WILSON-COOK BALLOON INFLATION DEVICE
Device Facts
| Record ID | K960415 |
|---|---|
| Device Name | WILSON-COOK BALLOON INFLATION DEVICE |
| Applicant | Wilson-Cook Medical, Inc. |
| Product Code | KNT · Gastroenterology, Urology |
| Decision Date | Apr 26, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5980 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
This device is a disposable device to be used to inflate, deflate and monitor pressure for gastrointestinal dilation balloons.
Device Story
Hand-held mechanical device; 20cc syringe; plunger mechanism; lever; extension line; stopcock; analog pressure gauge (300 psi/20atm). Used by clinicians during GI procedures to inflate/deflate dilation balloons; monitors pressure via analog gauge. Disposable; sterile or non-sterile. Benefits: provides controlled inflation/deflation and real-time pressure monitoring for GI dilation.
Clinical Evidence
Bench testing only. In-house functional testing demonstrated device integrity and adherence to specifications.
Technological Characteristics
Mechanical hand-held device; 20cc syringe; analog pressure gauge (300 psi/20atm); plunger/lever mechanism; extension line; stopcock. Sterilization: Validated EO cycle per AAMI Overkill Method (SAL 10⁻⁶).
Indications for Use
Indicated for inflation, deflation, and pressure monitoring of gastrointestinal dilation balloons in patients requiring GI dilation procedures.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Related Devices
- K991375 — BARD BALLOON INFLATION SYSTEM · C.R. Bard, Inc. · Sep 10, 1999
- K122321 — BASIXCONPAK ANALOG INFLATION SYRINGE · Merit Medical Ireland, Ltd. · Dec 19, 2012
- K972964 — ATRION MEDICAL BALLOON CATHETER INFLATION DEVICE · Atrion Medical Products, Inc. · Oct 10, 1997
- K140943 — MASTRO PLUS BALLOON IN-DEFLATION DEVICE · Beijing Demax Medical Technology Co.,Ltd · Mar 5, 2015
- K130566 — BASIXTOUCH · Merit Medical Systems, Inc. · Jun 19, 2013