The Warming System is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.
Device Story
Forced air warming system comprising a Medical Warming Unit (TMS-W101) and disposable warming blankets; used in clinical settings to regulate patient body temperature or provide thermal comfort. Unit takes ambient air, heats it via internal heater, and delivers it through a hose to a disposable blanket; blanket distributes air via internal channels and perforations to the patient. Operated by clinicians via front-panel buttons; includes LCD display for temperature settings and fault codes. Safety features include audible/visual alarms and automatic heater shut-off if outlet temperature exceeds 50°C. Benefits include prevention/treatment of hypothermia and maintenance of normothermia during clinical procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including performance testing (temperature control, accuracy, protection, noise), electrical safety (IEC 60601-1, 60601-1-2, 60601-2-35), biocompatibility (ISO 10993 series), and sterility/packaging integrity (ASTM F1886, F2096, F1929, F88, USP <71>).
Technological Characteristics
Forced air warming unit with mobile trolley; heater power 1000VA. Disposable blankets made of non-woven fabric and PE film. Temperature settings: 32°C, 38°C, 43°C (±1°C). Features replaceable 0.2-micron air filter, LCD display, and hose locking mechanism. Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-8, 60601-2-35. Sterilization via unspecified method for sterile variants.
Indications for Use
Indicated for hypothermic patients or normothermic patients requiring induced hyperthermia or localized temperature increase. Suitable for adult and pediatric patients.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 7, 2026
Hangzhou Tappa Medical Technology Co., Ltd.
Na Ban
Registration Specialist
No. 225#, Chutian Road, Binjiang District, Hangzhou
Hangzhou,
China
Re: K252861
Trade/Device Name: Warming System
Regulation Number: 21 CFR 870.5900
Regulation Name: Thermal Regulating System
Regulatory Class: Class II
Product Code: DWJ
Dated: March 6, 2026
Received: March 6, 2026
Dear Na Ban:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252861 - Na Ban
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K252861 - Na Ban
Page 3
Sincerely,
**Meaghan Erlewein -S**
For Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252861 | |
| Device Name Warming System | |
| Indications for Use (Describe) The Warming System is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K252861
# 510(k) Summary
## I. Submitter
Hangzhou Tappa Medical Technology Co., Ltd.
No. 225#, Chutian Road, Binjiang District, Hangzhou, China
Contact person: Ms. Banna
Tel.: +86-13985728517
E-mail: bann@tongpumed.com
Preparation date: September 8th, 2025
## II. Proposed Device
Device Trade Name: Warming System
Common name: system, thermal regulating
Regulation Number: 21 CFR 870.5900
Regulatory Class: Class II
Product code: DWJ
Review Panel: Cardiovascular
## III. Predicate Devices
510(k) Number: K231596
Trade name: IOB Temperature Management System
Common name: System, thermal regulating
Classification: Class II
Product Code: DWJ
Manufacturer: IOB Medical Inc
Page 1 of 9
{5}
Page 2 of 9
## IV. General description
The Warming System is a forced air warming system which is comprise of a Medical Warming Unit and variety models of Disposable Warming Blankets. It is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.
## Medical Warming Unit (TMS - W101)
The Medical Warming Unit is composed of a Medical Warming Unit and a mobile trolley (optional accessory).
The Medical Warming Unit has the function of physically warming the human body in vitro, achieving the goal of assisting in regulating human body temperature. It can also be used to provide patients thermal comfort when the conditions exist that may cause patients to become too cold. The Medical Warming Unit are suitable for both adult and pediatric patients. Its application part is a Disposable Warming Blanket.
The operation panel of Medical Warming Unit fully considers human-computer interaction. It is located on the inclined surface of the device and is simply operated by buttons. The desired temperature can be selected by pressing different temperature buttons.
In an emergency, the alarm will sound and the screen will flash and display a fault code. In the standby interface, the total running time of the device can be checked by pressing the call mode button (hidden button, under the product model character) and the 43°C button.
## Disposable Warming Blanket (Sterile or Non Sterile)
The Warming System is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.
## V. Indications for use
The Warming System is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated.
## VI. Comparison of technological characteristics with the predicate devices
The comparison and discussion between the subject device and the predicate devices are listed in below table:
{6}
Table 1 General Comparison
| Characteristics | Proposed device | Predicate device (K231596) | Discussion |
| --- | --- | --- | --- |
| Trade name | Warming System | IOB Temperature Management System | / |
| Classification Name | system, thermal regulating | system, thermal regulating | Same |
| Product Code | DWJ | DWJ | Same |
| Regulation Number | 21 CFR 870.5900 | 21 CFR 870.5900 | Same |
| Indications for use | The Warming System is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated. | The IOB Temperature Management System Model IOB-507 is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated. | Same |
| Temperature Settings | Average temperature at the end of the gas pipe: High temperature:43°C±1°C Medium temperature:38°C±1°C Low temperature:32°C±1°C | 43°C+/-2°C 38°C+/-2°C 32°C+/-2°C Ambient | Same |
| System Power | 100VAC 50Hz/60Hz 12A 110VAC 50Hz/60Hz 13.5A 120VAC 50Hz/60Hz 14A 127VAC 50Hz/60Hz 15A | 110-120 V, 60 Hz, 14 A | Similar |
| Heater Power | 1000VA(Nominal value) | 950 W | Similar |
| Dimensions | 33cm Length *27.5cm width *33cm height | 29.5 x 22 x 22 cm | Similar |
{7}
Page 4 of 9
| Weight | 6.5kg | 5.4 kg | Similar |
| --- | --- | --- | --- |
| EMI/EMC Compliant | IEC 60601-1, IEC 60601-1-2 | IEC 60601-1, IEC 60601-1-2 | Same |
| Forced Air Over Temperature | When the temperature of the outlet of the air pipe exceeds 50°C, the sound and light alarm will be turned on and the heating will stop. | Auto-shuts heater off at 47°C+/-2°C | Similar |
| Hose with Secure Locking | Yes | Yes | Same |
| Air Filter | Replaceable 0.2 micron | Replaceable 0.2 micron | Same |
| Temperature Display | Front panel LCD display | Front panel LCD display | Same |
Analysis 1 System Power
Clinical warming performance remains consistent across all supported voltages. This difference does not impact safety or effectiveness.
Analysis 2 Heater Power
The device meet EC 60601-1 requirements, this minor difference does not affect safety or effectiveness.
Analysis 3 Forced Air Over Temperature
Testing demonstrates equivalent patient protection. This difference does not compromise safety or effectiveness.
Table 2 General Comparison between Disposable Warming Blanket and the predicate device
| Characteristics | Proposed device | Predicate device (K231596) | Discussion |
| --- | --- | --- | --- |
| Indications For Use | The Warming System is indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized increase in temperature is clinically indicated. | The IOB Warming Blankets are indicated for hypothermic patients or normothermic patients for whom induced hyperthermia or localized | Same |
{8}
| | | increase in temperature is clinically indicated. | |
| --- | --- | --- | --- |
| Material Design | Consists of Non-woven fabric and PE film | Consists of two layers of nonwoven polypropylene fabric bonded to a fusion layer of polyethylene.
The layers are bonded together to form a distribution network of air delivery channels.
The warm air is distributed around the patient's body through the delivery channels and exits the blanket through a specially designed series of perforations in the patient side layer of the blanket.
The distribution of air is designed to minimize temperature differences of delivered air at different blanket locations. | Same |
| Shelf Life | 5 years | 3 years | Difference 1 |
| Sterility | Non-sterile and sterile | Non-sterile and sterile | Same |
| Blanket Dimensions (approximate) | WB-1101-R/WB-1101-N
210×100cm
WB-1102-R/WB-1102-N
190X70cm
WB-1103-R/WB-1103-N
200X100cm | IOB-001/IOB-001S
142×120cm
IOB-002/IOB-002S
142×120cm
IOB-003/IOB-003S 202×64cm
IOB-004/IOB-004S
210×120cm | Similar |
Page 5 of 9
{9}
| WB-1104-R/WB-1104-N | IOB-005/IOB-005S 100×80cm |
| --- | --- |
| 150X100cm | IOB-006/IOB-006S |
| WB-1105-R/WB-1105-N | 215×100cm |
| 130X100cm | IOB-007/IOB-007S |
| WB-1106-R/WB-1106-N | 170×100cm |
| 130X100cm | IOB-008/IOB-008S |
| WB-1107-R/WB-1107-N | 210×120cm |
| 190X100cm | IOB-009/IOB-009S 160×80cm |
| WB-1109-R/WB-1109-N | IOB-010/IOB-010S |
| WB-1110-R/WB-1110-N | 215×100cm |
| WB-1111-R/WB-1111-N | IOB-011/IOB-011S |
| WB-1112-R/WB-1112-N | 200×100cm |
| WB-1113-R/WB-1113-N | IOB-012/IOB-012S |
| WB-1114-R/WB-1114-N | 142×100cm |
| WB-1115-R/WB-1115-N | IOB-014/IOB-014S 122×64cm |
| WB-1116-R/WB-1116-N | IOB-015/IOB-015S 192×74cm |
| WB-1117-R/WB-1117-N | IOB-016/IOB-016S |
| WB-1118-R/WB-1118-N | 210×120cm |
| WB-1119-R/WB-1119-N | IOB-017/IOB-017S |
| WB-1120-R/WB-1120-N | 215×100cm |
| WB-1121-R/WB-1121-N | IOB-018/IOB-018S |
| WB-1122-R/WB-1122-N | 210×120cm |
| WB-1123-R/WB-1123-N | IOB-019/IOB-019S |
| WB-1124-R/WB-1124-N | 109×102cm |
| WB-1125-R/WB-1125-N | IOB-020/IOB-020S |
| WB-1126-R/WB-1126-N | 142×120cm |
| WB-1127-R/WB-1127-N | IOB-021/IOB-021S |
| WB-1128-R/WB-1128-N | 230×100cm |
| WB-1129-R/WB-1129-N | IOB-022/IOB-022S |
| WB-1130-R/WB-1130-N | 210×120cm |
| WB-1131-R/WB-1131-N | IOB-023/IOB-023S 162×74cm |
| WB-1132-R/WB-1132-N | IOB-024/IOB-024S |
| WB-1133-R/WB-1133-N | 180×100cm |
Page 6 of 9
{10}
Page 7 of 9
| | WB-1204-R/WB-1204-N
180X70cm
WB-1205-R/WB-1205-N
210X100cm
WB-1206-R/WB-1206-N
210X100cm
WB-2201-R/WB-2201-N
100X90cm
WB-2202-R/WB-2202-N
140X90cm | IOB-025/IOB-025S 120×80cm
IOB-026/IOB-026S 120×80cm
IOB-027/IOB-027S
220×120cm
IOB-028/IOB-028S
100×100cm
IOB-029/IOB-029S
200×100cm
IOB-030/IOB-030S 198×80cm
IOB-034/IOB-034S 140×64cm
IOB-301/IOB-301S
230×100cm
IOB-302/IOB-302S
280×100cm
IOB-303/IOB-303S
330×120cm
IOB-304/IOB-304S
384×120cm |
| --- | --- | --- |
Analysis 1
The shelf life has been verified. This difference does not compromise safety or effectiveness.
VII. Non-Clinical Testing
The device described in this summary, was tested and demonstrated to be in conformance with the following standards:
Performance testing:
The following tests were conducted on the Medical Warming Unit:
Appearance
Output temperature control
Temperature control accuracy
Nozzle temperature protection
Temperature protection of air duct compartment
Output pipeline length
Noise
{11}
Time to reach the set temperature
Safety
Software function
Labels and accompanying materials
The following tests were conducted on the Disposable Warming Blanket:
Appearance
Dimensions
Sterility
Welding Strength
In addition, combined use testing of the Medical Warming Unit and Disposable Warming Blanket was performed.
## Reprocessing, Sterility, and Shelf-Life
ASTM F1886/F1886M-16
ASTM F2096-11
ASTM F1929-23
ASTM F88/F88-23
USP<71>
ASTM F 1980
## Biocompatibility testing
Biocompatibility of the device was evaluated in accordance with the FDA guidance “Use of International Standard ISO 10993-1” The following testing was conducted:
- ISO 10993-1
- ISO 10993-5
- ISO 10993-10
- ISO 10993-23
## Software
Guidance for Industry and Food and Drug Administration Staff-Content of Premarket Submissions for Device Software Functions.
## Electrical Safety and EMC
- IEC 60601-1
- IEC 60601-1-2
- IEC 60601-4-2
Page 8 of 9
{12}
Page 9 of 9
- IEC 60601-1-6
- IEC 60601-1-8
- IEC 60601-2-35
- IEC 61000-4-3
## VIII. Clinical Testing
No clinical study is included in this submission.
## IX. Conclusion
The proposed device has the same indication for use and has similar design features and technological characteristic as the predicate device. Performance testing data demonstrates that the proposed device is as safe and as effective as the predicated device. Accordingly, the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.