Advanix Biliary Stent with NaviFlex RX Delivery System

K203162 · Boston Scientific Corporation · FGE · Dec 18, 2020 · Gastroenterology, Urology

Device Facts

Record IDK203162
Device NameAdvanix Biliary Stent with NaviFlex RX Delivery System
ApplicantBoston Scientific Corporation
Product CodeFGE · Gastroenterology, Urology
Decision DateDec 18, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5010
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.

Device Story

Biliary plastic stent and delivery system; used for biliary drainage, splinting, or maintaining patency in strictures/stones. System includes stent and delivery catheter with locking mechanism; stent features center or duodenal bend shapes, tapered leading tip, and rounded trailing end. Operated by clinicians in clinical settings. Stent prevents duct closure; delivery system facilitates placement via papilla. Device is provided sterile; material change implemented for specific 7F and 8.5F configurations.

Clinical Evidence

Bench testing only. Biocompatibility testing performed: MEM Elution Cytotoxicity, Guinea Pig Maximization, Intracutaneous Reactivity, Acute Systemic Injection, and Implantation. Bench performance testing included: drainage lumen ID, stent dimensions, shape, repositionability, deployment force, trackability, guidewire/duodenoscopy compatibility, barb flap cover compatibility, tensile strength (aged in bile), and lumen patency. All tests met acceptance criteria.

Technological Characteristics

Biliary plastic stent and delivery catheter. Materials evaluated per ISO 10993. Sterilization via Ethylene Oxide (ISO 11135-1, ISO 10993-7). Configurations: 7F and 8.5F, lengths 5-18cm, center or duodenal bend. Mechanical delivery system with locking mechanism.

Indications for Use

Indicated for patients requiring biliary tract drainage, bile duct splinting during healing, or maintenance of bile duct patency in the presence of a stricture or stone.

Regulatory Classification

Identification

A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.

Special Controls

*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that. December 18, 2020 Boston Scientific Corporation Elena Hennessey Fellow, Regulatory Affairs Specialist 100 Boston Scientific Wav Marlborough, MA 01752 Re: K203162 Trade/Device Name: Advanix Biliary Stent with NaviFlex RX Delivery System Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: Class II Product Code: FGE Dated: October 21, 2020 Received: October 23, 2020 # Dear Elena Hennessey: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thelma I. Valdes, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K203162 #### Device Name Advanix™ Biliary Stent with NaviFlex™ RX Delivery System #### Indications for Use (Describe) Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CER 801 Subpart D) | <span style="text-decoration: underline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | |-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------| |-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------| ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K203162 Page 1 of 2 ## SECTION 5 510(K) SUMMARY #### 510(k) SUMMARY 1. Submitter: Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Contact: Elena Hennessev Fellow Regulatory Affairs Specialist Tel: 508-683-4347 Date Prepared: December 15, 2020 ## 2. Proposed Device: Trade Name: Advanix™ Biliary Stent with NaviFlex™ RX Delivery System Common Name: Stents, Drains And Dilators For The Biliary Ducts Regulation Name: Biliary Catheters and Accessories Regulation Number: 876.5010 Product Code: FGE Classification: Class II ## 3. Predicate Device: ### Predicate Device: Trade Name: Advanix™ Biliary Stent with NaviFlex™ RX Delivery System Common Name: Stents, Drains And Dilators For The Biliary Ducts Regulation Name: Biliary Catheters and Accessories Regulation Number: 876.5010 Product Code: FGE Classification: Class II 510(k) Clearance Number: K101314 #### 4. Device Description: The Advanix™ Biliary Stent with NaviFlex™ RX Delivery System consists of biliary plastic stents and delivery system catheters. They are sold separately or in a pre-loaded configuration, in which a stent comes attached to the catheter via a suture. The system is comprised of two (2) primary components: stent and delivery catheter with a locking mechanism. The Advanix Biliary Stent allows for drainage of the biliary duct by preventing closure and maintaining the patency of the biliary duct. These biliary stents are provided in center bend and duodenal bend shapes and have leading and trailing bars, a tapered leading end tip to facilitate access through the papilla, and a rounded tailing end to match the profile of the push catheter portion of the delivery system. The Advanix Biliary Stent with NaviFlex RX Delivery System is currently cleared under K101314. This proposed material change will only impact the Advanix Biliary 7F and 8.5F stent singles and the 7F and 8.5F pre-loaded configurations as shown below in Table 4-1: {4}------------------------------------------------ | | Advanix Biliary Stent | | |---------------------|--------------------------------------------------------|--| | Shape | • Center Bend | | | | • Duodenal Bend | | | Sizes (Fr) | • 7F | | | | • 8.5F | | | Lengths (cm) | • 5cm | | | | • 7cm | | | | • 9cm | | | | • 12cm | | | | • 15cm | | | | • 18cm | | | Single or Stent Set | • Center Bend Single Stent (no delivery system) | | | | • Duodenal Bend Single Stent (no delivery system) | | | | • Center Bend Pre-loaded (stent and delivery system) | | | | • Duodenal Bend Pre-loaded (stent and delivery system) | | ## Table 4-1 - Advanix Biliary Stent Material Change Stent Configurations: ## 5. Indications for Use: The proposed Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone. ## 6. Technological Characteristics: The proposed Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is identical to the predicates in design, intended use and all materials with the exception of the stent material. ## 7. Performance Data: The proposed device meets the requirements of ISO 10993 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a risk management process", ISO 11135-1 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices", and ISO 10993-7 "Biological evaluation of medical devices -Part 7: ethylene oxide sterilization residuals". The following Biocompatibility tests were performed on the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System: MEM Elution Cytotoxicity; Guinea Pig Maximization; Intracutaneous Reactivity; Acute Systemic Injection; and Implantation. In addition, the results of chemical extractable studies are provided which meet FDA's Guidance document, entitled, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process." The proposed device bench testing is in alignment with FDA guidance document, "Recommended Content and Format of Non-Clinical Bench Performance Testing Information in Premarket Submissions". The following bench tests were performed on the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System: Drainage Lumen ID; Stent Length; Stent OD; Stent Shape; Repositionability (pre-loaded system only); Deployment Force; Device Trackability Force; Guidewire Compatibility; Duodenoscopy Compatibility; Barb Flap Cover Compatibility; Tensile: Aged in Bile; and Lumen Patency. All tests met required acceptance criteria, therefore supporting a finding of substantial equivalence. ## 8. Conclusion: The information Boston Scientific Corporation provided in this submission demonstrates that the proposed Advanix Biliary Stent with NaviFlex RX Delivery System is substantially equivalent to the currently cleared Advanix Biliary Stent with NaviFlex RX Delivery System (K101314).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...