K261964 · Elesta S.P.A · GEX · Aug 19, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261964
Device Name
ECHOLASER X4
Applicant
Elesta S.P.A
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 19, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ECHOLASER X4 laser system is indicated for use to cut, vaporize, ablate, necrotize or coagulate soft tissue through interstitial irradiation in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, plastic surgery, orthopedics, pulmonology, radiology, and urology, at a wavelength of 1064 nm. Echolaser X4 laser system includes the following specific indications: Discectomy: - Percutaneous Lumbar Discectomy Gastrointestinal: - Tissue Ablation: Benign and malignant Neoplasm Prostatectomy: - Soft Tissue Coagulation: Benign Prostatic Hyperplasia (BPH prostatectomy) Pulmonary Surgery: - Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system. Examples include: Benign and malignant pulmonary obstruction GENERAL SURGERY - Soft Tissue: Tumors and Lesions Tissue Ablation GENITOURINARY SURGERY Ablation and Hemostasis: - Superficial Urinary Bladder Tumors - Invasive Bladder Carcinoma GYNECOLOGICAL TISSUE ABLATION - Endometrial Ablation (menorrhagia) - Submucous Fibroids OTORHINOLARYNGOLOGY SURGERY - Soft Tissue: Tumors and Lesions Tissue Ablation
Device Story
ECHOLASER X4 is a diode laser system emitting 1064 nm light for interstitial soft tissue surgery. System utilizes up to 4 independent optical fibers for simultaneous delivery; each fiber features a colored connector with an identification microchip. Operator controls laser via touchscreen panel and footswitch; optical fibers are guided to target tissue using needles. Device used in clinical settings by surgeons to cut, vaporize, ablate, necrotize, or coagulate tissue. Output allows precise thermal destruction or coagulation of lesions, tumors, or hyperplastic tissue; benefits include minimally invasive surgical access and controlled tissue management.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on identical technological characteristics and performance data to the predicate device. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, and IEC 62304 standards demonstrated via declaration of conformity.
Technological Characteristics
Diode laser source; 1064 nm wavelength; 28W max power; continuous wave mode. Fiber-optic delivery (up to 4 fibers). Compliance with IEC 60601-1, 60601-1-2, 60601-2-22, 60825-1, and 62304. Touchscreen interface with microchip-based fiber identification.
Indications for Use
Indicated for patients requiring soft tissue cutting, vaporization, ablation, necrosis, or coagulation via interstitial irradiation across various surgical specialties including cardiovascular thoracic (excluding heart/pericardial vessels), dermatology, ENT, gastroenterology, general surgery, gynecology, head/neck, orthopedics, pulmonology, radiology, and urology. Specific applications include percutaneous lumbar discectomy, benign/malignant neoplasms, BPH, pulmonary obstructions, bladder tumors, and fibroids.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 19, 2026
Elesta S.P.A
% Jessica Czamanski
Regulatory, Quality, and Compliance Consultant
DuVal & Associates, P.A.
825 Nicollet Mall
Suite 1820
Minneapolis, Minnesota 55402
Re: K261964
Trade/Device Name: ECHOLASER X4
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: June 11, 2026
Received: June 11, 2026
Dear Jessica Czamanski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261964 - Jessica Czamanski
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K261964 - Jessica Czamanski
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2026.08.19 16:46:45 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261964
Device Name
ECHOLASER X4
Indications for Use (Describe)
The ECHOLASER X4 laser system is indicated for use to cut, vaporize, ablate, necrotize or coagulate soft tissue through interstitial irradiation in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, plastic surgery, orthopedics, pulmonology, radiology, and urology, at a wavelength of 1064 nm.
Echolaser X4 laser system includes the following specific indications:
Discectomy:
- Percutaneous Lumbar Discectomy
Gastrointestinal:
- Tissue Ablation: Benign and malignant Neoplasm
Prostatectomy:
- Soft Tissue Coagulation: Benign Prostatic Hyperplasia (BPH prostatectomy)
Pulmonary Surgery:
- Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system.
Examples include: Benign and malignant pulmonary obstruction
GENERAL SURGERY
- Soft Tissue:
Tumors and Lesions
Tissue Ablation
GENITOURINARY SURGERY Ablation and Hemostasis:
- Superficial Urinary Bladder Tumors
- Invasive Bladder Carcinoma
GYNECOLOGICAL TISSUE ABLATION
- Endometrial Ablation (menorrhagia)
- Submucous Fibroids
OTORHINOLARYNGOLOGY SURGERY
- Soft Tissue:
Tumors and Lesions
Tissue Ablation
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740
EF
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 2 of 2
{5}
ELESTA SPA
K261964
ECHOLASER X4
### 510(k) Summary K261964
ECHOLASER X4 Traditional 510(k)
Prepared on 08/06/2026
This 510(k) Summary is being submitted as required by 21 CFR 807.92.
#### 1. General Information
Applicant: ELESTA SPA
+39 055 776 0190
Legal Office: Via Baldanzese 17, Calenzano (FI) 50041, Italy
Operative Site: Via Arsiero Salvanti 41/43, Calenzano (FI) 50041, Italy
Establishment registration number: 3015077548
Contact for this application:
Jessica Czamanski, RAC
Regulatory, Quality, and Compliance Consultant, DuVal & Associates
825 Nicollet Mall, Suite 1820
Minneapolis, MN 55402
Office: 612.338.7170
Cell: 754.422.9101
email: jczamanski@duvalfdalaw.com
#### 2. Device
Trade name: ECHOLASER X4
Classification name: Powered Laser Surgical Instrument
Regulation Name & Number: Laser surgical instrument for use in general and plastic surgery and in dermatology (21 CFR 878.4810)
Product Code: GEX
Class: II
#### 3. Predicate Devices
ECHOLASER X4 is substantially equivalent to the following devices:
| Applicant | Device name | 510(k) Number | Use |
| --- | --- | --- | --- |
| ELESTA SPA | ECHOLASER X4 | K213594 | PRIMARY PREDICATE |
| QUANTA SYSTEM SPA | QUANTA1064Quanta Diode Laser Family | K100558 | SECONDARY PREDICATE |
| LAMBDA SPA | DOCTOR DIODE Laser Family | K102036 | REFERENCE |
The Subject Device has same technological characteristics, materials, chemical composition, principle of operation, energy source, and biocompatibility as the Predicate ECHOLASER X4 (K213594) and equivalent
Page 1 of 6
{6}
ELESTA SPA
K261964
ECHOLASER X4
technological characteristics, principle of operation, energy source and mode of use of Predicate QUANTA 1064 (K100558).
4. Comparison of Technological Characteristics
| | SUBJECT DEVICE (K261964) | PRIMARY PREDICATE DEVICE (K213594) | SECONDARY PREDICATE DEVICE (K100558) | REFERENCE DEVICE (K102036) | Comments |
| --- | --- | --- | --- | --- | --- |
| Product name | ECHOLASER X4 | ECHOLASER X4 | Quanta1064 - Quanta Diode Laser Family | Doctor Diode Laser Family | - |
| 510(k) | K261964 | K213594 | K213594 | K102036 | - |
| Applicant | ELESTA S.p.A. | ELESTA S.p.A. | QUANTA SYSTEM S.p.A. | LAMBDA S.p.A. | - |
| Classification | 21 CFR §878.4810 | 21 CFR §878.4810 | 21 CFR §878.4810 | 21 CFR §878.4810 | SAME |
| Product Code | GEX | GEX | GEX | GEX | SAME |
| Indications for Use | The ECHOLASER X4 laser system is indicated for use to cut, vaporize, ablate, necrotize or coagulate soft tissue through interstitial irradiation in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, plastic surgery, orthopedics, pulmonology, radiology, and urology, at a wavelength of 1064 nm.Echolaser X4 laser system includes the following specific indications:Discectomy- Percutaneous Lumbar DiscectomyGastrointestinal- Tissue Ablation: Benign and malignant NeoplasmProstatectomy- Soft Tissue Coagulation: Benign Prostatic Hyperplasia (BPH prostatectomy)Pulmonary Surgery:- Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system. | The ECHOLASER X4 laser system is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, plastic surgery, orthopedics, pulmonology, radiology, and urology, at a wavelength of 1064nm. | The Quanta System QUANTA Diode Laser Family, including the QUANTA532, QUANTA808, QUANTA940, QUANTA980,QUANTA1064, QUANTA1320, QUANTA1470 and QUANTA1950 (and all their double wavelength combination and their delivery accessories used to deliver optical energy) are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology (BPH), Genitourinary (Urology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery, Gastroenterology, Head/neck/ENT and Radiology, Endovascular coagulation, Oral Surgery and Dental procedures. [...]QUANTA1064 DERMATOLOGY/PLASTIC SURGERYPhotocoagulation: | Doctor Diode Laser Family including Doctor Smile, Doctor Surgery and Helios are available at ... 1064 wavelength with variable power of 500w, 2w, 2,5w , 5w, 7w, 8w, 10w and 15w (and their delivery accessories used to deliver optical energy) are indicated for use in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology (BPH), Genitourinary (Urology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery (PLDD), Gastroenterology, Head/neck/ENT and Radiology, Endovascular coagulation, Oral Surgery and Dental procedures [...]Doctor Diode Laser Family at 1064mm wavelength: DERMATOLOGY/PLASTIC SURGERYPhotocoagulation: Colored Vascular Lesions of Skin (only if Argon Laser is unsuccessful)Discectomy:Percutaneous Lumbar DiscectomyGastrointestinal:Tissue Ablation: | The subject IFU is based on the IFUs of the predicate devices and is further supported by the reference device's IFU. This update adds specificity to the indication for the same general intended use. |
Page 2 of 6
{7}
ELESTA SPA
K261964
ECHOLASER X4
| | SUBJECT DEVICE (K261964) | PRIMARY PREDICATE DEVICE (K213594) | SECONDARY PREDICATE DEVICE (K100558) | REFERENCE DEVICE (K102036) | Comments |
| --- | --- | --- | --- | --- | --- |
| | Examples include: Benign and malignant pulmonary obstruction GENERAL SURGERY - Soft Tissue: Tumors and Lesions Tissue Ablation GENITOURINARY SURGERY Ablation and Hemostasis: - Superficial Urinary Bladder Tumors - Invasive Bladder Carcinoma GYNECOLOGICAL TISSUE ABLATION - Endometrial Ablation (menorrhagia) - Submucous Fibroids OTORHINOLARYNGOLOGY SURGERY - Soft Tissue: Tumors and Lesions Tissue Ablation | | Colored Vascular Lesions of Skin (only if Argon Laser is unsuccessful) Discectomy: Percutaneous Lumbar Discectomy Gastrointestinal: Tissue Ablation: • Benign and malignant Neoplasm • Polyps • Colitis • Ulcers • Aniodysplasia • Hemorroids [...] GENERAL SURGERY Soft Tissue: • Skin Incision • Tissue Dissection • Excision (external tumors and lesions) • Resection of Internal Organs (complete or partial) • Tumors and Lesions • Tissue Ablation Vessel Coagulation GENITOURINARY SURGERY Ablation and Hemostasis: • Superficial Urinary Bladder Tumors • Invasive Bladder Carcinoma • Urethral Strictures • Lesions of the External Genitalia GYNECOLOGICAL TISSUE ABLATION • Endometrial Ablation (menorrhagia) • Soft Tissue Excisional Conization • Submucous Fibroids • Polyps • Septa NEUROSURGERY [...] OTORHINOLARYNGOLOGY SURGERY Soft Tissue: • Skin Incision • Tissue Dissection | • Benign and malignant Neoplasm • Polyps • Colitis • Ulcers • Aniodysplasia • Hemorroids Prostatectomy: Soft Tissue Coagulation: • Benign Prostatic Hyperplasia (BPH Prostatectomy) Pulmonary Surgery: Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system. Examples include: • Tracheobronchial malignancy or stricture • Benign and malignant pulmonary obstruction • Endoscopic pulmonary applications [...] Endovenous Occlusion of the Greater Saphenous Vein in Patient with Superficial Vein Reflux [...] | |
Page 3 of 5
{8}
ELESTA SPA
K261964
ECHOLASER X4
| | SUBJECT DEVICE (K261964) | PRIMARY PREDICATE DEVICE (K213594) | SECONDARY PREDICATE DEVICE (K100558) | REFERENCE DEVICE (K102036) | Comments |
| --- | --- | --- | --- | --- | --- |
| | | | - Excision (external tumors and lesions) - Resection of Internal Organs (complete or partial) - Tumors and Lesions - Tissue Ablation - Vessel Coagulation**Prostatectomy:** **Soft Tissue Coagulation:** - Benign Prostatic Hyperplasia (BPH Prostatectomy) POLMUNARY SURGERY Palliative Treatment: Benign and Malignant Pulmonary Airway Obstructions | | |
| **Laser Type** | Diode | Diode | Diode | Diode | SAME |
| **Wavelength** | 1064 nm | 1064 nm | 1064 nm | 1064 nm | SAME |
| **Maximum delivered power** | 28W | 28W | **30 W** | **15W** | SAME AS PRIMARY PREDICATE DEVICE and within the range of SECONDARY PREDICATE DEVICE and REFERENCE DEVICE |
| **Stability of the output power** | ±20% | ±20% | unknown | unknown | SAME AS PRIMARY PREDICATE DEVICE |
| **Mode of operation** | Continuous wave | Continuous wave | **Continuous wave, PW** | Continuous wave | SAME |
## 5. Device description
The ECHOLASER X4, is a medical device equipped with a diode laser source emitting at 1064nm wavelength in CW mode. The laser beam is transmitted via optical fibres. The operator can use up to 4 optical fibres simultaneously, one independent from the other in both activation and power adjustment. Laser activation is controlled by footswitch. The laser beam is transmitted via optical fibers, each identified by a colored connector containing an identification microchip; it is extremely easy to control from the touchscreen control panel which for each fiber displays a corresponding area in the same color. The optical fiber is guided to the target tissue with the aid of needle.
Page 4 of 6
{9}
ELESTA SPA
K261964
ECHOLASER X4
### 6. Indications for Use
The ECHOLASER X4 laser system is indicated for use to cut, vaporize, ablate, necrotize or coagulate soft tissue through interstitial irradiation in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, plastic surgery, orthopedics, pulmonology, radiology, and urology, at a wavelength of 1064 nm.
Echolaser X4 laser system includes the following specific indications:
Discectomy:
- Percutaneous Lumbar Discectomy
Gastrointestinal:
- Tissue Ablation: Benign and malignant Neoplasm
Prostatectomy:
- Soft Tissue Coagulation: Benign Prostatic Hyperplasia (BPH prostatectomy)
Pulmonary Surgery:
- Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system.
Examples include: Benign and malignant pulmonary obstruction
GENERAL SURGERY
- Soft Tissue:
- Tumors and Lesions
- Tissue Ablation
GENITOURINARY SURGERY Ablation and Hemostasis:
- Superficial Urinary Bladder Tumors
- Invasive Bladder Carcinoma
GYNECOLOGICAL TISSUE ABLATION
- Endometrial Ablation (menorrhagia)
- Submucous Fibroids
OTORHINOLARYNGOLOGY SURGERY
- Soft Tissue:
- Tumors and Lesions
- Tissue Ablation
### 7. Performance data (non-clinical)
No additional performance testing was needed since the subject and predicate devices have identical technological characteristics.
The Subject Device is compliant with the following Consensus Standards:
Page 5 of 6
{10}
ELESTA SPA
K261964
ECHOLASER X4
| Recognition Number | Standard Number and Version | Standard Title | Compliance |
| --- | --- | --- | --- |
| 19-49 | IEC 60601-1 Edition 3.2 2020-08 CONSOLIDATED VERSION | Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - Note: This standard is recognized with relevant US national differences applied see references #1 and #2 in the Relevant FDA Guidance and/or Supportive Publication section below. | Declaration of Conformity |
| 19-36 | IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION | Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)] | Declaration of Conformity |
| 12-356 | IEC 60601-2-22 Edition 4.0 2019-11 | Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical cosmetic therapeutic and diagnostic laser equipment | Declaration of Conformity |
| 12-273 | IEC 60825-1 Edition 2.0 2007-03 | Safety of laser products - Part 1: Equipment classification and requirements [Including: Technical Corrigendum 1 (2008) Interpretation Sheet 1 (2007) Interpretation Sheet 2 (2007)] | Declaration of Conformity |
| 13-79 | ANSI AAMI IEC 62304:2006/A1:2016 | Medical device software - Software life cycle processes [Including Amendment 1 (2016)] | Declaration of Conformity |
## 8. Conclusion
The ECHOLASER X4 is substantially equivalent to the cleared Predicate Devices K213594 and K100558.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.