K251329 · Risu Medical Technology Co., Ltd. · GEX · Jun 27, 2025 · General, Plastic Surgery
Device Facts
Record ID
K251329
Device Name
Diode laser device (RF3120-BI)
Applicant
Risu Medical Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 27, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diode laser device is intended for use in dermatology procedures requiring coagulation. The indications for use for the Trio-Wavelength Handpiece include: The Fast Hair Removal (FHR) Mode is intended for temporary hair reduction. The HR mode is intended for use in dermatology procedures requiring coagulation. The indications for use for the HR mode include: Benign vascular and vascular dependent lesions. The 810nm wavelength handpiece is intended for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regimen. Use on all skin types (Fitzpatrick I-VI), including tanned skin.(HR, and FHR Modes).
Device Story
Diode laser device (RF3120-BI) utilizes semiconductor laser technology to emit light for dermatological procedures. System comprises main unit, foot switch, power cord, and two handpieces: one 808nm single-band and one 755nm/808nm/1064nm three-band. Device operates in two modes: HR (Hair Removal/coagulation) and FHR (Fast Hair Removal). HR mode allows adjustable pulse frequency (1-10 Hz) and energy fluence (2-60 J/cm2); FHR mode uses fixed 10 Hz frequency with adjustable fluence (2-8 J/cm2). Operator selects mode via software interface; laser energy is delivered to target tissue to induce thermal effects for coagulation or hair follicle destruction. Used in clinical settings by trained professionals. Benefits include permanent hair reduction and treatment of vascular lesions. Device performance relies on precise energy delivery per unit area.
Clinical Evidence
No clinical data. Bench testing only. Compliance verified via IEC 60601-1, IEC 60825-1, IEC 60601-2-22, IEC 60601-1-2, ISO 10993-5, ISO 10993-10, and ISO 10993-23.
Technological Characteristics
Semiconductor laser (808nm or 755/808/1064nm); PN junction resonant cavity; HR/FHR modes; adjustable pulse frequency (1-10 Hz) and energy fluence (2-60 J/cm2); cooling system; software-controlled user interface; Class II device; electrical safety per IEC 60601-1/2-22; EMC per IEC 60601-1-2; biocompatibility per ISO 10993.
Indications for Use
Indicated for dermatology procedures requiring coagulation, temporary hair reduction (FHR mode), treatment of benign vascular/vascular-dependent lesions (HR mode), and permanent reduction in hair regrowth (810nm handpiece). Applicable to all skin types (Fitzpatrick I-VI), including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 27, 2025
RiSu Medical Technology Co., Ltd.
Tianqiang Yu
Management Representative
Rm 201, Bldg 11, Western Cloud Valley (Phase III)
intersection of Kangding Road and Tongwen Road
Fengxi New City of Xixian New Area, Shaanxi 712000
China
Re: K251329
Trade/Device Name: Diode laser device (RF3120-BI)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: GEX
Dated: April 30, 2025
Received: April 30, 2025
Dear Tianqiang Yu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{2}
K251329 - Tianqiang Yu
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K251329 - Tianqiang Yu
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA
L. HITHE-S
Digitally signed by
TANISHA L. HITHE-S
Date: 2025.06.27
11:06:25 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251329 | |
| Device Name Diode laser device (RF3120-BI) | |
| Indications for Use (Describe) The Diode laser device is intended for use in dermatology procedures requiring coagulation. The indications for use for the Trio-Wavelength Handpiece include: The Fast Hair Removal (FHR) Mode is intended for temporary hair reduction. The HR mode is intended for use in dermatology procedures requiring coagulation. The indications for use for the HR mode include: Benign vascular and vascular dependent lesions. The 810nm wavelength handpiece is intended for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regimen. Use on all skin types (Fitzpatrick I-VI), including tanned skin.(HR, and FHR Modes). | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
510(k) #: K251329
510(k) Summary
Prepared on: 2025-06-27
| Contact Details | 21 CFR 807.92(a)(1) |
| --- | --- |
| Applicant Name | RiSu Medical Technology Co., Ltd. |
| Applicant Address | Rm 201, Bldg 11, Western Cloud Valley (Phase III), The intersection of Kangding Road and Tongwen Road, Fengxi New City of Xixian New Area Shaanxi 712000 China |
| Applicant Contact Telephone | +86-18910677558 |
| Applicant Contact | Mr. Tianqiang Yu |
| Applicant Contact Email | information@believe-med.com |
| Device Name | 21 CFR 807.92(a)(2) |
| Device Trade Name | Diode laser device (RF3120-BI) |
| Common Name | Laser surgical instrument for use in general and plastic surgery and in dermatology |
| Classification Name | Powered Laser Surgical Instrument |
| Regulation Number | 878.4810 |
| Product Code(s) | GEX |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) |
| K222064 | The Alma Soprano Titanium |
| K230580 | Diode Laser System |
| K241951 | Lasya-Trinity |
| Product Code | GEX |
| Device Description Summary | 21 CFR 807.92(a)(4) |
| --- | --- |
| The Diode Laser device includes two handles with different laser band characteristics. One handle contains a single-band 808nm semiconductor laser, and the other handle contains a three-band (755nm/808nm/1064nm) semiconductor laser. The semiconductor lasers of the two handles are powered by the laser power supply in the host to emit laser of corresponding wavelength.
Working principle of semiconductor laser
Semiconductor laser uses semiconductor materials of different doping types as laser working substance, uses natural cleavage surface to form resonant cavity for laser oscillation and amplification, and adds forward voltage to PN junction area of semiconductor laser to form particle number inversion of non-equilibrium carrier between conduction band and valence band of semiconductor substance. When a large number of electrons and holes in particle number inversion state are recombined, excess energy will be released, and these energies will be expressed in the form of photons, that is, laser is formed. Due to resonance amplification of cleavage surface resonant cavity, stimulated feedback is realized, so that laser can be directionally emitted and output from semiconductor laser.
The Diode laser device consists of a main unit, a handheld, a foot switch and a power cord.
Mode Description
1) Trio FHR Mode
The Diode Laser device Trio-Wavelength module with FHR mode is intended for temporary hair reduction.
In this mode, there is a set of parameter setting ranges. The specific parameter ranges are as follows: Frequency is 10 Hz and energy | |
1/3
{6}
fluence is 2-8J/cm2. The operator can adjust the parameters within this range according to the actual situation.
2) Trio HR Mode
The Diode Laser device Trio-Wavelength module with HR mode is intended for use in dermatology procedures requiring coagulation. The indications for use for the HR mode include: Benign vascular and vascular dependent lesions.
In this mode, there is a set of parameter setting ranges. The specific parameter ranges are as follows: Frequency is 1-10 Hz and energy fluence is 2-60J/cm2. The operator can adjust the parameters within this range according to the actual situation.
3) 810nm wavelength
The 810nm wavelength handpiece is intended for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12months after the completion of a treatment regimen. Use on all skin types (Fitzpatrick I-VI)including tanned skin. (HR, and FHR Modes)
HR Mode: In this mode, there is a set of parameter setting ranges. The specific parameter ranges are as follows: Frequency is 1-10 Hz and energy fluence is 2-60J/cm2. The operator can adjust the parameters within this range according to the actual situation.
FHR Mode: In this mode, there is a set of parameter setting ranges. The specific parameter ranges are as follows: Frequency is 10 Hz and energy fluence is 2-8J/cm2. The operator can adjust the parameters within this range according to the actual situation.
The FHR Mode of the two handpiece (810nm& Trio-Wavelength) is the fast hair removal mode. This mode means a quick option for pulse frequency. In the FHR mode, the frequency is fixed at 10Hz and cannot be adjusted. The FHR mode of the 810nm handpiece is used for permanent reduction in hair regrowth defined as a long term, while the FHR mode of the Trio -wavelength handpiece is used for temporary hair reduction.
The HR mode of the two handpiece (810nm& Trio-Wavelength) is designed to differentiate from the FHR mode in terms of parameters. The HR mode covers all pulse frequency ranges (frequency: 1 - 10 Hz), and the pulse frequency can be adjusted. The HR mode of the 810 nm handpiece is used for permanent reduction in hair regrowth defined as a long term, while the FHR mode of the Trio -wavelength handpiece is used for benign vascular and vascular dependent lesions.
4) Operation Differences:
In terms of user operation, the software has two operation modes: HR mode and FHR mode. After startup, there will be a selection interface for these two modes. Users can click on different modes to enter different interfaces. In the HR mode interface, users can adjust the frequency and energy fluence through the up and down buttons. In the FHR mode interface, the frequency cannot be adjusted, and only the energy fluence can be adjusted through the up and down buttons. Other operations are the same for both modes.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
The Diode laser device is intended for use in dermatology procedures requiring coagulation. The indications for use for the Trio-Wavelength Handpiece include: The Fast Hair Removal (FHR) Mode is intended for temporary hair reduction. The HR mode is intended for use in dermatology procedures requiring coagulation. The indications for use for the HR mode include: Benign vascular and vascular dependent lesions.
The 810nm wavelength handpiece is intended for permanent reduction in hair regrowth defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9 and 12 months after the completion of a treatment regimen. Use on all skin types (Fitzpatrick I-VI), including tanned skin.(HR, and FHR Modes).
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The proposed device has the same indications for use with the predicate device.
## Technological Comparison
21 CFR 807.92(a)(6)
The proposed device shares same laser classification, laser module power, laser wavelength, pulse width, pulse frequency, power input and cooling method, but has minor differences in spot size (trio-wavelength handpiece), energy fluence (HR and FHR modes of trio-wavelength handpiece), spot size (808 nm handpiece), pulse frequency (HR mode of 808 nm handpiece), and energy fluence (HR and FHR mode of 808 nm handpiece).
For the difference about spot size (trio-wavelength handpiece):
The spot sizes between the proposed device (12 x 20mm) and the predicted device(2 cm2 and 4 cm2) are very similar.
The difference on spot size would not impact the safety and effectiveness, because the final safety and effectiveness about clinical indications will depends on the amount of energy output per unit area, which would produce thermal effects to the patient's skin area irradiated to achieve claimed indication for use. Therefore, this difference will not affect the safety and effectiveness.
For the difference about energy fluence (HR mode of trio-wavelength handpiece):
The energy fluence of predicate device K222064 is 2-120J/cm2, while the propose device is 2-60J/cm2. It is lower than 120J/cm2, the safety of the product is no problem. However, the effectiveness cannot be determined, so we selected reference device K241951 with the same indication for use for energy fluence comparison. The energy fluence of the reference device is Max 60 J/cm2, which is similar with the energy fluence of the proposed device, and the pulse width and the pulse frequency are comparable to proposed device, so the energy irradiated on the patient's skin is similar with that of proposed device.
Thus the energy fluence of proposed device could achieve its indication for use without effectiveness concerns.
2/3
{7}
Therefore, this difference will not affect the safety and effectiveness.
For the difference about energy fluence (FHR mode of trio-wavelength handpiece):
For the difference on energy fluence between the predicate and proposed device, The energy fluence of predicate device K222064 is 2-20J/cm², while the propose device is 2-8J/cm². It is lower than 20J/cm², the safety of the product is no problem. However, the effectiveness cannot be determined, so we selected reference device K230580 with the same indication for use for energy fluence comparison. The energy fluence of the reference device is 1-8J/cm², which is similar to the energy fluence of the proposed device, and the pulse width and the pulse frequency are comparable to proposed device, so the energy irradiated on the patient's skin is similar to the proposed device.
Thus the energy fluence of proposed could achieve its indication for use without effectiveness concerns.
Therefore, this difference will not affect the safety and effectiveness.
For the difference about spot size (808 nm handpiece):
The spot sizes between the proposed device (12 x 12mm) and the predicted device (1 cm² and 2 cm²) are very similar.
The difference on spot size would not impact the safety and effectiveness, because the final safety and effectiveness about clinical indications will depend on the amount of energy output per unit area, which would produce thermal effects to the patient's skin area irradiated to achieve claimed indication for use. Therefore, this difference will not affect the safety and effectiveness.
For the difference about pulse frequency (HR mode of 808 nm handpiece):
The HR mode pulse frequency of predicate device K222064 is 0.5-3Hz, while the propose device is 1-10Hz.
We selected reference device K230580 with the same indication for use for pulse frequency comparison, the pulse frequency of the propose device is within the frequency range of K230580, so the effectiveness of the product is not affected. Therefore, this difference will not affect the safety and effectiveness.
For the difference about energy fluence (HR mode of 808 nm handpiece):
For the difference on energy fluence between the predicate and proposed device, The fluence of predicate device K222064 is 2-120J/cm², while the propose device is 2-60J/cm². It is lower than 120J/cm², the safety of the product is no problem. However, the effectiveness cannot be determined, so we selected reference device K230580 with the same indication for use for fluence comparison. The energy fluence of the reference device is 1-50J/cm², which is similar to the fluence of the proposed device, and pulse width and pulse frequency are comparable to proposed device, so the energy irradiated on the patient's skin is similar to the proposed device; thus the energy fluence of proposed could achieve its indication for use without effectiveness concerns. Therefore, this difference will not affect the safety and effectiveness.
For the difference about energy fluence (FHR mode of 808 nm handpiece):
The energy fluence of the proposed device is within the range of that of the predicate device, which can justify that the difference in the parameter of energy fluence will not raise new safety issues of the proposed device. And the proposed device has passed the IEC 60601-1 test, IEC 60601-1-2 test, IEC 60601-2-22 test and IEC 60825-1 test, the safety and performance of the product can be ensured. Therefore, this difference will not affect the safety and effectiveness.
## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60601-1 Edition 3.2 2020-08, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
IEC 60825-1:2014, Safety of laser products - Part 1: Equipment classification, and requirements
IEC 60601-2-22: 2012 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
ISO 10993-10 Fourth edition 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
ISO 10993-23 First edition 2021-01, Biological evaluation of medical devices - Part 23: Tests for irritation
ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
IEC 60601-1-2 Edition 4.1 2020-09, Medical Electrical Equipment-Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements And Test
The clinical study is not applicable.
Based on the nonclinical tests performed, the subject device is as safe, as effective, and performance as well as the legally marketed predicate device, the Alma Soprano Titanium cleared under K222064.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.