Numen™ Helia Coil Embolization System; NumenFR™ Detachment System
Applicant
MicroPort NeuroTech (Shanghai) Co., Ltd.
Product Code
HCG · Neurology
Decision Date
May 20, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Numen Helia Coil Embolization System is intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. Numen Helia Coil Embolization System is indicated for endovascular embolization of: • Intracranial aneurysms • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae • Arterial and venous embolizations in the peripheral vasculature NumenFR Detachment System is intended for use with MicroPort NeuroTech detachable coils in the embolization of intracranial aneurysms and other vascular abnormalities of the neuro and peripheral vasculature
Device Story
System comprises Numen Helia Coil Embolization System and NumenFR Detachment System. Coil system includes introducer sheath, pusher, and permanent platinum/tungsten coil implant; used to occlude blood flow in vascular abnormalities. NumenFR is a sterile, handheld, battery-operated device used to trigger electrolytic detachment of coils. Used in neurovascular and peripheral procedures; operated by clinicians. Coil delivered via microcatheter to target lesion; detachment performed via NumenFR. Output is permanent mechanical occlusion of vessel/aneurysm. Benefits include minimally invasive treatment of vascular abnormalities.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing including visual inspection, dimensional verification, simulated use in tortuous models, fatigue testing, detachment reliability, friction testing, kink resistance, torque strength, and ISO 11070-compliant flexing/fracture tests. Biocompatibility evaluated via PTT and canine thrombogenicity studies per ISO 10993-4.
Indicated for endovascular embolization of intracranial aneurysms, neurovascular abnormalities (arteriovenous malformations, arteriovenous fistulae), and peripheral arterial/venous embolizations.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 20, 2026
MicroPort NeuroTech (Shanghai) Co., Ltd.
% John Skousen
Regulatory Consultant
MicroPort CRM USA, Inc.
5677 Airport Rd.
Arlington, Tennessee 38002
Re: K260351
Trade/Device Name: Numen Helia Coil Embolization System; NumenFR Detachment System
Regulation Number: 21 CFR 882.5950
Regulation Name: Neurovascular Embolization Device
Regulatory Class: Class II
Product Code: HCG, KRD
Dated: April 22, 2026
Received: April 22, 2026
Dear John Skousen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260351 - John Skousen
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260351 - John Skousen
Page 3
Sincerely,
Eda Gjika -S
For Sara S. Thompson, DVM
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260351 | |
| Device Name Numen Helia Coil Embolization System; NumenFR Detachment System | |
| Indications for Use (Describe) | |
| Numen Helia Coil Embolization System is intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. Numen Helia Coil Embolization System is indicated for endovascular embolization of: • Intracranial aneurysms • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae • Arterial and venous embolizations in the peripheral vasculature | |
| NumenFR Detachment System is intended for use with MicroPort NeuroTech detachable coils in the embolization of intracranial aneurysms and other vascular abnormalities of the neuro and peripheral vasculature | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K260351 Page 1 of 7
# 510(k) Summary
## Subject Device:
Numen™ Helia Coil Embolization System; NumenFR™ Detachment System
This 510(k) Summary is being submitted in accordance with the requirements of 21CFR § 807.92.
| Submitter Name and Address | MicroPort NeuroTech (Shanghai) Co., Ltd. Building#16, 222 Guangdan Road, Pudong New District, Shanghai, China |
| --- | --- |
| Contact Person | Name: Yuying Chen Email: YuYing.Chen@microport.com Phone:+86 13311583098 |
| Date Prepared | February 3, 2026 |
| Trade Name | Numen™ Helia Coil Embolization System; NumenFR™ Detachment System |
| Common Name | Detachable Coil |
| Classification Name | Neurovascular Embolization Device (HCG); Device, Vascular, for Promoting Embolization (KRD) |
| Regulation Number | 21 CFR 882.5950 (HCG); 21 CFR 870.3300 (KRD) |
| Product Code(s) | HCG, KRD |
| Classification | II |
| Review Panel | Neurology (HCG); Cardiovascular (KRD) |
| Use | Prescription Use Only |
| Legally Marketed Predicate Device | Numen™ Coil Embolization System; NumenFR™ Detachment System (K242154) |
510(k) Summary
{5}
K260351 Page 2 of 7
# 1. Device Description
MicroPort NeuroTech has developed the Numen™ Helia Coil Embolization System. The Numen™ Helia Coil Embolization System is designed to be used in conjunction with the NumenFR™ Detachment System (sold separately) for endovascular embolization of vascular abnormalities described in the intended use.
The Numen™ Helia Coil Embolization System is composed of two parts as described below:
- An introducer sheath: The function of the introducer sheath is to facilitate introduction of the coil into the microcatheter.
- The coil system: The coil system is composed of a pusher and coil implant. The coil is a permanent implant intended to occlude blood flow in vascular abnormalities. The pusher is used to deliver the coil implant to the target lesion.
The NumenFR™ Detachment System is a sterile, handheld, single-patient use device designed for use with the MicroPort NeuroTech detachable coils. The device is operated by two pre-loaded batteries.
# 2. Indications for Use
Numen™ Helia Coil Embolization System is intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Numen™ Helia Coil Embolization System is indicated for endovascular embolization of:
- Intracranial aneurysms
- Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
- Arterial and venous embolizations in the peripheral vasculature
NumenFR™ Detachment System is intended for use with MicroPort NeuroTech detachable coils in the embolization of intracranial aneurysms and other vascular abnormalities of the neuro and peripheral vasculature.
510(k) Summary
{6}
K260351 Page 3 of 7
# 3. Comparison of the Subject Device with the Predicate Device
Comparison for Numen™ Helia Coil Embolization System
| Characteristics | Numen™ Coil Embolization System (Predicate device, K242154) | Numen™ Helia Coil Embolization System (Subject device, K260351) |
| --- | --- | --- |
| Manufacturer | MicroPort NeuroTech (Shanghai) Co., Ltd | Same |
| Device Classification | Class II | Same |
| Regulation Number and Regulation Description | 21 CFR §870.3300, Device, Vascular, for Promoting Embolization 21 CFR §882.5950 Neurovascular Embolization Device | Same |
| Classification Product Code | HCG KRD | Same |
| Intended Use/Indications for Use | Numen™ Coil Embolization System is intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. Numen™ Coil Embolization System is indicated for endovascular embolization of: • Intracranial aneurysms • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae • Arterial and venous embolizations in the peripheral vasculature | Numen™ Helia Coil Embolization System is intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. Numen™ Helia Coil Embolization System is indicated for endovascular embolization of: • Intracranial aneurysms • Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae • Arterial and venous embolizations in the peripheral vasculature Same. |
| Types | MicroFrame, MicroFill, MicroFinish, SILK | Cube, Spiral Subject device introduces 2 additional types to MicroPort’s Numen Coil Product Family. |
| No. of Models Offered | 224 | 130 |
| Dimension/Shape of Coil Embolization System | | |
| Secondary Shape | 3D, Helical | Same |
| Coil Type | Stretch Resistance | Same |
| Coil Primary Diameter | 0.0100 inch-0.0140 inch | 0.0110 inch - 0.0115 inch The primary coil diameter of the subject device falls within the range of that of the predicate device. |
510(k) Summary
{7}
K260351 Page 4 of 7
| Characteristics | Numen™ Coil Embolization System (Predicate device, K242154) | Numen™ Helia Coil Embolization System (Subject device, K260351) |
| --- | --- | --- |
| Coil Secondary Diameter | 1-24 mm | 1-3.5mm The coil secondary diameter of the subject device falls within the range of that of the predicate device. |
| Coil Length | 1-70 cm | 2-60cm The coil length of the subject device falls within the range of that of the predicate device. |
| Pusher Length | 187.8 cm | Same |
| Material of Coil Embolization System | | |
| Primary Coil Wire | Pt (92%) / W (8%) | Same |
| Stretch Resistant Thread | Polypropylene | Same |
| Pusher (Body Hypotube) | SS 304 | Same |
| Introducer Sheath | HDPE | Same |
| Proximal Rod | Stainless Steel | Same |
| Adhesive | Epoxy 353ND | Same |
| Other | | |
| Detachment Mechanism | Electrolytic | Same |
| How Supplied | Sterile, for single use only | Same |
| Sterilization Method | Ethylene Oxide | Same |
| Compatible Microcatheter Inner Diameter | ≥ 0.0165 inch | ≥ 0.013 inch ≥ 0.0165 inch Different, Numen Helia is additionally compatible with microcatheters with an inner diameter of ≥ 0.013 inches. |
## Comparison for NumenFR™ Detachment System
The subject device and predicate device are identical, and there are no changes to the NumenFR™ Detachment System (K242154) except for a minor wording revision to its indication.
| Characteristics | NumenFR™ Detachment System (Predicate device, K242154) | NumenFR™ Detachment System (Subject device, K260351) |
| --- | --- | --- |
| Manufacturer | MicroPort NeuroTech (Shanghai) Co., Ltd | Same |
| Device Classification | Class II | Same |
| Regulation Number and Regulation Description | 21 CFR §870.3300, Device, Vascular, for Promoting Embolization | Same |
510(k) Summary
{8}
K260351 Page 5 of 7
| Characteristics | NumenFR™ Detachment System (Predicate device, K242154) | NumenFR™ Detachment System (Subject device, K260351) |
| --- | --- | --- |
| | 21 CFR §882.5950 Neurovascular Embolization Device | |
| Classification Product Code | HCG KRD | Same |
| Intended Use/Indications for Use | NumenFR™ Detachment System is intended for use with MicroPort NeuroTech Numen™ Coil Embolization System in the embolization of intracranial aneurysms and other vascular abnormalities of the neuro and peripheral vasculature. | NumenFR™ Detachment System is intended for use with MicroPort NeuroTech detachable coils in the embolization of intracranial aneurysms and other vascular abnormalities of the neuro and peripheral vasculature. Minor wording change to the indication of NumenFR™ Detachment System with the addition of the Numen™ Helia Coil Embolization System to the MicroPort NeuroTech Numen family coils |
# 4. Performance Testing
The following non-clinical bench testing was performed to evaluate the new Numen™ Helia Coil Embolization System and to demonstrate substantial equivalence of the subject device to the predicate device. The testing was performed on test units representative of final finished devices.
| Test | Test Method Summary | Test Results |
| --- | --- | --- |
| Visual Inspection | Examine the test sample surface under specific magnification. | Pass |
| Dimensional Verification | Implant coil, pusher and introducer sheath dimensions are measured to match the specifications. | Pass |
| Simulated Use | Verify that the coil embolization system performs as intended in a representative tortuous anatomical model. | Pass |
| Fatigue Testing | Verify the durability of the coil embolization system by repeating the simulated use six times, including coil retraction into microcatheter and re-deployment | Pass |
| Detachment Time and Detachment Reliability | Verify the reliability of intentional detachment as well as reliability of the coil attachment after fatigue testing of the coil embolization system in a representative tortuous anatomical model. | Pass |
510(k) Summary
{9}
K260351 Page 6 of 7
| Test | Test Method Summary | Test Results |
| --- | --- | --- |
| Delivery and Retraction Friction in Introducer Sheath | Measured by max friction force when advancing or retracting the coil system in introducer sheath. | Pass |
| Delivery, Deployment and Retraction Friction in Microcatheter | Measured by max friction force when advancing, deploying or retracting the coil system through microcatheter in a relevant, tortuous, anatomical model. | Pass |
| Kink Resistance | Demonstrate that the resistance to kinking of the device meets pre-specified acceptance criteria, and could withstand bending forces that the device may encounter in clinical usage. | Pass |
| Torque Strength | Verify the torque strength by rotating the proximal end of the device for 8 turns. | Pass |
| Flexing Test | Per ISO 11070, Annex G Test method for resistance of guidewires to damage by flexing | Pass |
| Fracture Test | Per ISO 11070, Annex F Test method Test method for fracture of guidewires | Pass |
## 5. Shelf-life
The Numen™ Helia Coil Embolization System is constructed using the same materials, packaging, and manufacturing process as the predicate device. The storage and transport conditions for both devices are also the same. There is no increased risk of device failure due to material degradation. Therefore, the shelf-life testing previously established for the predicate devices is also applicable to the subject device, and the shelf-life claims of three (3) years for the Numen™ Helia Coil Embolization System are the same as for the predicate device.
## 6. Biocompatibility
The materials and processing methods of the Numen™ Helia Coil Embolization System are the same as those of the predicate device. There are no material or contact duration changes to the components of the device that contact circulating blood. Therefore, part of the biocompatibility performances of Numen™ Helia could be adopted from the predicate device.
510(k) Summary
{10}
K260351 Page 7 of 7
To evaluate the biocompatibility impact of design modifications relative to the predicate device, the following biocompatibility tests were performed on the final finished device of Numen™ Helia.
| Test | Test Method | Results |
| --- | --- | --- |
| Partial Thromboplastin Time | Partial Thromboplastin Time per ISO 10993-4. | Pass |
| Thrombogenicity | Standard Thrombogenicity in Canine per ISO 10993-4. | Pass |
The Numen™ Helia Coil Embolization System has been evaluated to meet requirements specified in FDA guidance, “Use of International Standard ISO 10993-1, “Biocompatibility Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process”.
## 7. Conclusion
MicroPort NeuroTech has made minor modifications to the design of the implanted coil and pusher of the Numen™ Coil Embolization System. The modified device will be marketed under a new trade name: Numen™ Helia Coil Embolization System. The intended use and indications for use of the Numen™ Helia device remain unchanged in comparison with the predicate device, and the minor modifications do not affect the fundamental scientific technology of the predicate device. There is no modification to the NumenFR™ Detachment System except the minor wording revision to the indications for use of the NumenFR™ device because of the addition of the Numen™ Helia Coil Embolization System to the MicroPort NeuroTech Numen family.
A comprehensive risk assessment of the modification and successful verification testing have been performed, which did not raise any new questions regarding safety and effectiveness. Based on these evaluations, MicroPort NeuroTech has concluded that the subject device is substantially equivalent to the predicate device.
510(k) Summary
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.