Kiba CTO Crossing Catheter

K262013 · Reflow Medical, Inc. · PDU · Aug 28, 2026 · Cardiovascular

Device Facts

Record IDK262013
Device NameKiba CTO Crossing Catheter
ApplicantReflow Medical, Inc.
Product CodePDU · Cardiovascular
Decision DateAug 28, 2026
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The Kiba CTO Crossing Catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices, including facilitation of the intraluminal placement of diagnostic/ interventional devices beyond stenotic lesions (including chronic total occlusions [CTOs]) and provide a conduit for delivery of saline solutions or diagnostic/therapeutic agents. The Kiba CTO Crossing Catheter is contraindicated for use in the coronary and cerebral vasculature.

Device Story

Support catheter for peripheral vasculature; used with steerable guidewires to navigate stenotic lesions or chronic total occlusions (CTOs). Device features laser-cut stainless steel hypotube with radiopaque gold-plated beveled tip; PET heat-shrink outer jacket; luer hub handle. Operated by physician in clinical setting; provides conduit for guidewire exchange, saline, or diagnostic/therapeutic agents. Enhances guidewire support to facilitate crossing of complex lesions. Bench testing confirms performance equivalence to predicate.

Clinical Evidence

No clinical data submitted. Substantial equivalence supported by bench testing including simulated use, dimensional verification, radiopacity, leak, kink resistance, corrosion, component/bond integrity, particulate, torque, burst, lubricity/coating integrity, design validation, sterility, and biocompatibility assessment.

Technological Characteristics

304 stainless steel hypotube; gold-plated beveled tip; PET heat-shrink jacket; Lexin resin luer hub; Dynaflex G7960-1001 strain relief. Hydrophilic coating (40cm length). Sizes: 14, 18, 35. Sterilization: Ethylene Oxide (EO). Connectivity: None (mechanical device).

Indications for Use

Indicated for patients requiring peripheral vascular access to facilitate guidewire placement, exchange, and delivery of diagnostic/interventional devices beyond stenotic lesions or chronic total occlusions (CTOs). Contraindicated for coronary and cerebral vasculature.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0} **U.S. FOOD & DRUG** ADMINISTRATION August 28, 2026 Reflow Medical, Inc. Angela Lamprey Director, Regulatory Affairs 208 Avenida Fabricante Suite 100 San Clemente, California 92688 Re: K262013 Trade/Device Name: Kiba CTO Crossing Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: PDU Dated: June 15, 2026 Received: June 15, 2026 Dear Angela Lamprey: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K262013 - Angela Lamprey Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K262013 - Angela Lamprey Page 3 Sincerely, LYDIA S. Digitally signed by LYDIA S. GLAW -S GLAW -S Date: 2026.08.28 14:00:00 -04'00' Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # **Indications for Use** Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K262013 Device Name Kiba CTO Crossing Catheter Indications for Use (Describe) The Kiba CTO Crossing Catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices, including facilitation of the intraluminal placement of diagnostic/ interventional devices beyond stenotic lesions (including chronic total occlusions [CTOs]) and provide a conduit for delivery of saline solutions or diagnostic/therapeutic agents. The Kiba CTO Crossing Catheter is contraindicated for use in the coronary and cerebral vasculature. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) **CONTINUE ON A SEPARATE PAGE IF NEEDED.** This section applies only to requirements of the Paperwork Reduction Act of 1995. **\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {4} K262013 - Special 510(k) Page 1 of 2 SPECIAL 510(K) SUMMARY | Submitter | Reflow Medical, Inc.208 Avenida Fabricante, Suite 100San Clemente, CA 92672Contact person: Angela LampreyEmail: alamprey@reflowmedical.comPhone: (949)-481-0399 | | | --- | --- | --- | | Date Prepared | June 10, 2026 | | | Device | Name of the Device: Kiba CTO Crossing CatheterCommon Name: Percutaneous CatheterClassification Name: Percutaneous CatheterRegulatory Class: 2Product Code: PDU, DQY | | | Legally Marketed Device to which your firm is claiming equivalence | Predicate: K193596Device Name: Wingman Crossing CatheterNo reference devices were used in this submission. | | | Description of the device | The Kiba CTO Crossing Catheter is a device intended to provide additional support to a steerable guidewire when accessing discrete regions of the peripheral vasculature.The device consists of a support catheter with a radiopaque beveled guide-tip and a handle. | | | Intended use of the device | The Kiba CTO Crossing Catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices, including facilitation of the intraluminal placement of diagnostic/ interventional devices beyond stenotic lesions (including chronic total occlusions [CTOs]) and provide a conduit for delivery of saline solutions or diagnostic/therapeutic agents.The Kiba CTO Crossing Catheter is contraindicated for use in the coronary and cerebral vasculature. | | | Summary of the technological characteristics of your device compared to the predicate device | | | | The technological characteristics of the subject Kiba CTO Crossing Catheter are similar to the technological characteristics of the Wingman Crossing Catheter previously cleared under K193596.At a high level, the subject and predicate devices are based on the following same technological elements:Lasercut stainless steel hypotube with a beveled tipGold plated tip for radiopacity14, 18, and 35 sizesHydrophilic coatingThe following technological differences exist between the subject and predicate devices:Replaced outer polymer jacket with a PET heat-shrink jacketModify the handle from a nose and plunger to a luer hub | | | | | Predicate – (Wingman Crossing Catheter, K193596) | Subject – Kiba CTO Crossing Catheter | | Intended Use | The Wingman Crossing Catheters (14/14C/18/35) are intended to be used in conjunction with steerable guidewires to access discrete regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices, including | Same | Confidential Special 510(k) Summary Page 1 of 2 {5} K262013 - Special 510(k) Page 2 of 2 | | facilitation of the intraluminal placement of diagnostic/ interventional devices beyond stenotic lesions (including chronic total occlusions [CTOs]) and provide a conduit for delivery of saline solutions or diagnostic/therapeutic agents. | | | --- | --- | --- | | Guidewire Compatibility | 0.014", 0.018", 0.035" | 0.014", 0.018", 0.035" | | Sheath Compatibility | WM 14C: 3Fr minimum WM 18: 4Fr minimum WM 35: 5Fr minimum | Kiba 14/18/35 4Fr minimum | | Catheter Length | WM 14C: 90cm, 135cm, 150cm WM 18: 90cm, 135cm, 150cm WM 35: 65cm, 90cm, 135cm | Kiba 14: 90cm, 105cm, 135cm, 150cm Kiba 18: 90cm, 105cm, 135cm, 150cm Kiba 35: 90cm, 135cm | | Catheter Shaft OD | WM 14C: 0.035" maximum WM 18: 0.05" maximum WM 35: 0.06" maximum | Kiba 14: 0.023" maximum Kiba 18: 0.033" maximum Kiba 35: 0.053" maximum | | Component Materials | WM 14C: 17-7 Stainless Steel WM 18/35: 304 Stainless Steel Braided polyimide shaft Pebax extrusion ABS handle & plunger | Kiba 14/18/35 304 Stainless Steel Gold plating PET heatshrink Lexin resin luer Dynaflex G7960-1001 Strain Relief | | Tip | Gold-plated beveled tip | Gold-plated beveled tip | | Coating Material | Hydrophilic | Hydrophilic | | Coating Length | WM 14C: 30cm WM 18: 30cm WM 35: 40cm | Kiba 14/18/35 40cm | | Packaging Configuration | HDPE backer card and coil in single poly/Tyvek Pouch | HDPE backer card and coil in single poly/Tyvek Pouch | | Sterilization Method | EO sterilization | EO sterilization | # A brief discussion of the nonclinical tests submitted | The following bench testing was completed: | | | --- | --- | | • Simulated Use • Dimensional Verification • Radiopacity • Leak Testing • Kink Resistance • Corrosion Resistance • Component Integrity • Bond Integrity | • Particulate Testing • Torque Testing • Burst Testing • Lubricity and Coating Integrity Testing • Design Validation/Usability • Sterility Testing • Biocompatibility Assessment | The Kiba CTO Crossing Catheter met all specified criteria and did not raise new safety of performance questions. Based on performance and biocompatibility testing, the Kiba device was found to be substantially equivalent to the predicate device. # A brief discussion of the clinical data submitted | No clinical data is submitted. | | --- | | Conclusions | | The conclusions drawn from the non-clinical testing demonstrate that the Kiba CTO Crossing Catheter is substantially equivalent to the legally marketed predicate device. | Confidential Special 510(k) Summary Page 2 of 2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...