K262315 · Monteris Medical · ONO · Sep 3, 2026 · General, Plastic Surgery
Device Facts
Record ID
K262315
Device Name
NeuroBlate System
Applicant
Monteris Medical
Product Code
ONO · General, Plastic Surgery
Decision Date
Sep 3, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Monteris Medical NeuroBlate System is a neurosurgical tool and is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures (for example, brain tumor, radiation necrosis, and epileptogenic foci as identified by noninvasive and invasive neurodiagnostic testing, including imaging) through interstitial irradiation or thermal therapy in pediatrics and adults with 1064 nm lasers. The intended patients are adults and pediatric patients from the age of 2 years and older.
Device Story
NeuroBlate System is an MRI-compatible laser surgical tool for neurosurgery. System components: gas-cooled laser delivery probes (SideFire/FullFire); robotic probe driver for positioning/stabilization; interface platform for MRI table attachment; system electronics rack; control workstation with M-Vision software suite. Used in clinical settings by neurosurgeons. Surgeon uses workstation to plan trajectories, monitor thermal therapy in near real-time via MRI, and manipulate probes. Device delivers 1064 nm laser energy to target intracranial soft tissue. Output is thermal ablation of tissue. Benefits include precision thermal therapy for brain structures. No physical or technological changes in this submission; labeling clarification only.
Clinical Evidence
No new clinical or bench testing was performed. The submission relies on previously provided in-vitro (bench) and in-vivo (animal) data from predicate submissions (K260976, K240877), as the proposed changes are limited to labeling clarifications and do not alter the device's physical or technological characteristics.
Technological Characteristics
MRI-compatible laser surgical system. Energy source: 1064 nm laser. Components: gas-cooled laser delivery probes, robotic probe driver, interface platform, control workstation. Software: M-Vision, M-Vision Pro, M-Vision Fusion, Fusion-S. Connectivity: interfaces to MRI and hardware subsystems. Sterilization: not specified (standard for surgical instruments).
Indications for Use
Indicated for ablation, necrosis, or coagulation of intracranial soft tissue (e.g., brain tumors, radiation necrosis, epileptogenic foci) via interstitial 1064 nm laser thermal therapy. Patient population includes adults and pediatric patients aged 2 years and older.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 3, 2026
Monteris Medical
Amber Daly
Senior Quality Manager
131 Cheshire Ln., Suite 100
Minnetonka, Minnesota 55305
Re: K262315
Trade/Device Name: NeuroBlate System
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: ONO
Dated: July 8, 2026
Received: July 8, 2026
Dear Amber Daly:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K262315 - Amber Daly
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K262315 - Amber Daly
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# JAIME RABEN -S
for Julia Slocomb, PhD
Acting Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K262315
Please provide the device trade name(s).
NeuroBlate System
Please provide your Indications for Use below.
The Monteris Medical NeuroBlate System is a neurosurgical tool and is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures (for example, brain tumor, radiation necrosis, and epileptogenic foci as identified by noninvasive and invasive neurodiagnostic testing, including imaging) through interstitial irradiation or thermal therapy in pediatrics and adults with 1064 nm lasers.
The intended patients are adults and pediatric patients from the age of 2 years and older.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
Please select the age group(s) for which the device(s) is to be used.
☐ Neonates/Newborns (Birth to < 29 days old)
☐ Infants (29 days old to < 2 years old)
☑ Children (2 years old to < 12 years old)
☑ Adolescents (12 years old to < 22 years old)
☑ Adults (22 years old and greater)
?
?
{5}
MONTERIS MEDICAL
# 510(k) Summary: K262315
a. Device Information:
| Category | Comments |
| --- | --- |
| Sponsor / Submitter: | Monteris Medical 131 Cheshire Lane, suite 100 Minnetonka, MN 55305 866-799-7655 www.monteris.com |
| Correspondent Contact Information: | Amber Daly Senior Quality Manager Monteris Medical TEL: 763-647-8799 Email: adaly@monteris.com |
| Device Common Name: | Magnetic Resonance Image Guided Laser Thermal Therapy System |
| Device Classification Regulation & Name: | 21 CFR 878.4810 • Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Device Classification & Product Code: | Class II, ONO |
| Device Proprietary Name: | NeuroBlate® System |
Predicate Device Information:
| Manufacturer | Monteris Medical | REFERENCE PREDICATE: CLS |
| --- | --- | --- |
| Commercial Name | NeuroBlate System | TRANBERG®|Thermoguide Therapy System |
| Common Name: | Laser surgical instrument for use in general and plastic surgery and in dermatology; Stereotaxic instrument | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Premarket Notification # | K260976, K240877 | K214125 |
| Regulation | 21 CFR 878.4810 21 CFR 882.4560 | 21CFR878.4810 |
| Class/ Product Code | Class II; ONO, GEX, HAW | Class II; ONO, |
# b. Date Summary Prepared
07Jul2026
# c. Description of Device
The Monteris NeuroBlate System is a collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy in the practice of neurosurgery.
The NeuroBlate System components consist of:
Families of gas-cooled Laser Delivery Probe (Probe) (SideFire & FullFire) to deliver controlled energy to a target zone.
1
{6}
MONTERIS MEDICAL
Robotic Probe Driver which allow the surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
An Interface Platform, which attaches to the MRI system patient table and provides supporting electronics for the Robotic Probe Driver and interconnections for the Laser Delivery Probes;
A System Electronics Rack and Components, which includes necessary umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation,
A Control Workstation including the M-Vision™, M-Vision Pro™, M-Vision Fusion™, Fusion-S™ software, which includes a user interface for procedure planning, interactive monitoring of NeuroBlate procedures, and interfaces to the MRI and hardware subsystems.
The NeuroBlate System is utilized with stereotaxic frames and patient stabilization systems, such as:
The Monteris Mini-Bolt fixation components, and
The AtamA Stabilization System and MRI receive-only head coil, as well as, other optional accessories, including: drill bits, bolts, thumbscrews, instrument adaptors, accessory host adaptors, MRI trajectory wands, cranial screws, bone screws, fiducial markers, stereotactic manual driver with mandrel and T-handle, and other manual accessory instruments and tools.
### d. Indications for Use/ Intended Use
The Monteris Medical NeuroBlate System is a neurosurgical tool and is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures (for example, brain tumor, radiation necrosis, and epileptogenic foci as identified by noninvasive and invasive neurodiagnostic testing, including imaging) through interstitial irradiation or thermal therapy in pediatrics and adults with 1064 nm lasers.
The intended patients are adults and pediatric patients from the age of 2 years and older.
### e. Comparison to Predicate Device
The application for the Monteris Medical NeuroBlate System with the modified labeling is substantially equivalent to the predicate Monteris NeuroBlate System (K260976, K240877) and the other described predicates in intended use, technology, design and physician use. Reference Table 1.
This submission proposes no (zero) physical changes to the NeuroBlate System hardware, software, accessories, operating principles or functionality.
The proposed indication clarifications present no (zero) physical or technological changes to the predicate device. The only change proposed in this submission is the indication clarification as described in the NeuroBlate Instructions for Use (IFU), e.g., there are no changes to the NeuroBlate System's software, hardware, materials, laser probes, sterilization, packaging, technology or maximal performance, fundamental technology, technical modes of action and technical principles, or the composition, source, and use of patient contacting materials.
2
{7}
MONTERIS MEDICAL
Additionally, there are no material, or technology proposed changes to the NeuroBlate System. The technical modes of action and technical principles remain the same as the predicate NeuroBlate devices. Considering the proposed changes are labeling related, the previously provided in-vitro (bench) data remains applicable and no additional bench testing was required.
The fundamental functionality and technical characteristics of the proposed modified device is identical to the existing NeuroBlate System (K260976, K240877), as well as being substantially equivalent to the other described reference device. Thus, the proposed modified device is as safe as the described predicate devices.
### f. Summary of Supporting Data
This submission proposes no physical changes, materials changes, or technology changes to the NeuroBlate System. Considering the proposed changes are labeling, the previously provided in-vitro (bench), in-vivo (animal) data remains applicable and no additional testing was required.
Multiple FDA Guidance documents were utilized to ensure substantial equivalency, e.g.:
- United States Food and Drug Administration. Deciding When to Submit a 510(k) for a Change to an Existing Device. Washington, D.C.: FDA, 2017.
- United States Food and Drug Administration, Guidance for Industry and Food and Drug Administration Staff, General/Specific Intended Use. Washington, D.C.: FDA, 1998
- United States Food and Drug Administration. Premarket Assessment of Pediatric Medical Devices Guidance, Guidance for Industry and Food and Drug Administration Staff. March 24th edition. Washington, D.C. 2014
The proposed modifications to the NeuroBlate System's Indications For Use (IFU) are to provide clarification for the existing terminology utilized within the existing indications as the device is currently utilized. The proposed changes are not intended to provide specific indications for use which are different than the current language.
### g. Risk Analysis Summary
Monteris Medical's Quality Systems utilize various Risk Assessment and Risk Mitigation methodologies. As there are no physical changes, materials changes, or technology changes to the NeuroBlate System, and that the technical modes of action and technical principles remain the same as the predicate devices, and that the proposed changes are labeling related, the previously provided (K260976, K240877) Risk Assessment and Risk Mitigation documentation remains applicable.
3
{8}
MONTERIS MEDICAL
Table 1: Substantial Equivalency Indications For Use Comparison
| UNDER REVIEW: Proposed modified NeuroBlate System | PREDICATE: Monteris NeuroBlate System K240877 | REFERENCE PREDICATE: Visualase V2™ MRI Guided Laser Ablation K250307 |
| --- | --- | --- |
| • Note: Underlined text indicates modified text when compared to K240877 Indications for Use | • Note: Underlined text used as reference for comparison to proposed clarified | • Note: Underlined text used as reference for comparison to proposed clarified |
| PRIMARY FDA ProCode: ONO | PRIMARY FDA ProCode ONO | PRIMARY FDA ProCode ONO |
| SUBSEQUENT FDA ProCode: None | SUBSEQUENT FDA ProCode: GEX, HAW | SUBSEQUENT FDA ProCode: None |
| The Monteris Medical NeuroBlate System is a neurosurgical tool and is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures (for example, brain tumor, radiation necrosis, and epileptogenic foci as identified by noninvasive and invasive neurodiagnostic testing, including imaging) through interstitial irradiation or thermal therapy in pediatrics and adults with 1064 nm lasers. | The Monteris Medical NeuroBlate System is a neurosurgical tool and is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue, including brain structures (e.g., brain tumor, radiation necrosis, and epileptogenic foci as identified by non-invasive and invasive neurodiagnostic testing, including imaging), through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. | The Visualase V2™ MRI Guided Laser Ablation System is a neurosurgical tool and is indicated for use to ablate, necrotize, or coagulate intracranial soft tissue including brain structures (for example, brain tumor, radiation necrosis, and epileptic foci as identified by noninvasive and invasive neurodiagnostic testing, including imaging) through interstitial irradiation or thermal therapy in pediatrics and adults with 980 nm lasers. |
| Not Included | The Monteris Medical NeuroBlate System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI-based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate Laser Delivery Probes. It also provides near real-time thermographic analysis of selected MRI images | Not included |
| The intended patients are adults and pediatric patients from the age of 2 years and older. | When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate System analysis. The intended patients are adults and pediatric from the age of 2 years and older. | The intended patients are adults and pediatric patients from the age of 2 years and older. |
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.