K232186 · Ge Medical Systems Ultrasound and Primary Care Diagnostic, · IYN · Nov 3, 2023 · Radiology
Device Facts
Record ID
K232186
Device Name
LOGIQ e
Applicant
Ge Medical Systems Ultrasound and Primary Care Diagnostic,
Product Code
IYN · Radiology
Decision Date
Nov 3, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The LOGIQ e is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid. LOGIQ e is intended to be used in a hospital or medical clinic. LOGIQ e clinical application include: Ophthalmic, fetal/ob; abdominal (gyn & urology); pediatric; small organ (breast, testes, thyroid); neonatal and adult cephalic; cardiac (adult & pediatric); peripheral vascular; musculoskeletal conventional & superficial; transrectal; transvaginal; transesophageal; intraoperative (abdominal, thoracic and peripheral); thoracic/pleural for motion and fluid detection and imaging guidance of interventional procedures (e.g. Nerve block; vascular access). Mode of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/PWD, B/Color/PWD, B/Power/PWD.
Device Story
LOGIQ e is a compact, portable diagnostic ultrasound system for hospital/clinic use by trained professionals. It utilizes electronic array transducers (linear, curved, sector/phased, TEE) to acquire ultrasound data, which is processed and displayed on an integrated LCD. The system supports various imaging modes (B, M, Doppler, Harmonic, Coded Pulse). It features a new industrial design, 15.6-inch LCD, and optional secondary battery for extended scanning. Output includes real-time images, measurements, and analysis for clinical decision-making in diagnostic and interventional procedures (e.g., nerve blocks, vascular access). The system includes software tools for automated measurements (AutoEF), scan assistance, strain elastography, and transducer element checks. Data can be stored, printed, or transmitted via LAN/wireless connectivity. It benefits patients by providing high-quality imaging guidance and diagnostic data for diverse clinical applications.
Clinical Evidence
No clinical studies were required to support substantial equivalence. The device was evaluated through bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electromagnetic, and mechanical safety testing in accordance with recognized standards (e.g., IEC 60601-1, ISO 10993-1, IEC 62359).
Technological Characteristics
General-purpose ultrasound system with compact console, 15.6-inch LCD, and control panel. Utilizes electronic array transducers (linear, curved, phased, TEE). Power via AC/DC adapter or battery packs. Connectivity via LAN/wireless. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, IEC 62359, and DICOM standards. Software includes automated analysis tools and imaging assistance features.
Indications for Use
Indicated for qualified healthcare professionals to perform ultrasound imaging, measurement, and analysis of human body and fluid in hospital or clinic settings. Applications include ophthalmic, fetal/OB, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, peripheral vascular, musculoskeletal, transrectal, transvaginal, transesophageal, and intraoperative imaging, plus interventional guidance.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
GE Medical Systems Ultrasound and Primary Care Diagnostic c/o Lee Bush Regulatory Affairs Director, GE HealthCare LLC 9900 Innovation Dr. WAUWATOSA WI 53226
Re: K232186
Trade/Device Name: LOGIQ e Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: October 2, 2023 Received: October 2, 2023
Dear Lee Bush:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
November 3, 2023
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Yanna S. Kang -S
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of a circular emblem with a stylized "GE" monogram on the left. To the right of the emblem is the text "GE HealthCare" in a sans-serif font. The logo is purple.
### GE HealthCare 510(k) Premarket Notification Submission
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2023<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
Indications for Use
| 510(k) Number (if known) | K232186 | | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Device Name | LOGIQ e | | |
| Indications for Use (Describe) | The LOGIQ e is a general purpose diagnostic ultrasound system for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid.<br><br>LOGIQ e is intended to be used in a hospital or medical clinic.<br><br>LOGIQ e clinical application include: Ophthalmic, fetal/ob; abdominal (gyn & urology); pediatric; small organ (breast, testes, thyroid); neonatal and adult cephalic; cardiac (adult & pediatric); peripheral vascular; musculoskeletal conventional & superficial; transrectal; transvaginal; transesophageal; intraoperative (abdominal, thoracic and peripheral); thoracic/pleural for motion and fluid detection and imaging guidance of interventional procedures (e.g. Nerve block; vascular access). Mode of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/PWD, B/Color/PWD, B/Power/PWD. | | |
| Type of Use (Select one or both, as applicable) | <table><tr><td><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</td><td><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of the GE monogram in a circle on the left, followed by the text "GE HealthCare" on the right. The logo is purple in color. The text is in a sans-serif font.
#### 510(k) Summary
## K232186
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | 24 July, 2023 |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems Ultrasound and Primary Care Diagnostics<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 |
| Primary Contact Person: | Lee Bush<br>Regulatory Affairs Director<br>GE HealthCare<br>T: (262)309-9429 |
| Secondary Contact Person: | Yuan Wang<br>Regulatory Affairs Leader<br>GE HealthCare |
| Device Trade Name: | LOGIQ e |
| Common/Usual Name: | Diagnostic Ultrasound System |
| Classification Names: | Class II |
| Product Code(s): | Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550, 90- IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Predicate Device: | LOGIQ e (K151028), Diagnostic Ultrasound System |
| Classification Names:<br>Product Code(s): | Class II<br>Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the GE Healthcare logo. The logo consists of a circular emblem with a stylized "GE" monogram inside, followed by the text "GE HealthCare" in a simple, sans-serif font. The color scheme is a consistent shade of purple, giving the logo a modern and clean appearance.
| Reference Device: | Vivid iq (K221148), Diagnostic Ultrasound System |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Names: | Class II |
| Product Code(s): | Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Reference Device: | Versana Premier (K210438), Diagnostic Ultrasound System |
| Classification Names: | Class II |
| Product Code(s): | Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Reference Device: | Venue Go (K202233), Diagnostic Ultrasound System |
| Classification Names: | Class II |
| Product Code(s): | Ultrasonic Pulsed Doppler Imaging System, 21CFR 892.1550, 90-IYN;<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO;<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
#### Device Description:
The proposed LOGIQ e system is a general-purpose, Track 3, diagnostic ultrasound device, intended for ultrasound imaging, measurement, display and analysis of the human body and fluid.
It is an ultrasound imaging & analysis system, consisting of a compact console with LCD, and control panel with new keyboard.
The system has digital acquisition, processing and display capability and operates from an integrated battery or AC/DC power adapter. It has one battery pack as standard configuration and also with an optional battery pack to be provided for additional power for longer scanning time.
The system also has an optional height-adjustable cart for comfortable standing and sitting positions. The cart is approx. 510mm length, 510mm width and adjustable height from 830 to 1130mm with optional battery box consisting of a charger box and 4pcs battery Packs to supply the system power from the cart.
LOGIQ e utilizes a variety of electronic array transducers operating in linear, curved, sector/phased array, TEE.
The system includes electronics for transmit and receive of ultrasound data, ultrasound signal
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the GE HealthCare logo. The logo consists of the GE monogram enclosed in a circular shape on the left, followed by the text "GE HealthCare" on the right. The logo and text are both in a purple color.
processing, software computing, hardware for image storage, printing, and network access to the facility through both LAN and wireless (supported by use of a wireless LAN USB-adapter) connection.
The system supports electronic delivery of software. The system allows the user, in addition to service personnel, to update the SW by logging into a GEHC website to download SW available to them and install it on the system. Download from the GEHC website requires an account.
#### Intended Use/Indication For Use:
The LOGIO e is a general purpose diagnostic ultrasound system for use by qualified and trained HealthCare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid.
LOGIQ e is intended to be used in a hospital or medical clinic.
LOGIQ e clinical application include: Ophthalmic, fetal/ob; abdominal (gyn & urology); pediatric; small organ (breast, testes, thyroid); neonatal and adult cephalic; cardiac (adult & pediatric); peripheral vascular; musculoskeletal conventional & superficial; transrectal; transvaginal; transesophageal; intraoperative (abdominal, thoracic and peripheral); thoracic/pleural for motion and fluid detection and imaging guidance of interventional procedures (e.g. Nerve block; vascular access). Mode of operation include: B, M, PW Doppler, CW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse and Combined modes: B/M, B/PWD, B/Color/PWD, B/Power/PWD.
#### Technology:
The LOGIQ e employs the same fundamental scientific technology as its predicate device.
#### Determination of Substantial Equivalence:
The proposed LOGIQ e system is substantially equivalent to the predicate device LOGIQ e and reference devices with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness.
The following is an overview of the differences between the proposed LOGIQ e and its predicate device.
#### Indications for use:
The proposed LOGIQ e and predicate LOGIO e (K151028) have similar clinical indications for use. There are no additional clinical applications, however the indications for use wording has been modified to align with FDA's new ultrasound guidance issued on 21-Feb-
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of the GE monogram in purple on the left, followed by the text "GE HealthCare" in purple. The text is in a sans-serif font and is slightly smaller than the monogram.
2023, identify users similar to reference device Versana Premier (K210438), and provide information on the environments the system can be used in.
Transducer and modes:
- The proposed LOGIQ e and predicate LOGIQ e (K151028) have identical imaging modes.
- The proposed LOGIQ e and predicate LOGIQ e (K151028) transducers are similar, ● except for:
- Addition of L4-20t-RS, which was previous cleared on Vivid iq (K221148). The O clinical applications and imaging modes of L4-20t-RS are similar on the proposed LOGIQ e as they are on the reference device Vivid ig (K221148).
- o Addition of 12S-RS. which was previous cleared on Vivid ig (K221148). The clinical applications and imaging modes of 12S-RS are similar on the proposed LOGIQ e as they are on the reference device Vivid iq (K221148).
- The Thoracic/Pleural is added to the transducer C1-5-RS, based on the clearance of ● refence device Vivid iq (K221148).
- Pediatric are added to the transducer L10-22-RS, based on the clearance of reference ● device Venue Go (K202233).
#### Software:
- Added AutoEF cleared in Versana Premier (K210438) ●
- Added Scan assistant cleared in Versana Premier (K210438) ●
- Added Strain Elastography, cleared in Versana Premier (K210438) ●
- Added Probe check (Transducer Element Check), cleared in Versana Premier ● (K210438)
- Added Imaging Insights, cleared in Vivid iq (K221148) ●
- Added Tricefy Uplink, cleared in Vivid iq (K221148)
#### Hardware:
- New Product Industrial Design with 15.6inch LCD and new keyboard ●
- Additional option for secondary battery pack to support longer scanning time on ● batteries.
- Optimize thermal solution with low noise fan. ●
#### Accessories:
- Added compatible OEM biopsy guide accessory compatibility for the L4-20t-RS ● transducer
#### Summary of Non-Clinical Tests:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of the GE monogram in a circle on the left, followed by the words "GE HealthCare" on the right. The logo is purple in color. The text is slightly blurred.
LOGIQ e is evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electromagnetic, and mechanical safety, and have been found to comply with applicable medical device safety standards. The LOGIO e complies with voluntary standards:
- AAMI/ANSI ES60601-1, Medical Electrical Equipment -- Part 1: General Requirements for ● Safety and Essential Performance, 2005/ A2:2012
- ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021], Medical electrical equipment -. Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests [Including Amendment 1 (2021)]
- IEC 60601-2-37 Edition 2.1 2015 Medical electrical equipment Part 2-37; Particular . requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- IEC 62359 Edition 2.1 2017-09 Ultrasonics Field characterization Test methods for the . determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields
- ISO 14971, Application of risk management to medical devices, 2019
- NEMA PS 3.1 3.20, Digital Imaging and Communications in Medicine (DICOM) Set. ● (Radiology), 2021e
The following quality assurance measures are applied to the development of the system:
- Risk Analysis ●
- Requirements Reviews ●
- Design Reviews
- Testing on unit level (Module verification)
- Integration testing (System verification) ●
- Performance testing (Verification & Validation) ●
- Safety testing (Verification) ●
Transducer materials and other patient contact materials are biocompatible.
#### Summary of Clinical Tests:
The subject of this premarket submission, LOGIQ e, did not require clinical studies to support substantial equivalence.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for GE HealthCare. The logo consists of the GE monogram in a purple circle on the left, followed by the words "GE HealthCare" in purple. The text is in a sans-serif font and is slightly larger than the monogram.
#### Conclusion:
Based on the equipment design similarities, conformance to recognized performance standards, and performance testing, GE HealthCare considers the proposed LOGIQ e to be as safe, effective, and performs in a substantially equivalent manner as the predicate device LOGIQ e (K151028) and reference devices Vivid iq (K221148), Versana Premier (K214038) and Venue Go (K202233).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.