The iTero Lumina Pro NIRI functionality is a diagnostic aid for the detection of interproximal carious lesions above the gingiva and for monitoring the progress of such lesions.
Device Story
Intraoral imaging system captures 3D optical impressions, 2D color images, and Near Infrared (NIR) images. System uses 450nm/520nm lasers for 3D models, white LEDs for 2D color, and 850nm NIR LEDs for tooth structure imaging. Healthy enamel appears dark/translucent; dentin/caries appear bright/opaque. Operated by dental professionals in clinical settings. Software manages GUI, controls scanning cycles, and processes images for display. Enhancements refine post-processing and artifact reduction based on user loupe positioning. Output allows clinicians to visualize interproximal caries, aiding diagnostic decision-making and monitoring. Device uses single-use disposable wand barrier sleeves to prevent cross-contamination.
Clinical Evidence
Bench testing only. Performance attributes evaluated included software functionality, NIR image sharpness, and NIR signal-to-noise ratio. Biocompatibility of patient-contacting materials verified per ISO 10993 standards. Wand barrier sleeve performance verified per 21 CFR 878.4370. All results demonstrated performance equivalent to the predicate device.
Technological Characteristics
Integrated intraoral scanner; 850nm NIR LED illumination; 450nm/520nm laser for 3D. Materials: Aluminum, Copolyester, Glass (wand); HDPE/LDPE (barrier sleeve). Connectivity: USB/cable to computer platform. Software: Firmware and GUI-based application. Sterilization: Single-use disposable barrier sleeve. Software modifications focus on post-processing image optimization.
Indications for Use
Indicated for dental patients requiring detection of interproximal carious lesions above the gingiva and monitoring of lesion progression.
Regulatory Classification
Identification
A laser fluorescence caries detection device is a laser, a fluorescence detector housed in a dental handpiece, and a control console that performs device calibration, as well as variable tone emitting and fluorescence measurement functions. The intended use of the device is to aid in the detection of tooth decay by measuring increased laser induced fluorescence.
Special Controls
*Classification.* Class II, subject to the following special controls:(1) Sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter;
(2) Premarket notifications must include clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
(3) The labeling must include detailed use instructions with precautions that urge users to:
(i) Read and understand all directions before using the device,
(ii) Store probe tips under proper conditions,
(iii) Properly sterilize the emitter-detector handpick before each use, and
(iv) Properly maintain and handle the instrument in the specified manner and condition.
1. restrictions on the sale, distribution, and use of the device to prescription use in accordance with 21 CFR 801.109;
2. clinical studies, or other relevant information, that demonstrates that the device aids in the detection of tooth decay by measuring increased laser induced fluorescence; and
3. labeling that includes detailed use instructions with precautions that urge users to (i) read and understand all directions before using the device, (ii) store probe tips under proper conditions, (iii) properly sterilize the emitter-detector handpiece before each use, and (iv) properly maintain and handle the instrument in the specified manner and condition.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 2, 2026
Align Technology, Ltd.
Aviva Touati
Director Regulatory Affairs
1 Yitzhak Rabin Rd.
Petach Tikva, 4925110
ISRAEL
Re: K261847
Trade/Device Name: iTero Lumina™ Pro
Regulation Number: 21 CFR 872.1745
Regulation Name: Laser Fluorescence Caries Detection Device
Regulatory Class: Class II
Product Code: NTK
Dated: June 1, 2026
Received: June 3, 2026
Dear Aviva Touati:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261847 - Aviva Touati
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261847 - Aviva Touati
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**MICHAEL E. ADJODHA -S**
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261847
Device Name
iTero Lumina™ Pro
Indications for Use (Describe)
The iTero Lumina Pro NIRI functionality is a diagnostic aid for the detection of interproximal carious lesions above the gingiva and for monitoring the progress of such lesions.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
align
[Non-Text]
invisalign®
iTero®
# K261847 - 510(k) SUMMARY
## Submitter
Align Technology Ltd.
1 Yitzhak Rabin Rd.,
Petach Tikva, 4925110,
Israel
Phone: +972544622662
email: atouati@aligntech.com
Contact Person: Aviva Touati
Date Prepared: July 1 \( ^{st} \) , 2026
Name of Device: iTero Lumina Pro™
Common or Usual Name: Caries Detector, Laser Light, Transmission
Classification Name: Laser fluorescence caries detection device
Regulation Number and Class: 21 CFR 872.1745, Class II
Product Code: NTK
Predicate Device: iTero Lumina Pro™ (K240573), Align Technology Ltd.
## Device Description
The modified iTero Lumina Pro is an integrated intraoral imaging system capable of capturing 3D optical impressions for CAD/CAM of dental devices, 2D color images and Near Infrared (NIR) images. Scanning cycles are employed allowing the system to quickly alternate between the three modalities, creating a virtually simultaneous capture of the 3D optical impression, 2D color and NIR images.
The principles of operation of the three modalities are briefly described hereafter.
1) 3D Optical Impression
During the 3D optical impression scan, 450nm (Blue) and 520nm (Green) laser light dot patterns, are operated to create the 3D model.
2) 2D Color Imaging
To capture the 2D color image, the intra-oral cameras are operated while the white LEDs are turned ON.
3) NIR Imaging LEDs with a wavelength of 850nm transmit light through the tooth resulting in an image that translates the teeth structure (enamel and dentin) to different brightness levels. The healthy enamel appears dark and translucent, while dentin or caries lesions appear brighter and opaque.
The first two modalities describe device functions that are subject of regulation 872.3661 which is exempt from 510(k). The NIR Imaging modality is the focus of this submission.
Page 1 of 6
{5}
align
[Non-Text]
invisalign®
iTero®
The modified iTero Lumina Pro consists of the following main components:
A. Scanning Wand used for intraoral scanning.
B. Wand Cradle used to hold the Scanning Wand when not in use.
C. Application Software that includes Graphical User Interface ("GUI").
D. Computer platform
The scanning wand houses electronic, mechanical, and optical components, including projectors, LEDs and cameras which enables capturing 3D impression data as well as 2D color and NIRI images. The wand body is made of Aluminum, Copolyester and Glass. When not in use, it is stored in the Wand Cradle.
The modified iTero Lumina Pro features proprietary software that can be installed on the various configurations of the computer platform. The software is designed to manage the GUI, control the scanning process, display the scanned images, and connect to the stored patient data, including scanned data for display or export.
The modified iTero Lumina scanning wand is designed to be used with a single-use, non-sterile, disposable, wand barrier sleeve. The only patient-contacting materials are those contained within the wand barrier sleeve.
### Indications for Use
The iTero Lumina Pro NIRI functionality is a diagnostic aid for the detection of interproximal carious lesions above the gingiva and for monitoring the progress of such lesions.
### Summary of Technological Characteristics
The modified iTero Lumina Pro and the predicate device consist of the same hardware components including the same hand-held scanning unit and computer platform and involve the same principles of operation. The Substantial Equivalence Comparison Table provides a comparison of the fundamental technological characteristics of the modified iTero Lumina Pro System and those of its predicate device.
Both systems include the same optical components and integrate three optical modalities into one system: 3D digital impressions, 2D color images and NIR images. The devices use similar optical cycles and same illumination power.
The software in both systems consists of a product firmware and a computer-based software. The software in both systems is designed to; manage the graphical user interface (GUI) through which users interact with the system, select scanning modes, and manage scanned data (including sending or storing it), to control the scanning wand according to the selected scanning mode and, to connect to stored data for display or export. The scanning process remains unchanged in terms of hardware operation, optical paths used for acquiring NIR images, and scanning workflow.
Page 2 of 6
{6}
align
[Non-Text]
invisalign®
iTero®
The devices include similar software components designed to obtain and display NIR images for a professional dental health care provider's review. The NIRI modality workflow and user interaction remain consistent compared to the predicate device, performing the same sequence of tasks as previously defined (e.g., patient setup via Rx (prescription) Form filling, scanning, data acquisition, and displayed output (image) review / visualization).
The software modifications introduced in the subject device are intended for enhancement of the software performance, to further refine existing processing capabilities in order to optimize the displayed output based on the user positioning of the loupe while maintaining and adjusting the artifacts reduction feature. The modified software supports the same software functionalities as the predicate device. The software modifications affect only the post-processing step.
The differences in the modified iTero Lumina Pro's software do not alter the underlying scanning technology or system's intended use. The modified iTero Lumina Pro software enhancements do not alter the core functionality or clinical workflow of the device.
Both the subject and predicate devices are designed to be used with a substantially equivalent, single-use Wand Barrier Sleeve. Both barrier sleeves function as a physical barrier covering the wand tip and wand body to reduce the risk of cross-contamination while preserving the ability to obtain high-quality images. The subject and predicate Wand Barrier Sleeves consist of a rigid portion covering the wand tip and a flexible polyethylene sheath portion covering the wand body and overlapping the rigid sleeve, including coverage of the proximal portion of the tip, resulting in continuous barrier integrity.
The subject and predicate Wand Barrier Sleeves differ in (1) configuration as provided to the user, and (2) optical window and adhesive foam materials. These differences were evaluated and determined not to raise new or different questions of safety or effectiveness when compared to the predicate device and were verified using established, well-characterized bench testing methodologies, as described below.
Accordingly, the technological characteristics of the subject device and predicate are substantially equivalent.
Page 3 of 6
{7}
align
[Non-Text]
invisalign®
iTero®
The table below provides a comparison of the fundamental technological characteristics of the iTero Lumina Pro System and those of its predicate device.
| Modified Feature / Characteristic | Modified iTero Lumina Pro K261847 | iTero Lumina Pro K240573 (Predicate device) |
| --- | --- | --- |
| Main Device Components | Hand-held device with USB, cable, software and computer. | Hand-held device with USB, cable, software and computer. |
| Software | The software consists of a product firmware and a computer-based software, which controls the acquisition, selection and display of dental images, and allows interaction of the user with the application- | The software consists of a product firmware and a computer-based software, which controls the acquisition, selection and display of dental images, and allows interaction of the user with the application- |
| Cross contamination control | A single use, disposable, wand barrier sleeve composed of a rigid sleeve that covers the tip of the handpiece and a flexible PE sheath that covers the handpiece body. | A single use, disposable, wand barrier sleeve composed of a rigid sleeve that covers the tip of the handpiece adhered to a flexible PE sheath that covers the handpiece body. |
| Patient-Contacting Materials and Contact Type | Mucosal (Oral) Direct:PE: 70% HDPE – Purell GA7760, 30% LDPE – Purell PE 1840H;Inorganic borosilicate crown optical glass, B270LDPE film FL9106 2 mil thicknessMucosal (Oral)Indirect:3M Transfer Adhesive 1524 | Mucosal (Oral) Direct:PE: 70% HDPE – Purell GA7760, 30% LDPE – Purell PE 1840H;Cast polymethyl methacrylate (PMMA)LDPE film FL9106 2 mil thicknessMucosal (Oral)Indirect:Polyethylene Foam Tape MED 6654R White Polyethylene Foam Tape MED 5696R White |
| Single-Unit Packaging Configuration | Packaged as a single assembled unit in a LDPE zip bag.Each 20 packed single units are packed in a shelf box.Shipping box configuration includes 2 or 8 shelf boxes. | Rigid sleeve packaged in the same LDPE zip bag as the predicate; sheath packaged separately within protective paper and polyethylene covers.Each 20 packed single units of Sheathes and of Rigid Sleeves are packed in the |
Page 4 of 6
{8}
align
[Non-Text]
invisalign®
iTero®
| Modified Feature / Characteristic | Modified iTero Lumina Pro K261847 | iTero Lumina Pro K240573 (Predicate device) |
| --- | --- | --- |
| | | same shelf box as the predicate.The same Shipping boxes are used as the predicate are used for transportation. |
| Modified Feature / Characteristic | Modified iTero Lumina Pro (Subject device) | iTero Lumina Pro K240573 (Predicate device) |
| Main Device Components | Hand-held device with USB, cable, software and computer. | Hand-held device with USB, cable, software and computer. |
| Software | The software consists of a product firmware and a computer-based software, which controls the acquisition, selection and display of dental images, and allows interaction of the user with the application- | The software consists of a product firmware and a computer-based software, which controls the acquisition, selection and display of dental images, and allows interaction of the user with the application- |
| Cross contamination control | A single use, disposable, wand barrier sleeve composed of a rigid sleeve that covers the tip of the handpiece and a flexible PE sheath that covers the handpiece body. | A single use, disposable, wand barrier sleeve composed of a rigid sleeve that covers the tip of the handpiece adhered to a flexible PE sheath that covers the handpiece body. |
| Single-Unit Packaging Configuration | Packaged as a single assembled unit in a LDPE zip bag.Each 20 packed single units are packed in a shelf box.Shipping box configuration includes 2 or 8 shelf boxes. | Rigid sleeve packaged in the same LDPE zip bag as the predicate; sheath packaged separately within protective paper and polyethylene covers.Each 20 packed single units of Sheathes and of Rigid Sleeves are packed in the same shelf box as the predicate.The same Shipping boxes are used as the predicate are used for transportation. |
Page 5 of 6
{9}
align
[LOGO]
invisalign®
iTero®
## Performance Data
To support the substantial equivalence of the modified iTero Lumina Pro to the predicate device, the company has performed bench testing following well-established bench and software testing methods which were based on FDA-recognized voluntary consensus standard and on methods previously used for the clearance of the predicate device, consistent with FDA's Special 510(k) guidance.
For the Near Infrared Imaging (NIR) functionality, key performance attributes tested included: software testing, NIR image sharpness and NIR image signal to noise ratio. All testing showed similar performance to the predicate system.
Testing for the Lumina Pro wand barrier sleeve testing met requirements for dental barriers regulated under 21 C.F.R §878.4370, and demonstrated equivalent performance as the predicate device wand barrier sleeve.
Patient contacting components were shown to be biocompatible for their intended use per the recognized consensus standards in the ISO 10993-series.
## Conclusions
The modified iTero Lumina Pro System is substantially equivalent to the predicate device, the iTero Lumina Pro System. The modified iTero Lumina Pro System has the same intended use and indications for use. Both systems operate based on the same fundamental principles of operation and have substantially equivalent technological characteristics. The minor technological differences between the subject and predicate devices were evaluated using well-established, validated testing methodologies. These evaluations demonstrated that the differences do not raise new or different questions of safety or effectiveness when compared to the predicate device. Performance data supporting that the modified iTero Lumina Pro system is substantially equivalent to the predicate device, as previously summarized.
Based on this assessment, the modified iTero Lumina Pro system is therefore substantially equivalent to the iTero Lumina Pro system cleared under 510(k) number K240573.
Page 6 of 6
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.