Retrospective clinical study data; Historical control patient records
The study evaluated the safety and effectiveness of the device by comparing clinical outcomes (graft dysfunction, mechanical support, length of stay, survival) of patients receiving hearts preserved at 10 °C against a historical control group using conventional ice storage.
Severe primary graft dysfunction, postoperative mechanical circulatory support, renal replacement therapy, ICU length of stay, hospital length of stay, and short-term survival
Indications for Use
The X°Port Heart Preservation System is indicated for static hypothermic preservation of donor hearts at 4–10 °C for up to 4 hours using a del Nido (or equivalent) cold storage solution.
Device Story
X°Port Heart Preservation System provides static hypothermic storage and transport for donor hearts. Device utilizes insulated chamber, internal organ support, and phase-change material (PCM) packs (transition temperature ~8 °C) preconditioned at 4 °C. System maintains internal temperature between 4–10 °C for up to 4 hours. Device includes temperature probe and logger; data displayed on external screen and optional mobile application. Non-sterile, single-use device; does not contact organ directly. Used by transplant teams during organ procurement and transport. Provides stable thermal environment to preserve donor heart viability; enables transplantation of higher-risk donor organs.
Clinical Evidence
Retrospective, single-arm study with historical controls (n=143 study, n=299 control). Compared 10 °C hypothermic preservation to conventional ice storage. Endpoints included severe primary graft dysfunction, postoperative mechanical circulatory support, renal replacement therapy, ICU/hospital length of stay, and short-term survival. Results showed comparable outcomes between groups, despite the 10 °C group utilizing higher-risk donor organs (older donors, longer preservation times, greater geographic distances).
Technological Characteristics
Insulated chamber; phase-change material (PCM) cooling packs (transition ~8 °C); temperature probe and logger. Connectivity via external display and optional mobile application. Non-sterile, single-use. Software-based temperature monitoring and logging. Bench testing included electrical, mechanical, thermal, EMC, and cybersecurity validation.
Indications for Use
Indicated for static hypothermic preservation of donor hearts at 4–10 °C for up to 4 hours using a del Nido (or equivalent) cold storage solution.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
September 16, 2026
Traferox Technologies, Inc.
Ms. Anuradha Iyer
Regulatory Affairs Manager
3505 Laird Rd. Unit 16
Mississauga, ON L5L 5Y7
Canada
Re: K261913
Trade/Device Name: X°Port Heart Preservation System
Regulation Number: 21 CFR 876.5880
Regulation Name: Isolated kidney perfusion and transport system and accessories
Regulatory Class: Class II
Product Code: MSB
Dated: June 8, 2026
Received: June 8, 2026
Dear Ms. Iyer:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261913 - Ms. Anuradha Iyer
Page 2
guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261913 - Ms. Anuradha Iyer
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
NICOLE M. GILLETTE -S
Nicole Gillette
Assistant Director
DHT2B: Division of Circulatory Support, Structural, and Vascular Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
*See PRA Statement below.*
510(k) Number (*if known*)
K261913
Device Name
X°Port Heart Preservation System
Indications for Use (*Describe*)
The X°Port Heart Preservation System is indicated for static hypothermic preservation of donor hearts at 4–10 °C for up to 4 hours using a del Nido (or equivalent) cold storage solution.
Type of Use (*Select one or both, as applicable*)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# 510(k) Summary
(K261913)
# Contact Details
21 CFR 807.92(a)(1)
| Applicant Name: | Traferox Technologies Inc. |
| --- | --- |
| Applicant Address: | 3505 Laird Rd. Unit 16 Mississauga ON L5L 5Y7 Canada |
| Applicant Contact Telephone: | +1 437 983 8971 |
| Applicant Contact: | Ms. Anuradha Iyer |
| Applicant Contact Email: | aiyer@traferox.com |
# Device Name
21 CFR 807.92(a)(2)
| Device Trade Name: | X°Port Heart Preservation System |
| --- | --- |
| Common Name: | System and accessories, isolated heart, transport and preservation |
| Classification Name | Isolated kidney perfusion and transport system and accessories |
| Regulation Number | 876.5880 |
| Product Code( s) | MSB |
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K180194 | SherpaPak Cardiac Transport System | MSB |
# Device Description Summary
21 CFR 807.92(a)(4)
The X°Port Heart Preservation System is designed to store, preserve, and transport human hearts donated for transplantation at approximately 10 °C. The target temperature range maintained during storage is 4–10 °C. The internal device temperature is displayed and logged on an external screen. Additionally, the internal temperature may be monitored via an optional mobile application.
The system consists of an insulated chamber, specialized phase-change material with a transition temperature of approximately 8 °C, an internal support for the organ, and a temperature probe and logger. The system is a non-sterile single-use device and does not contact the organ. The X°Port Heart Preservation System is intended to be used with the del Nido (or equivalent) cold storage solution.
The X°Port Heart Preservation System operates on the principle of preconditioning and use of specialized phase-change material cooling packs. The cooling packs are preconditioned at 4 °C
1
{5}
in a temperature-regulated refrigerator for a minimum of 48 hours before setting up the X°Port Heart Preservation System.
### Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The X°Port Heart Preservation System is indicated for static hypothermic preservation of donor hearts at 4–10 °C for up to 4 hours using a del Nido (or equivalent) cold storage solution.
### Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device differs from the predicate in its indicated organ storage temperature and cold storage solution, specifying 4–10 °C with del Nido (or equivalent) cold storage solution. This difference does not constitute a new intended use, as the overall intended use — organ storage and transportation with a cold storage solution for up to 4 hours — remains the same.
### Technological Comparison
21 CFR 807.92(a)(6)
The subject and predicate devices share the same mode of operation, cooling mechanism, and intended organ storage time. The two devices differ in organ container design and organ storage temperature (4–10 °C vs. 4–8 °C). The differences in the technological characteristics between the subject and predicate devices do not lead to different questions of safety or effectiveness.
### Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
The following non-clinical bench testing was performed on the X°Port Heart Preservation System to demonstrate the device's safety and effectiveness:
- Basic safety (electrical, mechanical, thermal)
- Electromagnetic compatibility
- Physical testing
- Functional validation (thermal performance)
- Software and cybersecurity testing
- Packaging and shipping testing
- Lifetime testing
- Human factors testing
The clinical evaluation supporting this submission includes one retrospective, single-arm study with historical controls, conducted in the United States. The study assessed the safety and effectiveness of controlled hypothermic donor heart preservation at 10 °C compared with conventional ice preservation.
In total, 143 study patients were included in the 10 °C group and 299 patients were included in the control group. Baseline characteristics were generally similar in the two groups (10 °C:
2
{6}
average recipient age 54 years, 28.0% female, 69.2% White; control: average recipient age 56 years, 24.1% female, 71.9% White).
The clinical data demonstrate that heart transplant recipient rates of severe primary graft dysfunction, postoperative mechanical circulatory support, renal replacement therapy, intensive care unit length of stay, hospital length of stay, and short-term survival were comparable between the two groups of 10 °C preservation and conventional ice storage. Notably, the 10 °C group included higher-risk donor organs (i.e., older donor hearts, longer preservation times, and organs procured from greater geographic distances).
Overall, the non-clinical and clinical data support that the X°Port Heart Preservation System does not raise any new issues of safety and effectiveness and is substantially equivalent to the predicate device.
3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.