Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip; Vitoss® BiModal Bioactive Bone Graft Substitute; Vitoss BBTrauma® Bioactive Bone Graft Substitute; Vitoss® BA2X Bioactive Bone Graft Substitute; Vitoss® Bioactive Foam Bone Graft Substitute Pack; Vitoss® Bioactive Foam Bone Graft Substitute; Vitoss® Foam Bone Graft Substitute; Vitoss® Bone Graft Substitute Filled Canister; Vitoss® Bone Graft Substitute
Applicant
Stryker Spine
Product Code
MQV · Orthopedic
Decision Date
Apr 24, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
These implants are intended for use as bone void fillers for voids or gaps that are not intrinsic to the stability of the bony structure. They are indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. The implants are intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, pelvis, intervertebral disc space, and posterolateral spine) and may be combined with saline, autogenous blood, and/or bone marrow. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. When used in intervertebral body fusion procedures with an intervertebral body fusion device cleared by FDA for use with a bone void filler, Vitoss® can be mixed with saline, autogenous blood, bone marrow, and/or used on its own. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process.
Device Story
Resorbable calcium salt bone void filler; scaffold placed in skeletal osseous defects (extremities, pelvis, spine); resorbs and replaced by bone during healing. Used by surgeons in orthopedic/spine procedures; can be mixed with saline, autogenous blood, or bone marrow. Filled canister variant functions as piston syringe for aspiration/mixing of autogenous blood/marrow. Provides structural scaffold for bone regeneration; facilitates fusion in intervertebral body procedures when used with cleared fusion devices.
Clinical Evidence
No new performance or clinical testing was performed. Expanded indications are supported by clinical and performance testing rationales.
Technological Characteristics
Resorbable calcium salt bone void filler. Form factors include foam strips, bioactive foam, and filled canisters. Operates as a passive scaffold for bone ingrowth. No energy source. Sterilized product.
Indications for Use
Indicated for treatment of surgically created or traumatic osseous defects in adults and pediatric patients ≥ 6 years old. Used for filling bony voids or gaps in the skeletal system, including extremities, pelvis, and spine (intervertebral disc space and posterolateral fusion). May be combined with saline, autogenous blood, or bone marrow. Not for use in defects intrinsic to bony structure stability or large defects failing to heal spontaneously.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
Vitoss BiModal Bioactive Bone Graft Substitute Foam Strip, Vitoss BiModal Bioactive Bone Graft Substitute, Vitoss BBTrauma Bioactive Bone Graft Substitute, Vitoss BA2X Bioactive Bone Graft Substitute, Vitoss Bioactive Foam Bone Graft Substitute Pack, Vitoss Bioactive Foam Bone Graft Substitute, Vitoss Foam Bone Graft Substitute, Vitoss Bone Graft Substitute Filled Canister, and Vitoss Bone Graft Substitute (K242280)
Reference Devices
FIBERGRAFT™ BG Putty GPS Bone Graft Substitute, FIBERGRAFT™ BG Putty Bone Graft Substitute, FIBERGRAFT™ AERIDYAN™ Matrix Bone Graft Substitute, and FIBERGRAFT™ BG Matrix Bone Graft Substitute (K253147)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 24, 2026
Stryker Spine
Kristina Daoud
Senior Regulatory Affairs Specialist
2 Pearl Ct.
Allendale, New Jersey 07401
Re: K261008
Trade/Device Name:
Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip;
Vitoss® BiModal Bioactive Bone Graft Substitute;
Vitoss BBTrauma® Bioactive Bone Graft Substitute;
Vitoss® BA2X Bioactive Bone Graft Substitute;
Vitoss® Bioactive Foam Bone Graft Substitute Pack;
Vitoss® Bioactive Foam Bone Graft Substitute;
Vitoss® Foam Bone Graft Substitute;
Vitoss® Bone Graft Substitute Filled Canister;
Vitoss® Bone Graft Substitute
Regulation Number: 21 CFR 888.3045
Regulation Name: Resorbable Calcium Salt Bone Void Filler Device
Regulatory Class: Class II
Product Code: MQV
Dated: March 26, 2026
Received: March 26, 2026
Dear Kristina Daoud:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261008 - Kristina Daoud
Page 2
adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part
{2}
K261008 - Kristina Daoud
Page 3
803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JESSE MUIR-S
Digitally signed by JESSE MUIR-S
Date: 2026.04.24 09:36:42 -04'00'
Jesse Muir, Ph.D.
Assistant Director
DHT6C: Division of Restorative, Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K261008 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip;
Vitoss® BiModal Bioactive Bone Graft Substitute;
Vitoss BBTrauma® Bioactive Bone Graft Substitute;
Vitoss® BA2X Bioactive Bone Graft Substitute;
Vitoss® Bioactive Foam Bone Graft Substitute Pack;
Vitoss® Bioactive Foam Bone Graft Substitute;
Vitoss® Foam Bone Graft Substitute;
Vitoss® Bone Graft Substitute Filled Canister;
Vitoss® Bone Graft Substitute | | |
| Please provide your Indications for Use below. | | ? |
| Vitoss® BA2X Bioactive Bone Graft Substitute, Vitoss BBTrauma® Bioactive Bone Graft Substitute,
Vitoss® BiModal Bioactive Bone Graft Substitute, Vitoss® BiModal Bioactive Bone Graft Substitute Foam
Strip
These implants are intended for use as bone void fillers for voids or gaps that are not intrinsic to the stability
of the bony structure. They are indicated for use in the treatment of surgically created osseous defects or
osseous defects created from traumatic injury to the bone. The implants are intended to be used for filling
bony voids or gaps of the skeletal system (i.e., the extremities, pelvis, intervertebral disc space, and
posterolateral spine) and may be combined with saline, autogenous blood, and/or bone marrow. These
osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6
years old), are surgically created or the result of traumatic injury to the bone. When used in intervertebral
body fusion procedures with an intervertebral body fusion device cleared by FDA for use with a bone void
filler, Vitoss® can be mixed with saline, autogenous blood, bone marrow, and/or used on its own.
Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the
healing process.
Vitoss® Bioactive Bone Graft Substitute, Vitoss® Bioactive Bone Graft Substitute Pack Vitoss® Bioactive
Foam Bone Graft Substitute
Bioactive Foam Bone Graft Substitute is intended for use as a bone void filler for voids or gaps that are not
intrinsic to the stability of the bony structure. Vitoss® Bioactive Foam is indicated for use in the treatment of
surgically created osseous defects or osseous defects created from traumatic injury to the bone. Vitoss®
Bioactive Foam Bone Graft Substitute is intended to be used for filling bony voids or gaps of the skeletal
system (i.e., the extremities, pelvis, and spine, which includes intervertebral disc space and posterolateral
fusion procedures) and may be combined with saline, autogenous blood, and/or bone marrow. These
osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6
years old), are surgically created or the result of traumatic injury to the bone. When used in intervertebral
body fusion procedures with an intervertebral body fusion device cleared by FDA for use with a bone void
filler, Vitoss® can be mixed with saline, autogenous blood, bone marrow, and/or used on its own. Following
placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing
process.
Vitoss® Bone Graft Substitute, Vitoss® Foam Bone Graft Substitute, Vitoss® Bone Graft Substitute Filled
Canister | | |
{4}
These implants are intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure. These implants are indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. These implants should not be used to treat large defects that in the surgeon's opinion would fail to heal spontaneously. These implants are intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, pelvis, and spine, which includes intervertebral disc space and posterolateral fusion procedures). When used in intervertebral body fusion procedures with an intervertebral body fusion device cleared by FDA for use with a bone void filler, Vitoss® can be mixed with saline, autogenous blood, bone marrow, and/or used on its own. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process.
The Vitoss® Foam family of products may be mixed with saline in addition to autogenous blood or bone marrow.
The Vitoss® Filled Canister is intended for use as a piston syringe system for the aspiration of autogenous blood and/or bone marrow. This canister provides the surgeon with a convenient way to mix autogenous blood with Vitoss® and deliver the material to the orthopaedic surgical site.
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| --- | --- | --- |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old)
☐ Infants (29 days old to < 2 years old)
☑ Children (2 years old to < 12 years old)
☑ Adolescents (12 years old to < 22 years old)
☑ Adults (22 years old and greater) | ? |
{5}
K261008
VB
SPINE
| 510(k) Summary: Vitoss Bone Graft Substitute | |
| --- | --- |
| Submitter: | Stryker Spine
2 Pearl Court
Allendale, NJ 07401 |
| Contact Person: | Name: Kristina Daoud
Email: kristina.tawadros@vbspineco.com
Secondary Contact:
Name: Justin Eggleton
Email: justin.eggleton@vbspineco.com |
| Date Prepared: | April 23, 2026 |
| Trade Names: | 1. Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip
2. Vitoss® BiModal Bioactive Bone Graft Substitute
3. Vitoss BBTrauma® Bioactive Bone Graft Substitute
4. Vitoss® BA2X Bioactive Bone Graft Substitute
5. Vitoss® Bioactive Foam Bone Graft Substitute Pack
6. Vitoss® Bioactive Foam Bone Graft Substitute
7. Vitoss® Foam Bone Graft Substitute
8. Vitoss® Bone Graft Substitute Filled Canister
9. Vitoss® Bone Graft Substitute |
| Common Name: | Resorbable Calcium Salt Bone Void Filler Device |
| Proposed Class: | Class II |
| Classification Name: | Filler, Bone Void, Calcium Compound (21 CFR 888.3045) |
| Product Code: | MQV |
Page 1 of 4
{6}
K261008
| 510(k) Summary: Vitoss Bone Graft Substitute | |
| --- | --- |
| Predicate Devices: | **Primary Predicate:** K242280
Vitoss BiModal Bioactive Bone Graft Substitute Foam Strip, Vitoss BiModal Bioactive Bone Graft Substitute, Vitoss BBTrauma Bioactive Bone Graft Substitute, Vitoss BA2X Bioactive Bone Graft Substitute, Vitoss Bioactive Foam Bone Graft Substitute Pack, Vitoss Bioactive Foam Bone Graft Substitute, Vitoss Foam Bone Graft Substitute, Vitoss Bone Graft Substitute Filled Canister, and Vitoss Bone Graft Substitute
**Reference:** K253147
FIBERGRAFT™ BG Putty GPS Bone Graft Substitute, FIBERGRAFT™ BG Putty Bone Graft Substitute, FIBERGRAFT™ AERIDYAN™ Matrix Bone Graft Substitute, and FIBERGRAFT™ BG Matrix Bone Graft Substitute |
| Device Description: | Vitoss Bone Graft Substitutes are intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure in the skeletal system. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process. |
| Indications for Use: | **Vitoss® Bone Graft Substitute, Vitoss® Foam Bone Graft Substitute, Vitoss® Bone Graft Substitute Filled Canister**
These implants are intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure.
These implants are indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. These implants should not be used to treat large defects that in the surgeon’s opinion would fail to heal spontaneously.
These implants are intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, pelvis, and spine, which includes intervertebral disc space and posterolateral fusion procedures). When used in intervertebral body fusion procedures with an intervertebral body fusion device cleared by FDA for use with a bone void filler, Vitoss® can be mixed with saline, autogenous blood, bone marrow, and/or used on its own. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process.
The Vitoss® Foam family of products may be mixed with saline in addition to autogenous blood or bone marrow. |
Page 2 of 4
{7}
K261008
| 510(k) Summary: Vitoss Bone Graft Substitute | |
| --- | --- |
| | The Vitoss® Filled Canister is intended for use as a piston syringe system for the aspiration of autogenous blood and/or bone marrow. This canister provides the surgeon with a convenient way to mix autogenous blood with Vitoss® and deliver the material to the orthopaedic surgical site. |
| **Vitoss® Bioactive Bone Graft Substitute, Vitoss® Bioactive Bone Graft Substitute Pack, Vitoss® Bioactive Foam Bone Graft Substitute**
Vitoss® Bioactive Foam Bone Graft Substitute is intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure. Vitoss® Bioactive Foam is indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. Vitoss® Bioactive Foam Bone Graft Substitute is intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, pelvis, and spine, which includes intervertebral disc space and posterolateral fusion procedures) and may be combined with saline, autogenous blood, and/or bone marrow. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. When used in intervertebral body fusion procedures with an intervertebral body fusion device cleared by FDA for use with a bone void filler, Vitoss® can be mixed with saline, autogenous blood, bone marrow, and/or used on its own. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process. | |
| **Vitoss® BA2X Bioactive Bone Graft Substitute, Vitoss BBTrauma® Bioactive Bone Graft Substitute, Vitoss® BiModal Bioactive Bone Graft Substitute, Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip**
These implants are intended for use as bone void fillers for voids or gaps that are not intrinsic to the stability of the bony structure. | |
| They are indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. | |
| The implants are intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, pelvis, intervertebral disc space, and posterolateral spine) and may be combined with saline, autogenous blood, and/or bone marrow. These osseous defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 6 years old), are surgically created or the result of traumatic injury to the bone. When used in intervertebral body fusion procedures with an intervertebral body fusion device cleared by FDA for use with a bone void filler, Vitoss® can be mixed with saline, autogenous blood, bone marrow, and/or used on its own. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process. | |
Page 3 of 4
{8}
K261008
| 510(k) Summary: Vitoss Bone Graft Substitute | |
| --- | --- |
| Summary of the Technological Characteristics | The subject Vitoss Bone Graft Substitute devices are substantially equivalent to the predicate devices. The subject devices have the same intended use, design, materials, operating principles, and performance specifications as their respective predicates. The proposed modification is to expand the indications statement within the subject Vitoss Bone Graft Substitute devices to include the pediatric population of ≥ 6 years old in addition to the posterolateral spine and use of previously cleared FDA cages which will not impact the safety and efficacy of the subject devices. There is no physical change to the products resulting from this IFU change and no new or significantly changed risks identified as a result. Therefore, the subject devices are substantially equivalent to their respective predicate devices. |
| Summary of the Performance Data | The expanded indications for use are supported by a clinical and performance testing rationale. No new performance testing was performed. |
| Conclusion | The subject devices have the same intended use with additions to the indications for use to align with the reference device. Per the clinical rationales detailed in this submission, the updated indications for use for the subject devices are substantially equivalent to their predicate and reference devices. |
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.