Nasal Aspirator (TLC-N02); Electric NoseFrida Pro (12300400); Electric NoseFrida (12100401)

K262822 · Zhongshan GL Electronics Co. , Ltd. · BTA · Oct 7, 2026 · General, Plastic Surgery

Device Facts

Record IDK262822
Device NameNasal Aspirator (TLC-N02); Electric NoseFrida Pro (12300400); Electric NoseFrida (12100401)
ApplicantZhongshan GL Electronics Co. , Ltd.
Product CodeBTA · General, Plastic Surgery
Decision DateOct 7, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4780
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The nasal aspirator is intended for intermittent removal of nasal secretions and mucus in children aged 2-12 years old in a home environment.

Device Story

Lightweight, portable, battery-operated nasal aspirator; consists of main unit and suction portion. Motor pump generates negative pressure to remove nasal secretions/mucus; powered by rechargeable battery. Used in home environment by lay persons. User interface includes buttons and display for vacuum pressure selection. Device provided non-sterile; not intended for user sterilization. Benefits patient by facilitating clearance of nasal passages.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), and environmental testing for home healthcare (IEC 60601-1-11).

Technological Characteristics

Battery-operated powered suction pump. Materials evaluated for mucosal membrane contact (<24h). Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11. Features user-selectable vacuum pressure via buttons and display. Non-sterile.

Indications for Use

Indicated for intermittent removal of nasal secretions and mucus in children aged 2-12 years old in a home environment.

Regulatory Classification

Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

Predicate Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION October 7, 2026 Zhongshan GL Electronics Co., Ltd. Cassie Gao Marketing Responsible 3-5th Floor, Wanhua Bldg., #20, Dehe Rd., Jidongyi, Xiaolan Town Zhongshan, Guangdong 528416 China Re: K262822 Trade/Device Name: Nasal Aspirator (TLC-N02); Electric NoseFrida Pro (12300400); Electric NoseFrida (12100401) Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: August 8, 2026 Received: August 10, 2026 Dear Cassie Gao: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K262822 - Cassie Gao Page 2 may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {3} K262822 - Cassie Gao Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JOYCE C. LIN -S for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262822 | ? | | Please provide the device trade name(s). | | ? | | Nasal Aspirator (TLC-N02); Electric NoseFrida Pro (12300400); Electric NoseFrida (12100401) | | | | Please provide your Indications for Use below. | | ? | | The nasal aspirator is intended for intermittent removal of nasal secretions and mucus in children aged 2-12 years old in a home environment. | | | | Please select the types of uses. | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | | Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☐ Adults (22 years old and greater) | ? | {5} 510(k) #: # 510(k) Summary Prepared on: 2026-08-08 Contact Details 21 CFR 807.92(a)(1) | Applicant Name | ZHONGSHAN GL ELECTRONICS CO.,LTD | | --- | --- | | Applicant Address | 3-5th Floor, Wanhua Building, No. 20, Dehe Road, Jidongyi, Xiaolan Town Zhongshan Guangdong 528416 China | | Applicant Contact Telephone | 086 13824723147 | | Applicant Contact | Ms. Cassie Gao | | Applicant Contact Email | cassie@zstlcelectronics.com | Device Name 21 CFR 807.92(a)(2) | Device Trade Name | Nasal Aspirator (TLC-N02); Electric NoseFrida Pro (12300400); Electric NoseFrida (12100401) | | --- | --- | | Common Name | Powered suction pump | | Classification Name | Pump, Portable, Aspiration (Manual Or Powered) | | Regulation Number | 878.4780 | | Product Code(s) | BTA, N/A | Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | --- | --- | --- | | K241202 | HTD2601US Electric nasal aspirator | BTA | | K222547 | BC026 Electric nasal aspirator | BTA | Device Description Summary 21 CFR 807.92(a)(4) The Nasal Aspirator is a lightweight, portable and battery operated device, which consists of main unit, and suction portion working together as one unit. The device is intended for suction of nasal passages for children aged 2-12 years old. This device is intended for use in a home environment by lay person.. The motor pump provides a negative pressure which removes nasal secretions. The motor pump operates on a rechargeable battery. The rechargeable battery can be charged from the external power adapter (not included in this device) through the provided charging line. The user interface consists of buttons and display, and the user can choose the vacuum pressure through the button. The product is provided non-sterile, and not to be sterilized by the user prior to use. Intended Use/Indications for Use 21 CFR 807.92(a)(5) The nasal aspirator is intended for intermittent removal of nasal secretions and mucus in children aged 2-12 years old in a home environment. Indications for Use Comparison 21 CFR 807.92(a)(5) The subject device shares the same indications for use with the predicate device. {6} # Technological Comparison 21 CFR 807.92(a)(6) The subject and predicate devices have same design principle, but different design features and performance specifications. The different technological characteristics between the subject and predicate devices will not raise different questions of safety or effectiveness as demonstrated in the non-clinical evidence. The reference device has a similar flow rate. # Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) The following non-clinical testing data was provided in support of the substantial equivalence determination. Biocompatibility testing The biocompatibility evaluation was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The worst case of the whole system is considered surface-contacting/ mucosal membrane for duration of less than 24 hours. And the testing included the following tests, results of which demonstrate the biocompatibility of the subject device: - Cytotoxicity - Skin Sensitization - Skin Irritation Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted, and the results show that the subject device complies with the following international standards: - IEC 60601-1 or ANSI AAMI ES60601-1 - IEC 60601-1-2 - IEC 60601-1-11 Bench Testing The specifications of each model was verified and a comparison testing with the predicate was conducted to verify the suction level and flowback protect capacity, and the results show that the subject device work as defined. Also the usability is evaluated per IEC 60601-1-6. Usability Testing Usability testing was conducted and the results show that the subject device complies with the IEC 60601-1-6:2010+A1:2013+A2:2020 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability. Home-used medical equipment requirements and environmental test: Environmental testing was conducted and the test results show that the subject device complies with the IEC 60601-1-11:2015 +A1:2020 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. Software Verification and Validation Testing Software documentation including verification & validation was provided in accordance with FDA Guidance: Content of Premarket Submissions for Device Software Functions. Clinical study: N/A This nonclinical test demonstrates that the subject device is substantially equivalent compared to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...