SherpaPak Cardiac Transport System and SherpaPak Kidney Transport System
K180194 · Paragonix Technologies, Inc. · MSB · Mar 30, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K180194
Device Name
SherpaPak Cardiac Transport System and SherpaPak Kidney Transport System
Applicant
Paragonix Technologies, Inc.
Product Code
MSB · Gastroenterology, Urology
Decision Date
Mar 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5880
Device Class
Class 2
Indications for Use
Organ storage and preservation for transplantation.
Device Story
System provides static hypothermic storage for donor hearts/kidneys during transport. Components include outer EPS shipping container, inner/outer hard shell (Sherpa Canister/Shell), and temperature data logger. Organ is immersed in cold storage solution within rigid, sterile, double-barrier canister. Temperature probe provides direct measurement of storage solution; data transmitted to logger and viewable via mobile app. Used by transplant teams to maintain organ temperature between 4°C and 8°C. System ensures organ viability during transit, facilitating successful transplantation.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1), electrical safety/EMC (EN 60601-1, EN 60601-1-6, EN 62366), transportation (ASTM D4169-09), thermal qualification (maintaining 4°C-8°C for 4/24 hours), and helium leak testing for seal integrity.
Technological Characteristics
System uses static cold ischemic storage. Materials include polycarbonate (canister) and rigid EPS (shipping container). Features direct temperature probe, data logger, and mobile app connectivity. Sterilization applied to Sherpa Pak, Shell, and heart connector. Complies with EN 60601-1, EN 60601-1-6, EN 62366, and ASTM D4169-09.
Indications for Use
Indicated for static hypothermic preservation of donor hearts (up to 4 hours) or kidneys (up to 24 hours) during transportation for transplantation, using cold storage solutions. Hearts exceeding clinically accepted preservation times require surgeon evaluation.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with "ADMINISTRATION" written below in a smaller font.
March 30, 2018
Paragonix Technologies, Inc. % Leo L. Basta Official Correspondent Northstar Biomedical Associates 93 Benefit Street Providence, RI 02904
Re: K180194
> Trade/Device Name: SherpaPak Cardiac Transport System SherpaPak Kidney Transport System Regulation Number: 21 CFR§ 876.5880 Regulation Name: Isolated Kidney Perfusion and Transport System and Accessories Regulatory Class: II Product Code: MSB. KDN Dated: January 22, 2018 Received: January 24, 2018
Dear Leo L. Basta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a simple, sans-serif font. The text is black and appears to be against a white background. The letters are evenly spaced and the text is left-aligned.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180194
Device Name SherpaPak Cardiac Transport System SherpaPak Kidney Transport System
Indications for Use (Describe) SherpaPak Cardiac Transport System
The SherpaPak Cardiac Transport System is intended to be used for the static hypothermic preservation of hearts during transportation and eventual transplantation into a recipient using colutions indicated for use with the heart.
The intended organ storage time for the SherpaPak Cardiac Transport System is up to 4 hours.
Donor hearts exceeding clinically accepted static hypothermic preservation times should be evaluated by the transplant surgeon to determine transplantability in accordance with accepted clinical guidelines and in the best medical interest of the intended recipient.
SherpaPak Kidney Transport System
The SherpaPak Kidney Transport System is intended to be used for the static hypothermic preservation of kidney organs during transportation and eventual transplantation into a recipient using colutions indicated for use with this organ.
The SherpaPak Kidney Transport System can maintain the donor organ storage temperature between 4 C and 8 C through 24 hours.
Type of Use (Select one or both, as applicable)
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|---------------------------------------------------------------------------------------------------------------|
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary - K180194 SherpaPak Cardiac Transport System SherpaPak Kidney Transport System
| Submitter: | Paragonix Technologies Inc.<br>c/o Vaughn & Associates<br>639 Granite Street<br>Braintree, MA02184 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Leo Basta<br>Northstar Biomedical Associates<br>93 Benefit Street<br>Providence, RI, 02904<br>Phone: 617.834.9866<br>lbasta@northstarbiomedical.com |
| Date Prepared: | March 30, 2018 |
| Trade Name: | SherpaPak Cardiac Transport System<br>SherpaPak Kidney Transport System |
| Classification Name: | Isolated Kidney Perfusion and Transport System<br>and Accessories |
| Classification: | Class II |
| Regulation Number: | 21 CFR 876.5880 |
| Product Code: | MSB and KDN |
| Predicate Devices: | Sherpa Pak Cardiac Transport System (K133432)<br>And Sherpa Pak Kidney Transport System (K133694) |
#### Device Description:
The SherpaPak Cardiac Transport System and SherpaPak Kidney Transport System result from modifications made to the cleared Sherpa Pak Cardiac Transport System and Sherpa Pak Kidney Transport System most recently cleared under K133432 and K133694, respectively. The principle of operation of the modified devices is identical to that of the cleared devices, namely, cold static storage and transportation of donor hearts (Cardiac version) or donor kidneys (Kidney version), using cleared preservation solutions specific to the organ type being transported. Further, the modified designs have the same indications for use and utilize the same technology as the predicate devices.
{4}------------------------------------------------
The subject SherpaPak design (cardiac and kidney) like its predecessor, consists of 1) outer transport shipper which contains within it various non-ice based temperature controlled packaging elements, 2) an inner and outer hard shell assembly (i.e. Sherpa Canister/Sherpa Shell) which provides a double, rigid barrier shipper in which the heart/kidney is immersed and suspended in a cold storage solution cleared for use in storing and transporting donor hearts/kidneys, and 3) a data logger that monitors and displays the temperature of the cold storage solution in which the heart/kidney is stored during transport.
The following modifications were made from the predicate to subject devices:
- 1. Change to the exterior shipping container to incorporate insulating capabilities using rigid EPS material,
- 2. Elimination of separate EPS and PIR panels which are no longer necessary to maintain temperature during preservation and transport (reducing system bulk),
- 3. Modification of the Phase Change Material (PCM) cooling packs and configuration used to maintain temperature during preservation and transport,
- 4. Direct measurement of cold storage solution temperature using a temperature probe inserted within the Sherpa Canister,
- 5. Modifications to the Sherpa Canister and Shell to a) incorporate the ability to transmit temperature data from the temperature probe to the data logger and ergonomic changes providing easier handling during use (e.g., handle on the Sherpa Shell lid), and
- 6. Ability to use a mobile app to see temperature during transport.
#### Intended Use:
Organ storage and preservation for transplantation.
#### Indications for Use:
#### SherpaPak Cardiac Transport System
The SherpaPak Cardiac Transport System is intended to be used for the static hypothermic preservation of hearts during transportation and eventual transplantation into a recipient using cold storage solutions indicated for use with the heart.
The intended organ storage time for the SherpaPak Cardiac Transport System is up to 4 hours.
{5}------------------------------------------------
Donor hearts exceeding clinically accepted static hypothermic preservation times should be evaluated by the transplant surgeon to determine transplantability in accordance with accepted clinical guidelines and in the best medical interest of the intended recipient.
## SherpaPak Kidney Transport System
The SherpaPak Kidney Transport System is intended to be used for the static hypothermic preservation of kidney organs during transportation and eventual transplantation into a recipient using cold storage solutions indicated for use with this organ.
The SherpaPak Kidney Transport System can maintain the donor organ storage temperature between 4 C and 8 C through 24 hours.
## Functional Testing:
The premarket notification was submitted to implement modifications to the currently cleared Sherpa Pak Cardiac and Kidney Transport Systems. Testing performed related to the modifications made to the cleared devices included the following:
- . Biocompatibility testing of any new materials that contact the body
- The biocompatibility evaluation for the modified devices was conducted in o accordance with International Standard ISO 10993-1: 2009 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests: Cytotoxicity, Sensitization, Irritation, Acute Systemic toxicity, Material Mediated Pyrogenicity, Hemocompatibility, and Genotoxicity testing
- Electrical Safety and EMC testing in conformance with ●
- EN 60601-1: 2006 +A11:2011 + CORR.1:2006 + CORR.2:2007 + CORR.3: 2010 o MEDICAL ELECTRICAL EQUIPMENT PART 1: GENERAL REQUIREMENTS FOR SAFETY; EN 60601-1-6: 2010 + A1:2014 MEDICAL ELECTRICAL EQUIPMENT PART 1-6: GENERAL REQUIREMENTS FOR BASIC SAFETY AND ESSENTIAL PERFORMANCE - COLLATERAL STANDARD: USABILITY; EN 62366:2007 + A1:2014 MEDICAL DEVICES -APPLICATION OF USABILITY ENGINEERING TO MEDICAL DEVICES
- Transportation Test ●
- The purpose of this test was to evaluate the entire System when subjected to the o rigors of transportation. The System was tested using a modified protocol per ASTM D4169-09. The protocol tested vibration and altitude stresses on the system.
- Thermal qualification .
{6}------------------------------------------------
- The modified SherpaPak System is required to store and transport a donor heart o (or kidney) in cold storage solution between 4°C - 8°C for up to 4 hours (cardiac) and 24 hours kidney). This test was performed to demonstrate the system could meet this specification.
- Helium Leak Testing .
- The purpose of this test was to evaluate the seal integrity of the organ container O (Canister and Shell) that contains the donor organ and the cold storage solution. The Canister and Shell are both provided sterile and are intended to maintain sterility during transport of the donor heart in solution.
## Comparison of Technological Characteristics with the Predicate Device
The following table compares the SherpaPak Cardiac Transport System and SherpaPak Kidney Transport System with the predicated devices.
{7}------------------------------------------------
| Characteristic | Modified SherpaPak Cardiac/Kidney Transport System<br>K180194 | Sherpa Pak Cardiac Transport System Device and Sherpa Pak<br>Kidney Transport System - K133432, K133694, respectively |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Organ storage and preservation for transplantation. | Organ storage and preservation for transplantation. |
| Indications for<br>Use | Cardiac<br>The SherpPak Cardiac Transport System is intended to be used<br>for the static hypothermic preservation of hearts during<br>transportation and eventual transplantation into a recipient using<br>cold storage solution indicated for use with the heart.<br>The intended organ storage time for the Sherpa Pak Cardiac<br>Transport System is up to 4 hours.<br>Donor hearts exceeding clinically accepted static hypothermic<br>preservation times should be evaluated by the transplant surgeon<br>to determine transplantability in accordance with accepted<br>clinical guidelines and in the best medical interest of the<br>intended recipient.<br>Kidney<br>The SherpaPak Kidney Transport System is intended to be used<br>for the static hypothermic preservation of kidney organs during<br>transportation and eventual transplantation into a recipient using<br>cold storage solutions indicated for use with the organ.<br>The Sherpa Pak Kidney Transport System can maintain the | Cardiac<br>The SherpPak Cardiac Transport System is intended to be used for<br>the static hypothermic preservation of hearts during transportation<br>and eventual transplantation into a recipient using cold storage<br>solution indicated for use with the heart.<br>The intended organ storage time for the Sherpa Pak Cardiac<br>Transport System is up to 4 hours.<br>Donor hearts exceeding clinically accepted static hypothermic<br>preservation times should be evaluated by the transplant surgeon to<br>determine transplantability in accordance with accepted clinical<br>guidelines and in the best medical interest of the intended recipient.<br>Kidney<br>The Sherpa Pak Kidney Transport System is intended to be used for<br>the static hypothermic preservation of kidney organs during<br>transportation and eventual transplantation into a recipient using<br>cold storage solutions indicated for use with the organ.<br>The Sherpa Pak Kidney Transport System can maintain the donor<br>organ storage temperature between 4°C and 8°C through 24 hours. |
| | donor organ storage temperature between 4°C and 8°C through<br>24 hours. | |
| Regulation<br>Number | 878.5880 | 878.5880 |
| Product Code | MSB (cardiac), KDN (kidney) | MSB (cardiac), KDN (kidney) |
| Characteristic | Modified SherpaPak Cardiac/Kidney Transport System<br>K180194 | Sherpa Pak Cardiac Transport System Device and Sherpa Pak<br>Kidney Transport System - K133432, K133694, respectively |
| Device<br>Classification<br>Name | Device Classification Name - Isolated kidney perfusion and<br>transport system and accessories | Device Classification Name - Isolated kidney perfusion and transport<br>system and accessories |
| | And | And |
| | Device Classification Name - System & Accessories, Isolated<br>Heart, Transport & Preservation | Device Classification Name - System & Accessories, Isolated Heart,<br>Transport & Preservation |
| Mode of<br>Operation | Static cold ischemic storage | Static cold ischemic storage |
| Meets UNOS<br>Policy 51 | Yes | Yes |
| Organ container | Two rigid airtight containers one of which contains the cold<br>storage solution in which the organ is immersed. | Two rigid airtight containers one of which contains the cold storage<br>solution in which the heart is immersed. |
| | Material - Polycarbonate | Material - Polycarbonate |
| | Inner Canister with temperature probe slot for direct temperature<br>measurement of storage solution. | Outer Shell with slot for temperature probe for indirect temperature<br>measurement of storage solution. |
| Cooling | Temperature preconditioned storage solution and temperature<br>controlled packaging including preconditioned phase change<br>material cold packs. | Temperature preconditioned storage solution and temperature<br>controlled packaging including preconditioned phase change material<br>cold packs, PIR insulating panels, and Expandable Polystyrene<br>panels |
| Characteristic | Modified SherpaPak Cardiac/Kidney Transport System<br>K180194 | Sherpa Pak Cardiac Transport System Device and Sherpa Pak<br>Kidney Transport System - K133432, K133694, respectively |
| System<br>components | Image: Modified SherpaPak Cardiac/Kidney Transport System K180194<br>Outer molded EPS Shipping container with wheels and extending handle PCM Cold Pack Panels Sherpa Pak and Sherpa Pak Shell Temperature data logger Mobile app to monitor temperature | Image: Sherpa Pak Cardiac Transport System Device and Sherpa Pak Kidney Transport System - K133432, K133694, respectively<br>Outer plastic corrugated container (top and base with wheels) PIR insulating panels PCM Cold Pack Panels EPS panels Sherpa Pak and Sherpa Pak Shell Temperature data logger Separate Timer |
| Single<br>Use/Reuse | Entire system is single use/patient only. | Entire system is single use/patient only. |
| Sterilization | Sherpa Pak, Sherpa Pak Shell, and Heart connector (cardiac version) are sterile. All other components are non-sterile. | Sherpa Pak, Sherpa Pak Shell, and Heart connector (cardiac version) are sterile. All other components are non-sterile. |
| Biocompatibility | Direct and indirect heart contact materials have been tested for biocompatibility. | Direct and indirect heart contact materials have been tested for biocompatibility. |
| Characteristic | Modified SherpaPak Cardiac/Kidney Transport System<br>K180194 | Sherpa Pak Cardiac Transport System Device and Sherpa Pak<br>Kidney Transport System - K133432, K133694, respectively |
| Storage<br>Temperature<br>Duration | Device can maintain 4° C to 8° C through 44 hours. | Device can maintain 4° C to 8° C through 36 hours. |
{8}------------------------------------------------
¹ http://www.optn.transplant.hrsa.gov
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
## Conclusion:
The design, intended use, principles of operation, and technological characteristics of the SherpaPak Cardiac Transport System and SherpaPak Kidney Transport System are substantially equivalent to the previously cleared Sherpa Pak Cardiac Transport System and the Sherpa Pak Cardiac Transport System (K133432) and Sherpa Pak Kidney Transport System (K133694), respectively. All of the employed components are very similar to those of the cleared Sherpa Pak devices and utilize the same fundamental scientific technology. The modifications made were such to improve manufacturability and user ergonomics of the system and do not change the devices' intended use. Based upon the descriptive information provided and the testing conducted, the modified SherpaPak Cardiac and Kidney Transport Systems are substantially equivalent to their predicates.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.