preservation; Chapter 9 Peroxygen Compounds (pages 167-181), Lea & Febiger, 1991.
{1}
OCT-01-97 WED 10:05 AM MINNTECH
FAX NO. 6125533387
P. 03
Peract™ 20 Liquid Sterilant/Disinfectant should be used under the following contact conditions:
| | Time | Temperature | Minimum Effective Concentration Peracetic Acid |
| --- | --- | --- | --- |
| Sterilization | 8 hours | 68°F (20°C) | 500 ppm |
| High Level Disinfection | 25 minutes | 68°F (20°C) | 500 ppm |
3. Comparison to another Device in Commercial Distribution within the United States:
Peract™ 20 Liquid Sterilant/Disinfectant is comparable in its intended use to other liquid sterilants currently on the market in the U.S. Peract™ 20 is similar in use and product claims to Cottrell’s ProCide® NS and Johnson & Johnson’s Cidex®.
4. Summary
Minntech Corporation has performed testing to demonstrate that Peract™ 20 Liquid Sterilant/Disinfectant and Peract™ 20 Indicator Test Strips are safe and effective when used according to the respective instructions for use.
4.1 Efficacy Testing
The following efficacy testing was performed on Peract™ 20 with all of the following conditions: at the minimum of its specifications, at the end of its shelf life, stressed to the end of its 14 day reuse period and at its MEC (500ppm PAA). The testing showed the product to be sporicidal, tuberculocidal, virucidal, fungicidal, and bactericidal.
AOAC sporicidal testing was performed on three lots of Peract™ 20 with the above noted conditions. Sporicidal simulated use testing was also performed on endoscopes to show efficacy as a sterilant on actual devices.
Tuberculocidal testing was performed on three lots of Peract™ 20 with the above noted conditions. Tuberculocidal simulated use testing was also performed on endoscopes to show efficacy as a high level disinfectant on actual devices.
Peract™ 20 was determined to be virucidal when tested against Poliovirus Type 2, Influenza A₂, Human Immunodeficiency Virus Type 1, and Herpes Simplex Virus Type 1. Peract™ 20 was considered fungicidal when tested against Trichophyton mentagrophytes. Use dilution testing showed the efficacy of Peract™ 20 against Staphylococcus aureus, Salmonella choleraesuis, Pseudomonas aeruginosa.
{2}
OCT-01-97 WED 10:05 AM MINNTECH
FAX NO. 6125533387
P. 04
Clinical testing of used scopes further supports the efficacy of the germicide when used under the instructions of the directions for use. Testing determined residues of Peract™ 20 remaining on endoscopes after sterilization/disinfection and rinsing were not significant.
## 4.2 Biocompatibility Testing
Standard patient toxicity testing evaluated the effect of residues, cytotoxicity, hemolysis, acute toxicity, and vaginal (mucosal membrane) irritation.
All biocompatibility testing demonstrated that Peract™ 20 is safe for the patient when used according to the instructions for use.
## 4.3 Material Compatibility
Material compatibility testing demonstrates that Peract™ 20 can be used with a wide range of materials and endoscopes. Testing included soaking and cycling common materials and endoscopes for the estimated lifetime of the items.
Material compatibility testing demonstrated that Peract™ 20 is compatible with the materials and devices listed when used according to the instructions for use.
## 4.4 Stability
Peract™ 20 has a shelf life of one year. Stability studies were performed according to section (III)(F)(3) of the Liquid Chemical Germicide Document. Studies demonstrated that the chemical and physical stability of Peract™ 20 were within specifications at the expiration date.
## 4.5 Test Strips
Peract™ 20 Indicator Test Strips demonstrated to consistently and accurately test the germicide at its minimum effective concentration of 500 ppm peracetic acid when three test strips were used.
Testing of the indicator strips was accomplished by showing: the efficacy of the strips when they were exposed to Peract™ 20; that they were stable over the labeled shelf life; and were stable in the opened bottle for 30 days. All of these tests were performed on a minimum of three lots of strips using Peract™ 20 diluted to various concentrations of PAA from much lower to much higher than the MEC of 500ppm PAA. The testing was also performed on solutions close to the MEC of 500ppm PAA to ensure the strips performed appropriately at concentrations around the MEC. The results of the testing showed the strips
{3}
OCT-01-97 WED 10:06 AM MINNTECH
FAX NO. 6125533387
P. 05
performed appropriately when three strips were used as called out in the directions for use.
## 5.0 Summary of Substantial Equivalence
Minntech Corporation has provided the above information within the 510(k) to support that Peract™ 20 Liquid Sterilant/Disinfectant and Peract™ 20 Indicator Test Strips are safe and effective when used according to the respective directions for use.
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
Ms. Lynn Lueders
Director, Regulatory Affairs
Minntech Corporation
14605 28th Avenue, North
Minneapolis, Minnesota 55447
OCT - 1 1997
Re: K960513
Trade Name: Peract™20 Liquid Sterilant/Disinfectant
Regulatory Class: Unclassified
Product Code: MED
Dated: July 14, 1997
Received: July 15, 1997
Dear Ms. Lueders:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531
{5}
Page 2 - Ms. Lueders
through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Director
Division of Dental, Infection Control and General Hospital Devices
Office of Device Evaluation
Center for Devices and Radiological Health
,Enclosure
{6}
SEP-30-97 TUE 03:30 PM MINNTECH
FAX NO. 6125533387
P. 02
# Indications for Use
510(k) Number (if known): K960513
Device Name: Peract™ 20 Liquid Sterilant/Disinfectant and Peract™ 20 Indicator Test Strips
Intended Use:
Peract™ 20 Liquid Sterilant/Disinfectant is intended for sterilization or high level disinfection of medical and surgical instruments that require submersion.
Peract™ 20 Liquid Sterilant/Disinfectant should be used only with heat sensitive medical and surgical instruments that are not compatible with other sterilization or high level disinfection processes that can be biologically monitored.
Peract™ 20 Liquid Sterilant/Disinfectant should be used under the following contact conditions:
| | Time | Temperature | Minimum Effective Concentration Peracetic Acid |
| --- | --- | --- | --- |
| Sterilization | 8 hours | 68°F (20°C) | 500 ppm |
| High Level Disinfection | 25 minutes | 68°F (20°C) | 500 ppm |
Peract™ 20 Indicator Test Strips are intended for verifying the minimum effective concentration of peracetic acid in Peract™ 20 Liquid Sterilant/Disinfectant during reuse.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☐ OR Over-the counter use ☑
(Per 21 CFR 801.109) (Optional Format 1-2-96)
(Division Sign-Off)
Division of Dental, Infection Control, and General Hospital Devices
510(k) Number K960513
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.