BIOFIRE® FILMARRAY® Tropical Fever Panel Control Kit
Applicant
Biofire Defense, LLC
Product Code
PMN · Microbiology
Decision Date
Jul 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3920
Device Class
Class 2
Indications for Use
The BIOFIRE® FILMARRAY® Tropical Fever Panel Control Kit contains Positive and Negative External Controls intended for use as assayed quality controls to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of chikungunya virus, dengue virus (serotypes 1, 2, 3, and 4), Leptospira spp., and Plasmodium spp. (including species differentiation of Plasmodium falciparum and Plasmodium vivax/ovale) when using the BIOFIRE® FILMARRAY® Tropical Fever Panel on the BIOFIRE® FILMARRAY® 2.0 and BIOFIRE® FILMARRAY® TORCH Systems. The BIOFIRE® FILMARRAY® Tropical Fever Panel Control Kit is designed for and intended to be used solely with the BIOFIRE® FILMARRAY® Tropical Fever Panel. This product does not replace manufacturer internal controls provided as part of the BIOFIRE® FILMARRAY® Tropical Fever Panel. Both the Positive and Negative External Controls are provided in a Control Injection Vial format. The Positive Control Injection Vial contains dried synthetic nucleic acid segments in buffer and stabilizer to assess for the presence of each individual assay on the BIOFIRE® FILMARRAY® Tropical Fever Panel. The Negative Control Injection Vial is non-reactive with the BIOFIRE® FILMARRAY® Tropical Fever Panel. For In Vitro Diagnostic Use.
Device Story
External quality control kit for BIOFIRE FILMARRAY Tropical Fever (TF) Panel; consists of single-use Control Injection Vials containing dried synthetic RNA segments (positive) or no template (negative). Used in clinical laboratory settings by technicians to monitor assay performance; replaces clinical specimen with molecular grade water. Positive controls assess reverse transcription, amplification, detection, and identification of target pathogens; negative controls confirm lack of cross-reactivity. Instrument software automatically reports results as 'Detected' or 'Not Detected'. Successful runs confirm system integrity; failure indicates potential assay or instrument issues, prompting troubleshooting before clinical testing.
Clinical Evidence
Multi-site reproducibility study conducted across three sites using three control lots, three panel lots, and two instrument types (2.0 and TORCH). Total of 540 runs performed (270 Positive, 270 Negative). Positive control agreement was 99.3% (95% CI: 97.3-99.8%); Negative control agreement was 99.6% (95% CI: 97.9-99.9%). Overall agreement with expected results was 99.4% (95% CI: 98.4-99.8%). No clinical patient data; bench testing only.
Technological Characteristics
External quality control kit; Control Injection Vial format. Positive control: dried synthetic nucleic acid segments in buffer/stabilizer. Negative control: non-reactive. Compatible with BIOFIRE FILMARRAY 2.0 and TORCH systems. Storage: room temperature. Single-use. Monitors reverse transcription, amplification, detection, and identification steps of the associated diagnostic panel.
Indications for Use
Indicated for use as assayed quality controls to monitor performance of in vitro diagnostic nucleic acid testing for chikungunya virus, dengue virus (serotypes 1-4), Leptospira spp., and Plasmodium spp. (including P. falciparum and P. vivax/ovale) using the BIOFIRE FILMARRAY Tropical Fever Panel on BIOFIRE FILMARRAY 2.0 and TORCH systems. Prescription use only.
Regulatory Classification
Identification
An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.
Special Controls
An assayed quality control material for clinical microbiology assays must comply with the following special controls: (1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate: (i) Analyte concentration; Expected values: (ii) Analyte source: (iii) (iv) Base matrix; (v) Added components; (vi) Safety and handling information; and, (vii) Detailed instructions for use. (2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including: (i) Description of the process for value assignment and validation. (ii) Description of the protocol(s) used to establish stability. (iii) Line data establishing precision/reproducibility. (iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance. (v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method. (vi) Where applicable, detailed documentation related to studies for surrogate controls. (3) Premarket notification submissions must include an adequate mitigation (e.g., realtime stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling. (4) Your 21 CFR 809.10 compliant labeling must include the following: (i) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following: (A) Assayed control material analyte(s); (B) Whether the material is intended for quantitative or qualitative assays: (C) Stating if the material is a surrogate control; (D)The system(s), instrument(s), or test(s) for which the quality control material is intended. (ii) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following statement: "This product is not intended to replace manufacturer controls provided with the device." (iii)A limiting statement that reads "Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements."
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate:
(i) Analyte concentration;
(ii) Expected values;
(iii) Analyte source;
(iv) Base matrix;
(v) Added components;
(vi) Safety and handling information; and
(vii) Detailed instructions for use.
(2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including:
(i) Description of the process for value assignment and validation.
(ii) Description of the protocol(s) used to establish stability.
(iii) Line data establishing precision/reproducibility.
(iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance.
(v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method.
(vi) Where applicable, detailed documentation related to studies for surrogate controls.
(3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling.
(4) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following:
(A) Assayed control material analyte(s);
(B) Whether the material is intended for quantitative or qualitative assays;
(C) Stating if the material is a surrogate control; and
(D) The system(s), instrument(s), or test(s) for which the quality control material is intended.
(ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.”
(iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”
Predicate Devices
BioFire Global Fever Panel, BIOFIRE SHIELD Control Kit for the BioFire Global Fever Panel (K220870)
Submission Summary (Full Text)
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
# **510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY**
# **I Background Information:**
# **A 510(k) Number**
K261475
# **B Applicant**
Biofire Defense, LLC
# **C Proprietary and Established Names**
BIOFIRE FILMARRAY Tropical Fever Panel Control Kit
# **D Regulatory Information**
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PMN | Class II | 866.3920 | MI |
# **II Submission/Device Overview:**
# **A Purpose for Submission:**
To obtain a substantial equivalence determination for the BIOFIRE FILMARRAY Tropical Fever Panel Control Kit.
# **B Measurand:**
Multi-analyte external quality control materials for microbiology NAAT assays
# **C Type of Test:**
The BIOFIRE FILMARRAY Tropical Fever (BIOFIRE TF) Panel Control Kit consists of Positive External Controls and Negative External Controls provided in a Control Injection Vial format for direct use with the BIOFIRE TF Panel. The Positive External Controls contain non-infectious RNA segments in buffer and stabilizer embedded on the filter of a Control Injection Vial. The RNA segments are designed to be detected by each of the assays on the BIOFIRE TF
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
Panel. The Negative External Controls are provided as a Control Injection Vial with no template and are non-reactive with the BIOFIRE TF Panel. Each Control Injection Vial is individually packaged and intended for a single use. Vials of Surrogate Sample (molecular grade water) are also provided with the BIOFIRE TF Panel Control Kit which are used in place of a clinical specimen when running the controls.
### III Intended Use/Indications for Use:
#### A Intended Use(s):
See Indications for Use below.
#### B Indication(s) for Use:
The BIOFIRE FILMARRAY Tropical Fever Panel Control Kit contains Positive and Negative External Controls intended for use as assayed quality controls to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of chikungunya virus, dengue virus (serotypes 1, 2, 3, and 4), *Leptospira* spp., and *Plasmodium* spp. (including species differentiation of *Plasmodium falciparum* and *Plasmodium vivax/ovale*) when using the BIOFIRE FILMARRAY Tropical Fever Panel on the BIOFIRE FILMARRAY 2.0 and BIOFIRE FILMARRAY TORCH Systems. The BIOFIRE FILMARRAY Tropical Fever Panel Control Kit is designed for and intended to be used solely with the BIOFIRE FILMARRAY Tropical Fever Panel. This product does not replace manufacturer internal controls provided as part of the BIOFIRE FILMARRAY Tropical Fever Panel.
Both the Positive and Negative External Controls are provided in a Control Injection Vial format. The Positive Control Injection Vial contains dried synthetic nucleic acid segments in buffer and stabilizer to assess for the presence of each individual assay on the BIOFIRE FILMARRAY Tropical Fever Panel. The Negative Control Injection Vial is non-reactive with the BIOFIRE FILMARRAY Tropical Fever Panel.
For In Vitro Diagnostic Use.
#### C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
#### D Special Instrument Requirements:
BIOFIRE FILMARRAY 2.0 and BIOFIRE FILMARRAY TORCH systems
### IV Device/System Characteristics:
#### A Device Description:
The BIOFIRE TF Panel Control Kit is an assayed external quality control consisting of six Positive Controls and six Negative Controls for monitoring the performance of in vitro laboratory nucleic acid procedures when using the BIOFIRE TF Panel. The controls may be used for monitoring the BIOFIRE TF Panel performance when used with BIOFIRE FILMARRAY 2.0 and BIOFIRE FILMARRAY TORCH Systems.
K261475 - Page 2 of 9
{2}
The BIOFIRE TF Panel Control Kit consists of Positive External Controls and Negative External Controls provided in a Control Injection Vial format for direct use with the BIOFIRE TF Panel. The Positive External Controls contain non-infectious RNA segments in buffer and stabilizer embedded on the filter of a Control Injection Vial (Figure 1). The RNA segments are designed to be detected by each of the assays on the BIOFIRE TF Panel. The Negative External Controls are provided as a Control Injection Vial with no template and are non-reactive with the BIOFIRE TF Panel. Each Control Injection Vial is individually packaged and intended for a single use. Vials of Surrogate Sample (molecular grade water) are also provided with the BIOFIRE TF Panel Control Kit which are used in place of a clinical specimen when running the controls.
The BIOFIRE TF Panel Control Kit monitors detection and identification of all pathogens included in by the BIOFIRE TF Panel. A successful Positive External Control run is indicated by all BIOFIRE TF Panel assays reported as “Detected.” A successful Negative External Control run will report all results for the BIOFIRE TF Panel assays as “Not Detected” as shown in Table 1 below.
**Table 1: BIOFIRE TF Panel Control Kit Targets**
| Targets | Expected Result | |
| --- | --- | --- |
| | Positive Control | Negative Control |
| **Viruses** | | |
| Chikungunya virus | Detected | Not Detected |
| Dengue virus | Detected | Not Detected |
| **Parasites** | | |
| *Plasmodium* spp. | Detected | Not Detected |
| *Plasmodium falciparum* | Detected | Not Detected |
| *Plasmodium vivax/ovale* | Detected | Not Detected |
| **Bacteria** | | |
| *Leptospira* spp. | Detected | Not Detected |
## **B Principle of Operation:**
Each positive or negative control included in the BIOFIRE TF Panel Control Kit is processed separately according to the assay manufacturer’s Instructions for Use and monitors the detection and identification of the organisms detected by the BIOFIRE TF Panel. Results are reported automatically by instrument software as either “Detected” or “Not Detected.”
## **V Substantial Equivalence Information:**
### **A Predicate Device Name(s):**
BioFire Global Fever Panel, BIOFIRE SHIELD Control Kit for the BioFire Global Fever Panel
### **B Predicate 510(k) Number(s):**
K220870
K261475 - Page 3 of 9
{3}
# **C Comparison with Predicate(s):**
| Device & Predicate Device(s): | K261475 | K220870 |
| --- | --- | --- |
| Device Trade Name | BIOFIRE FILMARRAY Tropical Fever Panel Control Kit | BIOFIRE SHIELD Control Kit for the BioFire Global Fever Panel |
| **General Device Characteristic Similarities** | | |
| Intended Use/Indications For Use | The BIOFIRE FILMARRAY Tropical Fever Panel Control Kit contains Positive and Negative External Controls intended for use as assayed quality controls to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of chikungunya virus, dengue virus (serotypes 1, 2, 3, and 4), *Leptospira* spp., and *Plasmodium* spp. (including species differentiation of *Plasmodium falciparum* and *Plasmodium vivax/ovale*) when using the BIOFIRE FILMARRAY Tropical Fever Panel on the BIOFIRE FILMARRAY 2.0 and BIOFIRE FILMARRAY TORCH Systems. The BIOFIRE FILMARRAY Tropical Fever Panel Control Kit is designed for and | The BIOFIRE SHIELD Control Kit for the BioFire Global Fever Panel contains Positive and Negative External Controls intended for use as assayed quality controls to monitor the performance of in vitro diagnostic laboratory nucleic acid testing procedures for the qualitative detection of chikungunya virus, dengue virus (serotypes 1, 2, 3, and 4), *Leptospira* spp., and *Plasmodium* spp. (including species differentiation of *Plasmodium falciparum* and *Plasmodium vivax/ovale*) when using the BioFire Global Fever Panel on BioFire FilmArray 2.0 and BioFire FilmArray Torch Systems. The BIOFIRE SHIELD Control Kit |
K261475 - Page 4 of 9
{4}
| | intended to be used solely with the BIOFIRE FILMARRAY Tropical Fever Panel. This product does not replace manufacturer internal controls provided as part of the BIOFIRE FILMARRAY Tropical Fever Panel. Both the Positive and Negative External Controls are provided in a Control Injection Vial format. The Positive Control Injection Vial contains dried synthetic nucleic acid segments in buffer and stabilizer to assess for the presence of each individual assay on the BIOFIRE FILMARRAY Tropical Fever Panel. The Negative Control Injection Vial is non-reactive with the BIOFIRE FILMARRAY Tropical Fever Panel. | for the BioFire Global Fever Panel is designed for and intended to be used solely with the BioFire Global Fever Panel. This product does not replace manufacturer internal controls provided as part of the BioFire Global Fever Panel device. Both the Positive and Negative External Controls are provided in a FilmArray Control Injection Vial format. The Positive Control Injection Vial contains dried synthetic DNA segments in buffer and stabilizer to assess the presence of each individual assay on the BioFire Global Fever Panel. The Negative Control Injection Vial contains no DNA and is non-reactive with the BioFire Global Fever Panel assays. |
| --- | --- | --- |
| Physical format | Same | Control Injection Vial |
| Test System | Same | BIOFIRE FILMARRAY 2.0 or BIOFIRE FILMARRAY TORCH Systems |
K261475 - Page 5 of 9
{5}
| Targets | Same | Chikungunya virus, dengue virus (serotypes 1, 2, 3 and 4), Leptospira spp., and Plasmodium spp. (including species differentiation of Plasmodium falciparum and Plasmodium vivax/ovale). |
| --- | --- | --- |
| General Device Characteristic Differences | | |
| Assay steps monitored | Reverse transcription, Amplification, detection, identification. | Amplification, detection, identification. |
| Composition | Synthetic RNA templates (positive control only) in buffer and stabilizer. | Synthetic DNA segments (positive control only) in buffer and stabilizer. |
## VI Standards/Guidance Documents Referenced:
None
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
#### 1. Precision/Reproducibility:
A multi-site reproducibility study was conducted by testing 270 replicates of BIOFIRE TF Positive control and 270 replicates of BIOFIRE TF Negative control using three lots of each material. Testing with the BIOFIRE TF Panel was conducted across three sites with two operators at each site over five consecutive days and incorporating nine BIOFIRE FILMARRAY 2.0 and ten BIOFIRE FILMARRAY TORCH systems. The agreement between the observed and expected test results for the Positive and Negative Controls across sites is summarized in Table 2.
Table 2. Multi-site Reproducibility Study Results
| Control Type | Observed/Expected (Percent Agreement) [95% CI] | | | |
| --- | --- | --- | --- | --- |
| | Site 1 | Site 2 | Site 3 | All Sites |
K261475 - Page 6 of 9
{6}
| Positive | 90/90 (100%) [95.9 – 100.0] | 89/90 (98.9%) [94.0 – 99.8] | 89/90 (98.9%) [94.0 – 99.8] | 268/270 (99.3%) [97.3 – 99.8] |
| --- | --- | --- | --- | --- |
| Negative | 90/90 (100%) [95.9 – 100.0] | 89/90 (98.9%) [94.0 – 99.8] | 90/90 (100%) [95.9 – 100.0] | 269/270 (99.6%) [97.9 – 99.9] |
Additionally, to further evaluate the reproducibility of the Positive Control measurements, the standard deviation of the melt temperatures for each analyte included in the Positive Control material was determined, as shown in Table 3 below.
Table 3. Summary of Tm (°C) Analyses for BIOFIRE FILMARRAY Tropical Fever Panel Positive Controls
| Panel Target | Panel Assay | Observed Tm (°C) (Mean ± StDev) | | | |
| --- | --- | --- | --- | --- | --- |
| | | Site 1 | Site 2 | Site 3 | All Sites (N=268) |
| Chikungunya virus | CHIKV 1 | 85.2 ± 0.2 | 85.1 ± 0.2 | 85.0 ± 0.3 | 85.1 ± 0.3 |
| | CHIKV 2 | 85.1 ± 0.2 | 84.9 ± 0.3 | 84.9 ± 0.3 | 85.0 ± 0.3 |
| Dengue virus | DENV 1 | 83.3 ± 0.2 | 83.2 ± 0.2 | 83.2 ± 0.3 | 83.2 ± 0.3 |
| | DENV 2_1 | 80.5 ± 0.2 | 80.4 ± 0.3 | 80.3 ± 0.3 | 80.4 ± 0.3 |
| | DENV 2_2 | 81.3 ± 0.3 | 81.2 ± 0.2 | 81.1 ± 0.3 | 81.2 ± 0.3 |
| | DENV 3 | 84.6 ±0.2 | 84.5 ±0.3 | 84.5 ±0.3 | 84.5 ±0.3 |
| | DENV 4 | 83.4 ±0.2 | 83.3 ±0.2 | 83.2 ±0.3 | 83.3 ±0.3 |
| *Leptospira* spp. | Leptospira 1 | 81.9 ± 0.2 | 81.8 ± 0.2 | 81.8 ± 0.3 | 81.8 ± 0.3 |
| *Plasmodium* spp. | *Plasmodium* spp. | 80.4 ± 0.2 | 80.3 ± 0.3 | 80.1 ± 0.3 | 80.3 ± 0.3 |
| | *P. falciparum* | 74.8 ± 0.2 | 74.7 ± 0.3 | 74.6 ± 0.3 | 74.7 ± 0.3 |
| | *P. vivax/ovale* | 74.2 ± 0.2 | 74.1 ± 0.2 | 74.0 ± 0.3 | 74.1 ± 0.3 |
The results support the reproducibility of the BIOFIRE TF Panel Control Kit across three separately manufactured control lots, sites, and operators.
# 2. Linearity:
Not applicable
K261475 - Page 7 of 9
{7}
# 3. Analytical Specificity/Interference:
Not applicable
# 4. Detection Limit and Assay Reportable Range:
Not applicable
# 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Not applicable
# 6. Assay Cut-Off:
Not applicable
# **B Comparison Studies:**
# 1. Method Comparison with Predicate Device:
Not applicable
# 2. Matrix Comparison:
Not applicable
# **C Clinical Studies:**
# 1. Clinical Sensitivity:
Not applicable
# 2. Clinical Specificity:
Not applicable
# 3. Clinical Cut-Off:
Not applicable
# 4. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable
# **D Expected Values/Reference Range:**
K261475 - Page 8 of 9
{8}
The BIOFIRE TF Panel Control Kit is a qualitative control and the expected results for BIOFIRE TF Panel are listed in Tables 4 and 5 below.
**Table 4. BIOFIRE TF Panel Positive Control**
| FilmArray TF Positive Control | Expected Result |
| --- | --- |
| **Viruses** | |
| Chikungunya virus | Detected |
| Dengue virus | Detected |
| **Parasites** | |
| *Plasmodium* spp. | Detected |
| *Plasmodium falciparum* | Detected |
| *Plasmodium vivax/ovale* | Detected |
| **Bacteria** | |
| *Leptospira* spp. | Detected |
**Table 5. BIOFIRE TF Panel Negative Control**
| FilmArray TF Negative Control | Expected Result |
| --- | --- |
| **Viruses** | |
| Chikungunya virus | Not Detected |
| Dengue virus | Not Detected |
| **Parasites** | |
| *Plasmodium* spp. | Not Detected |
| *Plasmodium falciparum* | Not Detected |
| *Plasmodium vivax/ovale* | Not Detected |
| **Bacteria** | |
| *Leptospira* spp. | Not Detected |
### VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
### IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K261475 - Page 9 of 9
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.