SHERPA PAK CARDIAC TRANSPORT SYSTEM

K133432 · Paragonix Technologies, Inc. · MSB · Feb 21, 2014 · Gastroenterology, Urology

Device Facts

Record IDK133432
Device NameSHERPA PAK CARDIAC TRANSPORT SYSTEM
ApplicantParagonix Technologies, Inc.
Product CodeMSB · Gastroenterology, Urology
Decision DateFeb 21, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5880
Device ClassClass 2

Indications for Use

The Sherpa Pak Cardiac Transport System is intended to be used for the static hypothermic preservation of hearts during transportation and eventual transplantation into a recipient using cold storage solutions indicated for use with the heart. The intended organ storage time for the Sherpa Pak Cardiac Transport System is up to 4 hours. Donor hearts exceeding clinically accepted static hypothermic preservation times should be evaluated by the transplant surgeon to determine transplantability in accordance with accepted clinical guidelines and in the best medical interest of the intended recipient.

Device Story

System provides cold ischemic storage and transport for donor hearts. Components include outer shipper with non-ice temperature-controlled packaging; inner and outer hard shell containers for heart immersion/suspension in cold storage fluid; integrated temperature display and timer. Used by transplant teams during organ retrieval and transport. Device maintains hypothermic environment to preserve organ viability. Healthcare providers monitor temperature and elapsed time via display to ensure storage remains within clinically accepted limits. Benefits include consistent, safe transport environment for donor hearts.

Clinical Evidence

No clinical data. Premarket notification supported by bench testing from the original 510(k) clearance (K123326).

Technological Characteristics

System consists of outer shipper, inner/outer hard shell containers, and temperature display/timer. Operates via static hypothermic preservation. No changes to materials or design from predicate.

Indications for Use

Indicated for static hypothermic preservation of donor hearts during transport for transplantation in recipients, using cold storage solutions. Intended storage time up to 4 hours.

Regulatory Classification

Identification

An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ FEB 2 1 2014 . ## Paragonix Sherpa Pak Cardiac Transport System | Submitter: | Paragonix Technologies Inc.<br>c/o Vaughn & Associates<br>639 Granite Street<br>Braintree, MA02184 | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Leo Basta<br>NorthStar Biomedical Associates<br>93 Benefit Street<br>Providence, RI, 02904<br>Phone: 617.834.9866<br>lbasta@northstarbiomedical.com | | Date Prepared: | January 7, 2014 | | Trade Name: | Paragonix Sherpa Pak Cardiac Transport System | | Classification Name: | System & Accessories, Isolated Heart, Transport &<br>Preservation | | Regulation Number: | 21 CFR 876.5880 | | Product Code: | MSB | | Predicate Devices: | Sherpa Pak Cardiac Transport System (K123326)<br>Celsior (K991594) | | Device Description: | The Paragonix Sherpa Pak Cardiac Transport System is a<br>device intended to provide a safe, consistent method for<br>cold ischemic storage and transport of donor hearts to<br>recipients for transplantation. The Sherpa Pak System<br>consists of 1) an outer shipper which contains various non-<br>ice based temperature controlled packaging elements. 2) an<br>inner and outer hard shell container (i.e. Sherpa Pak/Sherpa<br>Pak Shell) which provides a double, rigid barrier container<br>in which the heart is immersed and suspended in a Cold<br>Storage Fluid cleared for use in storing and transporting<br>donor hearts and 3) a temperature display and timer to | Paragonix 510(k) Rev. A · {1}------------------------------------------------ monitor temperature and elapsed time of transport. respectively. The changes made to the currently cleared device are labeling changes only. The device subject to this notification is identical to that cleared under K123326. Intended Use: Organ storage and preservation for transplantation. Indications for Use: The Sherpa Pak Cardiac Transport System is intended to be used for the static hypothermic preservation of hearts during transportation and eventual transplantation into a recipient using cold storage solutions indicated for use with the heart. > The intended organ storage time for the Sherpa Pak Cardiac Transport System is up to 4 hours. > Donor hearts exceeding clinically accepted static hypothermic preservation times should be evaluated by the transplant surgeon to determine transplantability in accordance with accepted clinical guidelines and in the best medical interest of the intended recipient. Functional Testing: The premarket notification was submitted for a labeling change to the device. The device did not change in any respect otherwise. The testing performed in support of the original 510(k) continues to be supportive of the modified device. Summary of Substantial The design, intended use, principles of operation, and Equivalence: technological characteristics of the Sherpa Pak Cardiac Transport System are substantially equivalent to the previously cleared Sherpa Pak Cardiac Transport System. The labeling change does not alter the device's intended use. There were no changes to the device itself and it is identical to the device cleared under K123326. Paragonix 510(k) Rev. A 16 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure embracing a three-lined globe. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 21, 2014 Paragonix Technologies, Inc. % Leo Basta Owner NorthStar Biomedical Associates 93 Benefit Street Providence, RI 02904 Re: K133432 > Trade/Device Name: Paragonix Sherpa Pak Cardiac Transport System Regulation Number: 21 CFR§ 876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: II Product Code: MSB Dated: January 7, 2014 Received: January 10, 2014 Dear Leo Basta. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (sec above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register, Please be advised that FDA 's issuance of a substantial equivalence determination does not mean r heast of acressed a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {3}------------------------------------------------ Page 2 - Leo Basta You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CIDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Sincerely yours, Herbert P. Lerner -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## INDICATION FOR USE 510(k) Number (if known): K133432 Device Name: Paragonix Sherpa Pak Cardiac Transport System Indications for Use: The Sherpa Pak Cardiac Transport System is intended to be used for the static hypothermic preservation of hearts during transportation and eventual transplantation into a recipient using cold storage solutions indicated for use with the heart. The intended organ storage time for the Sherpa Pak Cardiac Transport System is up to 4 hours. Donor hearts exceeding clinically accepted static hypothermic preservation times should be evaluated by the transplant surgeon to determine transplantability in accordance with accepted clinical guidelines and in the best medical interest of the intended recipient. Prescription Use: ____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The Counter Use: (Per 21 CFR 801 Subpart D) ................................................................................................................................................... (Per 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 ## Herbert P. Lerner 2014.02.21 07:07:42 -05'00' Paragonix 510(k) - Response Rev. B
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...