The Symbia Pro.specta systems are radiological imaging systems that combines a single photon emission computed tomography (SPECT) camera system for nuclear medicine images, and a computed tomography (CT) camera system for x-ray images. The SPECT system is intended to detect the location and distribution of gamma ray radionuclides in the body and produce cross-sectional images through computer reconstruction of the data, and the CT system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. The nuclear medicine images and the x-ray images may be registered and displayed in a fused format (overlaid in the same orientation) for the anatomical localization of the nuclear medicine data (that is, distribution of radiopharmaceuticals). The SPECT and CT portions of the system may be used independently or in combination, and may include signal analysis and display equipment, patient and equipment support, radionuclide anatomical markers, component parts, and accessories. The SPECT and CT images may be transferred to other systems for radiation therapy planning or additional uses.
Device Story
Hybrid SPECT/CT imaging system; combines gamma camera (SPECT) and x-ray CT scanner. SPECT subsystem detects radiopharmaceutical distribution for metabolic/physiologic function; CT subsystem provides anatomical detail. Inputs: gamma ray emissions (SPECT) and x-ray transmission data (CT). Transformation: computer reconstruction of data into cross-sectional images; CT data used for SPECT attenuation correction and image fusion/registration. Used in clinical settings by physicians/technicians. Output: fused physiological/anatomical images for diagnosis, staging, and radiotherapy planning. Benefits: precise localization of metabolic activity; improved diagnostic accuracy via attenuation correction. System includes display, storage, and patient support equipment.
Clinical Evidence
Bench testing only. Performance verified against NEMA NU-1 standards for SPECT and IEC 60601-2-44 for CT. System integration and validation testing confirmed that all functional specifications and risk mitigations were met. No clinical data provided.
Technological Characteristics
Hybrid SPECT/CT system. SPECT: gamma camera, NEMA NU-1 compliant. CT: x-ray transmission, spiral/axial reconstruction, IEC 60601-2-44 compliant. Materials: standard medical imaging components. Connectivity: networked for image transfer. Software: SPECTsyngo acquisition/analysis package. Standards: IEC 60601-1 series, 21 CFR 1020.30/33, ISO 14971, IEC 62304. Sterilization: N/A (non-invasive).
Indications for Use
Indicated for use by trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for evaluation of diseases/disorders (e.g., cardiovascular, neurological, cancer). SPECT component images radionuclide distribution (planar, whole body, tomographic) up to 588 keV. CT component produces cross-sectional anatomical images. SPECT+CT allows attenuation correction and image registration. Software (SPECTsyngo) aids assessment and quantification of pathologies. CT component indicated for low-dose lung cancer screening in high-risk populations. No known contraindications. Used for all age groups (neonates to adults).
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 21, 2026
Siemens Medical Solutions USA, Inc.
Tabitha Estes
Senior Regulatory Affairs Professional
810 Innovation Dr.
Knoxville, Tennessee 37932
Re: K261315
Trade/Device Name: Symbia Pro.specta Q3 (11364751);
Symbia Pro.specta X3 (11364752);
Symbia Pro.specta X7 (11364753);
Symbia Pro.specta VA40 Family
Regulation Number: 21 CFR 892.1200
Regulation Name: Emission Computed Tomography System
Regulatory Class: Class II
Product Code: KPS, JAK
Dated: April 21, 2026
Received: April 21, 2026
Dear Tabitha Estes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261315 - Tabitha Estes
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K261315 - Tabitha Estes
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Daniel M. Krainak, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261315 | ? |
| Please provide the device trade name(s). | | ? |
| Symbia Pro.specta Q3 (11364751); Symbia Pro.specta X3 (11364752); Symbia Pro.specta X7 (11364753); Symbia Pro.specta VA40 Family | | |
| Please provide your Indications for Use below. | | ? |
| The Symbia Pro.specta systems are intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning or additional uses. SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques: planar imaging, whole body imaging, and tomographic imaging for isotopes with energies up to 588 keV. CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes taken at different angles. SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images. The SPECT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (SPECT) and anatomical (CT) images. Software: The SPECTsyngo software is an acquisition, display and analysis package intended to aid the clinician in the assessment and quantification of pathologies in images produced from SPECT, PET, CT, and other imaging modalities. This CT system can be used for low dose lung cancer screening in high risk populations.* *As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. There are no known contraindications. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☑ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
Siemens Symbia Pro.specta Q3, X3, X7,
Symbia Pro.specta VA40 Family
Special 510(k) Premarket Notification
# K261315
## 510(k) Summary
as required by 21 CFR Part 807.87(h)
## Identification of the Submitter
Submitter: Tabitha Estes
Senior Regulatory Affairs Specialist
Siemens Molecular Solutions USA, Inc.
810 Innovation Drive
Knoxville, TN 37932
Manufacturer: Siemens Medical Solutions USA, Inc.
2501 North Barrington Road
Hoffman Estates, IL 60192
Telephone Number: (865) 804-4553
Date of Submission: April 21st, 2026
## Identification of the product
Device Proprietary Name: Symbia Pro.specta Q3(11364751), Symbia Pro.specta X3 (11364752), Symbia Pro.specta X7 (11364753), Symbia Pro.specta VA40 Family
Common Name: Single-Photon Emission Computed Tomography (SPECT) System
Computed Tomography (CT) System
Classification Name: Emission Computed Tomography System per 21 CFR 892.1200
Computed Tomography X-Ray System per 21 CFR 892.1750
Product Code: KPS and JAK
Classification Panel: Radiology
{5}
Siemens Symbia Pro.specta Q3, X3, X7,
Symbia Pro.specta VA40 Family
Special 510(k) Premarket Notification
Device Class: Class II
## Marketed Devices to which Equivalence is claimed:
### Predicate Device:
Device Proprietary Name: Symbia Pro.specta VA30A Family Update
Common Name: Single-Photon Emission Computed Tomography (SPECT) System
Computed Tomography (CT) System
Classification Name: Emission Computed Tomography System per 21 CFR 892.1200
Computed Tomography X-Ray System per 21 CFR 892.1750
Manufacturer: Siemens Medical Solutions USA, Inc.
Product Code: KPS and JAK
Classification Panel: Radiology
Device Class: Class II
510(k) Number: K243186
### Reference Device:
Device Name and 510(k) numbers: SOMATOM Go.Platform
Common Name: Computed Tomography X-Ray System
Classification Name: Computed Tomography X-Ray System per 21 CFR 892.1750
Picture Archiving and Communications System per 21 CFR 892.2050
Manufacturer: Siemens Healthineers AG
{6}
Siemens Symbia Pro.specta Q3, X3, X7,
Symbia Pro.specta VA40 Family
Special 510(k) Premarket Notification
Product Code: LLZ, JAK
Classification Panel: Radiology
Device Class: Class II
510(k) Number: K250822
## Reference Device:
Device Name and 510(k) numbers: MI View&GO VA40A
Common Name: Image Processing Software
Classification Name: Picture Archiving and Communications System per 21 CFR 892.2050
Manufacturer: Siemens Medical Solution, USA, Inc.
Product Code: LLZ
Classification Panel: Radiology
Device Class: Class II
510(k) Number: K254016
## Device Description:
The Siemens Symbia Pro.specta Q3, X3 and X7 systems consists of Single-Photon Emission Computed Tomography (SPECT) scanner and integrated hybrid x-ray Computed Tomography (CT) and SPECT scanner.
The SPECT subsystem images and measures the distribution of radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and integrates CT's anatomical detail for precise reference of the location of the metabolic activity.
The CT component produces cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.
{7}
Siemens Symbia Pro.specta Q3, X3, X7, Symbia Pro.specta VA40 Family Special 510(k) Premarket Notification
The system can be used as an integrated SPECT and CT modality while also enabling independent functionality of SPECT and CT as stand-alone diagnostic imaging devices.
Siemens Symbia Pro.specta systems maintains the same intended use and indications for use as the commercially available Symbia Pro.specta VA30A systems (K243186).
Symbia Pro.specta systems are hybrid modality imaging systems comprised of two separate but integrated components: a gamma camera (SPECT) and a CT. The gamma camera is based on hardware and software features that generate nuclear medicine images based on the uptake of radioisotope tracers in a patient's body. The CT system (spiral CT) is designed to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.
The combination of SPECT and CT in a single device has several benefits. The SPECT subsystem images biochemical function while the CT subsystem images anatomy. The combination enables scans that not only indicate function, e.g., how active a tumor is, but precise localization, e.g., the precise location of that tumor in the body.
In addition, CT can be used to correct for the attenuation in SPECT acquisitions. Attenuation in SPECT is an unwanted side effect of the gamma rays scattering and being absorbed by tissue. This can lead to errors in the final image. The CT directly measures attenuation and can be used to create a 3D attenuation map of the patient which can be used to correct the SPECT images. The SPECT-CT scanner can be used to image and track how much dose was delivered to both the target and the surrounding tissue.
The systems consist of display equipment, data storage devices, patient and equipment supports and component parts and accessories.
Symbia Pro.specta VA40 release is the product name for the addition of features to the approved Symbia Pro.specta VA30A Family (K243186). The Symbia Pro.specta VA40A devices are based on the Symbia Pro.specta VA30A Family. The difference lies in the additional features/changes. The Intended Purpose and fundamental scientific technology remain unchanged.
Proposed modifications in Symbia Pro.specta VA40 include:
- Single Button Positioning for CT for daily QA
- IQ-SPECT cardiac improved image quality
- Retrospective 2D (formerly known as Enhanced QCI)
- Enhanced 2D Data Format (aka Legacy Data Format)
- Extended ECL for Low Energy Isotopes
- User Defined Isotopes for Broad Quantification
- 3 Peak Lu-177 MELP Broad Quantification
- User Modifications of Scatter Weights
- Time Activity Curves
- Workflow improvements
{8}
Siemens Symbia Pro.specta Q3, X3, X7,
Symbia Pro.specta VA40 Family
Special 510(k) Premarket Notification
- General Software Improvements
The table below summarizes the differences between the subject and predicate devices.
| Features | Symbia Pro.specta VA40 (Subject Device) | Symbia Pro.specta VA30 (Predicate Device K243186) |
| --- | --- | --- |
| Intended/Indication For Use | Same | Same |
| Biocompatibility | Same | Same |
| Productivity Features | Same | Same |
| Optional Pallets/Accessories/Collimators | Same | Same |
| MI View&GO updates | All changes are included in K254016. | N/A |
| Operating System | Same | Same |
| Computer Systems additions | Same | Same |
| Scan Planning Updates | Same | Same |
| SPECT Software updates | Proposed modifications in Symbia Pro.specta VA40 include: • Single Button Positioning for CT for daily QA • IQ-SPECT cardiac improved image quality • Retrospective 2D (formerly known as Enhanced QCI) • Enhanced 2D Data Format (aka Legacy Data Format) • Extended ECL for Low Energy Isotopes • User Defined Isotopes for Broad Quantification • 3 Peak Lu-177 MELP Broad Quantification • User Modifications of Scatter Weights | N/A |
{9}
Siemens Symbia Pro.specta Q3, X3, X7, Symbia Pro.specta VA40 Family Special 510(k) Premarket Notification
| | • Time Activity Curves • Workflow improvements • General Software Improvements | |
| --- | --- | --- |
| Export updates | Same | Same |
| CT Software updates | Additional features as included in SOMATOM VB20 K250822. | N/A |
| Connectivity | Same | Same |
| SPECT Detectors | Same | Same |
| SPECT Gantry | Same | Same |
| CT Gantry updates | No changes as compared to SOMATOM VB20 K250822. | N/A |
| Detector Performance and Collimator Specifications | Same | Same |
| GUI Changes | Same | Same |
| 3rdParty Software Options | Cedars (K172609) Corridor 4DM (K101279) | Cedars (K172609) Corridor 4DM (K101279) |
## Intended Use:
The Symbia Pro.specta systems are radiological imaging systems that combines a single photon emission computed tomography (SPECT) camera system for nuclear medicine images, and a computed tomography (CT) camera system for x-ray images.
The SPECT system is intended to detect the location and distribution of gamma ray radionuclides in the body and produce cross-sectional images through computer reconstruction of the data, and the CT system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. The nuclear medicine images and the x-ray images may be registered and displayed in a fused format (overlaid in the same orientation) for the anatomical localization of the nuclear medicine data (that is, distribution of radiopharmaceuticals). The SPECT and CT portions of the system may be used independently or in combination, and may include signal analysis and display equipment, patient and equipment support, radionuclide anatomical markers, component parts, and accessories. The SPECT and CT images may be transferred to other systems for radiation therapy planning or additional uses.
## Indications for Use:
The Symbia Pro.specta systems are intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited
{10}
Siemens Symbia Pro.specta Q3, X3, X7, Symbia Pro.specta VA40 Family Special 510(k) Premarket Notification
to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning or additional uses.
SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques: planar imaging, whole body imaging, and tomographic imaging for isotopes with energies up to 588 keV.
CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes taken at different angles.
SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images. The SPECT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patient's physiological (SPECT) and anatomical (CT) images.
Software: The SPECTsyngo software is an acquisition, display and analysis package intended to aid the clinician in the assessment and quantification of pathologies in images produced from SPECT, PET, CT, and other imaging modalities.
This CT system can be used for low dose lung cancer screening in high risk populations.*
*As defined by professional medical societies. Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
There are no known contraindications.
## Technological Characteristics:
## Performance Testing:
Performance testing for the CT subsystem was included in the premarket notification for the CT subsystems (K250822) and there have been no changes affecting this testing.
Each CT subsystem is tested and passes the Applicable Performance Standards prior to shipment:
- 21 CFR 1020.30 (a) Applicability
- 21 CFR 1020.30 (b)(36)(iii)-(v) Technique Factors
- 21 CFR 1020.30 (b)(58)-(62) CT, Scan, Scan Time, Tomogram, Dose
- 21 CFR 1020.30 (h)(3)(vi)-(viii) Information to be provided for users
- 21 CFR 1020.33 Computed Tomography (CT) equipment
- 21 CFR 1040.10 Laser Products
{11}
Siemens Symbia Pro.specta Q3, X3, X7,
Symbia Pro.specta VA40 Family
Special 510(k) Premarket Notification
- 21 CFR 1040.11 Specific purpose laser products
Symbia Pro.specta VA40A systems are designed in accordance with the 60601-1 series including all relevant collateral standards general (IEC 60601-1, 1-2, 1-3, etc.) and specific (IEC 60601-2-44). Performance testing is conducted according to NEMA NU-1. All Performance testing met the predetermined acceptance values.
| Detector Specifications | | |
| --- | --- | --- |
| Intrinsic spatial resolution - Tc99m | SPECT | 3/8” |
| FWHM in CFOV | SPECT | ≤3.84 mm |
| FWHM in UFOV | SPECT | ≤3.94 mm |
| FWTM in CFOV | SPECT | ≤7.54 mm |
| FWTM in UFOV | SPECT | ≤7.74 mm |
| Intrinsic spatial linearity - Tc99m | SPECT | |
| Differential in CFOV | SPECT | ≤0.24 mm |
| Differential in UFOV | SPECT | ≤0.24 mm |
| Absolute in CFOV | SPECT | ≤0.44 mm |
| Absolute in UFOV | SPECT | ≤0.7 mm |
| Intrinsic energy resolution | SPECT | |
| FWHM in CFOV | SPECT | ≤9.9% |
| Intrinsic flood field uniformity (uncorrected) - Tc99m | SPECT | |
| Differential in CFOV | SPECT | ≤2.5% |
| Differential in UFOV | SPECT | ≤2.7% |
| Integral in CFOV | SPECT | ≤2.9% |
| Integral in UFOV | SPECT | ≤3.7% |
Existing NEMA detector and collimator performance specifications do not change between the commercially available Symbia Pro.specta VA30A Family and proposed Symbia Pro.specta VA40A systems. There are no changes in the system design that could impact the SPECT performance specifications and the commercially available specifications still apply.
Bench testing performed on the CT subsystem is conducted in accordance with IEC 60601-2-44 and in accordance with US regulations including 21 CFR 1020.33, Computed Tomography (CT) equipment. This is unchanged from the commercially available SOMATOM.go CT systems (K250822).
{12}
Siemens Symbia Pro.specta Q3, X3, X7, Symbia Pro.specta VA40 Family Special 510(k) Premarket Notification
Risk analyses performed at Siemens Medical Solutions USA; Inc. are in compliance with the requirements of:
- IEC/ISO 14971 Medical Devices – Application of Risk Management to Medical Devices
- IEC 62304 Medical Device Software – Software Life-cycle Processes
There is no significant difference in the risks imposed by this system and the risks associated with the predicate devices. The product Risk Management Team has reviewed and verified that all risks identified in the Symbia Pro.specta VA40A Product Risk Analysis has been adequately mitigated and the individual and overall residual risks are reduced as far as possible.
Verification and validation of Siemens software is performed in accordance with documented procedures, test plans and specifications. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software verification, and system testing.
System and System Integration testing (validation) was carried out for all features of the project, and all planned test cases were executed.
Verification and Validation is performed to:
- ensure the functionality described in the specifications are met,
- ensure the quality in the planning documentation is adhered to,
- ensure mitigations required by risk analysis are implemented, and appropriate
- identify additional issues that may be related to patient / operator safety, or effectiveness
- assure that the specifications are appropriate to fulfill the intended use of the system.
## Safety and Effectiveness:
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk Management is ensured via a risk analysis in compliance with ISO 14971 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product.
Cybersecurity information in accordance with FDA Guidance documents issued on September 27, 2023, has been provided. The Symbia Pro.specta VA40A systems software has specific cybersecurity controls to prevent unauthorized access, modifications, misuse or denial of use.
Additionally, controls are enabled to prevent the unauthorized use of information that is stored, accessed or transferred between the Symbia Pro.specta VA40A systems and external devices.
Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards such as IEC 60601-1 series and 21 CFR 1020.30 and 21 CFR 1020.33 to minimize electrical, mechanical and radiation hazards.
Symbia Pro.specta VA40A systems conform to applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as required by the respective SPECT FDA Guidance Documents. SPECT detector and CT performance is conducted according to NEMA NU-1, and the performance does not change from the predicate device.
{13}
Siemens Symbia Pro.specta Q3, X3, X7, Symbia Pro.specta VA40 Family Special 510(k) Premarket Notification
Verification and validation of Siemens systems is performed in accordance with documented procedures, design and code reviews, test plans and specifications. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.
## Statement Regarding Substantial Equivalence:
Symbia Pro.specta VA40A systems are based on the commercially available Symbia Pro.specta VA30A systems and have the same indications for use as well as utilizes the same fundamental scientific technology as the predicate device. The software updates pose no new issues of safety and / or efficacy. Siemens considers the Symbia Pro.specta VA40A systems to be as safe and effective as the commercially available predicate device with substantially equivalent performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.