K260398 · GE Medical Systems Ultrasound and Primary Care Diagnostics · IYN · May 26, 2026 · Radiology
Device Facts
Record ID
K260398
Device Name
LOGIQ e
Applicant
GE Medical Systems Ultrasound and Primary Care Diagnostics
Product Code
IYN · Radiology
Decision Date
May 26, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The LOGIQ e is a general-purpose diagnostic ultrasound system for use by qualified and trained HealthCare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. LOGIQ e is intended to be used in a hospital or medical clinic. LOGIQ e clinical application include Ophthalmic, Fetal/OB, Abdominal (GYN & Urology), Pediatric, Small organ (Breast, Testes, Thyroid), Neonatal and Adult cephalic, Cardiac (Adult & Pediatric), Peripheral vascular, Musculoskeletal conventional & superficial, Transrectal, Transvaginal, Transesophageal, Intraoperative (Abdominal, Thoracic and Peripheral), Thoracic/pleural for motion and fluid detection and imaging guidance of interventional procedures (e.g. Nerve block; vascular access). Mode of operation include B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/PWD, B/Color/PWD, B/Power/ PWD.
Device Story
LOGIQ e is a compact, battery-powered diagnostic ultrasound console with LCD and control panel. It processes ultrasound signals from various electronic array transducers (linear, curved, sector/phased, TEE) to produce real-time images and measurements. Operated by trained clinicians in hospitals or clinics, it supports B, M, Doppler, and harmonic imaging modes. New features include voice control, remote fleet management, and connectivity wizards. Data is stored locally or transmitted via LAN/wireless. Output assists clinicians in diagnosis, hemodynamic assessment (B-Flow), fetal biometry (SonoBiometry), and cardiac function analysis (AFI). The system integrates with carts and printers, and supports software updates via GEHC cloud. It facilitates interventional guidance and clinical decision-making through high-resolution imaging and automated analysis tools.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence is based on bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal/electrical/electromagnetic safety, demonstrating conformance to recognized standards (e.g., IEC 60601-1, IEC 60601-2-37, ISO 10993-1).
Technological Characteristics
Compact ultrasound console with LCD, control panel, and battery/AC power. Supports linear, curved, sector/phased, and TEE electronic array transducers. Connectivity via LAN and wireless USB-adapter. OS: Windows 11. CPU: Intel I5-125U. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, and DICOM standards. Biocompatible patient-contact materials.
Indications for Use
Indicated for qualified healthcare professionals to perform ultrasound imaging, measurement, and analysis of the human body and fluid in hospital or clinic settings. Patient population includes neonates, infants, children, adolescents, and adults. Clinical applications cover ophthalmic, fetal/OB, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, peripheral vascular, musculoskeletal, transrectal, transvaginal, transesophageal, and intraoperative procedures, plus interventional guidance.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
GE Medical Systems Ultrasound and Primary Care Diagnostics
Lee Bush
Regulatory Affairs Director
3200 N Grandview Blvd.
Waukesha, Wisconsin 53188
May 26, 2026
Re: K260398
Trade/Device Name: LOGIQ e
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX
Dated: February 6, 2026
Received: February 6, 2026
Dear Lee Bush:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260398 - Lee Bush
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260398 - Lee Bush
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2026.05.26 10:10:15 -04'00" For
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260398 | ? |
| Please provide the device trade name(s). | | ? |
| LOGIQ e | | |
| Please provide your Indications for Use below. | | ? |
| The LOGIQ e is a general-purpose diagnostic ultrasound system for use by qualified and trained HealthCare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. LOGIQ e is intended to be used in a hospital or medical clinic. LOGIQ e clinical application include Ophthalmic, Fetal/OB, Abdominal (GYN & Urology), Pediatric, Small organ (Breast, Testes, Thyroid), Neonatal and Adult cephalic, Cardiac (Adult & Pediatric), Peripheral vascular, Musculoskeletal conventional & superficial, Transrectal, Transvaginal, Transesophageal, Intraoperative (Abdominal, Thoracic and Peripheral), Thoracic/pleural for motion and fluid detection and imaging guidance of interventional procedures (e.g. Nerve block; vascular access). Mode of operation include B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/PWD, B/Color/PWD, B/Power/ PWD. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☑ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
GE HealthCare
GE HealthCare
510(k) Premarket Notification Submission
510(k) Summary – K260398
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: May 4, 2026
Submitter: GE Medical Systems Ultrasound and Primary Care Diagnostics
3200 N Grandview Blvd
Waukesha, WI 53188, USA
Manufacturer: GE Medical Systems (China) Co., Ltd.
No.19, ChangJiang Road, WuXi National Hi-Tech Dev. Zone,
214028 Jiangsu China
Primary Contact Person: Lee Bush
Regulatory Affairs Director
GE HealthCare
T:(262) 309-9429
Alternate Contact Person: Cui Fang
Regulatory Affairs Specialist
GE HealthCare
Device Trade Name: LOGIQ e
Common/Usual Name: Diagnostic Ultrasound System
Classification Names: Class II
Product Code: IYN (primary), IYO, ITX (secondary)
Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550, 90-IYN;
Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;
Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX
Primary Predicate Device: K232186 LOGIQ e Diagnostic Ultrasound System
Reference Device(s): K242005 Versana Premier
K250087 Vscan Air
K251322 Venue
K223832 Vivid S70N, Vivid S60N
K200278 Vivid iq
K232381 LOGIQ Totus
K251963 LOGIQ E10s
Classification Names: Class II
Product Code: IYN (primary), IYO, ITX (secondary)
Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550, 90-IYN;
Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;
Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX
{5}
GE HealthCare
510(k) Premarket Notification Submission
Device Description: The proposed LOGIQ e system is a general-purpose, Track 3, diagnostic ultrasound device, intended for ultrasound imaging, measurement, display and analysis of the human body and fluid.
It is an ultrasound imaging & analysis system, consisting of a compact console with LCD, and control panel with new keyboard.
The system has digital acquisition, processing and display capability and operates from an integrated battery or AC/DC power adapter. It has one battery pack as standard configuration and also with an optional battery pack to be provided for additional power for longer scanning time.
The system also has two optional carts. The primary LOGIQ e cart is height-adjustable cart for comfortable standing and sitting positions. The cart is approx. 510mm length, 510mm width and adjustable height from 830 to 1130mm with optional battery box consisting of a charger box and 4pcs battery Packs to supply the system power from the cart. The LOGIQ e simple cart is approx. 555mm length, 545mm width and fixed height 875mm without powered electrical components or functions.
LOGIQ e utilizes a variety of electronic array transducers operating in linear, curved, sector/phased array, TEE.
The system includes electronics for transmit and receive of ultrasound data, ultrasound signal processing, software computing, hardware for image storage, printing, and network access to the facility through both LAN and wireless (supported by use of a wireless LAN USB-adapter) connection.
The system supports electronic delivery of software. The system allows the user, in addition to service personnel, to update the SW by logging into a GEHC website to download SW available to them and install it on the system. Download from the GEHC website requires an account.
{6}
GE HealthCare
510(k) Premarket Notification Submission
Indications for Use:
The LOGIQ e is a general-purpose diagnostic ultrasound system for use by qualified and trained HealthCare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid.
LOGIQ e is intended to be used in a hospital or medical clinic.
LOGIQ e clinical application include Ophthalmic, Fetal/OB, Abdominal (GYN & Urology), Pediatric, Small organ (Breast, Testes, Thyroid), Neonatal and Adult cephalic, Cardiac (Adult & Pediatric), Peripheral vascular, Musculoskeletal conventional & superficial, Transrectal, Transvaginal, Transesophageal, Intraoperative (Abdominal, Thoracic and Peripheral), Thoracic/pleural for motion and fluid detection and imaging guidance of interventional procedures (e.g. Nerve block; vascular access). Mode of operation include B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/PWD, B/Color/PWD, B/Power/PWD.
Technology: The LOGIQ e employs the same fundamental scientific technology as its predicate device(s).
Determination of Substantial
Comparison to Predicates
Equivalence: The proposed LOGIQ e is substantially equivalent to the predicate devices. The following is an overview of the differences between the LOGIQ e (K232186).
Transducers:
The proposed LOGIQ e and predicate LOGIQ e (K232186) are similar, except for:
- Addition of L6-12-RS previously cleared on Versana Premier (K242005). The clinical applications and imaging modes of L6-12-RS are similar to those available on the reference device Versana Premier (K242005).
- Addition of IC9-RS: previously cleared on Versana Premier (K242005). The clinical applications and imaging modes of IC9-RS are similar to those available on the reference device Versana Premier (K242005).
- Addition of C2-9-RS: previously cleared on Venue (K251322). The clinical applications and imaging modes of C2-9-RS are similar to those available on the reference device Venue (K251322). However, the additional application of Abdominal, Fetal/Obstetrics and Urology are also being added, these applications were cleared on Vivid S70N, Vivid S60N (K223832) with the C2-9-D probe. The C2-9-RS probe is the same as the C2-9-D except it uses an RS connector instead of a D connector. Furthermore, the application of
{7}
GE HealthCare
GE HealthCare
510(k) Premarket Notification Submission
Thoracic/Pleural is also being added, which was already cleared with a similar curved transducer C1-5-RS on predicate device LOGIQ e (K232186).
- Addition of Vscan Air CL and Vscan Air SL probes, which were cleared on Vscan Air (K250087). The clinical applications and imaging modes are similar with reference device Vscan Air (K250087). The Vscan Air CL and Vscan Air SL have also been cleared on several other consoles such as Venue (K251322)
- Addition of High-Res PDI imaging mode on existing probe ML6-15-RS, both imaging mode and probe are cleared on predicate device LOGIQ e (K232186)
# Software Features/Functionality:
- B Flow: same as cleared on Versana Premier (K242005), it is intended to provide a more intuitive representation of non-quantitative hemodynamics in vascular structures.
- AFI: similar to AFI on cleared device Vivid iq (K200278), it is a tool for global and regional assessment of the systolic function of the left ventricle (LV), right ventricle (RV), and left atrium (LA). AFI calculates the myocardial tissue deformation based on feature tracking on 2D grey scale loops.
- SonoBiometry: same as cleared on Versana Premier (K242005), it is an alternative to the common fetal biometry measurements.
- Voice Control: similar to Voice Control on cleared device LOGIQ Totus (K232381), it lets users activate certain functions by speaking recognized commands.
- MyPreset: same as cleared on Versana Premier (K242005), allows users to customize and save frequently used scanning presets.
- Fleet Management: same as cleared on LOGIQ E10s (K251963), is a remote device management tool that enables bi-directional management capabilities on the device which allows cloud management of system preset configurations to a fleet of systems on network, as well as one to one system preset configuration cloud backup and restore functions.
- Connection Manager: similar to Connection Manager cleared on LOGIQ E10s (K251963), it is a wizard tool to set up connectivity step-by-step, the configuration progresses through Network, Remote devices, Print Flow, Dataflow, and Local devices setup screens.
- APP API: same as cleared on Venue (K251322), it is a standardized interface tool to facilitate the integration of multiple software components.
- OS updated from Windows 10 to Windows 11
{8}
GE HealthCare
GE HealthCare
510(k) Premarket Notification Submission
## Hardware:
- CPU was updated to Intel I5-125U
- Added two-probe port option
- CWI board was updated to support Voice Control
## Accessories:
- Added Digital Expert Connect
- Added wireless chargers for Vscan Air CL and Vscan Air SL probes
- Added WiFi-Bluetooth USB adaptor to allow wireless connectivity with the Vscan Air CL and Vscan Air SL probes
- Added LOGIQ e Simple Cart
- Added compatible OEM biopsy guide accessory compatibility for the L6-12-RS, C2-9-RS and IC9-RS Transducers
- Added the SONY UP-D899MD printer
## Summary of Non-Clinical Tests:
The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The LOGIQ e complies with voluntary standards:
- AAMI/ANSI ES60601-1, Medical Electrical Equipment – Part 1: General Requirements for Safety, 2005/A2:2021
- IEC 60601-1-2 Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Disturbance - Requirements and Tests, Edition 4.1, 2020
- IEC 60601-2-37, Medical Electrical Equipment – Part 2-37: Particular Requirements for the Safety and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment, Edition3.0, 2024
- ISO 10993-1, Biological Evaluation of Medical Devices-Part 1: Evaluation and Testing Within Risk Management Process, Fifth edition, 2018
- ISO 14971, Application of risk management to medical devices, 2019
- NEMA PS 3.1 - 3.20, Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology), 2024e
- IEC 62359, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, Edition 2.1, 2017
- AAMI TIR69, Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems, 2017(R2020)
{9}
GE HealthCare
GE HealthCare
510(k) Premarket Notification Submission
The following quality assurance measures are applied to the development of the system:
- Risk Analysis
- Requirements Reviews
- Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification)
- Performance testing (Verification)
- Safety testing (Verification)
Transducer materials and other patient contact materials are biocompatible.
## Summary of Clinical Tests:
The subject of this premarket submission, LOGIQ e, did not require clinical studies to support substantial equivalence.
## Conclusion:
Based on the equipment design similarities, conformance to recognized performance standards, and performance testing, GE HealthCare considers the proposed LOGIQ e to be substantially equivalent to the predicate LOGIQ e (K232186).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.