K181859 · Medtronic Navigation, Inc. · LLZ · Mar 6, 2019 · Radiology
Device Facts
Record ID
K181859
Device Name
Visualase Thermal Therapy System
Applicant
Medtronic Navigation, Inc.
Product Code
LLZ · Radiology
Decision Date
Mar 6, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Visualase™ Therapy System is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy under magnetic resonance imaging (MRI) guidance in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, plastic surgery, orthopedics, pulmonology, radiology, and urology, for wavelengths 800nm through 1064mm. When therapy is performed under MRI guidance, and when data from compatible MRI sequences is available, the Visualase™ system can process images using proton resonance-frequency (PRF) shift analysis and image subtraction to relate changes in complex phase angle back to relative changes in tissue temperature during therapy. The image data may be manipulated and viewed in a number of different ways, and the values of data at certain selected points may be monitored and/or displayed over time. The Visualase™ Thermal Therapy System is compatible with General Electric Medical Systems Signa model MR scanners and with Siemens Medical Solutions Magnetom Espree systems. When interpreted by a trained physician, this device provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of Visualase™ analysis.
Device Story
Visualase Thermal Therapy System performs MRI-guided laser ablation of soft tissue. System inputs include MRI image data from compatible scanners (GE Signa, Siemens Magnetom Espree). Core operation involves a diode laser (800-1064nm) and saline-cooled laser applicators. Software processes MRI data using proton resonance-frequency (PRF) shift analysis and image subtraction to calculate relative tissue temperature changes. Output is displayed on dual monitors via a graphical user interface, allowing physicians to monitor ablation progress and set temperature limits. Used in clinical settings by physicians; provides real-time thermal feedback to guide laser energy delivery. Benefits include precise, image-guided tissue destruction. Includes non-clinical remote presence software for Medtronic use only.
Clinical Evidence
No clinical data; bench testing only. Submission is for labeling changes only with no modifications to device hardware or software.
Technological Characteristics
System includes diode laser (800-1064nm), saline-cooled laser applicators (sterile, single-use), coolant pump, workstation, and dual monitors. Interfaces with MRI host computer. Software provides PRF shift analysis for thermal monitoring. Connectivity via peripheral interconnections to MRI scanners.
Indications for Use
Indicated for patients requiring soft tissue necrosis or coagulation via interstitial laser irradiation or thermal therapy under MRI guidance across various surgical specialties including neurosurgery, orthopedics, and general surgery.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
March 6, 2019
Medtronic Navigation Inc. c/o Sharon McDermott Principal Regulatory Specialist 826 Coal Creek Circle Louisville, Colorado 80027
Re: K181859
Trade/Device Name: Visualase Thermal Therapy System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 11, 2018 Received: July 12, 2018
Dear Ms. McDermott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Matthew C. Krueger -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181859
### Device Name Visualase Thermal Therapy System
### Indications for Use (Describe)
The Visualase™ Therapy System is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy under magnetic resonance imaging (MRI) guidance in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, plastic surgery, orthopedics, pulmonology, radiology, and urology, for wavelengths 800nm through 1064mm.
When therapy is performed under MRI guidance, and when data from compatible MRI sequences is available, the Visualase™ system can process images using proton resonance-frequency (PRF) shift analysis and image subtraction to relate changes in complex phase angle back to relative changes in tissue temperature during therapy. The image data may be manipulated and viewed in a number of different ways, and the values of data at certain selected points may be monitored and/or displayed over time.
The Visualase™ Thermal Therapy System is compatible with General Electric Medical Systems Signa model MR scanners and with Siemens Medical Solutions Magnetom Espree systems. When interpreted by a trained physician, this device provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of Visualase™ analysis.
Type of Use (*Select one or both, as applicable*)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary 5
| Submitter/Sponsor: | Medtronic Navigation, Inc.<br>826 Coal Creek Circle<br>Louisville, CO 80027 |
|--------------------|-----------------------------------------------------------------------------|
| Contact Person: | Sharon McDermott<br>Phone: (720) 890-3461 Fax: (720) 890-3500 |
| E-mail: | sharon.l.mcdermott@medtronic.com |
| | Rishi Sinha (Alternate) |
| E-mail: | Phone: (720) 890-2485 Fax: (720) 890-3500<br>rishi.k.sinha@medtronic.com |
This 510(k) submission notifies FDA of labeling changes to the Visualase Thermal Therapy System Manual to add new warnings and information. No design changes are the subject of this submission.
### Date Summary Prepared: March 5, 2019
| Device Trade Name: | Visualase™ Thermal Therapy System |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Common Name: | Magnetic Resonance Image-guided Laser Thermal Therapy System |
| Device Classification: | Class II |
| Product Code: | GEX (Surgical Laser, Laser Applicator)<br>LLZ (Image Processing System)<br>FRN (Infusion Pump) |
| Device Regulation: | 21CFR 878.4810 - Surgical Laser Instrument/Applicator<br>21CFR 892.2050 – Picture Archiving and Communications System<br>21CFR 880.5725 – Infusion Pump |
| Predicate Device: | Visualase™ Thermal Therapy System |
| Predicate Manufacturer: | Medtronic Navigation, Inc. |
| Predicate 510(k): | K081656 (Biotex, Inc) |
| Predicate Product Code: | GEX (Surgical Laser, Laser Applicator)<br>LLZ (Image Processing System)<br>FRN (Infusion Pump) |
| Predicate Regulation: | 21CFR 878.4810 – Surgical Laser Instrument/Applicator<br>21CFR 892.2050 - Picture Archiving & Communications System<br>21CFR 880.5725 – Infusion Pump |
{5}------------------------------------------------
| Indications for Use: | The Visualase™ Thermal Therapy System is indicated for use to necrotize or coagulate<br>soft tissue through interstitial irradiation or thermal therapy under magnetic resonance<br>imaging (MRI) guidance in medicine and surgery in cardiovascular thoracic surgery<br>(excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-<br>throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery,<br>neurosurgery, plastic surgery, orthopedics, pulmonology, radiology, and urology, for<br>wavelengths 800nm through 1064nm. |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | When therapy is performed under MRI guidance, and when data from compatible MRI<br>sequences is available, the Visualase™ system can process images using proton<br>resonance-frequency (PRF) shift analysis and image subtraction to relate changes in<br>complex phase angle back to relative changes in tissue temperature during therapy. The<br>image data may be manipulated and viewed in a number of different ways, and the<br>values of data at certain selected points may be monitored and/or displayed over time. |
| | The Visualase™ Thermal Therapy System is compatible with General Electric Medical<br>Systems Signa model MR scanners and with Siemens Medical Solutions Magnetom<br>Espree systems. When interpreted by a trained physician, this device provides<br>information that may be useful in the determination or assessment of thermal therapy.<br>Patient management decisions should not be made solely on the basis of Visualase™<br>analysis. |
| Device Description: | The Visualase Thermal Therapy System comprises of hardware and software<br>components used in combination with three MR-compatible (conditional), sterile,<br>single-use, saline-cooled laser applicators with proprietary diffusing tips that deliver<br>controlled energy to the tissue of interest. The system consists of:<br>a diode laser (energy source); a coolant pump to circulate saline through the laser application; Visualase workstation which interfaces with MRI scanner's host computer Visualase software which provides the system's ability to visualize and<br>monitor relative changes in tissue temperature during ablation procedures,<br>set temperature limits and control the laser output; two monitors to display<br>all system imaging and laser ablation via a graphical user interface and<br>peripherals for interconnections. |
- Remote Presence software provides a non-clinical utility application for use by ● Medtronic only and is not accessible by the user.
## Comparison of Technological Characteristics with Predicate Device
The changes that are the subject of this 510(k) submission are labeling updates that do not change the indications for use or impact the device's technology. Along with grammatical corrections, the labeling changes encompass the addition of warnings and notes that were provided via a Field Corrective Action distributed to current users on June 1, 2018 and August 21, 2018.
#### Performance Characteristics
No new performance testing was required for this change as there are no changes to the Visualase™ Thermal Therapy System as a result of this labeling change.
{6}------------------------------------------------
### Conclusion
The Visualase™ Thermal Therapy System is substantially equivalent to the predicate Visualase™ Therapy System. The labeling changes do not affect the intended use or fundamental technology. Further, the labeling modifications do not raise any new questions of safety and effectiveness of the Visualase Thermal Therapy System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.