K954215 · Advanced Surgical Intervention, Inc. · MNG · May 8, 1996 · Gastroenterology, Urology
Device Facts
Record ID
K954215
Device Name
MINIGUARD
Applicant
Advanced Surgical Intervention, Inc.
Product Code
MNG · Gastroenterology, Urology
Decision Date
May 8, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5160
Device Class
Class 1
Attributes
Therapeutic, Expedited Review
Indications for Use
The Miniguard is an external female continence device that is applied to the urinary opening to prevent or decrease episodes of stress incontinence. Stress incontinence occurs when intra-abdominal pressure exceeds urethral resistance as a result of physical stress such as coughing, laughing, or lifting heavy objects. The Miniguard is designed for women who experience urine leakage under conditions of physical stress.
Device Story
Miniguard is an external female continence device; consists of a small foam pad coated with biocompatible polymer adhesive. Applied intralabially over the urinary opening to provide mechanical occlusion of the urethra; prevents or decreases urine leakage during physical stress. Used by patients for self-management of stress incontinence. Device provides a physical seal; reduces leakage episodes, perceived severity, and impact on quality of life. Benefits include improved continence control and daily living comfort. No complex electronics or software; purely mechanical/adhesive function.
Clinical Evidence
Prospective clinical study of 350 women across 12 centers over 12 weeks. Evaluated leakage episodes, perceived severity, quality of life, and urine volume. Results showed statistically significant improvements (p≤0.0001) across all metrics: leakage episodes reduced by 65.5%, perceived severity by 71.1%, quality of life impact by 71.4%, and leakage volume by 61.1%. Safety assessments included bladder function (cystometry, post-void residual), microbiology (vaginal/urine cultures), and dermatology (physical exams, cytology), showing no clinically significant adverse effects or increased infection rates.
Technological Characteristics
External mechanical device; foam pad backing with biocompatible polymer adhesive. Intralabial application. Materials have history in oral, ocular, vaginal, and wound care applications. No energy source, connectivity, or software. Biocompatibility tested per FDA tripartite and ISO 194 guidelines.
Indications for Use
Indicated for women experiencing stress urinary incontinence to prevent or decrease leakage episodes caused by physical stress (e.g., coughing, laughing, lifting).
Regulatory Classification
Identification
A urological clamp for males is a device used to close the urethra of a male to control urinary incontinence or to hold anesthetic or radiography contrast media in the urethra temporarily. It is an external clamp.
Predicate Devices
Dacomed C³ Male Continence Device
Diamed External Urinary Appliance
Xtramedics Fresh 'n Fit Padette
Submission Summary (Full Text)
{0}
K954215
MAY - 8 1996
# 510(k) Summary
## ASI Miniguard Female Continence Device
1. Name: Advanced Surgical Intervention
Address: P.O. Box 3134
Dana Point, CA 92629
Phone: (714) 240-5899
Contact: Robert Rosenblut, PhD
2. Device Name: Female Continence Device
Proprietary Name: Miniguard
Classification Name: Unknown
3. Predicate Devices:
Dacomed C³ Male Continence Device
Diamed External Urinary Appliance
Xtramedics Fresh 'n Fit Padette
4. Device Description
The Miniguard female continence device consists of a small pad that is coated on one side with a biocompatible polymer adhesive. The Miniguard is an external management device that is applied to the urinary opening to prevent or decrease episodes of stress incontinence. The Miniguard's function is to provide occlusion of the urinary opening by creating a seal over the urinary opening. The foam pad backing is intended to facilitate application and removal and for user comfort.
5. Intended Use
The Miniguard is an external female continence device that is applied to the urinary opening to prevent or decrease episodes of stress incontinence. Stress incontinence occurs when intra-abdominal pressure exceeds urethral resistance as a result of physical stress such as coughing, laughing, or lifting heavy objects. The Miniguard is designed for women who experience urine leakage under conditions of physical stress.
6. Technological Characteristics Comparison
The Miniguard female continence device is substantially equivalent in function to the Dacomed C³ Male Continence Device. Both the Miniguard and the Dacomed device achieve the intended use (to prevent or decrease episodes of incontinence) by occlusion of the urethra. The Miniguard is substantially equivalent in design characteristics to the Diamed External Urinary Appliance. The design characteristics are equivalent to the Diamed device in that both utilize an adhesive seal on the intralabial mucosa. The Miniguard also has technological similarities to the Xtramedics device in that both are worn intralabially, over the urethral opening. The materials used in the Miniguard have extensive histories in a variety of biomedical applications including oral, ocular, vaginal and wound care devices.
7. Nonclinical Tests
The Miniguard and it constituent materials were tested for biocompatibility, toxicity, cytotoxicity, bacteriostasis/fungistasis, and skin sensitivity. The testing exceeds the guidelines per the FDA tripartite and ISO 194 for a device that is chronically used on intact mucous membrane. The results of these tests demonstrated that the materials and the whole device are biocompatible, non-toxic and well tolerated by cutaneous and subcutaneous tissue. In addition, bacteriological testing indicated that the materials do not support growth of common urologic pathogens.
Section 8 - Page 1
{1}
# 8. Clinical Tests
The Miniguard device has been extensively tested for its safety and efficacy in decreasing the number of urinary leakage episodes in women with incontinence. All testing results indicate that use of the Miniguard imparts minimal risk, yet provides significant benefit for women in controlling urinary leakage.
Clinical testing on the Miniguard was completed on over 350 women from 12 investigative centers in the United States. Women used the Miniguard during a device usage period of 12 weeks to test four efficacy hypotheses which demonstrate the Miniguards ability to 1) decrease the number of leakage episodes; 2) reduce the perceived severity of urinary leakage; 3) reduce the impact of incontinence on quality of life; and 4) reduce the quantity of urine leakage. Testing materials used were: 1) a voiding diary; 2) a symptom questionnaire; 3) an incontinence impact questionnaire; and 4) a 12-hour home pad-weight test. All testing materials were derived from published sources on incontinence testing and management. Efficacy parameters were statistically analyzed using graphical analyses, paired-t analyses (Prob>|T|) and repeated-measures (MANOVA) analyses. Analyses' results demonstrate, in each of the four hypothesis, a statistically and clinically significant improvement for participants. The table below details the average values before using the Miniguard (Control) and after 12 weeks of device use (Trial 17).
| Hypothesis | Control | Trial 17 | % Improvement | Prob>|T| |
| --- | --- | --- | --- | --- | --- | --- |
| 1. No. of Leakage Episodes (per week) | 14.18 | 4.89 | 65.5% | ≤0.0001 |
| 2. Perceived Severity of Leakage | 11.02 | 3.18 | 71.1% | ≤0.0001 |
| 3. Impact on Quality of Life | 10.42 | 2.98 | 71.4% | ≤0.0001 |
| 4. Quantity of Leakage (grams/hr) | 1.31 | 0.51 | 61.1% | ≤0.0001 |
The women in this study had varying degrees of urinary incontinence. The incontinence was considered mild if the women had less than 8 episodes of incontinence per week, moderate for 8-21 episodes per week and severe for greater than 21 episodes per week. Based on this classification¹, 124 women had mild incontinence, 163 had moderate incontinence and 69 had severe incontinence.
# Effectiveness
The study showed that after three months of using the Miniguard, women experienced fewer leakage episodes and improved their overall quality of life. The severity of urine leakage is a measure of the patient's perception of the degree of leakage during various activities. Thirteen activities were rated for a total possible maximum score of 39. The impact of urinary incontinence on quality of life is a measure of the patient's perception of the degree to which leakage had a negative effective on various aspects of daily living. Twenty-six aspects were rated for a total possible maximum score of 78. Below are the results showing the average changes and improvement of the study participants.
Effectiveness Data for Patients Using Miniguard - Average Values
| Study Endpoints | Mild | | Moderate | | Severe | |
| --- | --- | --- | --- | --- | --- | --- |
| | Control | Trial* | Control | Trial* | Control | Trial* |
| Number of Leakage Episodes in One Week² | 4.48 | 2.20 | 13.14 | 4.75 | 34.06 | 9.84 |
| Severity of Urine Leakage During Various Physical Activities³ | 8.62 | 1.97 | 11.39 | 3.28 | 14.46 | 5.01 |
| Impact Urinary Incontinence has on Quality of Life⁴ | 6.73 | 1.76 | 10.85 | 2.84 | 16.04 | 5.43 |
| Volume of Urine Lost over a 12 Hour Period (grams/hr)⁵ | 0.57 | 0.35 | 1.10 | 0.59 | 3.16 | 0.59 |
*Trial information based on data collected after 12 weeks of Miniguard usage.
Section 8 - Page 2
{2}
To provide clinical safety assurance of the Miniguard, the effects of use on bladder function, local microbiology and dermatology were evaluated. Bladder function was assessed via post-void residual measurements and a cohort of patients underwent cystometry. There were no clinically significant changes noted and there was no indication that Miniguard use adversely effected bladder function. Microbiological testing included vaginal smears, vaginal cultures (for a cohort of subjects), urinalysis and urine culture. Results of the microbiological testing show no increased incidence of either vaginal infection or urinary tract infection associated with device usage. Dermatologic testing included physical examinations and vestibular cytology. Results of dermatologic testing show the Miniguard was well-tolerated by the intralabial mucosa and that there were no significant dermatologic effects resulting from device usage.
The sole adverse event which may be associated with Miniguard usage was a minor increase in the incidence of subject reported symptoms of irritation, characteristic of subjective perceptions of minor irritation associated with use of any topical device. Of the women who reported symptoms of vestibular irritation, the majority experienced one occurrence only and continued using the Miniguard with no further symptoms and no required intervention from the investigator.
To further assess the safety data, ASI submitted study data to teams of physicians for analysis. Bladder function data were assessed by leading urologists and urogynecologists; microbiology data were assessed by a core laboratory at the University of Washington specializing in urogenital microflora; dermatology data were assessed by an expert in dermatology and gynecology at Emory University. These independent reports confirmed ASI's analyses results and are included in the 510(k) in Sections 10-14. Statistical analyses were completed on bladder function data and to assess the incidence rates for bacterial vaginosis and urinary tract infection. The results of these analyses (t-tests and McNemar's analyses) showed no statistically significant differences between the control period and device usage periods.
## 9. Conclusions
The safety of the Miniguard female continence device was demonstrated by the extensive positive experience of the materials in medical applications and the nonclinical and clinical testing. Further, technological characteristics do not raise new types of safety and effectiveness questions relative to the predicate devices. The clinical investigation demonstrated that Miniguard usage did not effect bladder function and did not appear to effect the incidence rates for urinary tract or vaginal infection. The Miniguard was well tolerated by the intralabial mucosa and there were no significant dermatologic effects associated with device usage. The efficacy of the Miniguard was demonstrated by the clinical testing that included both subjective and objective measures. The clinical data showed that statistically significant improvement was obtained by study subjects when using the Miniguard. In summary, these data provide reasonable assurance that the Miniguard female continence device is a safe and effective alternative for women requiring stress incontinence management and is substantially equivalent to the predicate devices.
## References
1. Burns PA, et al: A Comparison of Effectiveness of Biofeedback and Pelvic Muscle Exercise Treatment of Stress Incontinence in Older Community-Dwelling Women. J Gerontol, Medical Sciences. 48 (4):M167-M174, 1993.
2. Ostergard DR, Bent AE, eds: Urogynecology & Urodynamics. Williams & Wilkins, Baltimore, MD. P209-211, 1991.
3. Jeter KF, Wagner DD: Incontinence in the American Home - A survey of 36,500 people. Journal of the American Geriatrics Society. 38:379-383, 1990.
4. Wyman JF, et al: Psychosocial Impact Urinary Incontinence in Women. Obstet Gynecol. Vol 70, P378, 1987.
5. Ostergard DR, Bent AE, eds: Urogynecology & Urodynamics. Williams & Wilkins, Baltimore, MD. P211-213, 1991.
Section 8 - Page 3
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
10903 New Hampshire Avenue
Document Control Room W-066-0609
Silver Spring, MD 20993-0002
APR 26 2010
Robert F. Rosenbluth, Ph.D.
Advanced Surgical Intervention
P.O. Box 3134
DANA POINT CA 92629
Re: K954215
Trade/Device Name: Miniguard
Regulation Number: 21 CFR§ 876.5160
Regulation Name: Urological clamp for males
Regulatory Class: I
Product Code: MNG
Dated: April 26, 1996
Received: April 29, 1996
Dear Dr. Rosenbluth:
This letter corrects our substantially equivalent letter of May 8, 1996.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}
Page 2 Robert F. Rosenbluth, Ph.D.
K954215
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health’s (CDRH’s) Office of Compliance. Also, please note the regulation entitled, “Misbranding by reference to premarket notification” (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.