Diode Laser Therapy System (PZ-BDT67-01, PZ-BDT67-02)
K261817 · Zhengzhou PZ Laser Slim Technology Co., Ltd. · GEX · Sep 17, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261817
Device Name
Diode Laser Therapy System (PZ-BDT67-01, PZ-BDT67-02)
Applicant
Zhengzhou PZ Laser Slim Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diode Laser Therapy System is intended for use in dermatological procedures requiring coagulation. The indications for use include: The Super Hair Removal (SHR) Mode is intended for temporary hair reduction. The HR Mode is intended for use in dermatological procedures requiring coagulation. The indications for use for the HR Mode include: Benign vascular and vascular dependent lesions.
Device Story
Diode Laser Therapy System (models PZ-BDT67-01, PZ-BDT67-02) is a Class IV laser device for dermatological coagulation and hair reduction. System comprises a host unit (laser power supply, water cooling, control board, display), treatment handle, and foot switch. Operator selects SHR or HR mode; device delivers 755/808/1064nm laser energy to target tissue via sapphire-tipped handpiece. Energy density and pulse duration are adjustable via interface. Used in clinical settings by trained professionals. Output energy induces thermal coagulation for vascular lesions or hair follicle destruction. Bench testing confirms safety and performance per IEC/ISO standards.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60601-2-22, IEC 60825-1), and biocompatibility (ISO 10993-5, -10, -23).
Technological Characteristics
Class IV diode laser (755/808/1064nm); 1000W laser module power; sapphire-tipped handpiece; water circulation cooling system; energy density 3-60 J/cm² (HR) and 3-8 J/cm² (SHR); pulse duration 7-183ms; frequency 1-10Hz. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 10993-5, ISO 10993-10, and ISO 10993-23.
Indications for Use
Indicated for adults requiring dermatological coagulation procedures, including benign vascular/vascular-dependent lesions and temporary hair reduction via SHR mode.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 17, 2026
Zhengzhou PZ Laser Slim Technology Co., Ltd.
Xinye Li
Official Applicant
3rd Floor, Bldg. 1, #101 Jinbai Rd.
High-Tech Development Zone
Zhengzhou, Henan 450001
China
Re: K261817
Trade/Device Name: Diode Laser Therapy System (PZ-BDT67-01, PZ-BDT67-02)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: August 18, 2026
Received: August 18, 2026
Dear Xinye Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K261817 - Xinye Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{3}
K261817 - Xinye Li
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2026.09.17 14:45:53 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261817 | ? |
| Please provide the device trade name(s). | | ? |
| Diode Laser Therapy System | | |
| Please provide your Indications for Use below. | | ? |
| The Diode Laser Therapy System is intended for use in dermatological procedures requiring coagulation. The indications for use include: The Super Hair Removal (SHR) Mode is intended for temporary hair reduction. The HR Mode is intended for use in dermatological procedures requiring coagulation. The indications for use for the HR Mode include: Benign vascular and vascular dependent lesions. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
510(k) Summary
The assigned 510(k) Number: K261817
## **510(k)Summary**
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
1. Date of preparation:May 30,2026
2. Sponsor Identification:
**Zhengzhou PZ Laser Slim Technology Co., Ltd.**
3rd Floor, Building 1, No. 101 Jinbai Road, High-tech Development Zone, 450001
Zhengzhou City, Henan Province, PEOPLE'S REPUBLIC OF CHINA.
Contact Person: Xiaojie Shi
Position: RA
Tel: +86-18703657355
Fax: +86-0371-55677886
Email: shixiaojie1225@163.com
3. Designated Submission Correspondent
**Zhengzhou PZ Laser Slim Technology Co., Ltd.**
3rd Floor, Building 1, No. 101 Jinbai Road, High-tech Development Zone, 450001
Zhengzhou City, Henan Province, PEOPLE'S REPUBLIC OF CHINA.
Contact Person: Xinye Li
Tel: +86-18652732821
Fax: +86-0371-55677886
Email: 316863169@qq.com
1 / 9
{6}
510(k) Summary---
#### 4. Identification of Proposed Device
Trade Name: Diode Laser Therapy System
Models: PZ-BDT67-01, PZ-BDT67-02
Common Name: Powered Laser Surgical Instrument
#### Regulatory Information
Classification Name: Powered Laser Surgical Instrument
Classification: II
Product Code: GEX
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology.
Review Panel: General & Plastic Surgery
#### Intended Use:
The Diode Laser Therapy System is intended for use in dermatological procedures requiring coagulation.
The indications for use include:
The Super Hair Removal (SHR) Mode is intended for temporary hair reduction.
The HR Mode is intended for use in dermatological procedures requiring coagulation. The indications for use for the HR Mode include: Benign vascular and vascular dependent lesions.
#### 5. Device Description:
The Diode Laser Therapy System is mainly composed of one host, treatment handle, power cord and treatment head. The host includes laser power supply, water circulation cooling system, control board and display screen.
There are two models included, PZ-BDT67-01 and PZ-BDT67-02, the two models have the same intended use, mechanism of action and principle. The difference only lies in the step length for adjusting energy density on the instrument interface. The detailed difference shown
2 / 9
{7}
510(k) Summary
as following:
Table 1 The difference of Models
| Item | PZ-BDT67-01 | PZ-BDT67-02 |
| --- | --- | --- |
| Energy density | Step size: 1J/cm\( ^{2} \) | Step size:2J/cm\( ^{2} \) |
## 6. Identification of Predicated Device
510(k) Number: K251329
Trade/Device Name: Diode laser device (RF3120-BI)
Manufacturer: RiSu Medical Technology Co., Ltd.
Identification of Reference Device
510(k) Number: K222064
Trade/Device Name: The Alma Soprano Titanium
Manufacturer: Alma Lasers Inc.
## 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1:2005+A1:2012+A2:2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2014+A1:2020, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic - Requirements and tests;
IEC/TS 60601-4-2:2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems.
IEC 60601-2-22:2019, Medical electrical equipment - Part 2-22: Particular requirements
3 / 9
{8}
510(k) Summary
for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
➤ IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification and requirements
➤ ISO 10993-5:2009, Biological Evaluation of Medical Devices, Part 5-Tests for in vitro cytotoxicity
➤ ISO 10993-10:2021, Biological Evaluation of Medical Devices, Part 10-Tests for skin sensitization
➤ ISO 10993-23:2021, Biological Evaluation of Medical Devices-Part 23: Tests for irritation
# 8. Clinical Test Conclusion
No clinical study is included in this submission.
4 / 9
{9}
510(k) Summary
# 9. Substantially Equivalent (SE) Comparison
Table 2 General Comparison
| Item | Proposed Device | Predicate Device K251329 | Reference Device K222064 |
| --- | --- | --- | --- |
| Device Name | Diode Laser Therapy System | Diode laser device (Trio- Wavelength Handpiece) | Soprano Titanium (Soprano Trio diode laser module) |
| Model | PZ-BDT67-01, PZ-BDT67-02 | RF3120-BI | / |
| Regulation Number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Regulation Medical Specialty | General & Plastic Surgery | General & Plastic Surgery | General & Plastic Surgery |
| Device Class | II | II | II |
| Product Code | GEX | GEX | GEX |
| Intended Use | The Diode Laser Therapy System is intended for use in dermatological procedures requiring coagulation. The indications for use include: The Super Hair Removal (SHR) Mode and SHR Stack Mode are intended for temporary hair reduction. The HR Mode is intended for use in dermatological | The Diode laser device is intended for use in dermatology procedures requiring coagulation. The indications for use for the Trio-Wavelength Handpiece include: The Fast Hair Removal (FHR) Mode is intended for temporary hair reduction. The HR mode is intended for use in dermatology procedures requiring coagulation. The indications for use for the | The Soprano Titanium diode laser module is intended for use in dermatology procedures requiring coagulation. The indications for use for the Soprano Trio diode laser module include: The Super Hair Removal (SHR) Mode is intended for temporary hair reduction. The Soprano Trio diode laser module HR mode is intended for use in |
5 / 9
{10}
510(k) Summary
| | procedures requiring coagulation. The indications for use for the HR Mode include: Benign vascular and vascular dependent lesions. | HR mode include: Benign vascular and vascular dependent lesions. | dermatology procedures requiring coagulation. The indications for use for the Soprano Trio diode laser HR module include: Benign vascular and vascular dependent lesions. |
| --- | --- | --- | --- |
| Configuration | Main Unit | Main unit | Main Unit |
| | Handpiece | Handpiece | Handpiece |
| | Foot switch | Foot pedal | Foot control |
Table 3 Performance Comparison
| Item | Proposed Device | | Predicate Device K251329 | | Reference Device K222064 | |
| --- | --- | --- | --- | --- | --- | --- |
| Laser Type | Diode Laser | | Diode Laser | | Diode Laser | |
| Laser Classification | Class IV | | Class IV | | Class IV | |
| Laser Wavelength | 755+808+1064nm | | 755&808&1064 nm | | Trio-wavelength 755+810+1064nm | |
| Laser Module Power | 1000W | | NA | | 600W/800W/1000W/1200W/1600W | |
| Handpiece tip material | Sapphire | | Gemstone Crystals | | Sapphire | |
| Spot Size | 12*19mm | | 12*20mm | | 4cm² | |
| Energy Density (J/cm²) | SHR | 3-8 | FHR | 2-8 | SHR | 1-8 |
| | HR | 3-60 | HR | 2-60 | HR | 2-120 |
| Frequency | 1Hz-10Hz | | 1Hz-10Hz | | Up to 10Hz | |
6 / 9
{11}
510(k) Summary
| Pulse Duration | 7-183ms | 5-200ms | Up to 200 ms |
| --- | --- | --- | --- |
| Power Supply | AC100-240V;50Hz/60Hz | AC100V-240V,50Hz/60Hz | 230 VAC, 16A, 50/60 Hz |
| Dimension | 620mm*400mm*450mm | 375mm*440mm*1170mm | 120 cm * 39 cm * 38 cm, |
| Weight | 39kg | 52kg | 84kg |
#### Analysis:
The proposed device shares the same intended use, laser type, laser wavelength and pulse frequency as the predicate device (K251329), with substantially consistent spot size and energy density.
#### Laser Module Power
The laser module power of the proposed device is the same as one of the laser module power configurations of the reference device (K222064). Its energy density is also within the parameter range of the reference device (K222064), which evidences that variations in laser module power will not raise new safety concerns for the proposed device. Meanwhile, effectiveness is directly correlated with energy density.
#### Spot Size and Energy Density
The spot size of the proposed device and the Predicate Device(K251329) is nearly identical, and the energy density for their corresponding modes is also substantially equivalent.
#### Pulse Duration
For the difference on pulse duration between the predicate (K251329) and proposed device, we can see that the pulse duration range of proposed device is within the range of the predicate device. There is only a minor difference in pulse duration between the proposed device and the predicate device. And this minor difference will not affect the effectiveness
7/9
{12}
510(k) Summary
and safety.
## Dimension and Weight
The proposed device is different in dimension and weight from the predicate device and the reference device. By complying with IEC 60601-1, the mechanical performance of the proposed device is determined to be accepted, therefore, this difference of dimension and weight has no effect on safety and effectiveness.
Table 4 Safety Comparison
| Item | Proposed Device | Predicate DeviceK251329 | Reference DeviceK222064 |
| --- | --- | --- | --- |
| EMC, Electrical and Laser Safety | | | |
| Electrical Safety | Comply with IEC 60601-1, IEC 60601-2-22 | Comply with IEC 60601-1, IEC 60601-2-22 | Comply with IEC 60601-1, IEC 60601-2-22 |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
| Laser Safety | Comply with IEC 60601-2-22, IEC 60825 | Comply with IEC 60601-2-22, IEC 60825 | Comply with IEC 60601-2-22, IEC 60825 |
| Patient Contact Materials and Biocompatibility | | | |
| Cytotoxicity | Comply with ISO 10993-5 | Comply with ISO 10993-5 | Comply with ISO 10993-1 |
| Sensitization | Comply with ISO 10993-10 | Comply with ISO 10993-10 | |
| Irritation | Comply with ISO 10993-23 | Comply with ISO 10993-23 | |
## 10. Substantially Equivalent (SE) Conclusion
8 / 9
{13}
510(k) Summary
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device (K251329).
9 / 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.