OtoCompanion software is intended to accurately detect the presence of middle ear fluid (e.g., otitis media with effusion) that may be associated with acute ear infection. The instrument is designed for use by doctors and other medical professionals on pediatric patients (6 months to 21 years).
Device Story
OtoCompanion is a standalone SaMD application for macOS; analyzes high-resolution otoscopic images (JPEG, PNG, TIFF, BMP) from Karl Storz IMAGE1 HD video endoscopy systems. Uses convolutional neural networks to classify images as 'Fluid Present' or 'Dry'; provides visual segmentation overlays for anatomical verification. Used by clinicians in clinical facilities as an adjunctive tool; does not replace clinical judgment. Device operates offline without cloud connectivity. Output assists clinicians in identifying middle ear pathology; supports clinical decision-making for pediatric patients. Benefits include objective, AI-assisted detection of middle ear fluid to aid diagnosis.
Clinical Evidence
Prospective, multi-site, image-based validation study (n=181 ears, 93 pediatric patients). Ground truth: surgical confirmation via myringotomy. Primary analysis (intention-to-diagnose, including 8 low-confidence outputs as errors): 97.8% sensitivity, 88.8% specificity, 93.4% accuracy. Supplementary analysis (excluding low-confidence outputs): 98.9% sensitivity, 95.5% specificity, 97.2% accuracy. Internal algorithm evaluation (n=460 images): 96.09% sensitivity, 96.96% specificity. Consistent performance across age, sex, and race/ethnicity subgroups.
Technological Characteristics
Standalone macOS software application; offline processing. Uses convolutional neural network (ML) algorithm. Inputs: high-resolution still otoscopic images (JPEG, PNG, TIFF, BMP). Outputs: classification ('Fluid Present'/'Dry') and segmentation overlays. Software Safety Class B per IEC 62304.
Indications for Use
Indicated for pediatric patients (6 months to 21 years) to detect middle ear fluid (e.g., otitis media with effusion) associated with acute ear infection. For use by doctors and medical professionals.
Regulatory Classification
Identification
Tyto Insights for Eardrum Bulging Detection is an over-the-counter web-based AI-enabled software analytics device intended to automatically detect the bulging of the eardrum in Oroscopic video recordings acquired by the compatible electronic otoscope, as identified in the product labeling, in patients aged 6 months and older. The device is not intended for diagnosis. A healthcare provider's advice is required to understand the meaning of the Tyto Insights for Eardrum Bulging Detection result. Healthcare providers should consider the device result in conjunction with otoscopic recordings and other relevant patient data. The device is not intended to detect other abnormal otoscopic findings.
Special Controls
In combination with the general controls of the FD&C Act, the ear, nose, and throat image analyzer is subject to the following special controls:
(1) Clinical performance testing must demonstrate the accuracy and precision of the device output to demonstrate that the device performs as intended under anticipated conditions of use in the intended patient population for the stated indications for use.
(2) Human factors/usability testing must demonstrate that the user can correctly use the device, based solely on reading the directions for use.
(3) Software verification, validation, and hazard analysis must be performed.
(4) Labeling must include:
(i) Hardware compatibility information;
(ii) A warning that the device is not intended to be used independently for diagnosis; and
(iii) A summary of the clinical performance testing conducted with the device, including the study endpoints and statistical confidence intervals.
Predicate Devices
Tyto Insights for Eardrum Bulging Detection (DEN250014)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
September 21, 2026
Ironsides Medical, Inc.
% Allison Komiyama
Senior Regulatory Strategist
DuVal & Associates
825 Nicollet Mall
Suite 1820
Minneapolis, Minnesota 55402
Re: K261725
Trade/Device Name: OtoCompanion
Regulation Number: 21 CFR 874.4775
Regulation Name: Ear, Nose, And Throat Image Analyzer
Regulatory Class: Class II
Product Code: SHL
Dated: August 18, 2026
Received: August 18, 2026
Dear Allison Komiyama:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261725 - Allison Komiyama
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K261725 - Allison Komiyama
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JOYCE C.
LIN -S
for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261725
Device Name
OtoCompanion
Indications for Use (Describe)
OtoCompanion software is intended to accurately detect the presence of middle ear fluid (e.g., otitis media with effusion) that may be associated with acute ear infection. The instrument is designed for use by doctors and other medical professionals on pediatric patients (6 months to 21 years).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
| Ironsides Medical Regulatory Document | | |
| --- | --- | --- |
| Document Title | Document Description | Version No. |
| IM-MEMO-0019 | 510(k) Summary | 5 |
## 510(k) Summary
*Pursuant to 21 CFR 807.92*
### 1. Submitter Information
| **Submitter Name** | Ironsides Medical Inc. |
| --- | --- |
| **Address** | 1 Mifflin Place, Suite 400, Cambridge, MA 02138 |
| **Contact Person** | Daphne Psacharopoulos, CEO |
| **Phone / Email** | (617) 513-1849 / daphne@ironsidesmedical.com |
| **Date of Summary** | September 2026 |
| **Consultant** | Allison Komiyama |
### 2. Device Information
| **Trade/Device Name** | OtoCompanion |
| --- | --- |
| **Common Name** | Otoscopic Image Analysis Software |
| **Classification Name** | Ear, nose, and throat image analyzer |
| **Product Code / CFR** | SHL / 21 CFR 874.4775 |
| **Device Class** | Class II |
| **Predicate Device** | Tyto Insights for Eardrum Bulging Detection (DEN250014, Tyto Care Ltd.) |
### 3. Predicate Device
The predicate device is Tyto Insights for Eardrum Bulging Detection (DEN250014), manufactured by Tyto Care Ltd., classified under product code SHL (21 CFR 874.4775, Ear, nose, and throat image analyzer) as a Class II device subject to general controls and the special controls established under the De Novo classification.
### 4. Device Description
OtoCompanion is a standalone software application (Software as a Medical Device, SaMD) that uses convolutional neural networks to analyze otoscopic images and classify them as "Fluid Present" or "Dry" based on the presence or absence of fluid behind the tympanic membrane. The software is intended for qualified medical professionals as an adjunctive tool to support clinical assessment in pediatric patients aged 6 months to 21 years.
The device runs locally on macOS without cloud connectivity, processes high-resolution otoscopic images captured by the Karl Storz IMAGE1 HD video endoscopy system (JPEG, PNG, TIFF, or BMP), and provides visual segmentation overlays that allow clinicians to verify the anatomical basis of each classification. OtoCompanion supplements clinical judgment; it does not replace it. Final clinical decisions remain the clinician's responsibility.
{5}
## 5. Indications for Use
OtoCompanion software is intended to accurately detect the presence of middle ear fluid (e.g., otitis media with effusion) that may be associated with acute ear infection. The instrument is designed for use by doctors and other medical professionals on pediatric patients (6 months to 21 years).
## 6. Technological Characteristics Comparison
Both devices are AI-enabled software-only medical devices that analyze otoscopic imagery of the tympanic membrane to identify findings associated with ear conditions. The technological characteristics are summarized below:
| Characteristic | OtoCompanion (Subject) | Tyto Insights (Predicate) | Comparison |
| --- | --- | --- | --- |
| Intended Use | AI-enabled SaMD to detect middle ear fluid in pediatric patients (6 months to 21 years) as an adjunctive aid | AI-enabled software to detect eardrum bulging in patients aged 6 months and older | Similar. Both analyze otoscopic imagery to identify findings associated with middle ear pathology. Subject detects fluid; predicate detects bulging. |
| Input | High-resolution still otoscopic images (JPEG, PNG, TIFF, or BMP) from the Karl Storz IMAGE1 HD video endoscopy system | Otoscopic video recordings from compatible electronic otoscope | Similar. Both analyze image-level data of the tympanic membrane. |
| Output | "Fluid Present" or "Dry" classification with annotated segmentation overlays | Detection result indicating presence or absence of eardrum bulging | Different. Subject reports middle ear fluid; predicate reports eardrum bulging. Both are AI-generated classifications of tympanic membrane appearance provided as adjunctive information, with the same risks and the same special-control mitigations. The subject output was validated by clinical performance testing against surgical ground truth per special control (1) (Section 7). Does not raise new safety or effectiveness questions. |
| Technology | Machine learning algorithm; standalone macOS application, offline processing | Machine learning algorithm; web-based application | Similar. Both software-only AI/ML devices for tympanic membrane image analysis. |
| Use Environment | Prescription use by clinicians in a clinical facility; adjunctive use labeling | Over-the-counter consumer use; healthcare provider interpretation required | Different. Subject: prescription, clinician-in-the-loop with adjunctive labeling. Predicate: OTC consumer use. Does not raise new safety questions. |
{6}
| Characteristic | OtoCompanion (Subject) | Tyto Insights (Predicate) | Comparison |
| --- | --- | --- | --- |
| Patient Population | Pediatric (6 months to 21 years) | Patients aged 6 months and older | Similar. Subject population is a subset of predicate population. |
The technological differences (input modality, output, software architecture, and use environment) do not raise different questions of safety or effectiveness. Both devices use AI to analyze image-level data of the tympanic membrane for indicators of middle ear pathology, and the risk profile is the same: incorrect results, overreliance on results, and software malfunction — all addressed through the same mitigation measures specified in the De Novo special controls. OtoCompanion meets all special controls of 21 CFR 874.4775, including special control (1), which requires clinical performance testing to demonstrate the accuracy and precision of the device output; this was met by the multi-site clinical performance validation, which validated the Fluid Present / Dry output against surgical ground truth in the intended pediatric population (Section 7).
## 7. Performance Data Summary
### Nonclinical software testing.
Software verification and validation were conducted in accordance with IEC 62304:2006/A1:2016 (Software Safety Class B) and applicable FDA guidance for software in medical devices. All pre-specified software requirements were verified as met.
### Algorithm characterization.
Internal algorithm evaluation on an independent hold-out test set (n = 460 images) demonstrated 96.09% sensitivity (95% CI: 92.7–98.2%) and 96.96% specificity (95% CI: 93.8–98.8%). All pre-specified acceptance criteria were met.
### Multi-site clinical performance validation.
A prospective, multi-site, image-based clinical performance validation study was conducted across six clinical sites in six U.S. states, five of which are institutionally independent of the training data source. The validation cohort included 181 evaluable ears from 93 pediatric patients. Ground truth was established by direct surgical confirmation of middle ear fluid status during myringotomy. Confidence intervals were computed using the Clopper-Pearson exact method.
Of the 181 ears, 173 received a Fluid Present or Dry classification and 8 (4.4%) returned a Low Confidence output. During a Low Confidence output, the device suppresses the classification and prompts the user to recapture the image.
Primary analysis (intention-to-diagnose; all 181 ears included, with the 8 Low Confidence outputs counted as classification errors):
- Sensitivity: \(97.8\%\) (95% CI: 92.4–99.7%)
• Specificity: 88.8% (95% CI: 80.3–94.5%)
• Overall accuracy: 93.4% (95% CI: 88.7–96.5%)
Supplementary analysis (the 8 Low Confidence outputs scored by the model's preliminary classification, which the device does not display):
- Sensitivity: \(98.9\%\) (95% CI: 94.1–100.0%)
• Specificity: 95.5% (95% CI: 88.9–98.8%)
• Overall accuracy: 97.2% (95% CI: 93.7–99.1%)
Performance at the five external (institutionally independent) sites (97.9% combined accuracy, supplementary analysis) was comparable to the primary site (97.0%), supporting generalizability beyond
{7}
the training data source. Subgroup analysis demonstrated consistent performance across age, sex, and race/ethnicity, with every subgroup meeting the \( \geq \) 90% accuracy threshold under both analyses.
### Adverse effects and complications.
The validation study was an image-based performance evaluation. Otoscopic images were obtained during standard-of-care myringotomy procedures; OtoCompanion output did not influence any clinical decision during the study. Given this study design, no device-attributable adverse effects or complications are applicable; none were observed or reported.
## 8. Conclusion
Based on the comparison of technological characteristics and the performance data summarized above, OtoCompanion does not raise different questions of safety or effectiveness compared to the predicate device. The same intended use, similar indications for use, comparable technological characteristics, and demonstrated performance characteristics support a determination that OtoCompanion is substantially equivalent to Tyto Insights for Eardrum Bulging Detection (DEN250014).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.