The Rusch Hydrophilic Intermittent catheters (Rusch FloCath Quick, Rusch FloCath Quick Kit, Rusch FloCath Intermittent Catheter, Rusch MMG H2O PVC Catheter Kit, Rusch MMG H2O Singles) are tubular devices that are inserted through the urethra to pass urine from the bladder.
Device Story
Single-use, sterile, disposable PVC intermittent catheters; shaft sizes 6-20 Fr; hydrophilic coating for lubrication. Variants include 'all-in-one' kits and closed systems with integrated collection bags and introducer tips to bypass the first 1.5" of the urethra. Used for bladder drainage; inserted through the urethra. Operated by clinicians or patients. Provides a pathway for urine to exit the bladder; reduces friction during insertion via hydrophilic coating; minimizes contamination risk in closed-system variants. Benefits include improved patient comfort and reduced risk of urethral trauma or infection during intermittent catheterization.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included visual inspection, flow rate, friction testing, and coating presence. Biocompatibility testing conducted per ISO 10993-1, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation and sensitization). Performance verified against ASTM F623 requirements.
Technological Characteristics
Materials: PVC (limited contact <24hr). Sensing/Actuation: Mechanical drainage. Energy: None. Dimensions: 6-20 Fr shaft, 40 cm length. Connectivity: None. Sterilization: Ethylene Oxide. Coating: Hydrophilic. Tip types: Nelaton, Olive, Tiemann, or introducer tip. Standards: ASTM F623, ISO 10993-1, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for adult and pediatric, male and female patients requiring urethral catheterization to pass urine from the bladder. Contraindicated in patients with insurmountable urethral obstructions, urethral injuries, or urethral inflammation.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 8, 2019
Teleflex Medical, Inc. Lori Pfohl Senior Regulatory Affairs Specialist 3015 Carrington Mill Blvd., Suite 600 North Morrisville, NC 27560
Re: K183461
> Trade/Device Name: Rusch Hydrophilic Intermittent catheters (Rusch FloCath Quick, Rusch FloCath Quick Kit, Rusch FloCath Intermittent Catheter, Rusch MMG H2O PVC Catheter Kit, Rusch MMG H2O Singles) Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: GBM Dated: April 5, 2019 Received: April 8, 2019
Dear Lori Pfohl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Acting, Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183461
#### Device Name
Rusch Hydrophilic Intermittent catheters (Rusch FloCath Quick Kit, Rusch FloCath Intermittent Catheter, Rusch MMG H2O PVC Catheter Kit, Rusch MMG H2O Singles).
#### Indications for Use (Describe)
The Rusch Hydrophilic Intermittent catheters (Rusch FloCath Quick, Rusch FloCath Quick Kit, Rusch FloCath Intermittent Catheter, Rusch MMG H2O PVC Catheter Kit, Rusch MMG H2O Singles) are tubular devices that are inserted through the urethra to pass urine from the bladder.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span> <span style="font-size:20px">✖</span> </span> </div> |
|----------------------------------------------|-------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span> </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
#### A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Inc. 3015 Carrington Mill Blvd Morrisville, NC 27560 Phone: 919-433-4831 919-361-3939 Fax:
### B. Contact Person
Lori Pfohl Senior Regulatory Affairs Specialist
### C. Date Prepared
May 7, 2019
### D. Device Name
Trade Name: Rusch Hydrophilic Intermittent catheters (Rusch FloCath Quick, Rusch FloCath Quick Kit, Rusch FloCath Intermittent Catheter, Rusch MMG H2O PVC Catheter Kit, Rusch MMG H2O Singles)
| Common Name: | Catheter, Urethral |
|-----------------------|-------------------------------------|
| Product Code: | GBM |
| Regulation Number: | CFR 876.5130 |
| Classification Name: | Urological catheter and accessories |
| Classification: | II |
| Classification Panel: | Gastroenterology/Urology |
# E. Predicate Device
This submission demonstrates substantial equivalence to the predicate device:
# Rusch® FloCath - K000070
# F. Device Description
The Teleflex Rusch Hydrophilic Intermittent catheters (Rusch FloCath Quick, Rusch FloCath Quick Kit, Rusch FloCath Intermittent Catheter, Rusch MMG H2O PVC Catheter Kit, Rusch MMG H2O Singles) are single use, disposable and sterile. The catheters are made of clear
Teleflex Medical, Inc.
{5}------------------------------------------------
polyvinylchloride (PVC) with vertically cut and softly rounded polished eyes and are coated with a hydrophilic coating. They are composed of a shaft with a proximal funnel and a tip. The funnel is made of PVC. The shaft size ranges from 6-20 French gauge. The funnel is color coded to facilitate size identification. The Rusch® FloCath™ Quick is an "all-in-one catheter" that allows for convenience and the Rusch® MMG H2O™ is a closed system with integrated collection bag that allows for "no touch" catheterization due to an introducer tip that helps bypass the first 1.5" of the urethra.
# G. Indications for Use
The Rusch Hydrophilic Intermittent catheters (Rusch FloCath Quick, Rusch FloCath Quick Kit, Rusch FloCath Intermittent Catheter, Rusch MMG H2O PVC Catheter Kit, Rusch MMG H2O Singles) are tubular devices that are inserted through the urethra to pass urine from the bladder.
# H. Contraindications
- Insurmountable urethral obstructions
- Urethral injuries
- Urethral inflammation
# I. Substantial Equivalence
The subject device Teleflex Rusch FloCath Quick Hydrophilic Intermittent Catheter is substantially equivalent to the predicate devices with respect to intended use, technology and construction. The differences between the predicates and the subject devices are minor and any risks have been mitigated through testing. Table 07.1 summarizes the differences between the subject and predicate device.
The subject devices are substantially equivalent to the predicate device:
| Features | Teleflex Medical<br>Rusch FloCath Quick<br>Urological Catheter<br>(Subject Devices 1) | Teleflex Medical<br>Rusch MMG<br>Urological Catheter<br>(Subject Device 2) | Rusch FloCath - Uncoated or<br>Coated Urological Catheter<br>(Predicate Device-K000070) |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| Classification<br>Name | Urological Catheter<br>and Accessories | Same | Same |
| Regulation<br>Number | 876.5130 | 876.5130 | Same |
| FDA Procode | GBM | GBM | KOD |
| Class | II | II | Same |
| Indication for Use | The Rusch<br>Hydrophilic<br>Intermittent | The Rusch Hydrophilic<br>Intermittent<br>catheters (Rusch | The FloCath catheter is a<br>tubular device that is<br>inserted through the urethra |
| | | | |
| | catheters (Rusch<br>FloCath Quick, Rusch<br>FloCath Quick Kit,<br>Rusch FloCath<br>Intermittent<br>Catheter, Rusch<br>MMG H2O PVC<br>Catheter Kit, Rusch<br>MMG H2O Singles)<br>are tubular devices<br>that are inserted<br>through the urethra<br>to pass urine from<br>the bladder. | FloCath Quick, Rusch<br>FloCath Quick Kit,<br>Rusch FloCath<br>Intermittent Catheter,<br>Rusch MMG H2O PVC<br>Catheter Kit, Rusch<br>MMG H2O Singles)<br>are tubular devices<br>that are inserted<br>through the urethra<br>to pass urine from the<br>bladder. | and is used to pass fluids to<br>or from urinary tracts. |
| Contraindications | - Insurmountable<br>urethral<br>obstructions-<br>Urethral injuries-<br>Urethral<br>inflammation | None | None |
| Single Use | Yes | Yes | Same |
| Population | Adult and Pediatric,<br>Male and Female | Adult and Pediatric,<br>Male and Female | Same |
| Size Range | 6-20 Fr. | 6-20 Fr. | 5-30 Fr. |
| Overall Length | 40 cm | 40cm | Same |
| Shaft | Tubular | Tubular | Same |
| Shaft Material | PVC (Mucosal and<br>skin, limited contact<br><24hr) | PVC (Mucosal and<br>skin, limited contact<br><24hr) | Same |
| Coating | Hydrophilic | Hydrophilic | Uncoated or Hydrogel/<br>Hydrophilic Coating |
| Biocompatibility | Evaluation and<br>testing within the<br>risk management<br>process ISO 10993-1:<br>Cytotoxicity ISO<br>10993-5<br>Sensitization ISO<br>10993-10<br>Skin Irritation ISO<br>10993-10 | Evaluation and testing<br>within the risk<br>management process<br>ISO 10993-1:<br>Cytotoxicity ISO<br>10993-5<br>Sensitization ISO<br>10993-10<br>Skin Irritation ISO<br>10993-10 | Same |
| Tip | Nelaton, Olive, or<br>Tiemann tip | Split introducer,<br>introducer tip, and<br>cap | Nelaton, Olive, or Tiemann<br>tip |
| Eyelets | Yes | Yes | Yes |
| | solution | solution | |
| Primary Packaging | Paper and film peel<br>back | Paper and film peel<br>back | Same |
| Sterile | Yes | Yes | Same |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| Shelf-life | 5 Years | 5 Years | Same |
# Table 07.1 - Comparison of Predicate vs. Subject Devices
{6}------------------------------------------------
Teleflex Medical, Inc.
{7}------------------------------------------------
### J. Comparison to the Predicate
Table 7.1 illustrates the similarities and differences between the subject and predicate device (K000070). The basic technological and operating principles are the same for both devices. Both the predicate and subject devices have same intended use.
Both the subject and predicate devices are intended for similar patient populations adult and pediatric, male and female. Both the subject and predicate devices are disposable, non-sterile, single patient use devices. As evidenced by comparison Table 7.1, above, the subject FloCath Quick Hydrophilic Intermittent Catheter is substantially equivalent to the predicate device. There are no significant differences that would affect safety and efficacy.
#### K. Performance Data
The bench testing performed verifies that the performance of the subject devices are substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include:
- Visual Inspection
- Flow Rate
- Friction Test
- Coating Presence
- Biocompatibility
- a. ISO 10993-1:2009 Biological Evaluation of Medical Devices -- Part 1: Evaluation and testing within a risk management process
- b. ISO 10993-5:2009 Biological Evaluation of Medical Devices -- Part 5: Tests for in Vitro Cytotoxicity
- c. ISO 10993-10:2010 Biological Evaluation of Medical Devices -- Part 10: Tests for Irritation and Skin Sensitization
The testing performed verifies that the performance of the subject devices are substantially equivalent in terms of critical performance characteristics to the predicate device.
The subject devices were tested to the requirements of ASTM F623.
{8}------------------------------------------------
Cytotoxicity, Sensitization, Irritation, were performed to demonstrate biocompatibility of the patient contacting materials. Overall, the results are comparable to the predicate and support a determination of substantial equivalence.
## L. Conclusion
The Rusch Hydrophilic Intermittent Catheters (Rusch FloCath Quick, Rusch FloCath Quick Kit, Rusch FloCath Intermittent Catheter, Rusch MMG H2O PVC Catheter Kit, Rusch MMG H2O Singles) are tubular devices that are inserted through the urethra to pass urine from the bladder.
The Rusch Hydrophilic Intermittent Catheters have the same intended use and technological characteristics as the predicate. Test results demonstrate that the subject devices meet their intended use and performs as well as or better than the legally marketed predicate device. It is for these reasons that the subject devices can be found substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.