K214125 · Clinical Laserthermia Systems, AB · ONO · Sep 22, 2022 · General, Plastic Surgery
Device Facts
Record ID
K214125
Device Name
TRANBERG Thermoguide Therapy System
Applicant
Clinical Laserthermia Systems, AB
Product Code
ONO · General, Plastic Surgery
Decision Date
Sep 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TRANBERG®|Thermoguide Therapy System is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy under magnetic resonance imaging (MRI) guidance in medicine and surgery in neurosurgery, for a wavelength of 1064nm. When therapy is performed under MRI guidance, and when data from compatible MRI sequences is available, the TRANBERG®|Thermoguide therapy system can process images using proton resonancefrequency (PRF) shift analysis and image subtraction to relate changes in complex phase angle back to relative changes in tissue temperature during therapy. The image data may be manipulated and viewed in a number of different ways, and the values of data at certain selected points may be monitored and/or displayed over time. The TRANBERG®|Thermoguide Therapy System is compatible with the following 3.0T MR scanner systems: Siemens MRI Magnetom and GE MRI Signa. When interpreted by a trained physician, this device provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of analysis using the TRANBERG®|Thermoguide Therapy System.
Device Story
System performs MRI-guided soft tissue laser ablation; consists of mobile laser unit (1064nm), laser applicator/introducer, and Thermoguide workstation. Laser energy transmitted via optical fiber to tissue; causes thermal necrosis/coagulation. Workstation processes MRI magnitude/phase maps using PRF shift analysis; calculates real-time temperature maps and thermal dose (CEM43). Used in MRI suite by trained medical professionals. Workstation provides GUI for workflow guidance, temperature monitoring, and safety interlocks. Safety features include real-time image update monitoring (min 5s), communication loss detection (1.5s), and RFID-based applicator validation. Physician uses output to assess thermal therapy progress; patient management decisions require clinical judgment beyond device analysis.
Clinical Evidence
No clinical data provided. Performance evaluated via bench testing (functional safety, IEC 60601-1-2, IEC 60601-2-22 compliance, software V&V, interoperability, thermometry algorithm accuracy) and a prospective GLP preclinical animal study evaluating MR thermometry and thermal damage estimation accuracy.
Indicated for necrotizing or coagulating soft tissue via interstitial irradiation or thermal therapy under MRI guidance in neurosurgery (1064nm wavelength). For use by trained physicians. Compatible with 3.0T Siemens Magnetom and GE Signa MRI scanners.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 22, 2022
Clinical Laserthermia Systems AB % John Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 13th Street NW Washington, District of Columbia 20004
Re: K214125
Trade/Device Name: TRANBERG| Thermoguide Therapy System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONO Dated: December 30, 2021 Received: December 30, 2021
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### -DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below
510(k) Number (if known)
#### K214125
Device Name
#### TRANBERG® | Thermoguide Therapy System
Indications for Use (Describe)
The TRANBERG®|Thermoguide Therapy System is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy under magnetic resonance imaging (MRI) guidance in medicine and surgery in neurosurgery, for a wavelength of 1064nm.
When therapy is performed under MRI guidance, and when data from compatible MRI sequences is available, the TRANBERG®|Thermoguide therapy system can process images using proton resonancefrequency (PRF) shift analysis and image subtraction to relate changes in complex phase angle back to relative changes in tissue temperature during therapy. The image data may be manipulated and viewed in a number of different ways, and the values of data at certain selected points may be monitored and/or displayed over time.
The TRANBERG®|Thermoguide Therapy System is compatible with the following 3.0T MR scanner systems: Siemens MRI Magnetom and GE MRI Signa. When interpreted by a trained physician, this device provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of analysis using the TRANBERG®|Thermoguide Therapy System.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) □ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff
> > PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
Clinical Laserthermia Systems' TRANBERG®|Thermoquide Therapy System
## Submitter
Clinical Laserthermia Systems, AB Scheelevagen 2 223 81 Lund, Sweden
Phone: +46-(0)70-590 11 40
Contact Person: Dan Mogren, CEO
Date Prepared: September 22, 2022
Name of Device: TRANBERG®|Thermoguide Therapy System
Common or Usual Name: TRANBERG®|Thermoguide Therapy System
Requlation Name: Laser surqical instrument for use in general and plastic surgery and in dermatology
Regulation: 21 CFR 878.4810
Product Code: ONO
Predicate Device
Medtronic Navigation Inc., Visualase Thermal Therapy System, K181859
### Reference Device
# Monteris Medical Inc., NeuroBlate System, K201056
### Device Description
The TRANBERG®|Thermoguide Therapy System is indicated for use in an MRI suite to perform soft tissue ablations under MRI guidance, it consists of three parts:
- 1) TRANBERG®|Mobile Laser Unit, cleared by K142216.
- 2) TRANBERG®|Laser applicator and introducer, cleared by K201466.
- 3) TRANBERG®|Thermoguide Workstation, article no. 1100-01, new in this submission.
The TRANBERG® Mobile Laser Unit includes a laser generator that operates at the wavelength of 1064nm, a continuous wave. The generated laser light is locally applied by means of a single use applicator kit (TRANBERG®)Laser applicator and introducer, cleared by K201466) through a minimally invasive surgical or percutaneous procedure. The energy from the laser generator is transmitted to tissue through the TRANBERG® Laser applicator and absorbed by the tissue surrounding the laser applicator, resulting in increased tissue temperature that necrotizes or coagulates soft tissue. The TRANBERG® Laser applicator is a 12m long optical fiber that allows the laser generator to be placed in the MRI control room. A workstation with software (TRANBERG®)Thermoguide Workstation) is used to extract temperature maps from magnetic resonance (MR) images and to calculate the thermal dose in treated tissue. Algorithms
{4}------------------------------------------------
used in the system to calculate temperature maps and thermal dose in tissue are well established and described in scientific literature.
The TRANBERG® Mobile Laser Unit has safety systems to prevent the use of a malfunctioning unit, including self-testing at startup and continuous monitoring of software and components that are critical for the unit and laser emission to function optimally. All laser safety requirements are met according to IEC 60601-2-22:2019.
The TRANBERG® Laser applicator utilizes an RFID tag which limits the maximum power and time (per applicator type) that can be used. It also ensures that an expired fiber, a reused fiber, or a fiber programmed for a different use cannot be used as a treatment fiber.
The TRANBERG®|Thermoguide Workstation has an interface for control of the TRANBERG®|Mobile Laser Unit output through the computer interface port of the laser control). It controls power and time settings on the laser unit, and it can start and stop the laser control and safety as per medical laser equipment requirements are managed by the TRANBERG®|Mobile Laser Unit.
Mandatory conditions must be satisfied to enable the laser unit and run a treatment. When one of more of these conditions are not met, the laser will not allow emitting laser radiation until all conditions are fulfilled:
- Real time images from the scanner are received at least every 5s. If Thermoguide Workstation ● detects update rates longer than 5s the laser emission is automatically interrupted.
- . Laser unit enabled and the connection is verified, any loss of communication within 1.5s between the Laser unit and Thermoquide workstation or data incoherency automatically stops the laser emission.
- . The use of the RFID tag is a mandatory condition to run a treatment and limits the maximum power and time (per applicator type).
- Laser Applicator type confirmed and received by TRANBERG®Thermoguide Workstation, the ● information is read on the RFID tag belonging to the fiber.
- Minimum 1 ROI (monitoring or guard) has been set.
- . Baseline temperature (e.g., core body temperature) is set and confirmed
- . B0 drift compensation: Reference baseline ROI is set and confirmed
- Placement of the High temperature guard ROI.
- . Test dose successful and confirmed
Safety guard functionality: TRANBERG®Thermoguide Workstation can be used to prescribe limits for the temperature at certain points (ROIs) in the image which can, in turn, be used to deactivate the laser if the limits are reached.
# Intended Use / Indications for Use
The TRANBERG® Thermoguide Therapy System is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy under magnetic resonance imaging (MRI) guidance in medicine and surgery in neurosurgery, for a wavelength of 1064nm.
When therapy is performed under MRI guidance, and when data from compatible MRI sequences is available, the TRANBERG®Thermoguide therapy system can process images using proton resonancefrequency (PRF) shift analysis and image subtraction to relate changes in complex phase angle back to relative changes in tissue temperature during therapy. The image data may be manipulated and viewed in a number of different ways, and the values of data at certain selected points may be monitored and/or displayed over time.
{5}------------------------------------------------
The TRANBERG® Thermoguide Therapy System is compatible with 3.0T MR scanner systems. When interpreted by a trained physician, this device provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of analysis using the TRANBERG®|Thermoguide Therapy System.
The Intended Use/Indications for Use has been limited as compared to the predicate device. Only a fixed wavelength of 1064nm (wavelength of the laser unit, part of TRANBERG®|Thermoguide Therapy System) and only neurosurgical interventions are part of the more limited Intended Use/Indications for Use. Also it is stated that only compatible scanners can be used with the system where the predicate names certain scanner types.
The differences are not critical to the intended therapeutic or surgical use of the device. Bench testing and animal testing according to the Intended Use/Indications for Use have been performed showing that the differences do not affect the safety and effectiveness of the device when used as labeled.
## Summary of Technological Characteristics
Soft tissue laser ablation and MRI thermometry is the technological principle for both the subject and predicate devices. The two systems are indicated for precise soft tissue laser ablation by medical professionals only and consist of a medical laser apparatus with single use sterile devices and software for extracting temperature maps from Magnetic Resonance (MR) images and calculation of thermal dose to the tissue. Generated laser light is locally applied within a tissue by means of a single use applicator. The energy within the laser light is absorbed by the tissue resulting in increased tissue temperature
At a high level, the subject and predicate devices are based on the following same technological elements:
- . Mobile laser unit using the same energy source.
- . Laser Applicator for transferring energy to tissue.
- . MR thermometry workstation that can calculate and display temperature and thermal damage maps.
- . Thresholds may be defined to deactivate laser power if the temperature in a target point is exceeded.
- . Primary user interface for supporting the workflow and controlling the treatment is performed in the MR thermometry workstation
The following technological differences exist between the subject and predicate devices:
- . The predicate operates in the optical wavelength range 980–1064 nm, whereas the subject device operates only at 1064 nm.
- . The predicate devices use a cooled fiber whereas the subject device uses a non-cooled fiber. However, the subject device fiber is already cleared by K201466.
- . The predicate device includes a cooling pump in its system, whereas no such pump is necessary for the subject device due to the use of a non-cooled fiber.
- . Different operating systems are used: Thermoguide uses Windows 10, Visualase Linux. But the setup of hardware and supporting SOUPs (e.g. drivers) are similar. The combination of hardware and software is supplied pre-installed by each manufacturer.
{6}------------------------------------------------
- . The predicate uses the Arrhenius model for calculation of thermal damage or thermal dose whereas the subject device uses the CEM43 algorithm. The CEM43 algorithm is used in the reference device. NeuroBlate System cleared under K201056.
A table comparing the key features of the subject and predicate devices is provided in Table 1 at the end of this 510(k) summary.
Labeling related to safety and mitigating actions of the TRANBERG Thermoguide System was compared to the predicate device, Visualase™ Therapy system and the reference device, NeuroBlate System. The minor differences in technical parameters between the subject device and predicate as well as reference device do not (1) raise different questions of safety and effectiveness; and (2) do not adversely affect the safety and effectiveness of the subject device compared to the predicate.
### Performance Data
### Bench Testing
The laser unit was designed and tested to comply with functional safety, and essential performance as well as laser safety requirements of IEC60601-1-2, IEC60601-2-22. Since the original clearance in K142216, the testing has been repeated due to changes in the hardware as well as to show compliance to updated versions of the IEC standards.
Biocompatibility data for the invasive devices is not included in this submission as it was part of the recent clearance under K201466. Full software V&V data is provided for both the TRANBERG®Mobile Laser Unit and the TRANBERG®|Thermoguide Workstation.
TRANBERG® Thermoguide Workstation used together with TRANBERG® Thermal Therapy System when used for treatments in MR was tested addressing:
- Intraoperability between the different devices of the TRANBERG® Thermoquide Therapy . System, i.e., Mobile laser unit, Thermoguide Workstation, Laser applicator, and MRI scanners.
- . Correct operation of the thermometry algorithm used in the Thermoquide Workstation as determined by correlation to physical measurements.
- . Evaluation of near real time behavior of temperature measurements. When using a commercially available scanner sequence, for the chosen scanner, determine offset from real time and update rate of temperature maps.
- . Verify that the products fulfil product requirement specifications.
### Pre-clinical Animal Study
The accuracy and performance of MR Thermometry and Thermal Damage Estimate were evaluated for the TRANBERG®Thermoguide Therapy System in a prospective preclinical animal study under GLP conditions.
No clinical data was provided in support of this submission.
In all instances, the TRANBERG® Thermoguide Therapy System functioned as intended and the performance observed was as expected.
### Conclusions
The TRANBERG®|Thermoguide Therapy System is as safe and effective as the Visualase Thermal Therapy System (K181859). The TRANBERG®)Thermoguide Therapy System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended therapeutic or surgical use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor
{7}------------------------------------------------
technological differences between the TRANBERG®|Thermoguide Therapy System and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the TRANBERG®Thermoguide Therapy System is as safe and effective as the Visualase Thermal Therapy System (K181859). Thus, the TRANBERG®|Thermoguide Therapy System is substantially equivalent.
{8}------------------------------------------------
# Table 1. Substantial equivalence of the TRANBERG®|Thermoguide Therapy System is claimed to
the predicate device Visualase® Thermal Therapy system, cleared under K181859.
| Parameter | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Image: TRANBERG Thermoguide Workstation | |
| | Image: TRANBERG Mobile Laser K142216, TRANBERG Laser applicator K201466, TRANBERG MR introducer K201466 FDA-cleared | |
| Product name | TRANBERG® Thermoguide Therapy System | Visualase™ Thermal Therapy system |
| Manufacturer | Clinical Laserthermia Systems CLS,<br>Sweden | Medtronic Navigation, Inc. |
| Intended use /<br>Indications for use | The TRANBERG® Thermoguide Therapy<br>System is indicated for use to necrotize or<br>coagulate soft tissue through interstitial<br>irradiation or thermal therapy under magnetic<br>resonance imaging (MRI) guidance in<br>medicine and surgery in neurosurgery, for a<br>wavelength of 1064nm.<br><br>When therapy is performed under MRI<br>guidance, and when data from compatible<br>MRI sequences is available, the<br>TRANBERG® Thermoguide therapy system<br>can process images using proton resonance<br>frequency (PRF) shift analysis and image<br>subtraction to relate changes in complex<br>phase angle back to relative changes in<br>tissue temperature during therapy. The image<br>data may be manipulated and viewed in a<br>number of different ways, and the values of<br>data at certain selected points may be<br>monitored and/or displayed over time.<br><br>The TRANBERG® Thermoguide Therapy<br>System is compatible with the following 3.0T<br>MR scanner systems: Siemens MRI<br>Magnetom and GE MRI Signa. When<br>interpreted by a trained physician, this | "The Visualase® Thermal Therapy System<br>is indicated for use to necrotize or coagulate<br>soft tissue through interstitial irradiation or<br>thermal therapy under magnetic resonance<br>imaging (MRI) guidance in medicine and<br>surgery in cardiovascular thoracic surgery<br>(excluding the heart and the vessels in the<br>pericardial sac), dermatology, ear-nose-<br>throat surgery, gastroenterology, general<br>surgery, gynecology, head and neck<br>surgery, neurosurgery, plastic surgery,<br>orthopedics, pulmonology, radiology, and<br>urology, for wavelengths 800nm through<br>1064nm.<br><br>When therapy is performed under MRI<br>guidance, and when data from compatible<br>MRI sequences is available, the Visualase®<br>system can process images using proton<br>resonance-frequency (PRF) shift analysis<br>and image subtraction to relate changes in<br>complex phase angle back to relative<br>changes in tissue temperature during<br>therapy. The image data may be<br>manipulated and viewed in a number of<br>different ways, and the values of data at |
| | device provides information that may be<br>useful in the determination or assessment of | certain selected points may be monitored<br>and/or displayed over time. |
| | thermal therapy.<br>Patient management decisions should not be<br>made solely on the basis of analysis using the<br>TRANBERG® Thermoguide Therapy System. | The Visualase® Thermal Therapy System is<br>compatible with General Electric Medical<br>Systems Signa model MR scanners and<br>with Siemens Medical Solutions Magnetom<br>Espree systems. When interpreted by a<br>trained physician, this device provides<br>information that may be useful in the<br>determination or assessment of thermal<br>therapy. Patient management decisions<br>should not be made solely on the basis of<br>Visualase® analysis." |
| Device Regulatory<br>Classification | 21 CFR 878.4810 (ΟΝΟ) | CFR892.2050 (LLZ)<br>CFR878.4810 (GEX)<br>CFR880.5725 (FRN) |
| Product code | ΟΝΟ (Neurosurgical Laser With MR<br>Thermography) | LLZ (Image Processing System)<br>GEX (Surgical Laser, Laser Applicator)<br>FRN (Infusion Pump) |
| Device class | 2 | 2 |
| 510(k) No | K214125 | K181859 |
| Diode laser generator | Cleared under K142216 | Cleared under K092197<br>Biotex, Inc.PHOTEX30 DIODE LASER |
| Laser wavelength | 1064 nm | 800 - 1064nm |
| Laser applicator /<br>handpiece | Cleared under K201466 | Cleared under K053087 |
| Thermometry<br>software system | New<br>TRANBERG® Thermoguide Workstation<br>software | Cleared under K063505<br>Visualase ENVISION Software system |
| Workstation OS | Windows | Linux |
| Software access | Installation of TRANBERG® Thermoguide<br>Workstation can only be performed by<br>qualified personnel authorized by CLS.<br>Installation controlled by License code.<br><br>Log into the system with login prompt and<br>password. | The Visualase® Thermal Therapy System<br>should only be assembled and installed by<br>a Medtronic Service Representative.<br><br>Log into the system with login prompt and<br>password. |
| GUI | Temperature and damage information is<br>displayed in near real-time during the<br>treatment<br><br>GUI with running Thermoguide software is the<br>primary user interface and control tool for the<br>entire TRANBERG® Thermoguide Therapy<br>System and provides a Workflow guidance for<br>the user. | Temperature and damage information is<br>displayed in near real-time during the<br>treatment<br><br>Visualase Console window is the primary<br>user interface and control tool for the entire<br>Visualase system. |
| Imaging modality | MR | MR |
| Connectivity | DICOM Import | DICOM Impo…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.