K091210 · Best Sanitizers, Inc. · MED · Feb 5, 2010 · General Hospital
Device Facts
Record ID
K091210
Device Name
ACECIDE-C HIGH LEVEL DISINFECTANT AND STERILANT
Applicant
Best Sanitizers, Inc.
Product Code
MED · General Hospital
Decision Date
Feb 5, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6885
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K091210 · Feb 5, 2010
ACECIDE-C HIGH LEVEL DISINFECTANT AND STERILANT
Best Sanitizers, Inc.
Clinical endoscopes (gastroscopes, colonoscopes, bronchoscopes) collected from patients at an endoscopy clinic and a hospital
In-use testing was performed on endoscopes retrieved from clinical settings to verify the disinfectant's efficacy on devices soiled during routine clinical procedures.
In-use testing; Endoscopes; Clinical setting
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
In-use tests with flexible endoscopes; Retrospective/Observational (In-use testing); Follow-up/Duration: 5.0 to 7.0 minutes exposure
Olympus gastroscopes, colonoscopes, and bronchoscopes used in clinical practice; Number of Sites: 2 (Endoscopy clinic and hospital)
Not applicable for this study
Microbial recovery (zero bacteria recovered)
Indications for Use
ACECIDE-C is a high level disinfectant intended to disinfect reusable clean heatsensitive semi-critical medical devices (such as flexible endoscopes) which contact mucous membranes when used at or above its minimum recommended concentration of 2000 ppm peracetic acid for 7.0 minutes at 20°C. ACECIDE-C is a sterilant intended to sterilize reusable clean heat-sensitive critical and semi-critical medical devices (such as flexible endoscopes) which contact and potentially penetrate into sterile body areas, for which there is no other practical method of sterilization, when used at or above its minimum recommended concentration of 2000 ppm peracetic acid for 2.0 hours at 20℃. ACECIDE Test Strips are used with ACECIDE-C High Level Disinfectant and Sterilant use solution to ensure that the level of peracetic acid is above the minimum recommended concentration of 2000 ppm.
Device Story
ACECIDE-C is a two-part liquid chemical system (peracetic acid concentrate and buffer concentrate) mixed with tap water at point-of-use to generate a peracetic acid solution (3100-3400 ppm). Used in clinical settings for high-level disinfection (7 minutes at 20°C) or sterilization (2 hours at 20°C) of heat-sensitive medical devices like flexible endoscopes. Operators use ACECIDE Test Strips to verify peracetic acid concentration remains above 2000 ppm before each use. The test strips utilize a colorimetric reaction (iodide to iodine with starch) to provide visual confirmation of efficacy. The device provides an alternative sterilization method for devices where other methods are impractical, ensuring patient safety by eliminating microbial contamination on reusable instruments.
Clinical Evidence
No clinical trials; evidence based on extensive bench testing. Studies included AOAC Use Dilution (955.14, 955.15, 964.02), Fungicidal (955.17), and Sporicidal (966.04) tests. Results confirmed 100% disinfection of bacterial-labeled cylinders and complete kill of fungi, viruses (Influenza A, HSV-1, Adenovirus, Poliovirus), and M. terrae. Simulated-use testing on bronchoscopes, gastroscopes, and colonoscopes demonstrated >6 log10 reduction of M. terrae and B. subtilis. In-use testing at clinical sites showed zero bacterial recovery from endoscopes after processing.
Technological Characteristics
Two-part liquid chemical system: Solution 1 (acetic acid, hydrogen peroxide, catalyst, stabilizer) and Solution 2 (sequestrant, anti-corrosive, buffer salts). Mixed 1:1:18 with tap water. Active ingredient: peracetic acid (3100-3400 ppm). Final pH 3.4-3.7. Uses colorimetric test strips for concentration verification. Non-electronic, chemical-based disinfection/sterilization.
Indications for Use
Indicated for high-level disinfection or sterilization of reusable, clean, heat-sensitive critical and semi-critical medical devices, including flexible endoscopes, that contact mucous membranes or penetrate sterile body areas.
Regulatory Classification
Identification
A liquid chemical sterilant/high level disinfectant is a germicide that is intended for use as the terminal step in processing critical and semicritical medical devices prior to patient use. Critical devices make contact with normally sterile tissue or body spaces during use. Semicritical devices make contact during use with mucous membranes or nonintact skin.
Special Controls
*Classification.* Class II (special controls). Guidance on the Content and Format of Premarket Notification (510(k)) Submissions for Liquid Chemical Sterilants/High Level Disinfectants, and user information and training.
Predicate Devices
Acecide High Level Disinfectant and Sterilant (K041984)
Submission Summary (Full Text)
{0}------------------------------------------------
## FEB - 5 2010
# Section C 510(k) Summary ACECIDE-C High-Level Disinfectant and Sterilant
Date of Application: April 23, 2009
- 1. Summary of the safety and effectiveness of ACECIDE-C High-Level Disinfectant and Sterilant, a liquid chemical high-level disinfectant and sterilant.
- Applicant/Sponsor a.
Laura King Best Sanitizers Inc. 154 Mullen Drive Walton, KY 41094
Phone: (859) 485-2293 Fax: (859) 485-3572
- b. Application Correspondent
Norman Miner, Ph.D. MicroChem Laboratory, Inc. 1107 - C South Airport Circle Euless (Dallas), TX 76040
Phone: (817) 283 - 9100 (817) 283 - 9110 Fax:
- Name of the Device C.
Trade Name: ACECIDE-C High-Level Disinfectant and Sterilant Common Name: Liquid Chemical High-Level Disinfectant and Sterilant Classification Name: Not Classified
- d. Predicate Name
Acecide High Level Disinfectant and Sterilant (K041984)
- Summary of the substantial equivalence (SE) of ACECIDE-C High-Level e. Disinfectant and Sterilant to the predicate Acecide High Level Disinfectant and Sterilant.
Both ACECIDE-C High-Level Disinfectant (HLD) and Sterilant and Acecide High Level Disinfectant and Sterilant are two part products which are mixed and diluted with water (ACECIDE-C only) at the time of use. Solution 1 for both ACECIDE-C
{1}------------------------------------------------
K091210
p2of5
and Acecide contains concentrations of the chemicals that react to produce peracetic acid (hydrogen peroxide and acetic acid) at low pH values of about 1.0 or less. ACECIDE-C Solution 2 is a concentrated buffer solution, while Acecide Solution 2 is a dilute buffer solution. The mixed use solutions contain antimicrobial concentrations of 3100 to 3800 ppm peracetic acid. ACECIDE-C has a final pH value of about 3.4 to 3.7 and Acecide has a final pH value of about 4.0 to 4.5. Both of these products are intended for the high-level disinfection and/or sterilization of clean. heat-sensitive devices for which there is no other practical method for disinfection or sterilization. Acecide is a legally marketed sterilant. For these reasons, Acecide is an appropriate predicate device for comparisons of substantial equivalence to ACECIDE-C.
#### f. Summary description of ACECIDE-C High-Level Disinfectant and Sterilant.
ACECIDE-C High-Level Disinfectant and Sterilant is packaged in two separate plastic bottles: ACECIDE-C 6.8% Peracetic Acid Concentrate Solution 1 is a concentrate of acetic acid and hydrogen peroxide, which react with each other in the presence of a catalyst and stabilizer to form 6.2 - 6.8 % w/w peracetic acid at a pH value of <1. ACECIDE-C Buffer Concentrate Solution 2 contains a sequestrant, an anti-corrosive, and buffer salts for a pH value of about 8.7 to 9.7. At the time of use, one part of ACECIDE-C 6.8% Peracetic Acid Concentrate Solution 1 and one part of the ACECIDE-C Buffer Concentrate Solution 2 is mixed with 18 parts of tap water at 20℃ for a final concentration of 3100 - 3400 ppm peracetic acid.
#### g. Summary of the intended use of ACECIDE-C High-Level Disinfectant and Sterilant.
ACECIDE-C is a high level disinfectant intended to disinfect reusable clean heatsensitive semi-critical medical devices (such as flexible endoscopes) which contact mucous membranes when used at or above its minimum recommended concentration of 2000 ppm peracetic acid for 7.0 minutes at 20°C.
ACECIDE-C is a sterilant intended to sterilize reusable clean heat-sensitive critical and semi-critical medical devices (such as flexible endoscopes) which contact and potentially penetrate into sterile body areas, for which there is no other practical method of sterilization, when used at or above its minimum recommended concentration of 2000 ppm peracetic acid for 2.0 hours at 20℃.
ACECIDE-C High-Level Disinfectant and Sterilant must be tested for peracetic acid concentration at the time of each use with ACECIDE Test Strips as manufactured by Merck KGaA (Darmstadt Germany). ACECIDE Test Strips are used with ACECIDE-C use solution to ensure that the level of peracetic acid is above the minimum recommended concentration of 2000 ppm.
{2}------------------------------------------------
K091210
page 3 of 5
- h. Summary of the technological characteristics of ACECIDE -C High-Level Disinfectant and Sterilant as Compared with the Predicate, Acecide High-Level Disinfectant and Sterilant.
- ACECIDE-C High-Level Disinfectant and Sterilant and Acecide High-Level i. Disinfectant and Sterilant contain similar concentrations of peracetic acid as an active ingredient, and the use dilution of each are similar. The Acecide use dilution is about 3300 to 3800 ppm peracetic acid and gradually declines through five days of use and re-use at 25°C until its concentration decreases to 1900 ppm peracetic acid. The use dilution of ACECIDE-C starts at 3100 to 3400 ppm peracetic acid and gradually declines through five days of use and re-use at 20°C until its concentration decreases to 2000 pom peracetic acid. In eight separate five day EPA Re-Use Tests at 20°C, the concentration of peracetic acid in ACECIDE-C never decreased below 2000 ppm before the conclusion of each study.
#### i-1. Summary of the AOAC Use Dilution Test using worst-case ACECIDE-C.
Stainless steel penicylinders were labeled with Staphylococcus aureus, Salmonella choleraesuis, and Pseudomonas aeruginosa containing 5% v/v animal serum. and tested according to the methods of the AOAC Use Dilution Test 955.14, 955.15, and 964.02. When exposed to one lot of ACECIDE-C at a worst-case peracetic acid concentration of 2000 ppm, from a 5-day EPA Reuse Test, all (100%) of the bacterialabeled cylinders were disinfected following exposures of 1.0, 3.0, and 5.0 minutes at 20°C.
### i-2. Summary of the AOAC Fungicidal Test 955.17 using worst-case ACECIDE-C.
Cultures of Trichophyton mentagrophytes, Aspergillus niger, and Candida albicans with 5% v/v animal serum were tested against one lot of worst-case ACECIDE-C at 2000 ppm peracetic acid from a 5-day EPA Reuse Test according to the methods of the AOAC Fungicidal Test 955.17. All of these surrogate fungi were killed with exposures of 1.0, 2.5, and 5.0 minutes at 20℃.
#### i-3. Summary of the virucidal tests using worst-case ACECIDE-C.
Influenza A Virus, Herpes Simplex Virus type 1, and Adenovirus type 2, all containing 5% v/v animal serum, were killed within the limits of detection with exposure for 5.0 minutes at 20℃ to ACECIDE-C at 2000 ppm peracetic acid from a 5-day EPA Reuse Test. Poliovirus type 1 was killed within the limits of detection with exposure for 7.0 minutes at 20℃. This Poliovirus test determined the exposure time of 7.0 minutes for high level disinfection for ACECIDE-C.
### i-4. Summary of the mycobactericidal tests using worst-case ACECIDE-C.
{3}------------------------------------------------
K091210
page 4 of 5
Worst-case ACECIDE-C from a 5-day EPA Reuse Test, further diluted to 2000 ppm peracetic acid, killed 6.0 log10 of M. terrae within about 4.0 min and about 8 log10 were killed within 5.0 min of exposure time at 20℃.
#### i-5. Summary of the studies to determine the sterilization exposure time for ACECIDE-C.
- One Lot of ACECIDE-C from a 5-day EPA Reuse Test, further diluted to 2000 ppm peracetic acid, was tested against 40 of each type of spore-labeled carrier with exposures of 15.0. 30.0, 60.0, 120.0, and 240.0 hours at 20°C according to the methods of the AOAC Sporicidal Test 966.04. All of the B. subtilis-labeled carriers were sterilized within 15.0 minutes and all of the C. sporogenes-labeled carriers were sterilized within 1.0 hour. A sterilization exposure time of 2.0 hours was selected to provide a margin of safety.
### i-6. Summary of the results of a full three lot AOAC Sporicidal Test 966.04 of ACECIDE-C.
Three Lots of ACECIDE-C from a 5-day EPA Reuse Test. further diluted to 2000 ppm peracetic acid, passed the AOAC Sporicidal Test 966.04 with an exposure of 2.0 hours at 20℃.
### i-7. Summary of the results of a Confirmatory Sporicidal Test with ACECIDE-C.
One Lot of ACECIDE-C from a 5-day EPA Reuse Test, further diluted to 2000 ppm peracetic acid, passed the Confirmatory AOAC Sporicidal Test 966.04 with an exposure of 2.0 hours at 20℃.
- i-8. Summary of the results of simulated-use tests with flexible endoscopes inoculated with M. terrae and disinfected with ACECIDE-C.
The interior channels and exterior surfaces of a bronchoscope, gastroscope, and colonoscope were inoculated with a culture of M. terrae with 5% v/v animal serum and dried for one hour. The M. terrae-inoculated endoscopes were then exposed to worst-case ACECIDE-C from a 5-day EPA Reuse Test, further diluted to 2000 ppm peracetic acid for 5.0 or 7.0 minutes at 20℃. At least 6 log10 of the M. terrae was killed by the ACECIDE-C within 7.0 minutes.
- i.9. Summary of the results of simulated-use tests with flexible endoscopes inoculated with B. subtilis and sterilized with ACECIDE-C.
The interior channels and exterior surfaces of a bronchoscope, gastroscope, and colonoscope were inoculated with a culture of B. subtilis with 5% v/v animal serum and dried for one hour. The B. subtilis-inoculated endoscopes were then exposed to worst-case ACECIDE-C from a 5-day EPA Reuse Test, further diluted to 2000 ppm peracetic acid for 2.0 hours at 20℃. At least 6 log10 of the B. subtilis was killed by the ACECIDE-C within 2.0 hours.
{4}------------------------------------------------
K091210
page 5 of 5
#### Summary description of in-use tests with ACECIDE-C. i.
Olympus gastroscopes and colonoscopes as received directly from patients at an endoscopy clinic, and cleaned, but not disinfected, according to standard cleaning procedures of the clinic, were exposed for 5.0 minutes at 20°C to worst-case ACECIDE-C from a 5-day EPA Reuse Test, further diluted to 2000 ppm peracetic acid. No (zero) bacteria were recovered from these endoscopes after the exposure to ACECIDE-C.
Olympus bronchoscopes as received directly from patients at a hospital, and processed as above, were exposed for 7.0 minutes to ACECIDE-C at 20°C as above. No (zero) bacteria were recovered from these endoscopes after the exposure to ACECIDE-C.
#### k. Summary of ACECIDE Test Strips
ACECIDE Test Strips are used with ACECIDE-C use solution to ensure that the level of peracetic acid is above the minimum recommended concentration of 2000 ppm.
The test strips are based on the same basic chemistry as other peracetic acid test strips and assay methods. Peracetic acid oxidizes iodide to iodine with starch forming a visible blue-gray/blue-black compound with iodine. This color change in the indicator paper is the visual basis for the tests strips and their reading.
ACECIDE-C must be tested for peracetic acid concentration at the time of each use with ACECIDE Test Strips as manufactured by Merck KGaA (Darmstadt Germany).
- Summary of the conclusions drawn from the tests that indicate that ACECIDE-C l. High Level Disinfectant and Sterilant is as safe and as effective or better, than Acecide High Level Disinfectant and Sterilant.
We conclude that ACECIDE-C High-Level Disinfectant and Sterilant is as safe as Acecide High Level Disinfectant and Sterilant because the two products contain similar concentrations of peracetic acid throughout the use life, and are used at similar exposure temperatures. We conclude that ACECIDE-C is as antimicrobial or more antimicrobial than Acecide based on similar high level disinfection exposure times and temperatures, and that ACECIDE-C has a faster sterilization time at a lower temperature (2.0 hours at 20°C versus 5.0 hours at 25°C).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol resembling an abstract human figure with three flowing lines extending from the head, representing growth and vitality.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Best Sanitizers, Incorporated C/O Norman Miner, Ph. D. President/Study Director MicroChem Laboratory, Incorporated 1107 - C South Airport Circle Euless (Dallas), Texas 76040
FEB - 5 2010
Re: K091210
Trade/Device Name: ACECIDE -C High-Level Disinfectant and Sterilant ACECIDE Test Strips Regulation Number: 21 CFR 880.6885 Regulation Name: Liquid Chemical Sterilants/ High Level Disinfectants Regulatory Class: II Product Code: MED Dated: January 7, 2010 Received: January 13, 2010
Dear Dr. Miner: .
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
| 510(k) Number (if known): | K091210 |
|---------------------------|-------------------------------------------------|
| Device Name: | ACECIDE-C High-Level Disinfectant and Sterilant |
Indications For Use:
ACECIDE-C is a high level disinfectant intended to disinfect reusable clean heatsensitive semi-critical medical devices (such as flexible endoscopes) which contact mucous membranes when used at or above its minimum recommended concentration of 2000 ppm peracetic acid for 7.0 minutes at 20°C.
ACECIDE-C is a sterilant intended to sterilize reusable clean heat-sensitive critical and semi-critical medical devices (such as flexible endoscopes) which contact and potentially penetrate into sterile body areas, for which there is no other practical method of sterilization, when used at or above its minimum recommended concentration of 2000 ppm peracetic acid for 2.0 hours at 20℃.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
\$
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of 1
510(k) Number: Ko 91710
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K091210
Device Name:
Indications For Use:
ACECIDE Test Strips are used with ACECIDE-C High Level Disinfectant and Sterilant use solution to ensure that the level of peracetic acid is above the minimum recommended concentration of 2000 ppm.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ × (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth D'Clannin-Wells.
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
510(k) Number: 人 09 | 乙| 0
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.