MEDIVATORS ADVANTAGE PLUS ENDOSCOPE REPROCESSOR

K102996 · Medivators Reprocessing Systems · FEB · Jan 13, 2011 · Gastroenterology, Urology

Device Facts

Record IDK102996
Device NameMEDIVATORS ADVANTAGE PLUS ENDOSCOPE REPROCESSOR
ApplicantMedivators Reprocessing Systems
Product CodeFEB · Gastroenterology, Urology
Decision DateJan 13, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K102996 · Jan 13, 2011MEDIVATORS ADVANTAGE PLUS ENDOSCOPE REPROCESSORMedivators Reprocessing SystemsUsed clinical endoscopesUsed clinical endoscopes were evaluated to demonstrate the efficacy of the cleaning cycle, supporting the claim that manual cleaning is not required prior to machine processing.Cleaning efficacy; Clinical scopes; Reprocessing

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Evaluation of used clinical endoscopesUsed clinical endoscopesNot applicable for this studyCleaning efficacy

Indications for Use

Medivators Advantage Plus Endoscope Reprocessing System tests, cleans, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The Advantage Plus system is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is not required prior to placement in the Advantage Plus system. The scopes must be precleaned immediately after use. The Advantage Plus Endoscope Reprocessing System uses Rapicide PA High Level Disinfectant to provide high level disinfection of endoscopes when used according to the directions for use. The system uses Intercept Detergent in its cleaning cycle at a concentration of 0.5%. Rapicide PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is above the minimum recommended concentration (MRC) of 850ppm peracetic acid; this ensures that the disinfectant was above MRC during the entire disinfection cycle. Rapicide PA should be used in the Advantage Plus System under the following contact conditions: Contact Time: 5 minutes Temperature: 30℃ MRC: 850ppm

Device Story

Electro-mechanical system for automated testing, cleaning, high-level disinfection, and rinsing of fiberoptic and video endoscopes. Operates by circulating Rapicide PA (peracetic acid-based) disinfectant and Intercept Detergent (0.5% concentration). Eliminates need for manual cleaning prior to machine placement, provided scopes are precleaned immediately post-use. Used in clinical settings by healthcare personnel. System verifies disinfectant concentration via Rapicide PA Test Strips (MRC 850ppm) post-cycle. Ensures endoscope safety for subsequent patient use.

Clinical Evidence

No clinical data. Evidence consists of bench testing, including simulated use testing with major endoscope brands, testing of used clinical scopes, and evaluation of soil recovery procedures to validate the efficacy of the cleaning cycle.

Technological Characteristics

Electro-mechanical endoscope washer/disinfector. Uses peracetic acid-based high-level disinfectant (Rapicide PA) and detergent (Intercept). Operates at 30°C with 5-minute contact time. Manual cleaning replaced by automated cleaning cycle. No specific materials or connectivity standards provided.

Indications for Use

Indicated for high-level disinfection of heat-sensitive semi-critical fiberoptic and video endoscopes and related accessories between patient uses. Requires precleaning of scopes immediately after use.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K102996 p 1 of 2 # Medivators Advantage Plus Endoscope Reprocessing System ## 510(k) Summary of Safety and Effectiveness Prepared December 27, 2010 JAN 1 3 2011 510(k) Number: K 102996 Manufacturer: Medivators Reprocessing Systems, a Division of Minntech Corporation Address: 14605 28th Avenue North Mpls., MN 55447 USA Official Contact: Lynn Lueders Director, Regulatory Affairs (763) 553-3300 Device Name: Trade or proprietary name: Medivators Advantage Plus Endoscope Reprocessing System Common or usual name: Endoscope washer/disinfector Medivators has supplied the following information to the U.S. Food and Drug Administration to support the substantial equivalence of a cleaning claim for the Advantage Endoscope Reprocessing System to other endoscope reprocessors currently marketed in the U.S. ## 1. Device Description The Advantage AER is an electro-mechanical system intended to test, clean, and high level disinfect fiberoptic and video endoscopes and their related accessories between uses. The Advantage system uses the peracetic acid based Rapicide PA High Level Disinfectant. The AER machine has the capability of cleaning endoscopes as part of the overall cleaning/disinfection cycle of the machine. The users may use the cleaning cycle to replace manual cleaning of the endoscope, however, the users are instructed that they need to preclean the endoscope immediately after use to the SGNA and facility guidelines. ### 2. Intended Use Medivators Advantage Plus Endoscope Reprocessing System tests, cleans, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The Advantage Plus system is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is not required prior to placement in the Advantage Plus system. The scopes must be precleaned immediately after use. {2}------------------------------------------------ K102996 p²of2 The Advantage Plus Endoscope Reprocessing System uses Rapicide PA High Level Disinfectant to provide high level disinfection of endoscopes when used according to the directions for use. The system uses Intercept Detergent in its cleaning cycle at a concentration of 0.5%. Rapicide PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is above the minimum recommended concentration (MRC) of 850ppm peracetic acid; this ensures that the disinfectant was above MRC during the entire disinfection cycle. Rapicide PA should be used in the Advantage Plus System under the following contact conditions: | Contact Time: | 5 minutes | |---------------|-----------| | Temperature: | 30℃ | | MRC: | 850ppm | ## 3. Comparison to Another Device in Commercial Distribution Within the United States The Advantage Plus with a cleaning claim is the same machine as the original Advantage Plus AER. The machine function has not been changed. It has been shown to be equivalent to the Advanced Sterilization Products' EvoTech™ Integrated Endoscope Disinfection System (K061899). The machines have the same indications for use and the same methods of providing disinfection. The ASP EvoTech system and the Advantage system cleans the endoscopes, allowing the placement of the endoscopes into the machine without manual cleaning. ### 4. Summary of Testing Medivators has provided testing to show that the Advantage Endoscope Reprocessing system cleaning cycle is safe and effective for its intended use. Testing was presented to show that the cleaning cycle is efficacious and allows users to place used endoscopes into the machine without prior manual cleaning (users do need to preclean the scopes immediately following use). This testing included simulated use testing with the major brands of endoscopes and testing of used clinical scopes. Testing also included the evaluation of our soil recovery procedures. ### Summary of Substantial Equivalence న. Medivators has provided the above information in the form of a 510(k) to support the claim that the Advantage Endoscope Reprocessing System will clean endoscopes in order to eliminate manual cleaning. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ms. Lynn Lueders Director, Regulatory Affairs Medivators Reprocessing Systems 14605 28th Avenue North Minneapolis, Minnesota 55447 JAN 1 3 2011 Re: K102996 Trade/Device Name: Medivators Advantage Plus Endoscope Reprocessing System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FEB Dated: December 28, 2010 Received: December 29, 2010 Dear Ms. Lueders: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. . {4}------------------------------------------------ Page 2- Ms. Lueders Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.goy/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Jeen I Robets Fez Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ # Indications for Use 510(k) Number (if Known): K 102996 Device Name: Medivators Advantage Plus Endoscope Reprocessing System Indications for Use: Medivators Advantage Plus Endoscope Reprocessing System tests, cleans, disinfects and rinses endoscopes, such as fiberoptic and video endoscopes between patient uses. The Advantage Plus system is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes and related accessories. Manual cleaning of endoscopes is not required prior to placement in the Advantage Plus system. The scopes must be precleaned immediatelv after use. The Advantage Plus Endoscope Reprocessing System uses Rapicide PA High Level Disinfectant to provide high level disinfection of endoscopes when used according to the directions for use. The system uses Intercept Detergent in its cleaning cycle at a concentration of 0.5%. Rapicide PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is above the minimum recommended concentration (MRC) of 850ppm peracetic acid; this ensures that the disinfectant was above MRC during the entire disinfection cycle. Rapicide PA should be used in the Advantage Plus System under the following contact conditions: Contact Time: Temperature: MRC: 5 minutes 30℃ 850ppm Prescription Use AND/OR (Part 21 CFR 801 Subpart D) Over-the Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Evelyth. D. Fleming-Will Concurrence of CDRH, Office of Device Eval Anesthesiology, General Hos; it's Infection Control, Dental Devices KID2996 510(k) Number: _
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...