K260332 · Pathwai, Inc. · QXO · Jun 8, 2026 · Cardiovascular
Device Facts
Record ID
K260332
Device Name
EchoNext
Applicant
Pathwai, Inc.
Product Code
QXO · Cardiovascular
Decision Date
Jun 8, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2380
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K260332 · Jun 8, 2026
EchoNext
Pathwai, Inc.
Electronic Health Records (EHR); Transthoracic echocardiography reports; 12-lead ECG recordings
Retrospective analysis of 102,219 patients across four health systems to validate the device's ability to detect structural heart disease (SHD) and six specific sub-labels using routine clinical data.
Retrospective cohort; Multicenter; EHR data; Structural heart disease
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective multicenter evaluation; Retrospective cohort study; Follow-up/Duration: Not applicable
102,219 unique patients (≥18 years) with a clinically ordered 12-lead ECG and a transthoracic echocardiogram within 30 days.; Sample Size: 102,219; Number of Sites: 4
Not applicable for this study
Sensitivity, specificity, PPV, and NPV for detection of SHD and six sub-labels (Left/Right Ventricular Systolic Dysfunction, Aortic Stenosis, Valvular Regurgitation, Pulmonary Hypertension, Left Ventricular Hypertrophy).
EchoNext is intended to be used in parallel to the standard of care to analyze recordings of 12-lead ECG (electrocardiogram) made on compatible ECG devices. EchoNext is capable of analyzing the ECG, detecting signs associated with structural heart disease (SHD), and allowing the user to view the ECG and analysis results. When evidence of SHD is detected, the device additionally may detect six sub-labels, which are specific forms of SHD. EchoNext is indicated for use on 12-lead ECG recordings collected from patients 18 years of age or older. EchoNext is not intended for use on patients with implanted pacemakers. EchoNext is limited to analysis of ECG data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis. EchoNext identifies patients for further SHD follow-up and is not intended to replace the current standard of care methods for diagnosis of SHD. The results of the device are not intended to rule-out SHD follow-up.
Device Story
EchoNext is a SaMD that analyzes digital 12-lead ECG waveforms and demographic data to identify structural heart disease (SHD). It uses a deep neural network to detect electrical patterns associated with cardiac structural abnormalities. The device provides a binary notification for SHD and six specific sub-labels: LVEF, LVH, AS, Valvular Regurgitation, RVEF, and Pulmonary Hypertension. It is intended for use in clinical settings alongside standard of care; it does not replace diagnostic evaluation. Clinicians use the output to identify patients requiring further follow-up. The device benefits patients by flagging potential structural abnormalities that might otherwise go undetected during routine ECG screening.
Clinical Evidence
Retrospective, multicenter study of 102,219 patients across four US/Canada health systems. Ground truth established via transthoracic echocardiography interpreted by board-certified cardiologists. Primary endpoint: sensitivity and specificity for composite SHD and six sub-labels. Results: SHD sensitivity 73.7% (95% CI: 73.2-74.1%), specificity 75.2% (95% CI: 74.9-75.5%). Sub-label sensitivities ranged from 48.2% (LVH) to 80.1% (Left Ventricular Systolic Dysfunction). All pre-specified acceptance criteria met.
Technological Characteristics
SaMD; deep neural network-based machine learning algorithm; inputs: 12-lead ECG waveform data and demographic inputs; outputs: binary notifications for SHD and six sub-labels; locked model; standalone software; no hardware components; cybersecurity controls include static/dynamic code analysis, fuzz testing, and penetration testing.
Indications for Use
Indicated for patients 18 years of age or older undergoing 12-lead ECG analysis to identify potential structural heart disease (SHD). Contraindicated for patients with implanted pacemakers.
Regulatory Classification
Identification
Viz HCM is a cardiovascular machine learning-based notification software intended to be used in parallel to the standard of care to analyze 12-lead ECG recordings from patients 18 years of age or older. It detects signs associated with hypertrophic cardiomyopathy (HCM) and allows the user to view the ECG and analysis results. It is not intended for use on patients with implanted pacemakers, does not replace standard diagnostic methods, and is not intended to rule out HCM or be used in lieu of a full patient evaluation.
Special Controls
In combination with the general controls of the FD&C Act, cardiovascular machine learningbased notification software is subject to the following special controls:
{1}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 8, 2026
Pathwai, Inc.
Jeffrey Ruhl
Chief of Staff
189 W 89th St.
Apt. 6e
New York, New York 10024
Re: K260332
Trade/Device Name: EchoNext
Regulation Number: 21 CFR 870.2380
Regulation Name: Cardiovascular Machine Learning-Based Notification Software
Regulatory Class: Class II
Product Code: QXO, QYE, SIK, SII, SIJ, SAT
Dated: June 3, 2026
Received: June 3, 2026
Dear Jeffrey Ruhl:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device'
{2}
K260332 - Jeffrey Ruhl
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{3}
K260332 - Jeffrey Ruhl
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
STEPHEN C. BROWNING -S
LCDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260332 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| EchoNext | | |
| Please provide your Indications for Use below. | | ? |
| EchoNext is intended to be used in parallel to the standard of care to analyze recordings of 12-lead ECG (electrocardiogram) made on compatible ECG devices. | | |
| EchoNext is capable of analyzing the ECG, detecting signs associated with structural heart disease (SHD), and allowing the user to view the ECG and analysis results. When evidence of SHD is detected, the device additionally may detect six sub-labels, which are specific forms of SHD. EchoNext is indicated for use on 12-lead ECG recordings collected from patients 18 years of age or older. | | |
| EchoNext is not intended for use on patients with implanted pacemakers. EchoNext is limited to analysis of ECG data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis. EchoNext identifies patients for further SHD follow-up and is not intended to replace the current standard of care methods for diagnosis of SHD. The results of the device are not intended to rule-out SHD follow-up. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
Page 1 of 1
{5}
K260332

# 510(k) Summary
## Submitter
| Applicant | Pathwai, Inc. 169 Madison Avenue, STE 31277, New York, NY 10016 |
| --- | --- |
| Contact Person | Pierre Elias, MD Chief Executive officer pierre@pathwaylabs.com |
| Date Prepared | June 8, 2026 |
## Subject Device
| Trade/ Device Name: | EchoNext |
| --- | --- |
| Regulation Number: | 21 CFR 870.2380 |
| Regulation Name: | Cardiovascular machine learning-based notification software |
| Regulatory Class: | II |
| Product Code: | QXO, QYE, SAT, SII, SIJ, SIK |
## Predicate Device
| Trade / Device Name: | Viz HCM (DEN230003) |
| --- | --- |
## Device Description
EchoNext is a software-as-a-medical-device (SaMD) intended to assist clinicians by analyzing standard resting 12-lead electrocardiogram (ECG) recordings to identify patients who may be suspected of having structural heart disease (SHD) and who may benefit from further diagnostic evaluation.
EchoNext functions by processing digital 12-lead ECG waveform data using a machine-learning algorithm based on deep neural network techniques. The scientific basis of the device is the established relationship between subtle electrical patterns present in ECG signals and underlying structural abnormalities of the heart, such as abnormalities of cardiac chambers, walls, or valves.
Pathwai, Inc.
1
K260332
Page 1 of 7
{6}
The device evaluates ECG waveform data in combination with basic demographic inputs to generate a notification indicating whether the ECG demonstrates patterns associated with suspicion of structural heart disease. When SHD is suspected, the device may also provide informational indicators corresponding to affected cardiac structures. The potential outputs of EchoNext and their product codes are:
| Output | Product code |
| --- | --- |
| SHD: Structural Heart Disease | QXO |
| LVEF: Left Ventricular Ejection Fraction | QYE |
| LVH: Left Ventricular Hypertrophy | SIK |
| AS: Aortic Stenosis | SII |
| Valvular Regurgitation | SIJ |
| RVEF: Right Ventricular Ejection Fraction | QYE |
| PH: Pulmonary Hypertension | SAT |
## Intended Use / Indications for Use
EchoNext is intended to be used in parallel to the standard of care to analyze recordings of 12-lead ECG (electrocardiogram) made on compatible ECG devices.
EchoNext is capable of analyzing the ECG, detecting signs associated with structural heart disease (SHD), and allowing the user to view the ECG and analysis results. When evidence of SHD is detected, the device additionally may detect six sub-labels, which are specific forms of SHD. EchoNext is indicated for use on 12-lead ECG recordings collected from patients 18 years of age or older.
EchoNext is not intended for use on patients with implanted pacemakers. EchoNext is limited to analysis of ECG data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis. EchoNext identifies patients for further SHD follow-up and is not intended to replace the current standard of care methods for diagnosis of SHD. The results of the device are not intended to rule-out SHD follow-up.
## Summary of Technological Characteristics
A comparison of technological similarities and differences between the subject and predicate devices is as follows:
Pathwai, Inc.
2
K260332
Page 2 of 7
{7}
| Characteristic | Subject Device: Pathwai EchoNext | Predicate Device: Viz.ai HCM Module (DEN230003) |
| --- | --- | --- |
| Regulation Number | 21 CFR 870.2380 | Same |
| Regulation Name | Cardiovascular Machine Learning Based Notification Software | Same |
| Product Codes | QXO, QYE, SAT, SII, SIJ, SIK | QXO |
| SaMD | Yes | Same |
| Indications for Use | EchoNext is intended to be used in parallel to the standard of care to analyze recordings of 12-lead ECG made on compatible ECG devices. EchoNext is capable of analyzing the ECG, detecting signs associated with structural heart disease (SHD), and allowing the user to view the ECG and analysis results. EchoNext is indicated for use on 12-lead ECG recordings collected from patients 18 years of age or older. EchoNext is not intended for use on patients with implanted pacemakers. EchoNext is limited to analysis of ECG data and should not be used in lieu of full patient evaluation or relied upon to make or confirm diagnosis. EchoNext identifies patients for further SHD follow-up and is not intended to replace the current standard of care methods for diagnosis of SHD. The results of the device are not intended to rule-out SHD follow-up. | Viz HCM is intended to be used in parallel to the standard of care to analyze recordings of 12-lead ECG made on compatible ECG devices. Viz HCM is capable of analyzing the ECG, detecting signs associated with hypertrophic cardiomyopathy (HCM), and allowing the user to view the ECG and analysis results. Viz HCM is indicated for use on 12-lead ECG recordings collected from patients 18 years of age or older. Viz HCM is not intended for use on patients with implanted pacemakers. Viz HCM is limited to analysis of ECG data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis. Viz HCM identifies patients for further HCM follow-up and does not replace the current standard of care methods for diagnosis of HCM. The results of the device are not intended to rule-out HCM follow-up. |
| Clinical Specialty | Cardiovascular | Same |
| Rx/OTC | Rx | Same |
| Patient Population | ≥18 years | Same |
| Use with Pacemakers? | No | Same |
Pathwai, Inc.
3
K260332
Page 3 of 7
{8}
| Characteristic | Subject Device: Pathwai EchoNext | Predicate Device: Viz.ai HCM Module (DEN230003) |
| --- | --- | --- |
| Locked Machine Learning Model | Yes | Same |
| Input | 12-lead ECG | Same |
| Output (Non-Diagnostic Notification) | Binary notification **Primary Indicator:** SHD: Structural Heart Disease **Sub-Labels:** - AS: Aortic Stenosis (moderate or greater) - Valvular Regurgitation (moderate or greater): Aortic Regurgitation, Mitral Regurgitation, Pulmonary Regurgitation, Tricuspid Regurgitation - LVH: Left Ventricular Hypertrophy (severe LVH) - LVEF: Left Ventricular Ejection Fraction ($\leq 45\%$) - RVEF: Right Ventricular Ejection Fraction (moderate or greater dysfunction) - PH: Pulmonary Hypertension (PASP $\geq 45\text{mmHg}$ or TR max velocity $\geq 3.2\text{ m/s}$) | Binary notification (HCM flag) |
| Built-In View | No | Yes |
Both subject and predicate devices have the same intended use, similar principles of operations, and technological characteristics. Any differences do not raise different questions of safety and effectiveness.
## Summary of Nonclinical Tests
The following nonclinical tests were conducted to support the safety and effectiveness of EchoNext.
Pathwai, Inc.
4
K260332
Page 4 of 7
{9}
- Software Verification and Validation: SW V&V activities included Unit Testing, Integration and System-level testing. All software requirements successfully met their predefined acceptance criteria.
- Cybersecurity Testing: Cybersecurity testing - including Static and Dynamic Code Analysis, Fuzz Testing, and Penetration testing - was successfully conducted, demonstrating cybersecurity risks have been properly mitigated.
- Human Factors Validation - Human factors testing with 15 participants representative of the intended user population was successfully conducted, demonstrating use-related risks have been properly mitigated.
## Summary of Clinical Performance Data
EchoNext was validated based on a retrospective, multicenter evaluation designed to assess performance in a real-world clinical setting. The analysis uses a large dataset comprising 102,219 unique consecutive patients, with one ECG per patient, drawn from four geographically distinct health systems in the United States and Canada. These sites span academic and community hospitals and represent urban, suburban, and diverse patient populations, supporting evaluation across varied care environments, demographic groups, and disease prevalence patterns. The dataset was assembled to maximize geographic and clinical diversity.
Clinical ground truth was established using transthoracic echocardiography, which serves as the diagnostic reference standard for structural heart disease. Echocardiograms were interpreted by board-certified cardiologists using FDA-cleared systems, and finalized reports were used to extract structured measurements and diagnostic assessments. From these reports, a composite structural heart disease label was derived, along with multiple clinically relevant sub-labels, including left and right ventricular systolic dysfunction, severe left ventricular hypertrophy, moderate or greater valvular regurgitation, moderate or greater aortic stenosis, and moderate or greater pulmonary hypertension. Sub-labels were assigned using guideline-based thresholds, and multiple sub-labels could be present for a single patient.
Sub-group analyses included stratifications of device performance by age, sex, ECG manufacturer, comorbidity, race and ethnicity, and baseline ECG to index TTE. In all circumstances, EchoNext met its acceptance criteria, demonstrating clinical acceptability of device performance. Additional testing completed for the EchoNext device includes software verification and validation, human factors validation, and cybersecurity testing.
## Study Design and Patient Population
The external validation dataset comprised 102,219 unique patients drawn from four geographically and demographically distinct health systems in the United States and Canada. Sites were selected to maximize geographic, demographic, and clinical diversity, spanning academic and community hospitals in urban, suburban, and rural settings. The median patient age was 66.3 years (IQR 57–79); 52.0% were male.
Pathwai, Inc.
5
K260332
Page 5 of 7
{10}
Race/ethnicity: 41.4% White, 17.4% Asian, 10.8% Hispanic, and 6.1% Black. SHD prevalence was 34.8% (35,548 / 102,219).
## Performance Analysis
EchoNext was validated in an external, prospectively enrolled cohort of 102,219 patients across four independent clinical sites. Patients who had a clinically ordered 12-lead ECG and a transthoracic echocardiogram within 30 days were included. Ground truth for each condition was established by board-certified cardiologists interpreting the echocardiogram report, blinded to ECG findings.
Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) with 95% confidence intervals were computed across the full external validation cohort for both the composite structural heart disease (SHD) output and each of the six sub-label outputs. Pre-specified acceptance criteria were defined separately for sensitivity and specificity for each output prior to testing.
EchoNext met all pre-specified acceptance criteria across the composite SHD output and all six sub-labels. EchoNext achieved a sensitivity of 73.7% (95% confidence interval: 73.2-74.1%) and a specificity of 75.2% (74.9-75.5%) in detecting SHD in the external validation. **Table 1 shows performance for EchoNext and each sub-label.**
**Table 1: Performance Characteristics of EchoNext in performance study dataset. The sensitivity, specificity, PPV, and NPV of EchoNext in detecting SHD and each sub-label**
| | Sensitivity | Specificity | PPV | NPV | Prevalence |
| --- | --- | --- | --- | --- | --- |
| SHD | 73.7 (73.2-74.1) | 75.2 (74.9-75.5) | 61.6 (61.2-62.1) | 84.6 (84.3-84.9) | 34.8% |
| Left Ventricular Systolic Dysfunction | 80.1 (79.4-80.9) | 80.4 (80.1-80.6) | 34.8 (34.2-35.3) | 97.0 (96.9-97.2) | 11.3% |
| Right Ventricular Systolic Dysfunction | 76.8 (75.3-78.3) | 82.5 (82.3-82.7) | 12.1 (11.6-12.5) | 99.2 (99.1-99.2) | 3.0% |
| Aortic Stenosis | 54.9 (53.4-56.3) | 80.1 (79.8-80.3) | 11.8 (11.4-12.3) | 97.4 (97.3-97.5) | 4.5% |
| Valvular Regurgitation | 58.2 (57.5-59.0) | 83.9 (83.7-84.2) | 42.2 (41.6-42.9) | 91.2 (91.0-91.4) | 16.4% |
| Pulmonary Hypertension | 64.1 (63.1-65.1) | 81.1 (80.8-81.4) | 25.3 (24.8-25.9) | 95.9 (95.8-96.0) | 8.9% |
| Left Ventricular Hypertrophy | 48.2 (46.9-49.4) | 81.3 (81.1-81.6) | 14.2 (13.8-14.7) | 96.2 (96.0-96.3) | 5.9% |
Pathwai, Inc.
6
K260332
Page 6 of 7
{11}
We additionally estimated how PPV and NPV would change at different prevalences for some sub-labels. For valvular regurgitation, the PPV and NPV would be 24.4% and 95.7% at a prevalence of 8.2%, and 16.0% and 97.4% at a prevalence of 5.0%. For pulmonary hypertension, the PPV and NPV would be 13.2% and 98.0% at a prevalence of 4.3%, and 9.5% and 98.6% at a prevalence of 3.0%.
## Conclusions
EchoNext performs as expected and in a manner that is substantially equivalent to the predicate device. EchoNext has the same intended use and similar indications for use, principles of operation, and technological characteristics as the predicate device. The differences do not raise new concerns of safety and effectiveness. Therefore, EchoNext is substantially equivalent to the predicate device.
Nonclinical testing, including software verification and validation (101/101 test cases passing), human factors engineering validation (zero use errors or close calls across 15 participants), and cybersecurity testing (no vulnerabilities identified) demonstrated that EchoNext meets applicable safety and performance requirements.
Clinical performance testing conducted on 102,219 patients across four geographically and demographically diverse health systems demonstrated that EchoNext met all pre-specified acceptance criteria for the composite SHD label and all six sub-labels, with consistent performance across all evaluated subgroups. Therefore, EchoNext is substantially equivalent to the predicate device.
Pathwai, Inc.
7
K260332
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.