K102036 · Lambda S.P.A. · GEX · Sep 28, 2010 · General, Plastic Surgery
Device Facts
Record ID
K102036
Device Name
DOCTOR SURGERY DIODE LASERS FAMILY
Applicant
Lambda S.P.A.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 28, 2010
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Doctor Diode Laser Family including Doctor Smile, Doctor Surgery and Helios are available at 810 wavelength with variable power of 500mw, 2w, 2,5w, 5w, 7w, 8w, 10w and 15w; at 940 wavelength with variable power of 500mw, 2w, 2,5w , 5w, 7w, 8w, 10w and 15w; at 980 wavelength with variable power of 500mw, 2w, 2,5w , 5w, 7w, 8w, 10w and 15w and at 1064 wavelength with variable power of 500mw, 2w, 2,5w, 5w, 7w, 8w, 10w and 15w (and their delivery) in surgical applications requiring the vaporization, incision, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology (BPH), Genitourinary (Urology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery (PLDD), Gastroenterology, Head/neck/ENT and Radiology, Endovascular coagulation, Oral Surgery and Dental procedures.
Device Story
Doctor Diode Laser Family consists of portable diode laser systems (810, 940, 980, 1064 nm) for soft tissue surgery. System includes console, fiber optic delivery system, handpieces, and optical tips. Operated by physicians/specialists in clinical/surgical settings. Laser energy delivered via contact/non-contact fiber optics to target tissue; enables vaporization, incision, excision, ablation, hemostasis, and coagulation. Used in conjunction with endoscopic equipment (laparoscopes, hysteroscopes, etc.) or open surgery. Output is controlled laser energy; provider monitors procedure visually or via endoscope. Benefits include precise tissue management, reduced bleeding, and minimally invasive access for various surgical/dental indications.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, and 21 CFR 1040.10/1040.11 standards demonstrated.
Indicated for soft tissue surgical applications (vaporization, incision, excision, ablation, cutting, hemostasis, coagulation) across multiple specialties including Urology (BPH), Genitourinary, Thoracic, Plastic, General, Pulmonary, Spinal, and Neurosurgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Gastroenterology, ENT, Radiology, Dermatology/Aesthetics (vascular lesions, hair removal), and Dental/Oral surgery. Specific patient populations include men >50 with BPH (28-85 cc prostate volume) and patients with superficial vein reflux.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for LAMBDA S p A. The logo consists of a stylized lambda symbol enclosed in a partial circle on the left. To the right of the symbol is the text "LAMBDA S p A" with a line underneath. The text is in a simple, sans-serif font.
# 510(K) SUMMARY
| | K102036 |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) owner and submitter | LAMBDA S.p.A.<br>SEP 28 2010<br>Via Dell'Impresa 1<br>36040 Brendola (Vicenza)<br>Italy<br>+39 0444 349 165<br>+39 0444 349 954 |
| Contact person | Mr Boschi Alessandro<br>+39 0444 349 165<br>boschi@lambdascientifica.com |
| The common name of the device | Diode laser |
| Trade name / Propietary names | Doctor Diode Laser Family |
| Classification name | Laser surgical instrument for use in general and plastic surgery and<br>dermatology |
| Product code | Gex |
| CFR Number | 21 CFR §878.4810 |
| Predicate Devices | Doctor Diode Laser Family:<br>- Velure Family of Diode Laser (K063396)<br>- Quanta System Polysurge Diode Laser Family (K083613)<br>- Doctor Smile A-810, B980 Lasers (K091562)<br>- Arc Laser GmbH Fox Q-810, Q-980, Q-1064 (K073322)<br>- Biolase Ezlase® (K082938)<br>- Styla MicrolaserTM (K081214)<br>- Odissey navigator (K062258)<br>- Sirona Sirolaser (K053161)<br>- Kavo Gentleray 980 (K072262)<br>- Ceralas 980 (K072779)<br>- Claros Nano Dental Laser System (K081652) |
| Product description | Doctor Diode Laser Family are the same standard diode medical<br>lasers respectively with 810, 940, 980, 1064 nm wavelength. |
| Intended Use | Doctor Diode Laser Family including Doctor Smile, Doctor<br>Surgery and Helios are available at 810 wavelength with<br>variable power of 500mw, 2w, 2,5w, 5w, 7w, 8w, 10w and<br>15w; at 940 wavelength with variable power of 500mw, 2w,<br>2,5w , 5w, 7w, 8w, 10w and 15w; at 980 wavelength with<br>variable power of 500mw, 2w, 2,5w , 5w, 7w, 8w, 10w and<br>15w and at 1064 wavelength with variable power of 500mw,<br>2w, 2,5w, 5w, 7w, 8w, 10w and 15w (and their delivery) |
| | in surgical applications requiring the vaporization, incision, |
| | excision, ablation, cutting and hemostasis, or coagulation of soft |
| | tissue in conjunction with endoscopic equipment for medical |
| | specialist including: Urology (BPH), Genitourinary (Urology), |
| | Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics |
| | including vascular lesions and hair removal, General Surgery, |
| | Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal |
| | Surgery, Gynecology, Pulmonary Surgery, Neurosurgery |
| | (PLDD), Gastroenterology, Head/neck/ENT and Radiology, |
| | Endovascular coagulation, Oral Surgery and Dental procedures. |
Laser Equipment - Sensori e Sistemi di Monitoraggio / Monitoring Systems and sensors 13 Via Dell'Impresa, 1 - 36040 Brendola (VI) Italia - Tel/fax +39-0444-349165 /349954 http://www.lambdascientifica.com - - E-Mail: info@lambdascientifica.com Cap. Soc. Euro 500.000,00 – C.F. e P. I.V.A. 02558810244 - Registro Imprese di Vicenza n. 199230/96 R.E.A. n. 240226. Direzione e Coordinamento Helios Group SpA
{1}------------------------------------------------
| Performance standards | The Doctor Diode Laser Family comply with the following standards:<br>IEC 60601-1:1998+A1:1991+A2:1995<br>IEC 60601-2-22:1995<br>IEC 60825-1:1993+A1:1997+A2:2001<br>21 CFR 1040.10 and 1040.11 |
|-------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device technological<br>characteristics and<br>Comparison to Predicate<br>device(s) | The Doctor Diode Laser Family are portable instruments which<br>transmit laser energy via a contact fiber optic delivery system and<br>each unit is composed of the following main components:<br>-a laser system console (including display panel, software and<br>control);<br>-a fiber optic delivery system;<br>-one or more handpieces<br>-protective eye ware<br>-optical tips<br>Doctor Diode Laser Family have the same intended use and the same<br>or substantially equivalent technical specifications and mechanism of<br>action as compared with the named predicate devices. |
| Conclusion: | The Doctor Diode Laser Family have identical or similar indications<br>for use, principles of operation, overall technical and functional<br>capabilities. Therefore is substantially equivalent to the predicate<br>devices. They have similar intended uses and comply with the same<br>safety and performance standards. |
**Review:**
Boschi Alessandro
14 Laser Equipment - Sensori e Sistemi di Monitoraggio / Monitoring Systems and sensors Via Dell'Impresa, 1 - 36040 Brendola (VI) Italia - Tel/fax +39-0444-349165 /349954 http://www.lambdascientifica.com - E-Mail: info@lambdascientifica.com Cap. Soc. Euro 500.000,00 – C.F. e P. I.V.A. 02558810244 - Registro Imprese di Vicenza n. 199230/96 R.E.A. n. 240226. Direzione e Coordinamento Helios Group SpA
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features the department's symbol, which is a stylized representation of a human form. The symbol is composed of three curved lines that form the shape of a person. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
LAMBDA S.p.A. % Mr. Boschi Alessandro Via Dell'Impresa n. 1 36040 Brendola (Vicenza) Italy
SEP 2 8 2010
Re: K102036
Trade/Device Name: Doctor Surgery Diode Lasers Family Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: September 23, 2010 Received: September 23, 2010
Dear Mr. Alessandro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Boschi Alessandro
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.bgm.
Sincerely vours.
incerely yours,
ark N. Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the Greek letter lambda inside of a circle on the left. To the right of the circle is the text "LAMBDA sp.A" with a horizontal line underneath. The text is in a sans-serif font and is black.
102036
## ATTACHMENT A COPY
Section 4
Indications for Use
SEP 2 8 2010
510(k) Number (if known): K102036
Device name: Doctor Surgery Diode Lasers Family
## Indications for use:
Doctor Diode Laser Family including Doctor Smile, Doctor Surgery and Helios are available at 810 wavelength with variable power of 500mw, 2w, 2,5w, 5w, 10w and 15w; at 940 wavelength with variable power of 500mw, 2w, 2,5w, 5w, 7w, 8w, 10w and 15w; at 980 wavelength with variable power of 500mw, 2w, 2,5w , 5w, 7w, 8w, 10w and 15w and at 1064 wavelength with variable power of 500mw, 2w, 2,5w, 5w, 7w, 8w, 10w and 15w (and their delivery accessories used to deliver optical energy) are indicated for use in surgical applications requiring the vaporization, excision, ablation, cutting and hemostasis, or coagulation of soft tissue in conjunction with endoscopic equipment for medical specialist including: Urology (BPH), Genitourinary (Urology), Thoracic Surgery, Plastic Surgery and Dermatology, Aesthetics including vascular lesions and hair removal, General Surgery, Ophthalmology, Orthopedics, Podiatry, Arthroscopy, Spinal Surgery, Gynecology, Pulmonary Surgery, Neurosurgery (PLDD), Gastroenterology, Head/neck/ENT and Radiology, Endovascular coagulation, Oral Surgery and Dental procedures.
## Doctor Diode Laser Family at 810nm's wavelength:
## Dental application:
- Excisional and incisional biopsies, Exposure of unerupted teeth, Fibroma removal, Frenctomy, Gingival troughing for crown impressions, Gingivoplasty, Gingival incision and excision Haemostasis and coagulation, Implant recovery, Incision and drainage of abscess Leukoplakia, Operculectomy, Oral papillectornies, Pulpotomy, Pulpotomy as an adjunct to root canal therapy, Reduction of gingival hypertrophy. Soft tissue crown lengthening, Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa, Vestibuloplasty. Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket. Sulcular debridement (removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival bleeding index, probe depth, attachment loss and tooth mobility.). Light activation of whitening/bleaching material.
## Doctor Diode Laser Family at 810nm and 980nm's wavelength
## Urology:
- Lesions of external genitalia -
- -Circumcision
- Condyloma ।
- Bladder tumors ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
- Bladder neck incisions
- Vaporizations of the prostate
Nelkl Seden Ls mkm
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102036
5
Apparecchiature Laser / Laser Equipment - Sensori e Sistemi di monitoraggio / Monitoring Systems and Sensors
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for LAMBDA Sp.A. The logo consists of the greek letter lambda inside of a circle. To the right of the circle is the text "LAMBDA Sp.A." with a line underneath.
... ..
## General Surgerv:
- Rectal and anal hemorrhoidectomy -
- -Mastectomy
- -Dermabrasion
- Appendectomy (open and laparscopic)
- Bowel resection (open and laparscopic) -
- Colectomy -
- -Liver resection
- Resection of organs .
- Thyriodectomy .
- Adhesiolysis -
- -Hepatobiliary tumors
- Thoracotomy
- Cholecystectomy (open and laparscopic) -
- Condyloma -
- -Breast biopsy
#### Neurosurgery:
- Percutaneous Disc Decompression (PLDD) -
- -Discectomy
- Hemostasis in conjunction with meningiomas -
## Gynecology: . .
- Cervical conization -
- Myomectomy -
- Endometrial ablation -
- -Ovarian cystectomy
- Appendectomy -
## Ophthalmology:
- Dacryocystorhinostomy transcanalicular -
- Open DCR -
- Tumor excision -
- -Blepharoplasty
#### Orthopedics:
- Dissect and coagulate -
## Gastroenterology:
- Hemostasis of colonoscopy -
- Hemostasis of esophageal varices -
- Excision of polyps -
## Arthroscopy:
- Chondromalacia -
- Synovectomy -
- Menisectomy -
## Thoracic Surgery:
- Thoracotomy 1
Neil M. Darden for nxn
(Division Signer)
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
102036 510(k) Number_
Apparecchiature Laser / Laser Equipment - Sensori e Sistemi di monitoraggio / Monitoring Systems and Sensors
ﻴﺔ ، ﻓ
Via dell'Impresa - 36040 Brendola - Vicenza - Italy ☎ +39-0444-349165 - Fax +39-0444-349954 E-Mail: info@lambdaspa.com http://www.lambdaspa.com -Cap. Soc. 500.000 € - Reg. Imp. di Vi., C.F., P. NA 02558810244 -- R.E.A. n. 240226 Direzione e coordinamento Helios Group Spa
{6}------------------------------------------------
- Pulmonary resection
- Hemostasis
- Pericardiectomy .
- Adhesiolysis -
- -Coagulation of blebs and bullae
## Pulmonology:
- Endoscopic pulmonary applications .
- -Tracheal bronchial lesions
- Benign an malignant pulmonary obstruction .
## Otolaryngology ENT:
- Removal of benign lesions from the ear, nose and throat -
- Excision of carcinoma of the larynx -
- Incision and excision of carcinoma in situ t
- Neck dissection -
- Laryngeal papillomectomy -
- Removal of vocal cord/fold nodules, polyps and cyst -
## Pulmonary Surgery:
.. ...
- Endoscopic pulmonary applications -
- Tracheal bronchial lesions -
- Benign and malignant pulmonary. or stricture -
## Cardiac Surgery:
- Coagulation and hemostasis of cardiac tissue -
## Dermatology/Aesthetics:
: ~, (
- Photocoagulation of vascular & dermatological lesions of the face and extremities -
- Photocoagulation of telangiectasia, veinulectasia of the legs and face -
- -Treatment of reticular veins and branch varicosities
- Pyrogenic granuloma, lymphangioma and lymphangiomatosis disease, angiofibromas -
- Superficial benign vascular lesions including Telangiectasias, Rosacea, Angioma, venous lakes -Couperosis, Cherry angioma, hemangioma, Port wine stains, angiokeratoma, and benign epidermal pigments lesions as lentigines. Epidermal nevi, spider nevi.
- Dermatological surgery: Condyloma acuminate, warts, small non malignant skon tumors, small semimalignant tumors as basalomas, Bowe, Kaposi sarcom. Warty leucoplasty and ulcers debridment.
- Seborrheic keratosis -
- Mixoid cyst 1
- Papillary varix -
- Acne treatment -
- Hair removal of unwanted hair from skin type I-V -
## Plastic Surgery:
- Cut, coagulation & vaporization -
- -Resurfacing non
- Blepharopasty -
## Vascular Surgery:
- Endoluminal or endovenous laser surgery for saphenous incompetent veins -
neiffel Di
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102036
Apparecchiature Laser / Laser Equipment - Sensori e Sistemi di monitoraggio / Monitoring Systems and Sensors
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Lambda S.p.A. The logo consists of the Greek letter lambda inside of a circle. To the right of the circle is the text "LAMBDA" followed by "s.p.A" in a smaller font size.
## Doctor Diode Laser Family at 940nm wavelength:
The Doctor Diode Laser Family(and their delivery accessories used to deliver optical energy) are indicated for use in medicine and surgery, in the following medical specialties: Urology. Plastic Surgery, Dermatology, Gynecology, Pulmonary Surgery, Gastroenterology, ENT, Radiology, Dentistry,
The Doctor Diode Laser Family(and their delivery accessories used to deliver optical energy) are intended for use in cutting, vaporization, ablation and coagulation of soft tissue in conjunction with endoscopic equipment (including laparoscopes, hysteroscopes, bronchoscopes, cystoscopes, and colonoscopies), in incision/excision, vaporization, ablation and coagulation of soft tissue in contact and non-contact open surgery(with or without a handpiece), in the treatment and/or removal of vascular lesions (turnors) and removal of unwanted hair, and for endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux.
The Doctor Diode Laser Family(and their delivery accessories used to deliver optical energy) are intended for use in the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostates with median and/or lateral lobes ranging in total volume from 28-85 cc.
The Doctor Diode Laser Family(and their delivery accessories used to deliver optical energy) are indicated for incision, excision, vaporization, ablation of oral soft tissue including marginal and inter-dental gingival and epithelial lining of free gingival and the following specific indications: Excisional biopsies Exposure of unerupted teeth; Fibroma removal; Frenotomy; Gingival troughing for crown impressions; Gingivectomy; Gingival incision and excision; Haemostasis and coagulation; Implant recovery; incision and drainage of abscess; Leukoplakia; Operculectomy; Oral papillectomies; Pulpotomny as an adjunct to root canal therapy; Reduction of gingival hypertrophy; Soft tissue crown lengthening; Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa; Vestibuloplasty:
The Doctor Diode Laser Family(and their delivery accessories used to deliver optical energy) are indicated for Laser Periodontal procedures, including: Laser soft tissue curettage; Laser removal of diseased, infected, inflamed and necrosed soft tissue within the periodontal pocket; Sulcular debridement(removal of diseased, infected, inflamed, and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingiyal index, gingival bleeding index, probe depth, attachment loss and tooth mobility).
## Doctor Diode Laser Family at 980nm wavelength:
The Doctor Diode Laser Family(and their delivery accessories used to deliver optical energy) are indicated for use in incision, excision, vaporization, ablation, hemostasis or coagulation of soft tissue in the ear, nose and throat and oral surgery (otolaryngology), arthroscopy, gastroenterology, general surgery, dermatology, plastic surgery, podiatry, urology, gynecology, neurosurgery (peripheral nervous system), pulmonary surgery, cardiothoracic surgery, dental applications, abd endovenous occlusion of the greater saphenous vein. The device is specifically indicated for use as follows:
## Ear. Nose and Throat and Oral Surgery (Otolaryngology):
Hemostasis, incision, excision ablation, coagulation, and vaporization of tissug from the ear, nose, throat and adjacent areas including soft tissue in the oral cavity. Examples include Division Sign-Off)
- Removal of benign lesions from the ear, nose and throat -
- Excision and vaporization of vocal cord nodules and polyps -
- Incision and excision of carcinoma in situ •
- Ablation and vaporization of hyperkeratosis -
Division of Surgical, Orthopedic, and Restorative Devices
Apparecchiature Laser / Laser Equipment - Sensori e Sistemi di monitoraggi Sensors
Via deli'Impresa - 36040 Brendola - Vicenza - Italy http://www.lambdaspa.com - - E-Mail: info@lambdaspa.com Cap. Soc. 500.000 € - Reg. Impr. di Vi., C.F., P. VA 02558810244 -- R.E.A. n. 240226 Direzione e coordinamento Helios Group Spa
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for LAMBDA spA. The logo consists of a lambda symbol inside of a circle. The text "LAMBDA spA" is to the right of the circle, with a line underneath the text.
స్ట్ స్టే
- Excision of carcinoma of the larynx
- Laryngeal papillomectomy
- Excision and vaporization of herpes simplex I and II
- Neck dissection
- -Tonsillectomy
- Thyroidectomy -
- -Vocal cord polypectomy
- Hemiglossectomy -
- Tracheal stenosis -
- Oral cavity lesions ' -
## Arthroscopy:
Hemostasis, incision, excision, coagulation, vaporization of joint tissues during arthroscopic surgery. Examples include:
- Menisectomy -
- Synovectomy -
- -Chondromalacia
## Gastroenterology:
:
Hemostasis, incision, excision, ablation, coagulation and vaporization of tissue in the upper and lower gastrointestinal tracts and also with endoscopic procedures. Examples include:
- Hemostasis of upper and lower GI bleeding -
- Excision and vaporization of colorectal carcinoma -
- ・・・・ Excision of polyps
## General Surgery, Dermatology, Plastic Surgery and Podiatry:
... .. ..
Excision, ablation, vaporization and photocoagulation of skin lesions, hemostasis, incision, vaporiztion ablation and debulking of soft tissue, abdominal, rectal, skin, fat or muscle tissue and dermabrasion. Examples include:
- Matrixectomy -
- -Excision of neuromas
- Excision of periungual and subungual warts .
- -Excision of plantar warts
- Excision of keloids -
- -Liver resection
- Excision of cutaneous lesions -
- Hemorrhoidectomy -
- Appendectomy -
- Debridement of decubitus ulcers -
- Hepatobiliary tumors -
- Mastectomy -
- Dermabrasion -
- Vaporization and hemostasis of capillary hemangioma .
- Excision, vaporization and hemostasis of abdominal tumors -
- Excision, vaporiztion and hemostasis of rectal pathology .
- Pilonidal cystectomy -
- Herniorapphy -
- Adhesiolysis -
- Parathyroidectomy .
- Laparoscopic cholecystectomy
Neil RP Ogle
Carman
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
102036 510(k) Number.
Apparecchiature Laser / Laser Equipment - Sensori e Sistemi di monitoraggio / Monitoring Systems and Sensors
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for LAMBDA S.p.A. The logo consists of the Greek letter lambda inside of a circle. To the right of the circle is the text "LAMBDA" followed by "S.p.A."
- Thyroidectomy
- Resection of organs -
- Debridement of wounds -
- -Photocoagulation of teleangectasia of the legs and face
- -Photocoagulation of vascular lesions of the face and extremities
- -Endovascular coagulation of the greater saphenous vein of the thigh in patients with superficial vein reflux
- . Treatment of reticular veins and branch varicosities
#### Urology:
Excision, vaporization, incision, coagulation, ablation and hemostasis of urogical tissues. Examples include:
- Vaporization of urethral tumors -
- Release of urethral stricture ー
- Removal of bladder neck obstruction -
- Excision and vaporization of condyoma -
- -Lesions of external genitalia
- Vaporization of the prostate to treat benign prostatic hyperplasia (BPH) -
#### Gynecology:
Ablation, excision, incision, coagulation, hemostasis and vaporization of gynecological tissue. Examples include: 2 - 2 - " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " "
- Endometrial ablation .. -
- Excision or vaporization of condylomata acuminata -
- Vaporization of cervical intraepithelial neoplasia -
- Cervical conization
- Menorrhagia -
- :
#### Neurosurgery:
Vaporization, coagulation, excision, incision, ablation and hemostasis of soft tissue. · Examples include: hemostasis in conjunction with menigiomas.
#### Cardiac Surgery:
Hemostasis and coagulation of soft tissue, including cardiac tissue.
#### Pulmonary Surgery:
Neil K. Ogden for nxm
K 102036
Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonar (Division Sign-Off) Examples include:
- Tracheobronchial malignancy or stricture -
- Benign and malignant pulmonary obstruction -
- Endoscopic pulmonary applications –
## Dental Applications:
The Doctor Diode Laser Family(and their delivery accessories used to deliver optical energy) are indicated for the following applications on intraoral and extraoral soft tissue (including marginal and interdental gingival and epithelial lining of free gingival): frenectomy, biopsy, operculectomy, implant recovery, gingivectomy, gingivoplasty, gingival troughing, crown lengthening, hemostasis of donor site, removal of granulation tissue, laser assisted flap surgery, debridement of diseased epithelial lining, incisions and draining of abscesses, tissue retraction for impressions, papillectomy, vestibuloplasy, excision of lesions, exposure of unerupted/partially erupted teeth, leukoplakia, removal of hyperplastic tissues, treatment of aphthous ulcers and
Apparecchiature Laser / Laser Equipment - Sensori e Sistemi di monitoraggio / Monitoring Systems and 10 Sensors
Via dell'Impresa - 36040 Brendola - Vicenza - Italy 窗 +39-0444-349165 - Fax +39-0444-349954 http://www.lambdaspa.com Cap. Soc. 500.000 € - Reg. Imp. di Vi., C.F., P. IVA 02558810244 -- R.E.A. n. 240226 Direzione e coordinamento Helios Group Spa
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
and Restorative Devices
510(k) Number_
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for LAMBDA spA. The logo consists of a lambda symbol inside of a circle on the left, and the text "LAMBDA spA" on the right. There is a line underneath the text.
sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket), pulpotomy, pulpotomy as an adjunct to root canal therapy and light activation of bleaching materials for teeth whitening.
## Endovenous Occlusion of the Greater Saphenous Vein in Patients with Superficial Vein Reflux:
Coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients with Superficial Vein Reflux.
## Doctor Diode Laser Family at 1064nm wavelength:
## Dermatology/Plastic Surgery:
Photocoagulation:
- Colored Vascular Lesions of Skin (only if Argon Laser is unsuccessful) -
#### Discectomy:
Percutaneous Lumbar Discectomy
#### Gastrointestinal:
Tissue Ablation:
- Benign and malignant Neoplasm -
- Polyps -
- Colitis -
- Ulcers -
- Aniodysplasia -
- -Hemorrhoids
## Hemostasis:
- Varices -
- -Esophangitis
- Esophageal Ulcer ﮯ
## Pulmonary Surgery:
Hemostasis, vaporization, coagulation, incision, excision and ablation of soft tissue in the pulmonary system. Examples include:
- Tracheobronchial malignancy or stricture -
- Benign and malignant pulmonary obstruction -
- Endoscopic pulmonary applications -
## Dental Applications:
The Doctor Diode Laser Family(and their delivery accessories used to deliver optical energy) are indicated for the following applications on intraoral and extraoral soft tissue (including marginal and interdental gingival and epithelial lining of free gingival): frenectomy, biopsy, operculectomy, implant recovery, gingivectomy, gingivoplasty, gingival troughing, crown lengthening, hemostasis of donor site, removal of granulation tissue, laser assisted flap surgery, debridement of diseased epithelial lining, incisions and draining of abscesses, tissue retraction for impressions, papillectomy, vestibuloplasy, excision of lesions, exposure of unerupted/partially erupted teeth, leukoplakia, removal of hyperplastic tissues, treatment of aphthous ulcers and sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket), pulpotomy, pulpotomy as an adjunct to root canal therapy and light activation of bleaching materials for teeth whitening.
## Endovenous Occlusion of the Greater Saphenous Vein in Patients with Superficial Vein Reflux:
Coagulation of the Greater Saphenous Vein (GSV) of the thigh in patients with Superficial Vein Reflux.
Apparecchiature Laser / Laser Equipment - Sensori e Sistemi di monitoraggio / Monitoring Systems and 11 Sensors
Via dell'Impresa - 36040 Brendola - Vicenza - Italy http://www.lambdaspa.com - E-Mail: info@lambdaspa.com Cap. Soc. 500.000 € - Reg. Imp. di Vi., C.F., P. NA 02558810244 — R.E.A. n. 240226 Direzione e coordinamento Helios Group Spa
Neil H. Ogden for nxn
(Division Sign-C Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K102036
{11}------------------------------------------------
AMBDA SPA
## Doctor Diode Laser Family at 1064nm wavelength:
## Dermatology/Plastic Surgery:
Photocoagulation:
- -Colored Vascular Lesions of Skin (only if Argon Laser is unsuccessful)
#### Discectomy:
Percutaneous Lumbar Discectomy
## Gastrointestinal:
Tissue Ablation:
- Benign and malignant Neoplasm -
- -Polyps
- Colitis -
- Ulcers -
- Aniodysplasia -
- -Hemorrhoids
#### Hemostasis:
- Varices ।
- -Esophangitis
- -Esophageal Ulcer
## Prostatectomy:
Soft Tissue Coagulation:
Benign Prostatic Hyperplasia (BPH Prostatectomy) -
## Pulmonary Surgery:
Palliative Treatment:
- Benign and Malignant Pulmonary Airway Obstructions -
## Dental procedures
Prescription use XX
and/or
:
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
Over-the-counter use
Their RP de
(Division Sign-Off)
510(k) Number_
and Restorative Devices
Division of Surgical, Orthopedic,
102036
(21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
**Review:**
Boschi Alessandro
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.