Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip, Vitoss® BiModal Bioactive Bone Graft Substitute, Vitoss BBTrauma® Bioactive Bone Graft Substitute,Vitoss® BA2X Bioactive Bone Graft Substitute, Vitoss® Bioactive Bone Graft Substitute Pack,Vitoss® Bioactive Bone Graft Substitute, Vitoss® Foam Bone Graft Substitute, Vitoss® Bone Graft Substitute Filled Canister, Vitoss® Bone Graft Substitute, Vitoss® Bone Graft Substitute - Synthetic Cancellous Chips
Applicant
Stryker Spine
Product Code
MQV · Orthopedic
Decision Date
Aug 29, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Vitoss® BA2X Bioactive Bone Graft Substitute, Vitoss BBTrauma® Bioactive Bone Graft Substitute, Vitoss® BiModal Bioactive Bone Graft Substitute, Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip These implants are intended for use as bone void fillers for voids or gaps that are not intrinsic to the stability of the bony structure. They are indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. The implants are intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, pelvis and posterolateral spine) and may be combined with saline, autogenous blood, and/or bone marrow. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process. Vitoss® Bioactive Bone Graft Substitute, Vitoss® Bioactive Bone Graft Substitute Pack Vitoss® Bioactive Foam Bone Graft Substitute is intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure. Vitoss® Bioactive Foam is indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. Vitoss® Bioactive Foam Bone Graft Substitute is intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, pelvis, and spine, which includes posterolateral fusion procedures) and may be combined with saline, autogenous blood, and/or bone marrow. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process. Vitoss® Bone Graft Substitute, Vitoss® Bone Graft Substitute - Synthetic Cancellous Chips, Vitoss® Foam Bone Graft Substitute. Vitoss® Bone Graft Substitute Filled Canister These implants are intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure. These implants are indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. These implants should not be used to treat large defects that in the surgeon's opinion would fail to heal spontaneously. These implants are intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, spine and pelvis) and may be combined with autogenous blood and/or bone marrow. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process. The Vitoss® Foam family of products may be mixed with saline in addition to autogenous blood or bone marrow. The Vitoss® Filled Canister is intended for use as a piston syringe system for the aspiration of autogenous blood and/or bone marrow. This canister provides the surgeon with a convenient way to mix autogenous blood with Vitos® and deliver the material to the orthopaedic surgical site.
Device Story
Resorbable calcium salt bone void fillers; used to fill non-structural skeletal gaps/voids (extremities, pelvis, spine). Scaffold resorbs and is replaced by bone during healing. Can be mixed with saline, autogenous blood, or bone marrow. Filled Canister variant functions as a piston syringe for aspiration and delivery of autogenous blood/bone marrow mixture. Used in orthopedic surgical settings by surgeons. Benefits include providing a scaffold for bone regeneration in surgically created or traumatic osseous defects.
Resorbable calcium salt bone void filler; MR Safe per ASTM F2503. Includes foam strips, chips, and piston syringe delivery systems.
Indications for Use
Indicated for patients requiring bone void fillers for non-structural osseous defects in the skeletal system (extremities, pelvis, spine/posterolateral fusion) resulting from surgery or trauma. Contraindicated for large defects where spontaneous healing is unlikely.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Predicate Devices
Vitoss BiModal Bone Graft Substitute Foam Strip (K153608)
Vitoss BA Bimodal Bioactive Bone Graft Substitute (K103173)
Vitoss Bioactive Foam Bone Graft Substitute Pack, Vitoss Bioactive Foam Bone Graft Substitute, Vitoss Foam Bone Graft Substitute, Vitoss Bone Graft Substitute Filled Canister, Vitoss Bone Graft Substitute (K083033)
Submission Summary (Full Text)
{0}------------------------------------------------
August 29, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Stryker Spine Fiorella Lopardo Staff Regulatory Specialist 2 Pearl Court Allendale, New Jersey 07401
## Re: K242280
Trade/Device Name: Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip Vitoss® BiModal Bioactive Bone Graft Substitute Vitoss BBTrauma® Bioactive Bone Graft Substitute Vitoss® BA2X Bioactive Bone Graft Substitute Vitoss® Bioactive Bone Graft Substitute Pack Vitoss® Bioactive Bone Graft Substitute Vitoss® Foam Bone Graft Substitute Vitoss® Bone Graft Substitute Filled Canister Vitoss® Bone Graft Substitute Vitoss® Bone Graft Substitute - Synthetic Cancellous Chips
Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: August 1, 2024 Received: August 2, 2024
Dear Fiorella Lopardo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
Sincerely, JESSE MUIR - Digitally signed by JESSE
MUIR - MUIR - MUIR -S Date: 2024.08.29 15:27:32 S -04'00' Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K242280
Device Name
Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip;
Vitoss® BiModal Bioactive Bone Graft Substitute ;
Vitoss BBTrauma® Bioactive Bone Graft Substitute ;
Vitoss® BA2X Bioactive Bone Graft Substitute ;
Vitoss® Bioactive Bone Graft Substitute Pack;
Vitoss® Bioactive Bone Graft Substitute;
Vitoss® Foam Bone Graft Substitute :
Vitoss® Bone Graft Substitute Filled Canister :
Vitoss® Bone Graft Substitute;
Vitoss® Bone Graft Substitute - Synthetic Cancellous Chips
Indications for Use (Describe)
Vitoss® BA2X Bioactive Bone Graft Substitute, Vitoss BBTrauma® Bioactive Bone Graft Substitute, Vitoss® BiModal Bioactive Bone Graft Substitute, Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip
These implants are intended for use as bone void fillers for voids or gaps that are not intrinsic to the stability of the bony structure.
They are indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone.
The implants are intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, pelvis and posterolateral spine) and may be combined with saline, autogenous blood, and/or bone marrow. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process.
Vitoss® Bioactive Bone Graft Substitute, Vitoss® Bioactive Bone Graft Substitute Pack Vitoss® Bioactive Foam Bone Graft Substitute is intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure. Vitoss® Bioactive Foam is indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. Vitoss® Bioactive Foam Bone Graft Substitute is intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, pelvis, and spine, which includes posterolateral fusion procedures) and may be combined with saline, autogenous blood, and/or bone marrow. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process.
Vitoss® Bone Graft Substitute, Vitoss® Bone Graft Substitute - Synthetic Cancellous Chips, Vitoss® Foam Bone Graft Substitute. Vitoss® Bone Graft Substitute Filled Canister These implants are intended for use as a bone void filler for voids or gaps that are not intrinsic to the stability of the bony structure.
These implants are indicated for use in the treatment of surgically created osseous defects or osseous defects created from traumatic injury to the bone. These implants should not be used to treat large defects that in the surgeon's opinion would fail to heal spontaneously.
{4}------------------------------------------------
These implants are intended to be used for filling bony voids or gaps of the skeletal system (i.e., the extremities, spine and pelvis) and may be combined with autogenous blood and/or bone marrow. Following placement in the bony void or gap, the scaffold resorbs and is replaced with bone during the healing process.
The Vitoss® Foam family of products may be mixed with saline in addition to autogenous blood or bone marrow.
The Vitoss® Filled Canister is intended for use as a piston syringe system for the aspiration of autogenous blood and/or bone marrow. This canister provides the surgeon with a convenient way to mix autogenous blood with Vitos® and deliver the material to the orthopaedic surgical site.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase and black. The background is white.
| 510(k) Summary | | |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Submitter: | Stryker Spine<br>2 Pearl Court<br>Allendale, NJ 07401 | |
| Contact Person : | Name: Fiorella Lopardo<br>Phone: 954-871-5460<br>Email: fiorella.lopardo@stryker.com<br>Secondary Contact:<br>Name: Shraddha Moore<br>Phone: 201-831-5804<br>Email: shraddha.more@stryker.com | |
| Date Prepared: | August 29, 2024 | |
| Trade Names: | 1. Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip<br>2. Vitoss® BiModal Bioactive Bone Graft Substitute<br>3. Vitoss BBTrauma® Bioactive Bone Graft Substitute<br>4. Vitoss® BA2X Bioactive Bone Graft Substitute<br>5. Vitoss® Bioactive Bone Graft Substitute Pack<br>6. Vitoss® Bioactive Bone Graft Substitute<br>7. Vitoss® Foam Bone Graft Substitute<br>8. Vitoss® Bone Graft Substitute Filled Canister<br>9. Vitoss® Bone Graft Substitute<br>10. Vitoss® Bone Graft Substitute - Synthetic Cancellous Chips | |
| Common Name: | Resorbable Calcium Salt Bone Void Filler Device | |
| Proposed Class: | Class II | |
| Classification Name: | Filler, Bone Void, Calcium Compound (21 CFR 888.3045) | |
| Product Code: | MQV | |
| Predicate Devices: | Primary Predicate: Vitoss BiModal Bone Graft Substitute Foam Strip<br>(K153608)<br>Additional Predicates:<br>Vitoss BA Bimodal Bioactive Bone Graft Substitute (K103173)<br>FM-02 Bone Graft Substitute (K102545)<br>Vitoss Bioactive Foam Bone Graft Substitute Pack, Vitoss Bioactive Foam<br>Bone Graft Substitute, Vitoss Foam Bone Graft Substitute, Vitoss Bone Graft<br>Substitute Filled Canister, Vitoss Bone Graft Substitute (K083033) | |
| 510(k) Summary | | |
| Device Description: | Vitoss® Bone Graft Substitutes are intended for use as a bone void filler for<br>voids or gaps that are not intrinsic to the stability of the bony structure in the<br>skeletal system. Following placement in the bony void or gap, the scaffold<br>resorbs and is replaced with bone during the healing process. | |
| Indications for Use: | Vitoss® BA2X Bioactive Bone Graft Substitute, Vitoss BBTrauma®<br>Bioactive Bone Graft Substitute, Vitoss® BiModal Bioactive Bone Graft<br>Substitute, Vitoss® BiModal Bioactive Bone Graft Substitute Foam Strip<br>These implants are intended for use as bone void fillers for voids or gaps that<br>are not intrinsic to the stability of the bony structure.<br>They are indicated for use in the treatment of surgically created osseous<br>defects or osseous defects created from traumatic injury to the bone.<br>The implants are intended to be used for filling bony voids or gaps of the<br>skeletal system (i.e., the extremities, pelvis and posterolateral spine) and may<br>be combined with saline, autogenous blood, and/or bone marrow. Following<br>placement in the bony void or gap, the scaffold resorbs and is replaced with<br>bone during the healing process. | |
| | Vitoss® Bioactive Bone Graft Substitute, Vitoss® Bioactive Bone Graft<br>Substitute Pack<br>Vitoss® Bioactive Foam Bone Graft Substitute is intended for use as a bone void<br>filler for voids or gaps that are not intrinsic to the stability of the bony<br>structure. Vitoss® Bioactive Foam is indicated for use in the treatment of<br>surgically created osseous defects or osseous defects created from traumatic<br>injury to the bone. Vitoss® Bioactive Foam Bone Graft Substitute is intended to<br>be used for filling bony voids or gaps of the skeletal system (i.e., the extremities,<br>pelvis, and spine, which includes posterolateral fusion procedures) and may be<br>combined with saline, autogenous blood, and/or bone marrow. Following<br>placement in the bony void or gap, the scaffold resorbs and is replaced with<br>bone during the healing process. | |
| | Vitoss® Bone Graft Substitute, Vitoss® Bone Graft Substitute - Synthetic<br>Cancellous Chips, Vitoss® Foam Bone Graft Substitute, Vitoss® Bone Graft<br>Substitute Filled Canister<br>These implants are intended for use as a bone void filler for voids or gaps that<br>are not intrinsic to the stability of the bony structure.<br>These implants are indicated for use in the treatment of surgically created<br>osseous defects or osseous defects created from traumatic injury to the bone.<br>These implants should not be used to treat large defects that in the surgeon's<br>opinion would fail to heal spontaneously.<br>These implants are intended to be used for filling bony voids or gaps of the | |
| | skeletal system (i.e., the extremities, spine and pelvis) and may be combined<br>with autogenous blood and/or bone marrow. Following placement in the bony. | |
| | 510(k) Summary | |
| | void or gap, the scaffold resorbs and is replaced with bone during the healing<br>process.<br>The Vitoss® Foam family of products may be mixed with saline in addition to<br>autogenous blood or bone marrow.<br>The Vitoss® Filled Canister is intended for use as a piston syringe system for<br>the aspiration of autogenous blood and/or bone marrow. This canister<br>provides the surgeon with a convenient way to mix autogenous blood with | |
| Summary of the<br>Technological<br>Characteristics | Vitoss® and deliver the material to the orthopaedic surgical site.<br>The subject devices have similar technological characteristics as their<br>predicate devices. Changes implemented since the last clearance of the<br>subject devices were assessed and detailed in this submission. Minor<br>differences in technological characteristics of the subject devices do not<br>impact safety and efficacy of the devices. Therefore, the fundamental scientific<br>technology of the subject devices is the same as previously cleared devices. | |
| Summary of the<br>Performance Data | The device materials have been analyzed in accordance with ASTM F2503<br>to confirm they meet the definition of MR Safe. | |
| Conclusion | The subject devices have the same intended use with minor differences in the<br>indications for use and technological characteristics as the predicate devices.<br>Per the assessments detailed in this submission, the minor differences in<br>indications for use and technological characteristics of the subject devices and<br>updates to the labeling do not impact safety and efficacy of the devices.<br>Therefore, the subject devices are substantially equivalent to their predicate<br>devices. | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are all lowercase and connected, giving the word a modern and streamlined appearance. The color of the text is black, and the background is white, providing a high contrast that makes the word easily readable.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase and black. The background is white.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.