MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor

K261791 · Steris · FEB · Oct 2, 2026 · Gastroenterology, Urology

Device Facts

Record IDK261791
Device NameMEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor
ApplicantSteris
Product CodeFEB · Gastroenterology, Urology
Decision DateOct 2, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The MEDIVATORS™ ADVANTAGE NXT™ Endoscope Reprocessing System tests, cleans, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The MEDIVATORS™ ADVANTAGE NXT™ Reprocessor is indicated to provide high-level disinfection of heat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is not required prior to placement in the MEDIVATORS™ ADVANTAGE NXT™ Reprocessor. The endoscopes must be pre-cleaned immediately after use. The validated cleaning process replaces cleaning for endoscopes other than duodenoscopes and all endoscopes with an elevator guide wire lever. Manual cleaning of duodenoscopes and all endoscopes with an elevator guide wire lever according to the manufacturer's written instructions for use is required prior to placement. The MEDIVATORS™ ADVANTAGE NXT™ Endoscope Reprocessing System uses RAPICIDE PA High-Level Disinfectant to provide high-level disinfection of endoscopes when used according to the directions for use. The system uses Revital-Ox Detergent in its cleaning cycle at a concentration of 0.5%. RAPICIDE PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is at or above the minimum recommended concentration (MRC) of 850 ppm peracetic acid; this ensures that the disinfectant was above MRC during the entire disinfection cycle. RAPICIDE PA High-Level Disinfectant should be used in the MEDIVATORS™ ADVANTAGE NXT™ Reprocessor under the following contact conditions: - Contact time: 5 minutes - Temperature: 30 °C - MRC: 850 ppm

Device Story

Automated endoscope reprocessor (AER) for cleaning, high-level disinfection, and rinsing of flexible endoscopes; relocatable unit with dual processing basins. Inputs: endoscopes (pre-cleaned at point of use), RAPICIDE PA disinfectant, Revital-Ox detergent, water. Operation: automated leak testing, detergent wash, rinse, and peracetic acid-based high-level disinfection; system monitors pressure, flow, temperature, and fluid levels via integrated sensors. Output: reprocessed endoscope ready for clinical use; cycle documentation via PC interface. Used in clinical settings by healthcare staff. Benefits: standardized, repeatable reprocessing; automated safety interlocks and monitoring ensure effective disinfection; reduces manual labor and variability in cleaning/disinfection process.

Clinical Evidence

Bench testing only. Simulated-use testing demonstrated cleaning efficacy (protein < 6.4 μg/cm², TOC < 12 μg/cm²) and HLD efficacy (≥6 log10 reduction of M. terrae). In-use testing on clinically used endoscopes showed no viable microorganisms recovered. Residue evaluation confirmed chemical residues were below toxicological limits. Cytotoxicity testing (Neutral Red Uptake) confirmed non-cytotoxicity. Electrical safety and software verification passed.

Technological Characteristics

Dual-basin stainless steel system; automated fluid delivery via spray arms and manifolds. Uses peracetic acid-based RAPICIDE PA and Revital-Ox enzymatic detergent. Features HEPA-filtered air purging, automated leak testing, and PC-based control interface. Complies with IEC 60601-1-2 (EMC) and IEC 61010-1 (Safety). Software is C# based. Sterilization/Disinfection: High-level disinfection via automated chemical cycle.

Indications for Use

Indicated for high-level disinfection of heat-sensitive semi-critical endoscopes (fiberoptic and video) and related accessories between patient uses. Requires pre-cleaning immediately after use. Manual cleaning required for duodenoscopes and endoscopes with elevator guide wire levers.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1} **U.S. FOOD & DRUG** ADMINISTRATION October 2, 2026 STERIS Steve Elliott Sr. Manager, Regulatory Affairs 5960 Heisley Road Mentor, Ohio 44060 Re: K261791 Trade/Device Name: MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FEB Dated: May 29, 2026 Received: May 29, 2026 Dear Steve Elliott: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K261791 - Steve Elliott Page 2 When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {3} K261791 - Steve Elliott Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, KATHARINE SEGARS -S Katharine Segars, Ph.D. Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K261791 Device Name MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor Indications for Use (Describe) The MEDIVATORS™ ADVANTAGE NXT™ Endoscope Reprocessing System tests, cleans, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The MEDIVATORS™ ADVANTAGE NXT™ Reprocessor is indicated to provide high-level disinfection of heat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is not required prior to placement in the MEDIVATORS™ ADVANTAGE NXT™ Reprocessor. The endoscopes must be pre-cleaned immediately after use. The validated cleaning process replaces cleaning for endoscopes other than duodenoscopes and all endoscopes with an elevator guide wire lever. Manual cleaning of duodenoscopes and all endoscopes with an elevator guide wire lever according to the manufacturer's written instructions for use is required prior to placement. The MEDIVATORS™ ADVANTAGE NXT™ Endoscope Reprocessing System uses RAPICIDE PA High-Level Disinfectant to provide high-level disinfection of endoscopes when used according to the directions for use. The system uses Revital-Ox Detergent in its cleaning cycle at a concentration of 0.5%. RAPICIDE PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is at or above the minimum recommended concentration (MRC) of 850 ppm peracetic acid; this ensures that the disinfectant was above MRC during the entire disinfection cycle. RAPICIDE PA High-Level Disinfectant should be used in the MEDIVATORS™ ADVANTAGE NXT™ Reprocessor under the following contact conditions: - Contact time: 5 minutes - Temperature: 30 °C - MRC: 850 ppm Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF {5} STERIS® # K261791 # 510(k) Summary # MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Contact: Steve Elliott Senior Manager, Regulatory Affairs Tel: 223-267-9228 Summary Date: October 2, 2026 STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600 {6} STERIS Traditional 510(k) PREMARKET NOTIFICATION - K261791 MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor # 1. Device Name | Trade Name: | MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor | | --- | --- | | Device Classification: | Class II | | Common/usual Name: | Endoscope Cleaner and Reprocessor | | Classification Name: | Endoscope and Accessories | | Classification Number: | 21 CFR 876.1500 | | Product Code: | FEB | # 2. Predicate Device Medivators Advantage Plus Endoscope Reprocessing System, K102996 # 3. Description of Device The MEDIVATORS™ ADVANTAGE NXT™ Endoscope Reprocessing System (AER) is a Class II medical device designed to automate the testing, cleaning, high-level disinfection, and rinsing of flexible fiberoptic and video endoscopes and related accessories between patient uses. The system is a relocatable unit equipped with lockable caster wheels and is intended for use in dedicated endoscope reprocessing rooms. It features two independently operating processing basins, each capable of reprocessing up to two dual-lumen endoscopes or a single multi-lumen device per cycle, enabling efficient and flexible workflow. The AER performs a fully automated reprocessing cycle consisting of sequential phases including leak testing, detergent washing, rinsing and high-level disinfection. Prior to reprocessing, endoscopes are pre-cleaned at the point of use according to manufacturer instructions. Once connected, the system conducts automated pressure leak testing and channel connectivity verification before initiating the cleaning and disinfection process. During operation, the system delivers detergent and disinfectant solutions through both internal channels and external surfaces using spray arms, manifolds, and controlled fluid circulation, while continuously monitoring pressure, temperature, and flow conditions. High-level disinfection is achieved using RAPICIDE™ PA High-Level Disinfectant, a peracetic acid-based chemistry generated within the system by mixing two concentrated components with water immediately prior to use. Following disinfection, the system performs filtered water rinsing and optional alcohol flushing. The device incorporates multiple safety and monitoring features, including automatic alarms, cycle interruption in case of faults, and continuous verification of critical parameters to ensure effective reprocessing. The system is controlled by an integrated PC-based interface that allows operators to monitor cycle status, system parameters, and error conditions in real time. It also October 2, 2026 Page 2 of 9 {7} STERIS Traditional 510(k) PREMARKET NOTIFICATION - K261791 MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor supports documentation and traceability by recording cycle data such as endoscope identification, processing steps, and test results. # 4. Indications for Use The MEDIVATORS™ ADVANTAGE NXT™ Endoscope Reprocessing System tests, cleans, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The MEDIVATORS™ ADVANTAGE NXT™ Reprocessor is indicated to provide high-level disinfection of heat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is not required prior to placement in the MEDIVATORS™ ADVANTAGE NXT™ Reprocessor. The endoscopes must be pre-cleaned immediately after use. The validated cleaning process replaces cleaning for endoscopes other than duodenoscopes and all endoscopes with an elevator guide wire lever. Manual cleaning of duodenoscopes and all endoscopes with an elevator guide wire lever according to the manufacturer's written instructions for use is required prior to placement. The MEDIVATORS™ ADVANTAGE NXT™ Endoscope Reprocessing System uses RAPICIDE PA High-Level Disinfectant to provide high-level disinfection of endoscopes when used according to the directions for use. The system uses Revital-Ox Detergent in its cleaning cycle at a concentration of 0.5%. RAPICIDE PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is at or above the minimum recommended concentration (MRC) of 850 ppm peracetic acid; this ensures that the disinfectant was above MRC during the entire disinfection cycle. RAPICIDE PA High-Level Disinfectant should be used in the MEDIVATORS™ ADVANTAGE NXT™ Reprocessor under the following contact conditions: - Contact time: 5 minutes - Temperature: 30 °C - MRC: 850 ppm October 2, 2026 Page 3 of 9 {8} STERIS Traditional 510(k) PREMARKET NOTIFICATION - K261791 MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor # 5. Technological Characteristic Comparison Tables Table 1. Predicate Device Comparison Table | Feature | Proposed MEDIVATORS™ ADVANTAGE NXT™ AUTOMATED ENDOSCOPE REPROCESSOR | Predicate (K102996) Medivators Advantage Plus Endoscope Reprocessing System | Comparison | | --- | --- | --- | --- | | Indications for Use | The MEDIVATORS™ ADVANTAGE NXT™ Endoscope Reprocessing System tests, cleans, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The MEDIVATORS™ ADVANTAGE NXT™ Reprocessor is indicated to provide high-level disinfection of heat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is not required prior to placement in the MEDIVATORS™ ADVANTAGE NXT™ Reprocessor. The endoscopes must be pre-cleaned immediately after use. The validated cleaning process replaces cleaning for endoscopes other than duodenoscopes and all endoscopes with an elevator guide wire lever. Manual cleaning of duodenoscopes and all endoscopes with an elevator guide wire lever according to the manufacturer's written instructions for use is required prior to placement. The MEDIVATORS™ ADVANTAGE NXT™ Endoscope Reprocessing System uses RAPICIDE PA High-Level Disinfectant to provide high-level disinfection of endoscopes when used according to the directions for use. The system uses Revital-Ox Detergent in its cleaning cycle at a concentration of 0.5%. RAPICIDE PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is at or above the minimum recommended concentration (MRC) of 850 ppm peracetic acid; | Medivators Advantage Plus Endoscope Reprocessing System tests, cleans, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The Advantage Plus system is indicated to provide high level disinfection of heat sensitive semi- critical endoscopes and related accessories. Manual cleaning of endoscopes is not required prior to placement in the Advantage Plus system. The scopes must be precleaned immediately after use . The Advantage Plus Endoscope Reprocessing System uses Rapicide PA High Level Disinfectant to provide high level disinfection of endoscopes when used according to the directions for use. The system uses Intercept Detergent in its cleaning cycle at a concentration of 0.5%. Rapicide PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is above the minimum recommended concentration (MRC) of 850 ppm peracetic acid; this ensures that the disinfectant was above MRC during the entire disinfection cycle. Rapicide PA should be used in the Advantage Plus System under the following contact conditions: Contact Time: 5 minutes Temperature: 30°C MRC: 850 ppm | Similar – both systems provide cleaning and high level disinfection of flexible endoscopes. The subject device limits claims to exclude duodenoscopes from solely using automated cleaning and uses a different detergent for cleaning | October 2, 2026 Page 4 of 9 {9} # STERIS Traditional 510(k) PREMARKET NOTIFICATION - K261791 # MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor | Feature | Proposed MEDIVATORS™ ADVANTAGE NXT™ AUTOMATED ENDOSCOPE REPROCESSOR | Predicate (K102996) Medivators Advantage Plus Endoscope Reprocessing System | Comparison | | --- | --- | --- | --- | | | this ensures that the disinfectant was above MRC during the entire disinfection cycle. RAPICIDE PA High-Level Disinfectant should be used in the MEDIVATORS™ ADVANTAGE NXT™ Reprocessor under the following contact conditions: • Contact time: 5 minutes • Temperature: 30 °C • MRC: 850 ppm | | | | Operating Principles/Technology | • Automates a validated cycle to test, clean, rinse, and high-level disinfect flexible endoscopes. • Cleans internal channels and external surfaces using enzymatic detergent delivered through spray arms and channel manifolds. • Rinses endoscopes in multiple stages using filtered water. • Provides high-level disinfection via single-use, metered dosing of a peracetic acid-based disinfectant under validated temperature and time conditions. Continuously monitors pressure, flow, and temperature throughout the cycle. • Drains disinfectant after use, performs a final rinse, and optionally injects alcohol. • Removes residual moisture using HEPA-filtered air purging. • Uses integrated safety controls, alarms, and disinfectant concentration checks to ensure only properly reprocessed endoscopes are released for patient use. | • Automates a validated cycle to test, clean, rinse, and high-level disinfect flexible endoscopes. • Cleans internal channels and external surfaces using enzymatic detergent delivered through spray arms and channel manifolds. • Rinses endoscopes in multiple stages using filtered water. • Provides high-level disinfection via single-use, metered dosing of a peracetic acid-based disinfectant under validated temperature and time conditions. Continuously monitors pressure, flow, and temperature throughout the cycle. • Drains disinfectant after use, performs a final rinse, and optionally injects alcohol. • Removes residual moisture using HEPA-filtered air purging. • Uses integrated safety controls, alarms, and disinfectant concentration checks to ensure only properly reprocessed endoscopes are released for patient use. | Identical | | Process Parameters | Standardized cycle parameters cannot be altered by operator. The critical process parameters are: • Contact Time: 5 minutes • Temperature: 30 °C • Minimum Recommended Concentration (MRC): 850 ppm peracetic acid | Standardized cycle parameters cannot be altered by operator. The critical process parameters are: • Contact Time: 5 minutes • Temperature: 30 °C • Minimum Recommended Concentration (MRC): 850 ppm peracetic acid | Identical | October 2, 2026 Page 5 of 9 {10} # **STERIS Traditional 510(k) PREMARKET NOTIFICATION - K261791** **MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor** | Feature | Proposed MEDIVATORS™ ADVANTAGE NXT™ AUTOMATED ENDOSCOPE REPROCESSOR | Predicate (K102996) Medivators Advantage Plus Endoscope Reprocessing System | Comparison | | --- | --- | --- | --- | | | - Disinfectant Use: Single use; solution drained after cycle | - Disinfectant Use: Single use; solution drained after cycle | | | **Process Monitors** | - Automated sensors and gauges monitor key parameters such as pressure, flow, temperature, fluid levels, spray arm motion, and lid status during each cycle. - The software provides real-time displays of cycle progress, endoscope status, chemistry availability, time remaining, and warnings or failures. - Alarms, interlocks, and the Safety Control Unit (SCU) automatically stop cycles and alert users when operating limits are exceeded. - effectiveness is verified by controlling dosing, contact time, and temperature, with post-cycle MRC testing using RAPICIDE™ PA test strips. | - Automated sensors and gauges monitor key parameters such as pressure, flow, temperature, fluid levels, spray arm motion, and lid status during each cycle. - The software provides real-time displays of cycle progress, endoscope status, chemistry availability, time remaining, and warnings or failures. - Alarms, interlocks, and the Safety Control Unit (SCU) automatically stop cycles and alert users when operating limits are exceeded. - effectiveness is verified by controlling dosing, contact time, and temperature, with post-cycle MRC testing using RAPICIDE™ PA test strips. | Identical | | **Major Design Features** | - Automated, dual-basin system for cleaning and high-level disinfection of heat-sensitive endoscopes. - Fully enclosed, sensor-controlled unit for consistent, repeatable processing. - Two independent stainless-steel basins with automated lids, spray arms, and channel-specific hookup connections. - Automated dosing and mixing of disinfectant, detergent, and optional alcohol with concentration verification. - Filtered water system with continuous monitoring of temperature, pressure, flow, volume, and leaks. - Touchscreen PC with barcode tracking, role-based access, alarms, and cycle documentation. | - Automated, dual-basin system for cleaning and high-level disinfection of heat-sensitive endoscopes. - Fully enclosed, sensor-controlled unit for consistent, repeatable processing. - Two independent stainless-steel basins with automated lids, spray arms, and channel-specific hookup connections. - Automated dosing and mixing of disinfectant, detergent, and optional alcohol with concentration verification. - Filtered water system with continuous monitoring of temperature, pressure, flow, volume, and leaks. - Touchscreen PC with barcode tracking, role-based access, alarms, and cycle documentation. | Identical | | **Chemistry Uptake** | Weighted, flexible chemical uptake assembly | Rigid tubing uptake | Similar – both draw chemistry. The change in the subject device is intended to be a design | October 2, 2026 Page 6 of 9 {11} # **STERIS Traditional 510(k) PREMARKET NOTIFICATION - K261791** **MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor** | Feature | Proposed MEDIVATORS™ ADVANTAGE NXT™ AUTOMATED ENDOSCOPE REPROCESSOR | Predicate (K102996) Medivators Advantage Plus Endoscope Reprocessing System | Comparison | | --- | --- | --- | --- | | | | | improvement. The design change was validated with performance testing. | | **Operational Software** | LIO application (C# based) | LIO application (Delphi based) | Similar – Both control system functions. Change in subject device is intended to support cybersecurity hardening. The design change was validated with performance testing. | | **Accessories** | | | | | **Disinfectant** | RAPICIDE PA High-Level Disinfectant | RAPICIDE PA High-Level Disinfectant | Identical | | **Cleaning Chemistry** | Revital-Ox™ Enzymatic Detergent | Intercept™ Detergent | Similar – both are cleaning detergents; The subject detergent is enzymatic. The efficacy of the new detergent was validated with performance testing. | | **Endoscope Connectors** | Uses Hookups to attach endoscope lumens to the automated endoscope reprocessor | Uses Hookups to attach endoscope lumens to the automated endoscope reprocessor | Identical | | **Chemical Indicator** | RAPICIDE PA Test Strips | RAPICIDE PA Test Strips | Identical | | **Clinical In-Use** | Disinfection: - Clinically used flexible endoscopes, including bronchoscopes, colonoscopes, gastroscopes, and duodenoscopes after cleaning. - Endoscopes were reprocessed using the standard Sigma cycle and aseptically sampled from both lumens and external surfaces - No viable microorganisms were recovered (0 CFU). | - Testing conducted on clinically used gastroscopes, bronchoscopes, and colonoscopes obtained from healthcare facilities - Testing was conducted evaluating disinfection only and full cycles - No viable microorganisms were recovered from any endoscope following processing. | Similar – Subject submission evaluates a broader range of endoscopes. Current claims validated with performance testing. | October 2, 2026 Page 7 of 9 {12} STERIS Traditional 510(k) PREMARKET NOTIFICATION - K261791 MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor ### 6. Summary of Non-Clinical Testing Shown in Table 2 is the testing that was performed to demonstrate substantial equivalence of the MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor. Table 2. Summary of Verification Activities. | Test | Acceptance Criteria | Result | | --- | --- | --- | | Simulated Use Testing | Cleaning: Worst-case devices were soiled and processed in triplicate trials using the cleaning phase of the processing cycle, then examined visually and sampled for quantitation of two soil markers: protein < 6.4 μg/cm² and TOC < 12 μg/cm². Meets requirements of ANSI AAMI ST98:2022 | PASS | | | High Level Disinfection: Worst case devices were inoculated with Mycobacterium. terrae in triplicate trials and processed using the HLD phase of the processing cycle, then sampled to demonstrate ≥6 log₁₀ reduction per channel of M. terrae. Meets requirements of FDA Guidance: “Content and Format of Premarket Notification [510(k)] Submissions for Liquid Chemical Sterilants/ High Level Disinfectants” | PASS | | In Use Testing | Cleaning: Clinically used devices were placed into the subject device and exposed to a full processing cycle. At the end of the cycle, the devices were examined visually and sampled for quantitation of two soil markers: protein < 6.4 μg/cm² and TOC < 12 μg/cm². Meets applicable requirements of ANSI AAMI ST98:2022 | PASS | | | High Level Disinfection: Clinically used devices were placed into the subject device and exposed to a full processing cycle. At the end of the cycle, the devices were sampled to demonstrate no growth of organisms. Meets requirements of FDA Guidance: “Content and Format of Premarket Notification [510(k)] Submissions for Liquid Chemical Sterilants/ High Level Disinfectants” | PASS | | Residue Evaluation | Endoscope Test Representative worst-case endoscopes (a duodenoscope and a colonoscope) were evaluated after ten consecutive full cleaning and HLD cycles. Residual hydrogen peroxide, peracetic acid, benzotriazole, and tolyltriazole were quantitatively measured. All detected residues on endoscopes were well below defined toxicological allowable limits, with no evidence of accumulation after repeated cycles. Rinse Water Test After the water line disinfection cycle, peracetic acid and hydrogen peroxide levels in final rinse water were consistently below conservative NOAEL-based safety limits | PASS | October 2, 2026 Page 8 of 9 {13} # STERIS Traditional 510(k) PREMARKET NOTIFICATION - K261791 # MEDIVATORS™ ADVANTAGE NXT™ Automated Endoscope Reprocessor | Test | Acceptance Criteria | Result | | --- | --- | --- | | Cytotoxicity | An Olympus TJF-Q180V duodenoscope was subjected to 10 consecutive wash/disinfection cycles at 31–35 °C, then extracted and evaluated using the Neutral Red Uptake cytotoxicity assay. The undiluted extract from the processed endoscope was non-cytotoxic after a 24 ± 2-hour extraction period. Meets requirements of ISO 10993-5 Third edition 2009-06-01 | PASS | | Material Compatibility | Representative devices were exposed to multiple processing cycles and evaluated for physical changes to demonstrate material compatibility with the process and chemistries used. | PASS | | Electrical Safety Conformance | The device was tested against applicable IEC/UL/CSA standards, covering electrical safety, mechanical hazards, thermal performance, fluid handling, and fault conditions, with results demonstrating compliance with applicable requirements. Meets Requirements of IEC 60601-1-2 Edition 4.1 2020-09 and IEC 61010-1 Edition 3.1 2017-01 | PASS | | Software | Testing results, including regression and unit testing, demonstrated that the software meets requirements with no identified safety or performance issues. | PASS | ### 7. Conclusion Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well or better than the legally marketed predicate device (K102996), Class II (21 CFR 876.1500), product code FEB. October 2, 2026 Page 9 of 9
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...