These catheters are indicated for the routine drainage of urine from the bladder. These catheters are not intended or designed for indwelling use. Catheters are indicated for use in male and female adults and pediatric patients in the following pediatric subgroups: Children (2 years old to less than 12 years old), Adolescents (12 years old to less than 18 years old), and Transitional Adolescents (18 years old to less than 22 years old but treated like adult).
Device Story
Liquick X-treme and Liquick X-treme Plus are pre-activated hydrophilic-coated intermittent PVC catheters; Plus version includes pre-connected urine bag. Used for routine bladder drainage; not for indwelling use. Operated by patients or clinicians in various settings. Catheter features protective sleeve, introduction aid, and various tip styles (Nelaton, Ergothan, Tiemann). Device facilitates urine drainage via urethral insertion; hydrophilic coating reduces friction during insertion. Single-use, disposable, sterile. Benefits include ease of use and reduced risk of urethral trauma during intermittent catheterization.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), sterilization validation (ISO 11137), package integrity (ISO 11607, ASTM F1980, ASTM D4169), and performance testing (ISO 20696:2018) covering connector security, kink stability, tensile force, flow rate, and friction.
Technological Characteristics
PVC construction; hydrophilic coating; pre-activated with 0.9% sterile saline. Available in 20-43 cm lengths, 6-16 Fr diameters. Tip styles: Nelaton, Ergothan, Tiemann. Sterile, single-use, disposable. Gamma irradiation sterilization. Complies with ISO 10993-1 (biocompatibility), ISO 11137 (sterilization), ISO 11607 (packaging), and ISO 20696 (performance).
Indications for Use
Indicated for routine bladder urine drainage in male and female adults and pediatric patients (ages 2 to <22 years). Not for indwelling use.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{1}

September 23, 2026
Teleflex Medical
Aisling Foley
Senior Manager Regulatory Affairs
Ida Business & Technology Park
Garrycastle, Dublin Rd.
Athlone, Co. Westmeath
IRELAND
Re: K261418
Trade/Device Name: Liquick X-treme, Liquick X-treme Plus
Regulation Number: 21 CFR 876.5130
Regulation Name: Urological Catheter and accessories
Regulatory Class: II
Product Code: EZD
Dated: August 25, 2026
Received: August 25, 2026
Dear Aisling Foley:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K261418 - Aisling Foley
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}
K261418 - Aisling Foley
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANDREW M. FALES -S
for Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261418 | ? |
| Please provide the device trade name(s). | | ? |
| Liquick X-treme, Liquick X-treme Plus | | |
| Please provide your Indications for Use below. | | ? |
| These catheters are indicated for the routine drainage of urine from the bladder. These catheters are not intended or designed for indwelling use.Catheters are indicated for use in male and female adults and pediatric patients in the following pediatric subgroups:Children (2 years old to less than 12 years old),Adolescents (12 years old to less than 18 years old), andTransitional Adolescents (18 years old to less than 22 years old but treated like adult). | | |
| Please select the types of uses (select one or both, as applicable). | Prescription Use (21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | Neonates/Newborns (Birth to < 29 days old)Infants (29 days old to < 2 years old)Children (2 years old to < 12 years old)Adolescents (12 years old to < 22 years old)Adults (22 years old and greater) | ? |
{5}
510(k) Summary
K261418
Page 1 of 6
Liquick X-treme
# 510(k) Summary
A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical Europe Ltd
IDA Business and Technology Park, Dublin Road,
Athlone
Co Westmeath
Ireland
Phone: +353 (0) 9 06 46 08 00
Fax: +353 (0) 14 37 07 73
B. Contact Person
Aisling Foley
Sr. Manager Regulatory Affairs – Product Management
C. Date Prepared
September 23, 2026
D. Device Name
Trade Name: Liquick X-treme
Common Name: Urological Catheter and Accessories
Product Code: EZD
Regulation Number: CFR 876.5130
Classification: II
Classification Panel: Gastroenterology/Urology
E. Predicate Device
This submission demonstrates substantial equivalence to the predicate device, Rusch Flocath Quick Hydrophilic Intermittent catheters, K183461. The predicate device has not been subject to a design-related recall.
Teleflex Medical Europe Ltd.
Page 1 of 6
{6}
K261418
Page 2 of 6
**Liquick X-treme**---
## 510(k) Summary
### F. Device Description
Liquick X-treme consists of a pre-activated hydrophilic-coated intermittent catheter, covered by a protective sleeve. Liquick X-treme Plus is identical to Liquick X-treme, with the addition of a pre-connected urine bag. Device dimensions range from French gauge (Fr.) / Charrière (Ch) Ch06 to Ch16 in outer diameter (OD). The catheter is available in lengths of 20 cm, 30 cm and 43 cm, with an Ergothan, Nelaton or Tiemann tip. Liquick X-treme is manufactured of PVC, provided sterile, single use, and disposable, with proximal funnel and protective sleeve and introduction aid.
### G. Indications for Use
These catheters are indicated for the routine drainage of urine from the bladder. These catheters are not intended or designed for indwelling use.
Catheters are indicated for use in male and female adults and pediatric patients in the following pediatric subgroups:
Children (2 years old to less than 12 years old),
Adolescents (12 years old to less than 18 years old), and
Transitional Adolescents (18 years old to less than 22 years old but treated like adult).
### H. Substantial Equivalence
The subject device is substantially equivalent to the predicate device with respect to intended use, technology and construction. The difference between the predicate and the subject device is minor and any risks have been mitigated through testing. **Table 1** summarizes the similarities and differences between the subject and predicate device.
---Teleflex Medical Europe Ltd.
Page 2 of 6
{7}
K261418
Page 3 of 6
Liquick X-treme
510(k) Summary
1: Comparison of Predicate Device vs. Subject Device
| Features | Teleflex Medical Rusch FloCath Quick Urological Catheter (Predicate Device) | Liquick X-treme & Liquick X-treme Plus (Subject device) | Similarities and differences |
| --- | --- | --- | --- |
| Classification Name | Catheter, Urethral | Urological Catheter and Accessories | Classification name amended by FDA for new marketing applications |
| Regulation Number | 876.5130 | Same | Same |
| FDA Procode | GBM | EZD | GBM code obsoleted by FDA for new marketing applications |
| Class | II | Same | Same |
| Indication for Use | The Rusch Intermittent catheter is a tubular device that is inserted through the urethra and is used to pass urine from the bladder | These catheters are indicated for the routine drainage of urine from the bladder. These catheters are not intended or designed for indwelling use. Catheters are indicated for use in male and female adults and pediatric patients in the following pediatric subgroups: Children (2 years old to less than 12 years old), Adolescents (12 years old to less than 18 years old), and Transitional Adolescents (18 years old to less than 22 years old but treated like adult). | Equivalent, with inclusion of target population. |
| Single Use | Yes | Yes | Same |
| Population | Adult and Pediatric, Male and Female | Adult and Pediatric, Male and Female | Same |
| Size Range | 6-20 Fr. | 6-16 Fr. | The subject devices are within the size range cleared in the predicate |
| Overall Length | 40 cm | 20-43cm | The length of catheter is selected based on patient anatomy. The difference in length does not introduce new risks. |
| Shaft | Tubular | Tubular | Same |
Teleflex Medical Europe Ltd.
Page 3 of 6
{8}
K261418
Page 4 of 6
Liquick X-treme
# 510(k) Summary
| Features | Teleflex Medical Rusch FloCath Quick Urological Catheter (Predicate Device) | Liquick X-treme & Liquick X-treme Plus (Subject device) | Similarities and differences |
| --- | --- | --- | --- |
| **Shaft Material** | PVC | PVC | Same |
| **Coating** | Hydrophilic | Hydrophilic | Same |
| **Biocompatibility** | Complies with ISO 10993-1 | Complies with ISO 10993-1 | Same |
| **Tip** | Nelaton, Olive, or Tiemann tip | Nelaton, Ergothan or Tiemann tip | Equivalent |
| **Eyelets** | Yes | Yes | Same |
| **Liquid for Wetting** | 0.9% Sterile saline solution | 0.9% Sterile saline solution | Same |
| **Primary Packaging** | Paper and film peel back | Aluminium foil pouch | Material difference does not introduce new risks |
| **Sterile** | Yes | Yes | Same |
| **Sterilization Method** | Ethylene Oxide | Gamma Irradiation | Meets requirements. Same SAL 10^{-6} |
| **Shelf-life** | 5 Years | 2 Years | Shelf life within the shelf life of the predicate device. |
# **I. Comparison to the Predicate**
Table 1 illustrates the similarities and differences between the subject device, Liquick X-treme, and the predicate device. The basic technological and operating principles are the same for the devices.
Both the subject device and predicate device are intended for the same patient populations, adult and pediatric, male and female. Both the subject device and predicate device are disposable, single use devices. As evidenced by comparison Table 1, there are no significant differences that would affect safety and efficacy.
Teleflex Medical Europe Ltd.
Page 4 of 6
{9}
510(k) Summary
K261418
Page 5 of 6
Liquick X-treme
### J. Summary of Non-Clinical Testing
- Biocompatibility was conducted per ISO 10993-1– Biological Evaluation of Medical Devices -Part 1: Evaluation and testing within a risk management process:
- Cytotoxicity
- Sensitization
- Irritation
- Acute Systemic Toxicity
- Subacute Systemic Toxicity
- Material Mediated Pyrogenicity
- Sterilization validation was conducted per the following FDA-recognized standards:
- ISO 11137-1 – Sterilization of healthcare products – Radiation – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices.
- ISO 11137-2 – Sterilization of healthcare products – Radiation – Part 2: Establishing the sterilization dose.
- Transportation simulation, package integrity and shelf-life testing was conducted following the requirements of:
- ISO 11607-1:2019, Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems.
- ISO 11607-2:2019, Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes.
- ASTM F1980, Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices
- ASTM D4169-22, Standard Practice for Performance Testing of Shipping Containers and Systems.
- ASTM D4332-14, Standard Practice for Conditioning Containers Packages or Packaging Components for Testing.
- ASTM F1929-15, Standard test method for detecting seal leaks in porous medical packaging by dye penetration.
- ASTM F88/F88M-23, Standard Test Method for Seal Strength of Flexible Barrier Materials.
- ASTM F1886/F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
Teleflex Medical Europe Ltd.
Page 5 of 6
{10}
K261418
Page 6 of 6
510(k) Summary
Liquick X-treme
- Non-clinical Performance and shelf life testing was conducted per ISO 20696:2018, Sterile urethral catheters for single use and ASTM F1980, Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices.
o Connector security was verified by the test method in Annex B of ISO 20696:2018.
o Kink stability was verified by the test method in Annex G of ISO 20696:2018.
o Peak tensile force was verified by the test method in Annex H of ISO 20696:2018.
o Flow rate was verified by the test method in Annex E of ISO 20696:2018.
o Friction was verified by internal test method.
### K. Conclusion
The Liquick X-treme devices have the same intended use and technological characteristics as the predicate. Test results demonstrate that the subject devices meet their intended use and perform as well as the legally marketed predicate device. Any differences in technological characteristics do not raise different questions of safety and effectiveness. The subject devices are substantially equivalent to the predicate device.
Teleflex Medical Europe Ltd.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.