EVOTECH ECR

K140977 · Advanced Sterilization Products · FEB · Dec 22, 2014 · Gastroenterology, Urology

Device Facts

Record IDK140977
Device NameEVOTECH ECR
ApplicantAdvanced Sterilization Products
Product CodeFEB · Gastroenterology, Urology
Decision DateDec 22, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K140977 · Dec 22, 2014EVOTECH ECRAdvanced Sterilization ProductsClinically used endoscopes (post-procedure)Endoscopes used in clinical procedures were processed through the device to measure residual organic material (protein and TOC) and verify cleaning/disinfection efficacy in a real-world clinical environment.In-use testing; Clinical environment; Residual soil analysis

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
In-use clinical cleaning and disinfection validation; Observational/Performance validation using clinically used devicesEndoscopes used in clinical proceduresNot applicable for this studyResidual protein and total organic carbon (TOC) levels; sterility testing

Indications for Use

The EVOTECH® ECR Endoscope Cleaner & Reprocessor is designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes. High-level disinfection requires that the system be used with CIDEX® OPA Concentrate Solution

Device Story

Automated washer/disinfector for flexible, submersible video or fiber-optic endoscopes; utilizes enzymatic detergent and CIDEX OPA Concentrate high-level disinfectant. Device features two-basin design; processes endoscopes by pumping detergent and disinfectant through lumens and spraying surfaces. System monitors water temperature (50°C), disinfectant concentration, spray arm rotation, and detergent dispensing. Includes leak testing and endoscope connection verification. Onboard printer generates cycle summaries. Safety features cancel cycles and notify users if parameters are not met. Used in clinical settings to automate cleaning/disinfection, reducing manual labor requirements and ensuring standardized reprocessing. Benefits include consistent high-level disinfection and cleaning of heat-sensitive semi-critical endoscopes.

Clinical Evidence

Bench testing only. Simulated use studies demonstrated 6-log10 reduction of Mycobacterium terrae. In-use clinical studies showed no growth after reprocessing. Biocompatibility testing confirmed OPA residuals are within safe limits. Material compatibility testing showed minimal effects over multiple cycles. Non-inferiority trials compared automated wash cycle to SGNA manual cleaning procedures, showing non-inferiority in residual protein and TOC removal (<8.5ug/cm²). Tuberculocidal testing confirmed efficacy of 0.042% OPA concentration at 50-52°C.

Technological Characteristics

Two-basin washer/disinfector. Materials: powder-coated aluminum, glass, stainless steel, polyethylene, Neoprene, polypropylene, polyamide, Ultem, polycarbonate. Sensing: in-line optical monitor for disinfectant concentration, spray arm rotation sensor, detergent dispensing monitor, leak/connection test. Energy: electrical (pump/control). Connectivity: onboard printer for cycle records. Software-controlled. Sterilization: high-level disinfection via chemical process (CIDEX OPA).

Indications for Use

Indicated for use with high-level disinfectant CIDEX® OPA Concentrate and enzymatic detergent (CIDEZYME GI) to achieve cleaning and high-level disinfection of heat-sensitive (>60°C) semi-critical endoscopes. Manual cleaning is not required prior to placement in the system when selecting cycles containing a wash stage; manual cleaning is required when selecting 'Disinfect only' or 'Disinfect/Alcohol Flush' cycles.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 22,2014 Advanced Sterilization Products Mr. Reuben Lawson Sr. Manager, Regulatory Affairs 33 Technology Drive Irvine, California 92618 Re: K140977 Trade/Device Name: EVOTECH® ECR Endoscope Cleaner & Reprocessor Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscopes and Accessories Regulatory Class: II Product Code: FEB Dated: November 24, 2014 Received: November 25, 2014 Dear Mr. Lawson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Mr. Lawson Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Susan Runno DDS, MA Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {3}------------------------------------------------ #### Indications for Use 510(k) Number (if known): K140977 #### Device Name: EVOTECH® ECR Endoscope Cleaner and Reprocessor The EVOTECH® ECR Endoscope Cleaner and Reprocessor, a washer/disinfector, is indicated for use with high-level disinfectant CIDEX® OPA Concentrate and an enzymatic detergent (CIDEZYME GI) to achieve cleaning and high level disinfection of heat sensitive (>60 0C) semi-critical endoscopes. Manual cleaning of medical devices (endoscopes) is not required prior to placement in the EVOTECH® ECR System when selecting those cycles that contain a wash stage. (Manual cleaning of medical devices (endoscopes) is required when selecting the Disinfect only or Disinfect/Alcohol Flush Cycle.) Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 {4}------------------------------------------------ # 510(k) Summary #### Applicant's Name, Address, Telephone, FAX, Contact Person Advanced Sterilization Products, Division of Ethicon, Inc. a Johnson & Johnson company 33 Technology Drive Irvine, CA 92618 #### Contact Person Reuben Lawson Senior Manager, Regulatory Affairs Tel: (949) 789-8545 Fax: (949) 789-3900 #### SUMMARY DATE December 18, 2014 ### 1. CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME | Trade Name: | EVOTECH® ECR Endoscope Cleaner &<br>Reprocessor | |-------------------------|---------------------------------------------------------------| | Common/Usual Name: | Endoscope Cleaner & Reprocessor | | Classification Name: | Endoscope & Accessories (21 CFR 876.1500<br>Product Code FEB) | | Product Classification: | Class II | #### 2. PREDICATE DEVICES - . EVOTECH® Integrated Washer Disinfector (K082392, cleared November 13, 2008) ### 3. DESCRIPTION OF DEVICE The EVOTECH® ECR Endoscope Cleaner & Reprocessor is a two-basin washer/disinfector utilizing an enzymatic detergent and a concentrated high-level disinfectant, CIDEX OPA Concentrate Solution. Both the detergent and high-level disinfectant are diluted in the system to in-use concentrations. The EVOTECH® ECR Endoscope Cleaner & Reprocessor is capable of cleaning endoscopes that have not been manually cleaned prior to placing in the system. {5}------------------------------------------------ Endoscopes that have been qualified for processing in the EVOTECH® ECR Endoscope Cleaner & Reprocessor must go through manual bedside pre-cleaning according to SGNA and facility guidelines prior to being placed in the EVOTECH® ECR Endoscope Cleaner & Reprocessor for automated cleaning and/or high level disinfection. If automatic cleaning is not available or desired for the device by the user, the endoscope must also be manually cleaned by the user prior to being placed in the EVOTECH® ECR Endoscope Cleaner & Reprocessor. After the endoscopes are connected to the EVOTECH® ECR Endoscope Cleaner & Reprocessor, the detergent is sprayed onto the surface of the device, and pumped through the lumens of the device, at sufficient pressure levels to remove any soil present on or in the device. The disinfectant is then mixed with incoming water, which is monitored to ensure that the water temperature is then sprayed onto the surface of the device, and pumped through the lumens of the device, at sufficient time and temperature to achieve high level disinfection of any microorganisms. The temperature is monitored to ensure operating parameters of 50°C are met. A sample of the diluted CIDEX OPA Concentrate solution is extracted to ensure minimum effective concentration levels have been achieved. Following disinfection, the endoscopes are rinsed by the machine and air pumped through the lumens to facilitate drying of the endoscope. If selected by the user, an alcohol flush cycle may also be run to facilitate drying of the endoscope. The endoscope may then be removed, dried and used. The onboard printer generates a cycle printout summarizing the cycle results for the user's records. The machine has built in safety features which cancel the cycle and notify the user if parameters are not met. These notifications, their causes, and next steps for the user. are defined in the EVOTECH® ECR Endoscope Cleaner & Reprocessor User Guide. ## 4. INTENDED USE The EVOTECH® ECR Endoscope Cleaner & Reprocessor is designed to automatically clean and high-level disinfect flexible, submersible video or fiber-optic endoscopes. High-level disinfection requires that the system be used with CIDEX® OPA Concentrate Solution ## 5. INDICATIONS FOR USE The EVOTECH® ECR Endoscope Cleaner and Reprocessor, a washer/disinfector, is indicated for use with high-level disinfectant CIDEX® OPA Concentrate and an enzymatic detergent (CIDEZYME GI) to achieve cleaning and high level disinfection of heat sensitive (>60 0C) semi-critical endoscopes. Manual cleaning of medical devices (endoscopes) is not required prior to placement in the EVOTECH® ECR Endoscope Cleaner & Reprocessor when selecting those cycles that contain a wash {6}------------------------------------------------ stage. (Manual cleaning of medical devices (endoscopes) is required when selecting the Disinfect only or Disinfect/Alcohol Flush Cycle.) ## 6. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS The EVOTECH® ECR Endoscope Cleaner & Reprocessor that is this subject of this submission is the same model as the predicate cleared in K082392. The comparison between the predicate device and the EVOTECH® ECR Endoscope Cleaner & Reprocessor shows the device to meet the definition of substantially equivalence because: - . the devices have the same intended use - and the same technological characteristics ● {7}------------------------------------------------ | Feature | Predicate Device<br>EVOTECH® ECR Endoscope<br>Cleaner and Reprocessor | New Device<br>EVOTECH® ECR Endoscope<br>Cleaner and Reprocessor | |---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------| | Intended Use | The EVOTECH® ECR Endoscope<br>Cleaner and Reprocessor is intended<br>for use as a washer / disinfector for<br>cleaning and high level disinfection of<br>flexible endoscopes that do not contact<br>normally sterile areas of the body. | Same as Predicate Device | | Indications for Use | The EVOTECH® ECR Endoscope<br>Cleaner and Reprocessor, a<br>washer/disinfector, indicated is for use<br>with high-level disinfectant CIDEX®<br>OPA Concentrate and an enzymatic<br>detergent (CIDEZYME GI) to achieve<br>cleaning and high level disinfection of<br>heat sensitive (>60 0C) semi-critical<br>endoscopes. Manual cleaning of<br>medical devices (endoscopes) is not<br>required prior to placement in the<br>EVOTECH System when selecting<br>those cycles that contain a wash stage.<br>(Manual cleaning of medical devices<br>(endoscopes) is required when selecting<br>the Disinfect only or Disinfect/Alcohol<br>Flush Cycle.) | Same as Predicate Device | | Disinfectant Used | CIDEX OPA Concentrate Solution<br>(Single Use) | Same as Predicate Device | | Identification of<br>materials that come<br>into contact with<br>fluid pathways | Endoscopes and materials comprising<br>the EVOTECH® ECR Endoscope<br>Cleaner and Reprocessor: powder<br>coated aluminum glass, stainless steel,<br>polyethylene, Neoprene,<br>polypropylene, polyamide, Ultem,<br>polycarbonate, polypropylene, CIDEX<br>OPA Concentrate (in-use solution) | Same as Predicate Device | | Process Monitors | In line optical monitor for disinfectant<br>concentration, spray arm rotation<br>sensor, detergent dispensing monitor.<br>Same as predicate device with the<br>addition of an endoscope connection<br>test as part of the leak test. | Same as Predicate Device | | Process Parameters | Temperature of rinse water has been<br>increased to 49.5°C (from 45°C).<br>Flow rate in all channels has increased<br>due to the use of larger peristaltic pump | Same as Predicate Device | | Software /<br>firmware controls | Software controlled | Same as Predicate Device | | Cycle comparisons | Cleans endoscopes and achieves high-<br>level disinfection in 5 minutes at a<br>minimum of 50°C. | Same as Predicate Device | | Channels | Six channels for cleaning endoscopes<br>with up to six channels. | Same as Predicate Device | | Accessories | CIDEX® OPA Concentrate high level<br>disinfectant and an enzymatic detergent<br>(CIDEZYME GI) for cleaning medical<br>instruments | Same as Predicate Device | {8}------------------------------------------------ ## 7. SUMMARY OF NONCLINICAL TESTS The EVOTECH® ECR Endoscope Cleaner & Reprocessor is a washer/disinfector for use with an enzymatic detergent and the high-level disinfectant CIDEX® OPA Concentrate Solution at 50° with an exposure time of 5 min to achieve cleaning and high-level disinfection of flexible endoscopes. The EVOTECH® ECR Endoscope Cleaner & Reprocessor was tested in K040883 using the standard array of tests defined in the FDA Guidance on Premarket Notification [510(k)] Submission for Automated Endoscope Washers, Washer / Disinfectors, and Disinfectors Intended for Use in Health Care Facilities, August, 1993. For simulated use testing, the high level disinfectant indicated for use with the EVOTECH® ECR Endoscope Cleaner & Reprocessor was tested for efficacy in a simulated use environment. A 6 Log10 reduction of Mycobacterium terrae was achieved when flexible endoscopes, contaminated with Mycobacterium terrae and artificial soil and were exposed to CIDEX OPA Concentrate In Use solution without cleaning. The diluted CIDEX® OPA Concentrate at an MEC of 0.055% OPA concentration at 50°C is effective against Mycobacterium terrae in artificial soil. For in use testing, endoscopes used in a clinical environment were reprocessed in the EVOTECH® ECR Endoscope Cleaner & Reprocessor without manual cleaning. Sterility testing demonstrated no growth. For biocompatibility, an endoscope reprocessed in the EVOTECH® ECR Endoscope Cleaner & Reprocessor was evaluated for residue levels. The analysis indicates that the level of OPA residual remaining on an endoscope is not likely to cause toxic effects in humans. For material compatibility, the EVOTECH® ECR Endoscope Cleaner & Reprocessor using CIDEX OPA Concentrate In-Use solution was evaluated for its effect on materials commonly used in medical devices. Multiple disinfection cycles over extended periods of time resulted in minimal effect on the test articles. The effects {9}------------------------------------------------ seen were similar to those seen with the predicate device for the high level disinfectant - CIDEX OPA Solution K991487. The EVOTECH® ECR Endoscope Cleaner & Reprocessor was further tested in K061889 to determine its ability to high level disinfect and clean endoscopes. Both simulated use and in-use studies were completed. In both instances, endoscopes were contaminated and then processed in the "wash only" cycle in the EVOTECH® ECR Endoscope Cleaner & Reprocessor. Residual soil was quantified for both protein and total organic carbon (TOC). In all instances the residuals were below the predefined limit of 8.5ug/cm². Additionally, because the wash cycle is intended to eliminate the need for manual cleaning. studies were completed comparing the residual soil after manual cleaning according to the procedures of the Society for Gastroenterology Nurses and Associates (SGNA) and cleaning in the EVOTECH® ECR Endoscope Cleaner & Reprocessor. In a non-inferiority trial, the washing of endoscopes in the EVOTECH® ECR Endoscope Cleaner & Reprocessor was determined to be noninferior to washing of endoscopes following the SGNA procedure. Endoscopes used in clinical procedures were processed through the "wash only" cycle of the EVOTECH® ECR Endoscope Cleaner & Reprocessor and then extracted for residual organic material. Processing in the EVOTECH® ECR Endoscope Cleaner & Reprocessor reduced the residual protein and TOC in all channels and surfaces to less than the predefined acceptance criteria, indicating that the EVOTECH® ECR Endoscope Cleaner & Reprocessor can clean clinically used endoscopes. High-level disinfection of cleaned endoscopes was achieved. Endoscopes were contaminated with Mycobacterium terrae imbedded in soil. Soil was inoculated in channels and on the surfaces of endoscopes so that each endoscopes contained ≥8.5ug/cm² protein with >10' cfu/mL M. terrae. Endoscopes were disinfected in the EVOTECH® ECR Endoscope Cleaner & Reprocessor without manual or automated washing. After processing the contaminated endoscopes through the disinfect cycle only in the EVOTECH® ECR Endoscope Cleaner & Reprocessor, there was a >10° reduction in M. terrae. Further high level disinfection testing and quantitative tuberculocidal testing have been completed for this submission demonstrating effectiveness of the OPAC solution at 0.042% concentration in the presence of 5% fetal bovine serum organic soil load within 5 minutes at 50-52°C using CIDEX OPA-C solution. ## 8. OVERALL PERFORMANCE CONCLUSIONS Data presented in this submission indicate that the EVOTECH® ECR Endoscope Cleaner & Reprocessor is substantially equivalent to the Predicate Device. {10}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...