K262089 · Hangzhou Tappa Medical Technology Co., Ltd. · DWJ · Aug 21, 2026 · Cardiovascular
Device Facts
Record ID
K262089
Device Name
Disposable Warming Blanket
Applicant
Hangzhou Tappa Medical Technology Co., Ltd.
Product Code
DWJ · Cardiovascular
Decision Date
Aug 21, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5900
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Disposable Warming Blanket is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.
Device Story
Disposable Warming Blanket; used in conjunction with Medical Warming Unit (K252861) to provide physical warming to patients. Device consists of non-woven fabric and PE film; functions by receiving thermal energy from the warming unit to regulate patient body temperature. Used in clinical settings; operated by healthcare professionals. Benefits include maintenance of normothermia or therapeutic warming for hypothermic patients. Output is physical heat transfer to the patient.
Clinical Evidence
No clinical data. Bench testing only, including performance testing (appearance, dimensions, welding strength), sterility, shelf-life (ASTM F1980), and biocompatibility (ISO 10993-1, -5, -10, -23). Combined use testing with the Medical Warming Unit was performed.
Technological Characteristics
Materials: Non-woven fabric and PE film. Dimensions: Various models ranging from 100x70cm to 210x100cm. Energy source: External (Medical Warming Unit). Sterilization: Available in sterile and non-sterile versions. Standards: ASTM F1886, F2096, F1929, F88, F1980; USP <71>; ISO 10993.
Indications for Use
Indicated for hypothermic or normothermic patients requiring induced hyperthermia or localized temperature increase. Applicable to all age groups: neonates, infants, children, adolescents, and adults.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
Hangzhou Tappa Medical Technology Co., Ltd.
Dan Cao
Rm. 2002, Bldg. 1, Chutian Technology Bldg., # 225 Chutian Rd., Xixing St., Binjiang District, Hangzhou, Zhejiang, 310000, China
Hangzhou,
China
Re: K262089
Trade/Device Name: Disposable Warming Blanket
Regulation Number: 21 CFR 870.5900
Regulation Name: Thermal Regulating System
Regulatory Class: Class II
Product Code: DWJ
Dated: June 21, 2026
Received: June 22, 2026
Dear Dan Cao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262089 - Dan Cao
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K262089 - Dan Cao
Page 3
Sincerely,
# KATHLEEN M.
GRUNDER -S
for Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support,
Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262089 | ? |
| Please provide the device trade name(s). | | ? |
| Disposable Warming Blanket | | |
| Please provide your Indications for Use below. | | ? |
| The Disposable Warming Blanket is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☑ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
K262089
# 510(k) Summary
## I. Submitter
Hangzhou Tappa Medical Technology Co., Ltd.
Room 2002, Building 1, Chutian Technology Building, No. 225 Chutian Road, Xixing
Street, Binjiang District, Hangzhou, Zhejiang, 310000, China
Contact person: Ms.Caodan
Tel.:+86-13484165183
E-mail: caod@tongpumed.com
Preparation date: June 11, 2026
## II. Proposed Device
Device Trade Name: Disposable Warming Blanket
Common name: system, thermal regulating
Regulation Number: 21 CFR 870.5900
Regulatory Class: Class II
Product code: DWJ
Review Panel: Cardiovascular
## III. Predicate Devices
510(k) Number: K190221
Trade name: IOB Warming Blankets
Common name: System, thermal regulating
Classification: Class II
Product Code: DWJ
Manufacturer IOB Medical Inc
Page 1 of 5
{5}
#### IV. Device description
The Disposable Warming Blanket is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.
Combined with Medical Warming Unit (K252861), the disposable warming blanket provides physical warming to the patient by controlling the temperature of the blanket, thereby assisting in regulating the patient's body temperature.
#### V. Indications for use
The Disposable Warming Blanket is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.
#### VI. Comparison of technological characteristics with the predicate devices
The comparison and discussion between the subject device and the predicate devices are listed in below table.
Table1 General Comparison between Disposable Warming Blanket and the predicate device
| Characteristics | Proposed device | Predicate device (K190221) | Discussion |
| --- | --- | --- | --- |
| Indications For Use | The Disposable Warming Blanket is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated. | The IOB Warming Blankets are indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated. | Same |
| Material Design | Consists of Non-woven fabric and PE film | Each blanket consists of two layers of non-woven polypropylene fabric coated with a layer of polyethylene. | Same |
| Shelf Life | 5 years | 3 years | Difference 1 |
| Sterility | Non-sterile and sterile | Non-sterile and sterile | Same |
| Blanket | WB-1101-R/WB-1101-N | IOB-024 240cm×150cm | Similar |
Page 2 of 5
{6}
| Dimensions (approximate) | 210×100cm WB-1102-R/WB-1102-N 190X70cm WB-1103-R/WB-1103-N 200X100cm WB-1104-R/WB-1104-N 150X100cm WB-1105-R/WB-1105-N 130X100cm WB-1106-R/WB-1106-N 130X100cm WB-1105-R/WB-1107-N 190X100cm WB-1109-R/WB-1109-N 210X100cm WB-1110-R/WB-1110-N 210X100cm WB-1111-R/WB-1111-N 210X100cm WB-1113-R/WB-1113-N 210X100cm WB-1114-R/WB-1114-N 210X100cm WB-1115-R/WB-1115-N 210X100cm WB-2101-R/WB-2101-N 160X100cm WB-2102-R/WB-2102-N 100X70cm WB-1201-R/WB-1201-N | IOB-025 120cm×80cm IOB-026 150cm×120cm IOB-027 220cm×120cm IOB-028 100cm×100cm IOB-029 200cm×100cm IOB-301 170cm×100cm IOB-302 170cm×100cm IOB-303 180cm×120cm IOB-304 182cm×120cm | |
| --- | --- | --- | --- |
Page 3 of 5
{7}
| | 210X100cm WB-1202-R/WB-1202-N 210X100cm WB-1203-R/WB-1203-N 180X100cm WB-1204-R/WB-1204-N 180X70cm WB-1205-R/WB-1205-N 210X100cm WB-1206-R/WB-1206-N 210X100cm WB-2201-R/WB-2201-N 100X90cm WB-2202-R/WB-2202-N 140X90cm | | |
| --- | --- | --- | --- |
# Analysis 1
The shelf life has been verified. This difference does not compromise safety or effectiveness.
# VII. Non-Clinical Testing
The device described in this summary was tested in accordance with applicable standards and internal performance specifications. The following tests were conducted, and the results met the acceptance criteria:
# Performance testing:
The following tests were conducted on the Disposable Warming Blanket:
Appearance
Dimensions
Sterility
Page 4 of 5
{8}
Welding Strength
In addition, combined use testing of the Medical Warming Unit and Disposable Warming Blanket was performed.
# **Reprocessing, Sterility, and Shelf-Life**
ASTM F1886/F1886M-16
ASTM F2096-11
ASTM F1929-23
ASTM F88/F88-23
USP<71>
ASTM F 1980
# **Biocompatibility testing**
Biocompatibility of the device was evaluated in accordance with the FDA guidance “Use of International Standard ISO 10993-1” The following testing was conducted:
- ISO 10993-1
- ISO 10993-5
- ISO 10993-10
- ISO 10993-23
# **VIII. Clinical Testing**
No clinical study is included in this submission.
# **IX. Conclusion**
The proposed device shares the same indications for use as the predicate device and has similar design features and technological characteristics. Performance testing data have demonstrated that the proposed device is as safe and as effective as the predicate device. Therefore, the proposed device is determined to be substantially equivalent to the predicate device.
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.