K260835 · Q3 Medical USA, LLC (C/O Q3 Medical Devices Ltd.) · FGE · Jul 27, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K260835
Device Name
ARCHIMEDES Biodegradable Stent (Medium)
Applicant
Q3 Medical USA, LLC (C/O Q3 Medical Devices Ltd.)
Product Code
FGE · Gastroenterology, Urology
Decision Date
Jul 27, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K260835 · Jul 27, 2026
ARCHIMEDES Biodegradable Stent (Medium)
Q3 Medical USA, LLC (C/O Q3 Medical Devices Ltd.)
Retrospective clinical study cohorts
The sponsor utilized data from five clinical studies, including retrospective designs, to support the safety, efficacy, and clinical performance of the ARCHIMEDES Biodegradable Stent (Medium) in biliary and pancreatic duct drainage.
Five clinical studies (collective); Includes retrospective and prospective, randomized and non-randomized designs
Patients with various pancreatic and biliary conditions; Number of Sites: Single-center and multicenter
Not applicable for this study
Safety, efficacy, clinical performance, surgical and endoscopic placement, degradation profile, and technical success
Indications for Use
The ARCHIMEDES Biodegradable Stent (Medium) is intended to drain biliary ducts in patients indicated for biliary duct stenting. The ARCHIMEDES Biodegradable Stent (Medium) is intended to drain pancreatic ducts in patients indicated for pancreatic duct stenting.
Device Story
Single-use, sterile, biodegradable polymeric stent; delivered endoscopically via duodenoscope working channel. Fluted cross-sectional profile with two spiral drainage channels and one central hollow lumen; allows fluid flow around and through stent. Curved form fits ductal anatomy; features flaps and split-ends to minimize migration. Tapered tips facilitate atraumatic insertion. Visible under fluoroscopy. Stent biodegrades in-situ, eliminating need for removal procedure. Minimal strength retention (MSR) time is 20 days; full degradation occurs within <161 days. Used by gastroenterologists in clinical settings to maintain ductal patency. Benefits include reduced procedural burden by avoiding secondary removal.
Clinical Evidence
Supported by five clinical studies (single-center/multicenter, prospective/retrospective, randomized/non-randomized). Studies evaluated safety, efficacy, clinical performance, endoscopic placement, degradation profile, and technical success across diverse patient populations with various pancreatic and biliary conditions.
Indicated for patients requiring biliary or pancreatic duct stenting for drainage.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Advanix Biliary Stent with NaviFlex RX Delivery System (K203162)
Submission Summary (Full Text)
{0}

July 27, 2026
Q3 Medical USA, LLC (C/O Q3 Medical Devices Ltd.)
% John J. Smith
Partner
Hogan Lovells US LLP
Columbia Sq.
555 Thirteenth St., NW
Washington, D.C., D.C. 20004
Re: K260835
Trade/Device Name: ARCHIMEDES Biodegradable Stent (Medium)
Regulation Number: 21 CFR 876.5010
Regulation Name: Biliary Catheter And Accessories
Regulatory Class: Class II
Product Code: FGE
Dated: June 18, 2026
Received: June 22, 2026
Dear John J. Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260835 - John J. Smith
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260835 - John J. Smith
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# ANTHONY LEE -S
Anthony Lee, Ph.D., MBA
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K260835
Device Name
ARCHIMEDES Biodegradable Stent (Medium)
Indications for Use (Describe)
The ARCHIMEDES Biodegradable Stent (Medium) is intended to drain biliary ducts in patients indicated for biliary duct stenting.
The ARCHIMEDES Biodegradable Stent (Medium) is intended to drain pancreatic ducts in patients indicated for pancreatic duct stenting.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
K260835 page 1 of 12
# 510(k) Summary
# ARCHIMEDES™ Biodegradable Stent (Medium)
# 1. Submitter
Company Name: Q3 Medical USA, LLC
Company Address: 1800 Camden Rd., Suite 107
Charlotte, NC 28203
Contact: Eric K. Mangiardi, MSc.
President and CEO
(704) 712-1331
emangiardi@q3medical.com
Date Prepared: March 13, 2026
# 2. Device
Trade Name: ARCHIMEDES™ Biodegradable Stent (Medium)
Common Name: Biliary Stent, Pancreatic Stent
Product Code: FGE
Device Class: Class II
Device Panel: Gastroenterology / Urology
Classification Regulation: 21 CRF 876.5010, Biliary catheter and accessories
# 3. Predicate Device (Pancreatic indication)
Trade Name: ARCHIMEDES™ Biodegradable Pancreatic Stent (Fast)
Clearance Number: K243412
Common Name: Pancreatic Stent
Product Code: FGE
Device Class: Class II
Device Panel: Gastroenterology / Urology
Classification Regulation: 21 CRF 876.5010, Biliary catheter and accessories
{5}
K260835 page 2 of 12
#### 4. Predicate Device (Biliary indication)
| Trade Name: | Advanix™ Biliary Stent with NaviFlex™ RX Delivery System |
| --- | --- |
| Clearance Number: | K203162 |
| Common Name: | Stents, Drains, and Dilators for the Biliary Ducts |
| Product Code: | FGE |
| Device Class: | Class II |
| Device Panel: | Gastroenterology / Urology |
| Classification Regulation: | 21 CRF 876.5010, Biliary catheter and accessories |
#### 5. Device Description (Pancreatic and biliary indications)
The ARCHIMEDES™ Biodegradable Stent - Medium is a single use, sterile, biodegradable stent implant intended for use in the biliary and pancreatic duct. The device is intended to be delivered surgically or endoscopically through the working channel of a duodenoscope.
The ARCHIMEDES™ Medium has a fluted cross-sectional profile and provides three drainage channels, two of which spiral in a double helical pattern down the length of the stent and one of which is the hollow inner lumen of the device tube. This unique design permits fluids to flow both around and through the stent. It has a curved form design to fit the natural curvature of the biliary and pancreatic duct and contains flap and split-end features to minimize spontaneous migration of the stent while in situ. The tips of the stent are tapered to facilitate atraumatic insertion through the papilla. The stent is visible under fluoroscopy. The 8F and 10F stent design contains flaps on each end. The 6F stent design has a flap on one end and a split-end feature on the other end.
The stent is offered in outer diameters of 2.0mm (6F), 2.6mm (8F), or 3.4mm (10F), and in various lengths to accommodate variations in biliary and pancreatic anatomy across individuals (40, 60, 80, 100, 125, 150, 175, 200, and 225 mm). The ARCHIMEDES™ is designed to be an alternative to plastic stents that maintain the same clinical purpose. In particular, it is used in a manner similar to current FDA-cleared plastic pancreatic stents. The primary difference is that the ARCHIMEDES™ biodegrades in-situ rather than requiring an additional procedure for removal needed of plastic stents.
The Minimal Strength Retention (MSR) time for the material - i.e., the minimum amount of time for which the device retains at least 10% of an initial strength parameter and remains intact with no breaks is 20 days. Full-degradation or no stent presence is reached within <161 days.
{6}
K260835 page 3 of 12
# 6. Indications for Use (Pancreatic indication)
The ARCHIMEDES™ Biodegradable Stent (Medium) is intended to drain pancreatic ducts in patients indicated for pancreatic duct stenting.
# 7. Indications for Use) Biliary indication)
The ARCHIMEDES™ Biodegradable Stent (Medium) is intended to drain biliary ducts in patients indicated for biliary duct stenting.
# 8. Comparison to the Predicate Device (Pancreatic indication)
The ARCHIMEDES™ Biodegradable Stent (Medium) is substantially equivalent to the currently marketed ARCHIMEDES™ Biodegradable Pancreatic Stent (Fast) (K243412) as follows: same intended use and principles of operations, same packaging and sterilization methods; and similar technological characteristics.
The ARCHIMEDES™ Medium stent shares key technological characteristics with the predicate device in that they are both implantable stents intended to drain pancreatic ducts.
The main technological difference between the ARCHIMEDES™ Medium and the predicate device is material composition and degradation profile. The ARCHIMEDES™ Medium stent is designed to reach minimal strength retention in about 20 days whereas the predicate device reaches minimal strength retention in about 12 days. This difference does not raise new questions of safety or effectiveness. The fundamental mechanism of action, maintenance of ductal patency and facilitation of physiological drainage, remains the same.
| Device Properties | ARCHIMEDES Biodegradable Stent (MEDIUM) (Subject Device) | ARCHIMEDES Biodegradable Pancreatic Stent (FAST) (Predicate Device) K243412 |
| --- | --- | --- |
| Intended Use/Indications for Use | The ARCHIMEDES Biodegradable Pancreatic Stent – Medium is intended to drain pancreatic ducts in patients indicated for pancreatic duct stenting. | The ARCHIMEDES Biodegradable Pancreatic Stent – Fast is intended to drain pancreatic ducts in patients indicated for pancreatic duct stenting. The ARCHIMEDES Biodegradable Pancreatic Stent - Fast is intended to be delivered through the working channel of a therapeutic endoscope using a guidewire with |
{7}
K260835 page 4 of 12
| Device Properties | ARCHIMEDES Biodegradable Stent (MEDIUM) (Subject Device) | ARCHIMEDES Biodegradable Pancreatic Stent (FAST) (Predicate Device) K243412 |
| --- | --- | --- |
| | | a maximum outer diameter of 0.035 inches and a pushing catheter of between 5 to 10 French. |
| **Classification/ Regulation** | Class II 21 CFR 876.5010 | Class II 21 CFR 876.5010 |
| **Product Code** | FGE | FGE |
| **Device Description Summary** | The ARCHIMEDES Biodegradable Pancreatic Stent - Medium device is a single use, synthetic, biodegradable, polymeric implant, composed of polymers used extensively in implantable and absorbable medical devices. It comprises Polydioxanone (PDO) and Barium Sulphate (BaSO_{4}). | The ARCHIMEDES Biodegradable Pancreatic Stent - Fast device is a single use, synthetic, biodegradable, polymeric implant, composed of polymers used extensively in implantable and absorbable medical devices. It comprises Polydioxanone (PDO), Polyethylene Glycol (PEG) and Barium Sulphate (BaSO_{4}). |
| **Technological Characteristics** | Material: 85% wt Polydioxanone 15% wt Barium Sulfate Degradation: Minimum strength retention of 10% is reached in 20 days Design: Dual drainage – 1. Two external spiral channels that form a double helical twist. 2. Central internal guidewire lumen. Both allow for ductal flow on both the external surfaces and through the internal lumen of the stent. The stent has a form designed to fit the natural curvature of the anatomy. The | Material: 68% wt Polydioxanone 17% wt Polyethylene Glycol 15% wt Barium Sulfate Degradation: Minimum strength retention of 10% is reached in 12 days Design: Dual drainage – 1. Two external spiral channels that form a double helical twist. 2. Central internal guidewire lumen. Both allow for ductal flow on both the external surfaces and through the internal lumen of the stent. The stent has a form designed to fit the natural curvature of the anatomy. The |
{8}
K260835 page 5 of 12
| Device Properties | ARCHIMEDES Biodegradable Stent (MEDIUM) (Subject Device) | ARCHIMEDES Biodegradable Pancreatic Stent (FAST) (Predicate Device) K243412 |
| --- | --- | --- |
| | stents contain tapered tips to allow for atraumatic insertion into the papilla. The stent contains flap and split end features to minimize migration: the 6 Fr (2.0 mm diameter) model contains a flap and split end feature and the 8 Fr (2.6 mm) and 10 Fr (3.4 mm) models contain a flap on each end. Diameters: 2.0 mm (6Fr), 2.6 mm (8Fr), 3.4 mm (10Fr) Lengths: 2.0 mm Ø – 40, 60, 80, 100, 125, 150, 175 mm 2.6 and 3.4 mm Ø – 40, 60, 80, 100, 125, 150, 175, 200, 225 mm | stents contain tapered tips to allow for atraumatic insertion into the papilla. The stent contains flap and split end features to minimize migration: the 6 Fr (2.0 mm diameter) model contains a flap and split end feature and the 8 Fr (2.6 mm) and 10 Fr (3.4 mm) models contain a flap on each end. Diameters: 2.0 mm (6Fr), 2.6 mm (8Fr), 3.4 mm (10Fr) Lengths: 2.0 mm Ø – 40, 60, 80, 100, 125, 150, 175 mm 2.6 and 3.4 mm Ø – 40, 60, 80, 100, 125, 150, 175, 200, 225 mm |
| **Accessories** | The following accessories are required for use of the ARCHIMEDES MEDIUM stent. Of the accessories listed below, only the introducer sleeve is supplied with the ARCHIMEDES stent. - Introducer sleeve (supplied with ARCHIMEDES) - Therapeutic duodenoscope with a 4.2mm working channel - Guidewire: up to 0.035' - Pushing catheter: size dependent on physician's preference - Fluoroscopy | The following accessories are required for use of the ARCHIMEDES FAST stent. Of the accessories listed below, only the introducer sleeve is supplied with the ARCHIMEDES stent. - Introducer sleeve (supplied with ARCHIMEDES) - Therapeutic duodenoscope with a 4.2mm working channel - Guidewire: up to 0.035' - Pushing catheter: size dependent on physician's preference - Fluoroscopy |
| **Dimensions** | The stent is offered in outer diameters of 2.0mm, 2.6mm, or 3.4 | The stent is offered in outer diameters of 2.0mm, 2.6mm, or 3.4 |
{9}
K260835 page 6 of 12
| Device Properties | ARCHIMEDES Biodegradable Stent (MEDIUM) (Subject Device) | ARCHIMEDES Biodegradable Pancreatic Stent (FAST) (Predicate Device) K243412 |
| --- | --- | --- |
| | mm, and in various lengths to accommodate variations in pancreatic anatomy across individuals (40, 60, 80, 100, 125, 150, 175, 200, and 225 mm). | mm, and in various lengths to accommodate variations in pancreatic anatomy across individuals (40, 60, 80, 100, 125, 150, 175, 200, and 225 mm). |
| **Biocompatibility** | Stent is contact with tissue/bone for a permanent duration (>30 days). Accordingly, the device has been tested per GLP for the following biocompatibility endpoints in accordance with ISO 10993-1: 2009: - Cytotoxicity per ISO 10993-5; - Sensitization per ISO 10993-10; - Irritation or intracutaneous reactivity per ISO 10993-10; - Acute systemic toxicity per ISO 10993-11; - Subacute/subchronic toxicity per ISO 10993-11; - Implantation per ISO 10993-6; and - Material-mediated pyrogenicity per ISO 10993-11. - 28 Day Dual Route IV/IP Systemic Toxicity per ISO 10993-11 - Chemical Extractable Studies and Toxicological Risk Assessment per FDA Guidance “Use of International Standard ISO 10993-1” | Stent is contact with tissue/bone for a permanent duration (>30 days). Accordingly, the device has been tested per GLP for the following biocompatibility endpoints in accordance with ISO 10993-1: 2009: - Cytotoxicity per ISO 10993-5; - Sensitization per ISO 10993-10; - Irritation or intracutaneous reactivity per ISO 10993-10; - Acute systemic toxicity per ISO 10993-11; - Subacute/subchronic toxicity per ISO 10993-11; - Implantation per ISO 10993-6; and - Material-mediated pyrogenicity per ISO 10993-11. - 28 Day Dual Route IV/IP Systemic Toxicity per ISO 10993-11 - Chemical Extractable Studies and Toxicological Risk Assessment per FDA Guidance “Use of International Standard ISO 10993-1” |
| **Sterilization** | The device meets the requirements of ISO 11135-1 “Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and | The device meets the requirements of ISO 11135-1 “Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and |
{10}
K260835 page 7 of 12
| Device Properties | ARCHIMEDES Biodegradable Stent (MEDIUM) (Subject Device) | ARCHIMEDES Biodegradable Pancreatic Stent (FAST) (Predicate Device) K243412 |
| --- | --- | --- |
| | Routine Control of Sterilization processes for Medical Devices', and ISO 10993-7 'Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals'. | Routine Control of Sterilization processes for Medical Devices', and ISO 10993-7 'Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals'. |
| **Singe Use/Reuse** | Single use only | Single use only |
### 9. Comparison to the Predicate Device (Biliary indication)
The ARCHIMEDES™ Biodegradable Stent (Medium) is substantially equivalent to the currently marketed Advanix™ Biliary Stent with NaviFlex™ RX Delivery System (K203162) as follows: same intended use and principles of operations, same packaging and sterilization methods; and similar technological characteristics.
The ARCHIMEDES™ Medium stent shares key technological characteristics with the predicate device in that they are both implantable stents intended to drain biliary ducts.
The main technological difference between the ARCHIMEDES™ Medium and the proposed predicate device is the stent material. The ARCHIMEDES stent is made of a biodegradable polymer that degrades in the patient's body over time, and thus does not require removal. The Advanix™ stent is also made of a polymer, but it is not biodegradable and therefore must be removed. However, this difference does not raise new questions of safety or effectiveness. The ARCHIMEDES™ Medium stent operates in the same manner as existing non-biodegradable biliary stents for which the risks are well understood and adequately mitigated through general and special controls.
| Device Properties | ARCHIMEDES Biodegradable Biliary Stent – MEDIUM degrading (Subject Device) | Advanix Biliary Stent with NaviFlex RX Delivery System (Primary Predicate) K203162 |
| --- | --- | --- |
| **Intended Use/Indications for Use** | The ARCHIMEDES Biodegradable Biliary Stent – Medium is intended to drain biliary ducts in patients indicated for biliary duct stenting. | Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct |
{11}
K260835 page 8 of 12
| Device Properties | ARCHIMEDES Biodegradable Biliary Stent – MEDIUM degrading (Subject Device) | Advanix Biliary Stent with NaviFlex RX Delivery System (Primary Predicate) K203162 |
| --- | --- | --- |
| | | during healing, or for providing bile duct patency in a stricture or past a stone. |
| **Classification/ Regulation** | Class II 21 CFR 876.5010 | Class II 21 CFR 876.5010 |
| **Product Code** | FGE | FGE |
| **Device Description Summary** | The ARCHIMEDES Biodegradable Biliary Stent – Medium device is a single use, synthetic, biodegradable, polymeric implant, composed of polymers used extensively in implantable and absorbable medical devices. It comprises of Polydioxanone (PDO) and Barium Sulphate (BaSO_{4}). | The Advanix™ Biliary Stent with NaviFlex™ RX Delivery System consists of biliary plastic stents and delivery system catheters. They are sold separately or in a pre-loaded configuration, in which a stent comes attached to the catheter via a suture. The system is comprised of two (2) primary components: stent and delivery catheter with a locking mechanism. The Advanix Biliary Stent allows for drainage of the biliary duct by preventing closure and maintaining the patency of the biliary duct. These biliary stents are provided in center bend and duodenal bend shapes and have leading and trailing barbs, a tapered leading end tip to facilitate access through the papilla, and a rounded tailing end to match the profile of the push catheter portion of the delivery system. |
| **Technological Characteristics** | Material: 85% wt Polydioxanone 15% wt Barium Sulfate Degradation: | Material: Stent is constructed of a radiopaque (RO) polymer. Tips: |
{12}
K260835 page 9 of 12
| Device Properties | ARCHIMEDES Biodegradable Biliary Stent – MEDIUM degrading (Subject Device) | Advanix Biliary Stent with NaviFlex RX Delivery System (Primary Predicate) K203162 |
| --- | --- | --- |
| | Minimum strength retention of 10% is reached in 20 days Design: Dual drainage – 1. Two external spiral channels that form a double helical twist. 2. Central internal guidewire lumen. Both allow for ductal flow on both the external surfaces and through the internal lumen of the stent. The stent has a form designed to fit the natural curvature of the anatomy. The stents contain tapered tips to allow for atraumatic insertion into the papilla. The stent contains flap and split end features to minimize migration: the 6 Fr (2.0 mm diameter) model contains a flap and split end feature and the 8 Fr (2.6 mm) and 10 Fr (3.4 mm) models contain a flap on each end. Diameters: 2.0 mm (6Fr), 2.6 mm (8Fr), 3.4 mm (10Fr) Lengths: 2.0 mm Ø – 40, 60, 80, 100, 125, 150, 175 mm 2.6 and 3.4 mm Ø – 40, 60, 80, 100, 125, 150, 175, 200, 225 mm | The leading end has a tapered tip and the trailing end of the stent has rounded tip. Markers: Trailing end of stent has RO markers for placement guidance. Shapes: Center bend, duodenal bend, and double pigtail Barbs: Stents have leading and trailing barbs except for pigtail shape. Diameters: 7 F, 8.5 F, and 10F Lengths: 5 cm up to 18 cm Delivery system: NaviFlex™ RX delivery catheter has an ergonomic handle and locking mechanism for controlled deployment and repositionability. Compatibility with standard guidewires and endoscope working channels. Fluoroscopic and endoscopic markers on the delivery system to assist accurate placement. |
| **Accessories** | The following accessories are required for use of the ARCHIMEDES Medium stent. Of the accessories listed below, only the | The following accessories are required for use of the Advanix Biliary Stent with NaviFlex RX Delivery System. |
{13}
K260835 page 10 of 12
| Device Properties | ARCHIMEDES Biodegradable Biliary Stent – MEDIUM degrading (Subject Device) | Advanix Biliary Stent with NaviFlex RX Delivery System (Primary Predicate) K203162 |
| --- | --- | --- |
| | introducer sleeve is supplied with the ARCHIMEDES Medium stent. - Introducer sleeve (supplied with ARCHIMEDES) - Therapeutic duodenoscope with a 4.2mm working channel - Guidewire: up to 0.035” - Pushing catheter: size dependent on physician’s preference - Fluoroscopy | - Therapeutic duodenoscope with a 4.2mm working channel - Guidewire: up to 0.035” - RX Locking device - Fluoroscopy |
| **Dimensions** | The stent is offered in outer diameters of 2.0 mm (6 F), 2.6 mm (8 F), or 3.4 mm (10 F), and in various lengths to accommodate variations in pancreatic anatomy across individuals (40, 60, 80, 100, 125, 150, 175, 200, and 225 mm). | The stent is offered in outer diameters of 7 F, 8.5 F, and 10F, and in various lengths to accommodate variations in biliary anatomy across individuals (5 cm up to 18 cm ). |
| **Biocompatibility** | Stent is contact with tissue/bone for a permanent duration (>30 days). Accordingly, the device has been tested per GLP for the following biocompatibility endpoints in accordance with ISO 10993-1: 2009: - Cytotoxicity per ISO 10993-5; - Sensitization per ISO 10993-10; - Irritation or intracutaneous reactivity per ISO 10993-10; - Acute systemic toxicity per ISO 10993-11; - Subacute/subchronic toxicity per ISO 10993-11; - Implantation per ISO 10993-6; and - Material-mediated pyrogenicity per ISO 10993-11. | The device meets the requirements of ISO 10993 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a risk management process”. The following Biocompatibility tests were performed on the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System: MEM Elution Cytotoxicity; Guinea Pig Maximization Sensitization; Intracutaneous Reactivity; Acute Systemic Injection; and Implantation. In addition, the results of chemical extractable studies are provided which meet FDA’s Guidance document, entitled, “Use of International |
{14}
K260835 page 11 of 12
| Device Properties | ARCHIMEDES Biodegradable Biliary Stent – MEDIUM degrading (Subject Device) | Advanix Biliary Stent with NaviFlex RX Delivery System (Primary Predicate) K203162 |
| --- | --- | --- |
| | - 28 Day Dual Route IV/IP Systemic Toxicity per ISO 10993-11 - Chemical Extractable Studies and Toxicological Risk Assessment per FDA Guidance “Use of International Standard ISO 10993-1” | Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.” |
| **Sterilization** | The device meets the requirements of ISO 11135-1 “Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices”, and ISO 10993-7 “Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals”. | The device meets the requirements of ISO 11135-1 “Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices”, and ISO 10993-7 “Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals”. |
| **Singe Use/Reuse** | Single use only | Single use only |
### 10. Performance Data
Non-clinical performance
The proposed device meets the requirements of ISO 10993 “Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a risk management process”, ISO 11135-1 “Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices”, and ISO 10993-7 “Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals”.
The following Biocompatibility tests were performed: MEM Elution Cytotoxicity; Implantation; Guinea Pig Maximization Sensitization; 28 Day Dual Route IV/IP Systemic Toxicity; Acute Systemic Injection; Intracutaneous Reactivity; and Material Mediated Pyrogen. In addition, the results of chemical extractable studies and Toxicological Risk
{15}
K260835 page 12 of 12
Assessment are provided which meet FDA’s Guidance document, entitled, “Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process.”
The proposed device bench testing is in alignment with FDA guidance document, “Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions”. The following bench tests were performed on the ARCHIMEDES™ Biodegradable Stent (Medium): Visual Inspection; Outer diameter, Length; Flow Rate; Inherent Viscosity; Guidewire Compatibility; Endoscope Compatibility; Introducer Sleeve Compatibility; Simulated Use; Trackability; Pushability; Flexibility/Kink Resistance; Retraction Force; Tensile Strength; Fluoroscopic Visibility; and Degradation.
All tests met required acceptance criteria, therefore supporting the finding of substantial equivalence.
#### Clinical studies
The safety and efficacy of ARCHIMEDES™ Biodegradable Stent (Medium) is also supported by five clinical studies. The studies were conducted in various settings, including single-center and multicenter, prospective and retrospective, randomized and non-randomized designs. They included a diverse range of patients with different indications for stent placement and a variety of pancreatic and biliary conditions. The studies collectively evaluated various aspects of the ARCHIMEDES™ Biodegradable Stent (Medium), including its safety, efficacy, clinical performance, surgical and endoscopic placement, degradation profile, and technical success.
### 11. Conclusion
Q3 Medical LLC has demonstrated through its bench and clinical studies that the proposed ARCHIMEDES™ Biodegradable Stent (Medium) is substantially equivalent to the legally marketed predicate devices ARCHIMEDES™ Biodegradable Pancreatic Stent (Fast) (K243412) and Advanix™ Biliary Stent with NaviFlex™ RX Delivery System (K203162).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.