ORTHO-MUNE OKT 4A (CD4) - MONOCLONAL ANTIBODY (MURINE) FITC CONJUGATE

K951459 · Ortho Diagnostic Systems, Inc. · GKZ · Jul 14, 1997 · Hematology

Device Facts

Record IDK951459
Device NameORTHO-MUNE OKT 4A (CD4) - MONOCLONAL ANTIBODY (MURINE) FITC CONJUGATE
ApplicantOrtho Diagnostic Systems, Inc.
Product CodeGKZ · Hematology
Decision DateJul 14, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

Ortho-mune OKT4A (CD4) Monoclonal Antibody (Murine) FITC Conjugate is used to identify and enumerate the percentage of CD4+ human T lymphocytes in whole blood by flow cytometry. Identification and enumeration of abnormal levels of CD4 lymphocytes may be clinically significant in the prognosis of secondary immunodeficiency diseases (acquired immunodeficiency syndrome [AIDS]). AIDS is characterized by a severe reduction in T helper (CD3+CD4+) lymphocytes and a decrease in the CD4:CD8 ratio. The CD4:CD8 ratio has been used as a predictor of time to the development of AIDS. CD4 counts should also be determined using a CD3/CD4 combination reagent to eliminate monocyte contamination (CD3-CD4+dim). CD4+lymphocytes (expressed as an absolute number, a percentage of lymphocytes or as a ratio of CD4+ to CD8+ T lymphocytes) represents the best single predictor of the progression of AIDS. CD4+ lymphocyte numbers usually start to decline relatively soon after human immunodeficiency virus (HIV) infection.

Device Story

Ortho-mune OKT4A (CD4) FITC Conjugate is a murine monoclonal antibody conjugated to fluorescein isothiocyanate (FITC). It is used in clinical laboratories by trained personnel to stain whole blood samples. The reagent binds to CD4+ T lymphocytes, which are then detected and enumerated using a flow cytometer (e.g., ORTHO CYTORONABSOLUTE). The device provides quantitative data on the percentage of CD4+ cells. Clinicians use these results to assess the progression of HIV/AIDS, as CD4+ lymphocyte counts are a primary predictor of disease status. The device aids in clinical decision-making by providing objective measurements of immune cell populations, helping monitor patients with immunodeficiency.

Clinical Evidence

Clinical performance evaluated at three external sites using 206 normal donors and 88 AIDS/ARC patients. Comparison against predicate device showed equivalent mean percentages of CD4+ cells. Within-laboratory and between-laboratory reproducibility studies (N=10) demonstrated low coefficients of variation (CV) across low, normal, and high CD4 concentrations. Linearity studies confirmed performance across a range of 20 to 9000 cells/uL with slopes near 1.0 and R values of 1.000.

Technological Characteristics

Murine monoclonal antibody (OKT4A) conjugated to fluorescein isothiocyanate (FITC). Designed for use with flow cytometry systems. Reagent is an in vitro diagnostic immunophenotyping tool.

Indications for Use

Indicated for identification and enumeration of CD4+ human T lymphocytes in whole blood via flow cytometry to assist in the prognosis of secondary immunodeficiency diseases, specifically AIDS, in patients where CD4+ lymphocyte levels are clinically relevant.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0} Ortho Diagnostic Systems Inc. Raritan, NJ 08869-0606 Ortho-mune OKT4A (CD4) FITC Conjugate Ref. No. K951459 - Submitted July, 1997 K951459 510(k) Summary JUL 14 1997 SUBMITTER: Ortho Diagnostic Systems Inc. 1001 U.S. Highway 202 Raritan, NJ 08869-0606 CONTACT: Gail Kromer Tel: (908) 218-8179 Fax: (908) 218-8168 DEVICE NAME: Ortho-mune™ OKT™4A (CD4) Monoclonal Antibody (Murine) FITC Conjugate PREDICT: CD4(Leu™-3a) FITC DATE: July 8, 1997 ## DEVICE DESCRIPTION Ortho-mune OKT4A Monoclonal Antibody (Murine) FITC Conjugate contains the purified monoclonal antibody OKT4A conjugated to the fluorochrome fluorescein isothiocyanate. ## INTENDED USE: Ortho-mune OKT4A FITC Conjugate is intended for use in identification and enumeration of CD4+ human T lymphocytes in whole blood by flow cytometry. The intended use is the same as the intended use of the predicate device, CD4 (Leu-3a) FITC commercially distributed by Becton Dickinson Immunocytometry Systems. ## TECHNOLOGICAL CHARACTERISTICS Both Ortho-mune OKT4A Monoclonal Antibody (Murine) FITC Conjugate and CD4 (Leu-3a) FITC utilize monoclonal antibodies specific for human helper/inducer T cells OKT4A/Leu-3a) respectively, conjugated to the same fluorochrome, fluorescein isothiocyanate. c:\data\word\CD4 Page 1 {1} Ortho Diagnostic Systems Inc. Raritan, NJ 08869-0606 Ortho-mune OKT4A (CD4) FITC Conjugate Ref. No. K951459 - Submitted July, 1997 # PERFORMANCE CHARACTERISTICS Performance of Ortho-mune OKT4A Monoclonal Antibody (Murine) FITC Conjugate was compared with that of CD4 (Leu-3a) FITC at three external, geographically distinct sites. Whole blood specimens from 206 normal donors, and 88 AIDS/ARC patients were stained and analyzed using the ORTHO CYTORONABSOLUTE™ flow cytometer, Ortho Diagnostic Systems Inc. For each specimen, the percentage of gated cells which showed positive by each marker was calculated. The mean and range of the percent CD4+ cells for the normal donor and AIDS/ARC population are shown in Table 1 and Table 2 respectively. TABLE 1 | PERCENT POSITIVE STAINED CELLS IN NORMAL DONORS DETECTED BY OKT4A AND LEU-3A ASSAYED ON THE CYTORONABSOLUTE N=206 | | | | | | | --- | --- | --- | --- | --- | --- | | Ortho-mune Reagent | Mean % | Range % | BD Reagent | Mean % | Range % | | OKT4A (CD4)+ | 47.0 | 12.4 - 68.7 | LEU-3a (CD4)+ | 45.1 | 14.6 - 64.6 | TABLE 2 | PERCENT POSITIVE STAINED CELLS IN AIDS/ARC PATIENTS DETECTED BY OKT4A AND LEU-3A ASSAYED ON THE CYTORONABSOLUTE N=88 | | | | | | | --- | --- | --- | --- | --- | --- | | Ortho-mune Reagent | Mean % | Range % | BD Reagent | Mean % | Range % | | OKT4 (CD4)+ | 18.3 | 1.0 - 55.1 | LEU-3a (CD4)+ | 17.2 | 0.6 - 51.6 | c:\data\word\CD4 Page 2 {2} Ortho Diagnostic Systems Inc. Raritan, NJ 08869-0606 Ortho-mune OKT4A (CD4) FITC Conjugate Ref. No. K951459 - Submitted July, 1997 Linear regression analysis of total percent CD4+ cells from the combined normal and AIDS/ARC populations is found in Chart 1. # CHART 1 ## Ortho-mune OKT4A (CD4) FITC vs CD4 (Leu-3a) FITC ![img-0.jpeg](img-0.jpeg) Ortho-mune OKT4A(FITC) This study demonstrates that the performance of Ortho-mune OKT4A (CD4) Monoclonal Antibody (Murine) FITC Conjugate is equivalent to CD4 (Leu-3a) FITC reagent in identification and enumeration CD4+ human lymphocytes in whole blood by flow cytometry. c:\data\word\CD4 Page 3 {3} Ortho Diagnostic Systems Inc. Raritan, NJ 08869-0606 Ortho-mune OKT4A (CD4) FITC Conjugate Ref. No. K951459 - Submitted July, 1997 # WITHIN-LABORATORY REPRODUCIBILITY Ten normal donors were used in a within-laboratory reproducibility study at three independent laboratories. The samples were processed using antibody-coated magnetic microbeads to produce low, normal and high percent positive CD4 populations to simulate leukopenia and leukocytosis. The samples were stained with Ortho-mune OKT4A (CD4) FITC Conjugate and run on the ORTHO CYTORONABSOLUTE. All samples were analyzed in replicates of ten. The coefficients of variation, calculated from the standard deviation between replicates, were compared and demonstrated excellent within-laboratory reproducibility at all concentrations tested. The 95% confidence intervals were calculated from the standard error of the site mean which includes variance components from replicate and donor variability minimized by the number of donors (10). The comparison of normal, concentrated (high) and diluted (low) samples is contained in Table 3. | TABLE 3: N = 10 WITHIN-LABORATORY REPRODUCIBILITY Ortho-mune OKT4A (CD4) FITC CONJUGATE PERCENT POSITIVE RESULTS | | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | CD4 Level | Low | | | Normal | | | High | | | | Subset: OKT4A (CD4) | Site 1 | Site 2 | Site 3 | Site 1 | Site 2 | Site 3 | Site 1 | Site 2 | Site 3 | | Mean Percent Positive | 9.60 | 9.36 | 8.86 | 47.38 | 47.19 | 48.54 | 56.13 | 55.83 | 56.50 | | CV | 7.60 | 6.12 | 5.47 | 2.33 | 2.00 | 2.21 | 1.93 | 1.79 | 2.09 | | +/- 95% Confidence Interval | 4.18 | 3.53 | 3.45 | 3.74 | 3.84 | 4.22 | 4.93 | 4.91 | 5.15 | c:\data\word\CD4 Page 4 {4} Ortho Diagnostic Systems Inc. Raritan, NJ 08869-0606 Ortho-mune OKT4A (CD4) FITC Conjugate Ref. No. K951459 - Submitted July, 1997 # BETWEEN-LABORATORY REPRODUCIBILITY Ten normal donor samples were compared between three independent laboratories. The samples were processed using antibody-coated magnetic microbeads to produce low, normal and high percent positive CD4 populations. The samples were stained with Ortho-mune OKT4A (CD4) FITC Conjugate and run on the ORTHO CYTORONABSOLUTE. All samples were analyzed in replicates of ten. The between-laboratory coefficient of variation (CV) and the 95% confidence intervals were calculated from the standard deviation between site means. The data collected from all sites are shown in Table 4. The data show excellent between-laboratory reproducibility. | TABLE 4: N = 10 BETWEEN-LABORATORY REPRODUCIBILITY Ortho-mune OKT4A (CD4) FITC CONJUGATE PERCENT POSITIVE RESULTS | | | | | --- | --- | --- | --- | | CD4 Level | Low | Normal | High | | Mean Percent Positive | 9.25 | 47.70 | 56.15 | | CV | 4.09 | 1.53 | 0.60 | | +/- 95% Confidence Interval | 1.63 | 3.14 | 1.46 | Ortho-mune OKT4A (CD4) Monoclonal Antibody (Murine) FITC Conjugate immunophenotyping reagent shows acceptable within and between laboratory reproducibility for determination of CD4+ lymphocyte percentages. c:\data\word\CD4 Page 5 {5} Ortho Diagnostic Systems Inc. Raritan, NJ 08869-0606 Ortho-mune OKT4A (CD4) FITC Conjugate Ref. No. K951459 - Submitted July, 1997 A linearity study was performed using an automated hematology analyzer to determine total lymphocyte count, and the CYTORONABSOLUTE flow cytometer to determine the percent positive CDx cells. Specimens from four normal donors (whole blood, EDTA) were processed to produce samples with low, normal and high numbers of lymphocyte subsets. Each whole blood specimen was concentrated by harvesting the buffy coat to obtain a white blood cell count between 20,000 and 40,000 cells/ul and then diluting to produce samples of high, normal and low numbers of lymphocyte subsets. A portion of each sample was stained in triplicate using Ortho-mune OKT4A (CD4) FITC Conjugate immunophenotyping reagent and analyzed using the CYTORONABSOLUTE flow cytometer. The total lymphocyte count of the concentrated sample for each donor was obtained using an automated hematology analyzer. Linear regression analyses were performed as follows. The expected (X axis) values were calculated by multiplying the corresponding serial dilutions by the hematology analyzer derived buffy coat lymphocyte count and by the CYTORONABSOLUTE derived lymphocyte subset percent positive. The observed (Y axis) values were determined as the total lymphocyte count calculated from the hematology derived value of the concentrated sample times the CYTORONABSOLUTE derived lymphocyte subset percent positive at each dilution. The Ortho-mune OKT4A (CD4) FITC Conjugate reagent demonstrated linear performance for total CD4+ lymphocyte subsets across a lymphocyte count range of 20 cells/uL to 9000 cells/uL as demonstrated with slopes indistinguishable from 1 and R values of 1.000. Linear regression analyses of observed versus expected values for total percent CD4+ cells for each donor specimen are shown in. Regression analysis statistics are provided in Table 5. c:\data\word\CD4 Page 6 {6} Ortho Diagnostic Systems Inc. Raritan, NJ 08869-0606 Ortho-mune OKT4A (CD4) FITC Conjugate Ref. No. K951459 - Submitted July, 1997 # CHART 2 ![img-1.jpeg](img-1.jpeg) Ortho-mune OKT4A(FITC) c:\data\word\CD4 Page 7 {7} Ortho Diagnostic Systems Inc. Raritan, NJ 08869-0606 Ortho-mune OKT4A (CD4) FITC Conjugate Ref. No. K951459 - Submitted July, 1997 TABLE 5 | CONCENTRATION ORTHO-MUNE OKT4A (CD4) FITC Conjugate | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | Ortho-mune | Donor | SLOPE | CI | INTERCEPT | CI | R | | TOTAL CD4 | 1 | 0.999 | 0.003 | 6.099 | 5.242 | 1.000 | | TOTAL CD4 | 2 | 1.000 | 0.001 | 0.395 | 2.226 | 1.000 | | TOTAL CD4 | 3 | 0.999 | 0.003 | 4.449 | 4.176 | 1.000 | | TOTAL CD4 | 4 | 1.000 | 0.001 | 0.285 | 1.316 | 1.000 | | TOTAL CD4 | All | 0.999 | 0.001 | 2.777 | 1.655 | 1.000 | ## CONCLUSION Performance of Ortho-mune OKT4A (CD4) Monoclonal Antibody (Murine) FITC Conjugate is substantially equivalent to CD4 (Leu-3a) FITC reagent in the identification and enumeration of CD4+ human T lymphocytes in whole blood by flow cytometry. c:\data\word\CD4 Page 8 {8} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Gail Kromer Manager, Regulatory Affairs Ortho Diagnostic Systems, Inc. 1001 U.S. Highway 202 Raritan, New Jersey 09969-0606 Re: K951459/S3 Trade Name: Ortho-mune™ OKT™4A (CD4) Monoclonal Antibody (Murine) FITC Conjugate Regulatory Class: II Product Code: GKZ Dated: April 21, 1997 Received: April 23, 1997 Dear Ms. Kromer: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. {9} Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {10} Page 1 of 1 # Indications for Use Statement 510(k) Number (if known) K951459 Device Name: Ortho-mune™OKT™4A (CD4) Monoclonal Antibody (Murine) FITC Conjugate Indications for Use: Ortho-mune OKT4A (CD4) Monoclonal Antibody (Murine) FITC Conjugate is used to identify and enumerate the percentage of CD4+ human T lymphocytes in whole blood by flow cytometry. Identification and enumeration of abnormal levels of CD4 lymphocytes may be clinically significant in the prognosis of secondary immunodeficiency diseases (acquired immunodeficiency syndrome [AIDS]). AIDS is characterized by a severe reduction in T helper (CD3+CD4+) lymphocytes and a decrease in the CD4:CD8 ratio. The CD4:CD8 ratio has been used as a predictor of time to the development of AIDS. CD4 counts should also be determined using a CD3/CD4 combination reagent to eliminate monocyte contamination (CD3-CD4+dim). CD4+lymphocytes (expressed as an absolute number, a percentage of lymphocytes or as a ratio of CD4+ to CD8+ T lymphocytes) represents the best single predictor of the progression of AIDS. CD4+ lymphocyte numbers usually start to decline relatively soon after human immunodeficiency virus (HIV) infection. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Division of Clinical Laboratory Devices 510(k) Number Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter-Use ☐ (Optional Format 1-2-96) Ortho Diagnostic Systems Inc. Ortho-mune™OKT™4A (CD4) FITC Conjugate Ref. No. K951459 Additional Information - Submitted April, 1997 Page 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...