DEN230003 · Viz. Ai, Inc. · QXO · Aug 3, 2023 · Cardiovascular
Device Facts
Record ID
DEN230003
Device Name
Viz HCM
Applicant
Viz. Ai, Inc.
Product Code
QXO · Cardiovascular
Decision Date
Aug 3, 2023
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 870.2380
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN230003 · Aug 3, 2023
Viz HCM
Viz. Ai, Inc.
Retrospective medical records; Historical patient ECG data; Cardiologist chart and imaging review
Retrospective clinical study used to assess the performance (sensitivity, specificity, PPV) of the Viz HCM algorithm in identifying suspected HCM findings in ECGs compared to clinical findings established by historical chart and imaging review.
Pivotal retrospective performance assessment; Retrospective study using historical patient data; Follow-up/Duration: July 1, 2017, to June 30, 2022; Study Period: July 1, 2017 - June 30, 2022
Patients 18 years of age or older with 12-lead ECG recordings; Sample Size: 3,196 (291 HCM-Positive and 2,905 HCM-Negative); Number of Sites: 3
Not applicable for this study
Sensitivity, specificity, and device positive value (PPV)
Training set (80%): 301,106 patients (4,470 HCM positive, 298,394 HCM negative); Internal validation set (20%): 301,106 patients (4,470 HCM positive, 298,394 HCM negative).
—
Retrospective pivotal study: 3,196 ECG cases (291 HCM-Positive and 2905 HCM-Negative) from 3 hospitals.
1 (cardiologist) + 1 (cardiologist)
Indications for Use
Viz HCM is intended to be used in parallel to the standard of care to analyze recordings of 12-lead ECG made on compatible ECG devices. Viz HCM is capable of analyzing the ECG, detecting signs associated with hypertrophic cardiomyopathy (HCM), and allowing the user to view the ECG and analysis results. Viz HCM is indicated for use on 12-lead ECG recordings collected from patients 18 years of age or older. Viz HCM is not intended for use on patients with implanted pacemakers. Viz HCM is limited to analysis of ECG data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis. Viz HCM identifies patients for further HCM follow-up and does not replace the current standard of care methods for diagnosis of HCM. The results of the device are not intended to rule-out HCM follow-up.
Device Story
Viz HCM is a prescription software device that analyzes 12-lead ECG recordings to detect signs associated with hypertrophic cardiomyopathy (HCM). It operates in parallel to the standard of care; it does not replace clinical evaluation or diagnostic methods. The device processes ECG data to identify patients who may require further HCM follow-up. It is not intended to rule out HCM or detect arrhythmias. The software provides results to the user, who views the ECG and analysis. The device is intended to assist in identifying patients for referral or diagnostic follow-up; it is not for diagnostic-quality output. It is not intended for use on patients with implanted pacemakers.
Clinical Evidence
Retrospective study of 3,196 ECG cases (291 HCM-positive, 2,905 HCM-negative) from 3 hospitals. Ground truth established by cardiologist chart and imaging review using Cornell or Sokolow-Lyon criteria. Primary results: Sensitivity 68.4% (95% CI: 62.8%–73.5%), Specificity 99.1% (95% CI: 98.7%–99.4%), PPV 13.7% (95% CI: 10.1%–19.9%). Subgroup analyses performed for age, gender, race, ethnicity, and HCM type (obstructive vs. non-obstructive).
Technological Characteristics
SaMD consisting of an ECG analysis algorithm and mobile application. Inputs: 12-lead ECG data. Processing: Machine learning-based software algorithm. Connectivity: Networked/mobile interface. Software level of concern: Moderate. Cybersecurity controls implemented per FDA guidance.
Indications for Use
Indicated for patients 18 years or older to analyze 12-lead ECG recordings for signs associated with hypertrophic cardiomyopathy (HCM). Contraindicated for patients with implanted pacemakers.
Regulatory Classification
Identification
Viz HCM is a cardiovascular machine learning-based notification software intended to be used in parallel to the standard of care to analyze 12-lead ECG recordings from patients 18 years of age or older. It detects signs associated with hypertrophic cardiomyopathy (HCM) and allows the user to view the ECG and analysis results. It is not intended for use on patients with implanted pacemakers, does not replace standard diagnostic methods, and is not intended to rule out HCM or be used in lieu of a full patient evaluation.
Special Controls
In combination with the general controls of the FD&C Act, cardiovascular machine learningbased notification software is subject to the following special controls:
Submission Summary (Full Text)
{0}------------------------------------------------
# DE NOVO CLASSIFICATION REQUEST FOR VIZ HCM
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Cardiovascular machine learning-based notification software. Cardiovascular machine learning-based notification software employs machine learning techniques to suggest the likelihood of a cardiovascular disease or condition for further referral or diagnostic follow-up. The software identifies a single condition based on one or more non-invasive physiological inputs as part of routine medical care. It is intended as the basis for further testing and is not intended to provide diagnostic quality output. It is not intended to identify or detect arrhythmias.
NEW REGULATION NUMBER: 21 CFR 870.2380
CLASSIFICATION: Class II
PRODUCT CODE: QXO
#### BACKGROUND
DEVICE NAME: Viz HCM
SUBMISSION NUMBER: DEN230003
DATE DE NOVO RECEIVED: January 10, 2023
#### SPONSOR INFORMATION:
Viz.ai, Inc. 201 Mission St., 12th Floor San Francisco, California 94105
#### INDICATIONS FOR USE
The Viz HCM is indicated as follows:
Viz HCM is intended to be used in parallel to the standard of care to analyze recordings of 12-lead ECG made on compatible ECG devices. Viz HCM is capable of analyzing the ECG, detecting signs associated with hypertrophic cardiomyopathy (HCM), and allowing the user to view the ECG and analysis results. Viz HCM is indicated for use on 12-lead ECG recordings collected from patients 18 years of age or older. Viz HCM is not intended for use on patients with implanted pacemakers. Viz HCM is limited to analysis of ECG data and should not be used in-lieu of full patient evaluation or relied upon to
{1}------------------------------------------------
make or confirm diagnosis. Viz HCM identifies patients for further HCM follow-up and does not replace the current standard of care methods for diagnosis of HCM. The results of the device are not intended to rule-out HCM follow-up.
# LIMITATIONS
The sale, distribution, and use of the Viz HCM are restricted to prescription use in accordance with 21 CFR 801.109.
The device identifies patients for further HCM follow-up and does not replace the current standard of care methods for diagnosis of HCM. The results of the device are not intended to rule-out HCM follow-up.
The device is not intended for use on patients with implanted pacemakers.
The results of the device should not be used in lieu of full patient evaluation.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
# DEVICE DESCRIPTION
The Viz HCM ECG Analysis Algorithm (HCM Algorithm) is a machine learning-based software algorithm that analyzes 12-lead electrocardiograms (ECGs) for characteristics suggestive of hypertrophic cardiomyopathy (HCM). The mobile software module enables the end user to receive and toggle notifications for ECGs determined by the Viz HCM ECG Analysis Algorithm to contain signs suggestive of HCM.
Image /page/1/Figure/9 description: This image is a flowchart outlining the steps for a Viz HCM analysis. The process begins with a review of the patient's ECG, which is then analyzed by Viz HCM. The results are available on a mobile interface. The flowchart continues with steps such as patient assessment, review of patient history, physical exams, and follow-up imaging, ultimately leading to a determination of whether HCM is confirmed and further patient assessment for risk factors and treatment.
Figure 1. Clinical Workflow Diagram
{2}------------------------------------------------
The Viz HCM is a Software as a Medical Device (SaMD) intended to analyze ECG signals collected as part of a routine clinical assessment, independently and in parallel to the standard of care. Viz HCM is a combination of software modules that consists of an ECG analysis software algorithm and mobile application software module.
## SUMMARY OF NONCLINICAL/BENCH STUDIES
Nonclinical studies conducted for the Viz HCM system are summarized below.
## SOFTWARE
The software was reviewed according to the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 11, 2005. Appropriate software documentation consistent with a "Moderate" level of software concern were provided.
The software documentation included a detailed description of the machine learning model, the model inputs and outputs, and the supported patient population. Integration testing was conducted in the intended software system. Testing accounted for the impact of and variability between different ECG acquisition hardware. The ECG hardware controls included a description of input ECG signal control measures and mitigations for user error and system components on output accuracy.
Cybersecurity was reviewed in accordance with the FDA guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" dated October 2, 2014.
# PERFORMANCE TESTING-MODEL DEVELOPMENT AND INTERNAL VALIDATION
The data for algorithm development was collected from different US and Non-US (OUS) sources. The data contains both HCM Positive (obstructive and nonobstructive) and HCM Negative examples including random ECG samples (random control) and enrichment for conditions differential for and associated with HCM (negative controls), The data is diverse with respect to the age, sex, and health status of the patient, as well as the data source from which the ECG data was taken. Efforts were made to sample the data from data sources from ethnically diverse regions so as to promote ethnic diversity in the training and internal validation datasets.
{3}------------------------------------------------
Image /page/3/Figure/0 description: The image shows a diagram of an algorithm development set. The algorithm development set includes 301,106 patients and 831,329 ECG exams. The set is divided into HCM positive patients (4,470) and HCM negative patients (298,394). The data is then split into a training set (80%) and an internal validation set (20%) for algorithm development and internal evaluation.
Figure 2. Distribution and separation of training and internal validation data
Refer to Table 9 in the labeling section of this document for full demographics of training and internal validation datasets.
The development dataset was split into disjoint and secured training and internal validation partitions. The partitions were created using a hash function on a patientanonymized unique identifier (patient UID) to ensure consistency across all development phases and across different models and datasets and prevents different studies of the same patient being allocated to more than one partition.
# PERFORMANCE TESTING - USABILITY
The usability of the Viz HCM system was assessed per EN 62366-1 "Medical Devices-Part 1: Application of usability engineering to medical devices" dated 2015 + A1:2020 and the FDA guidance document "Applying Human Factors and Usability Engineering to Medical Devices". A use-related risk analysis (URRA) and system risk analysis (RA) were performed to identify use-related hazards and critical tasks. The study recruited 16 representative clinical users who were assessed within two (2) use scenarios for task completion according to correct use, user error, close calls, or use difficulty.
- . There were no use errors during the study
- All participants demonstrated full understanding of the intended use, device . output (i.e., the HCM flag), and that they would not rely on the device to make a diagnosis
- . 15/16 (93%) participants acknowledged the HCM flag during the session to review the ECG record
This testing demonstrated that the intended users of the product can perform the product's intended use in the expected use environment. It also demonstrated that the intended users could adequately comprehend the labeling.
# SUMMARY OF CLINICAL INFORMATION
{4}------------------------------------------------
## Overview
Model testing was performed on a dataset which was acquired from a retrospective study to assess the performance of Viz HCM in the identification of suspected HCM findings in ECG as compared to the clinical finding of HCM as established by cardiologist chart and imaging review of historical patient data. The objectives of the study were to assess the performance of the device in terms of sensitivity, specificity, and device positive value (PPV), and conduct additional analyses to assess the device performance for detection of suspected HCM in different sub-populations. This was a non-significant risk study using historical patient data and did not involve enrollment of any human beings. Due to the retrospective nature of the study, no adverse events were expected or observed. The Institutional Review Board (IRB) overseeing the study at participating hospitals waived the need for patient informed consent.
A total of 3.196 (291 HCM-Positive and 2905 HCM-Negative) ECG cases were included in the performance assessment for the pivotal study. Patient cases were selected from 3 hospitals representing a combination of academic and community hospitals between July 1. 2017, and June 30, 2022. Two (2) of the three (3) sites providing data were from the Boston. Massachusetts area which are racially and ethnically diverse in terms of the local African American. Asian, and Multi-Race (two or more) populations. The third site was in Salem, Massachusetts which was predominantly Caucasian or Latino. All three sites provided data from geographic areas with similar proportions of Hispanic or Latino individuals in the local population, which were also similar to the proportion of the Hispanic or Latino population in the USA overall. The proportions of obstructive and non-obstructive HCM were roughly equal. See Table 2 for full demographic information.
# TRUTHING PROCESS
For each HCM-Positive or HCM-Negative case, a single cardiologist performed a chart and imaging review (where available) for the patient to confirm the presence of HCM according to predefined guidelines using either the Cornell criteria or the Sokolow-Lyon criteria. In addition, ECGs were annotated for the presence of different features and pathologies. ECGs determined to contain a pacemaker or corrupt lead were excluded during ECG annotations.
For HCM-Positive cases, the cardiologist assessed the patient chart and imaging to confirm the presence of HCM and if confirmed, the degree of obstruction (either non-obstructive or obstructive). HCM-Negative cases were reviewed for the presence or mention of HCM in the patient chart, along with any available imaging to rule out HCM. If the patient chart and imaging confirmed the presence of HCM, the patient was moved from the HCM-Negative cohort to the HCM-Positive cohort. As part of a secondary assessment, a selection of 60 cases (30 HCM-Positive cases and 30 HCM-Negative) were truthed by a second cardiologist to perform an analysis of agreement/consistency in confirmation of HCM.
In the study, the ICD-10 Code was used to sample HCM-Positive and HCM-Negative patients prior to truthing. During the truthing process, HCM-Negative patient cases were confirmed if there were no notes related to HCM in the patient chart. These would be determined as HCM-Negative by the lack of ICD-9/10 code for HCM as was the case with algorithm development. For HCM-Negative patient cases with available imaging or HCM-Positive cases, the additional chart review and review of imaging provided more confidence into the label with imaging
{5}------------------------------------------------
evidence as established by the initial sampling of the ICD-9/10 codes associated with the patient diagnosis (or lack of HCM diagnosis).
The study results are in Table 1 below.
| Performance Measure | Results |
|---------------------------|-------------------------------|
| Sensitivity | 68.4% (95% CI: 62.8% - 73.5%) |
| Specificity | 99.1% (95% CI: 98.7% - 99.4%) |
| PPV (prevalence of 0.002) | 13.7% (95% CI: 10.1% - 19.9%) |
# Table 1. Clinical Testing Results
Refer to Table 9 for full demographics of training and internal validation datasets.
In addition, subgroup analyses were conducted by:
- Hospital site .
- ECG device make/model .
- . Gender
- . Age
- Race .
- . Ethnicity
- HCM Characterization (i.e., obstructive vs. non-obstructive) .
| Subgroup | AUC<br>(95% CI) | True<br>positive | False<br>negative | Sensitivity<br>(95% CI) | True<br>negative | False<br>positive | Specificity<br>(95% CI) |
|----------|------------------------|------------------|-------------------|-------------------------|------------------|-------------------|-------------------------|
| BWH | 0.982<br>(0.971,0.990) | 82 | 29 | 73.9<br>(65.0,81.2) | 957 | 11 | 98.9<br>(98.0,99.4) |
| MGH | 0.986<br>(0.980,0.992) | 75 | 32 | 70.1<br>(60.8,78.0) | 974 | 4 | 99.6<br>(98.9,99.9) |
| SH | 0.948<br>(0.915,0.970) | 42 | 31 | 57.5<br>(46.1,68.2) | 949 | 10 | 99.0<br>(98.1,99.5) |
### Table 2: Subgroup Analysis by Site
| Subgroup | True<br>positive | False<br>negative | Sensitivity<br>(95% CI) |
|-----------------|------------------|-------------------|-------------------------|
| Obstructive | 97 | 45 | 68.3<br>(60.2,75.4) |
| Non-obstructive | 102 | 47 | 68.5<br>(60.6,75.4) |
# Table 3: Subgroup Analysis by HCM type
{6}------------------------------------------------
| Subgroup | AUC<br>(95% CI) | True<br>positive | False<br>negative | Sensitivity<br>(95% CI) | True<br>negative | False<br>positive | Specificity<br>(95% CI) |
|----------|------------------------|------------------|-------------------|-------------------------|------------------|-------------------|-------------------------|
| Female | 0.979<br>(0.971,0.985) | 85 | 54 | 61.2<br>(52.8,68.9) | 1494 | 7 | 99.5<br>(99.0,99.8) |
| Male | 0.971<br>(0.954,0.984) | 114 | 38 | 75.0<br>(67.5,81.2) | 1386 | 18 | 98.7<br>(98.0,99.2) |
Table 4: Subgroup Analysis by gender
| Subgroup | AUC<br>(95% CI) | True<br>positive | False<br>negative | Sensitivity<br>(95% CI) | True<br>negative | False<br>positive | Specificity<br>(95% CI) |
|----------------------------------------------|------------------------|------------------|-------------------|-------------------------|------------------|-------------------|-------------------------|
| < 40 years | 0.977<br>(0.932,1.000) | 32 | 6 | 84.2<br>(69.2,92.9) | 485 | 1 | 99.8<br>(98.7,100.1) |
| 40 - 65 years | 0.973<br>(0.963,0.982) | 84 | 47 | 64.1<br>(55.6,71.8) | 1166 | 12 | 99.0<br>(98.2,99.4) |
| > 65 years | 0.973<br>(0.959,0.983) | 83 | 39 | 68.0<br>(59.3,75.7) | 1229 | 12 | 99.0<br>(98.3,99.5) |
| Table 5: Subgroup Analysis by age | | | | | | | |
| Subgroup | AUC<br>(95% CI) | True<br>positive | False<br>negative | Sensitivity<br>(95% CI) | True<br>negative | False<br>positive | Specificity<br>(95% CT) |
| American Indian or<br>Alaska Native | - | 0 | 0 | - | 2 | 0 | 100.0<br>(29.0.105.2) |
| Asian | 0.987<br>(0.962.0.999) | 11 | 4 | 73.3<br>(47.6,89.5) | 93 | 1 | 98.9<br>(93.6.100.4) |
| Black | 0.981<br>(0.961,0.994) | 20 | 11 | 64.5<br>(46.9,79.0) | 262 | 2 | 99.2<br>(97.1,100.0) |
| Native Hawaiian or<br>Other Pacific Islander | - | 0 | 0 | - | 1 | 0 | 100.0<br>(16.7.103.9) |
| White | 0.972<br>(0.959,0.981) | 149 | 74 | 66.8<br>(60.4,72.7) | 2222 | 21 | 99.1<br>(98.6,99.4) |
| Other | 1.000<br>(0.997,1.000) | 9 | 0 | 100.0<br>(65.5.104.5) | 192 | 1 | 99.5<br>(96.8.100.2) |
| Two or More | 1.000<br>(1.000,1.000) | 3 | 1 | 75.0<br>(28.9,96.6) | 21 | 0 | 100.0<br>(81.8,102.8) |
| Declined | 1.000<br>(0.956,1.000) | 3 | 1 | 75.0<br>(28.9,96.6) | 34 | 0 | 100.0<br>(87.9.101.9) |
| Table 6: Subgroup Analysis by race | |
|------------------------------------|--|
|------------------------------------|--|
l
80.0
(36.0,98.0)
53
0
4
0-996
(0.966,1.000)
Unknown
100.0
(91.9,101.3)
{7}------------------------------------------------
| Subgroup | AUC<br>(95% CI) | True<br>positive | False<br>negative | Sensitivity<br>(95% CI) | True<br>negative | False<br>positive | Specificity<br>(95% CI) |
|--------------|------------------------|------------------|-------------------|-------------------------|------------------|-------------------|-------------------------|
| Hispanic | 1.000<br>(0.995,1.000) | 7 | 0 | 100.0<br>(59.6,105.0) | 167 | 3 | 98.2<br>(94.7,99.6) |
| Non Hispanic | 0.973<br>(0.962,0.982) | 166 | 80 | 67.5<br>(61.4,73.0) | 2485 | 21 | 99.2<br>(98.7,99.5) |
| Declined | 0.989<br>(0.964,1.000) | 11 | 6 | 64.7<br>(41.2,82.8) | 120 | 0 | 100.0<br>(96.3,100.6) |
| Unknown | 0.964<br>(0.923,0.990) | 15 | 6 | 71.4<br>(49.8,86.4) | 108 | 1 | 99.1<br>(94.5,100.3) |
### Table 7: Subgroup Analysis by ethnicity
| Subgroup | True<br>positive | False<br>negative | Sensitivity<br>(95% CI) |
|-----------------|------------------|-------------------|-------------------------|
| Obstructive | 97 | 45 | 68.3<br>(60.2,75.4) |
| Non-obstructive | 102 | 47 | 68.5<br>(60.6,75.4) |
Table 8: Subgroup Analysis by HCM type
Further stratification was conducted for:
- Concurrent ECG findings and anomalies, and .
- Comorbid conditions .
### Pediatric Extrapolation
For medical devices, the FD&C Act defines patients before their 22nd birthday as pediatric patients. In this De Novo request, data from patients between 18-21 were used to support the use of the device in patients over the age of 18.
# LABELING
The labeling supports the decision, including information on all required and/or compatible parts, to grant the De Novo request for this device. The labeling includes a detailed description of the device, description of the patient population for which the device is indicated for use, a description of the intended user population, and instructions for use.
The labeling reflects the following critical intended use and limitations of the Viz HCM device:
- The device identifies patients for further HCM follow-up and does not replace the current . standard of care methods for diagnosis of HCM. The results of the device are not intended to rule-out HCM follow-up.
- The device is not intended for use on patients with implanted pacemakers. .
- The results of the device should not be used in lieu of full patient evaluation. .
{8}------------------------------------------------
The following baseline demographic information is included in the labeling for age, race, ethnicity, sex, and HCM type.
| Training (N=301,106) | | | Testing (N=3196) | | |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------|---------------------------------------|------|-------|
| Sex | | | Sex | | |
| Female | 166,885 | 55.4% | Female | 1640 | 51.3% |
| Male | 134,221 | 44.6% | Male | 1556 | 48.7% |
| Age | | | Age | | |
| < 40 | 73,099 | 24.3% | < 40 | 524 | 16.4% |
| 40-65 | 127,551 | 42.4% | 40-65 | 1309 | 41.0% |
| > 65 | 100,456 | 33.4% | > 65 | 1363 | 42.6% |
| Race | | | Race | | |
| American Indian /<br>Alaska Native | *Data collected from multiple regions<br>(USA and OUS) representing different<br>races:<br>- United States<br>- Israel<br>- Germany<br>- Brazil | | American Indian /<br>Alaska Native | 2 | 0.1% |
| Asian | | | Asian | 109 | 3.4% |
| Black | | | Black | 295 | 9.2% |
| Native Hawaiian /<br>Pacific Islander | | | Native Hawaiian /<br>Pacific Islander | 1 | <0.1% |
| White | | | White | 2466 | 77.2% |
| Other | | | Other | 202 | 6.3% |
| Two or More | | | Two or More | 25 | 0.8% |
| Declined | | | Declined | 38 | 1.2% |
| Unknown | | | Unknown | 58 | 1.8% |
| Ethnicity | | | Ethnicity | | |
| Hispanic | *See notes for racial distribution. | | Hispanic | 177 | 5.5% |
| Non-Hispanic | | | Non-Hispanic | 2752 | 86.1% |
| Declined | | | Declined | 137 | 4.3% |
| Unknown | | | Unknown | 130 | 4.1% |
| HCM Presence | | | HCM Presence | | |
| Obstructive | 3,682 | 1.2% | Obstructive | 142 | 4.4% |
| Non-Obstructive | 892 | 0.3% | Non-Obstructive | 149 | 4.7% |
| No HCM (Negative) | 296,530 | 98.5% | No HCM (Negative) | 2905 | 90.9% |
Table 9. Training and Test set demographics
# RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of cardiovascular machine learning-based notification software.
| Identified Risks to Health | Mitigation Measures |
|--------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| False positive or false negative<br>leading to incorrect treatment or<br>diagnosis | Clinical performance testing<br>Non-clinical performance testing<br>Labeling |
| Incorrect treatment or diagnosis due<br>to model bias or failure to adequately<br>generalize to the intended use<br>population | Clinical performance testing<br>Labeling |
| Device used in unsupported patient<br>population or with unsupported<br>input/hardware | Labeling<br>Human factors assessment<br>Software verification, validation, and hazard analysis |
Table 10 - Identified Risks to Health and Mitigation Measures
{9}------------------------------------------------
| Overreliance on device output for | Human factors assessment |
|-----------------------------------|--------------------------|
| follow-up | Labeling |
# SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, cardiovascular machine learningbased notification software is subject to the following special controls:
- (1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be met:
- Clinical validation must use a test dataset of real-world data acquired from a (i) representative patient population. Data must be representative of the range of data sources and data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment. The test dataset must be independent from data used in training/development and contain sufficient numbers of cases from important cohorts (e.g., demographic populations, subsets defined by clinically relevant confounders, comorbidities, and subsets defined by hardware and acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and acquisition systems (e.g., acquisition hardware or preprocessing software). Study protocols must include a description of the adjudication process(es) for determining ground truth of training and test datasets;
- (ii) Data must be provided within the clinical validation study or using equivalent datasets to demonstrate the consistency of the output over the full range of inputs:
- Performance goals used to determine success of clinical validation must be (iii) justified in the context of risks associated with follow-up testing;
- Objective performance measures (e.g., sensitivity, specificity, positive predictive (iv) value or negative predictive value) must be reported with relevant descriptive or developmental performance measures. Summary level demographic information and sub-group analyses must be provided for each study site, relevant demographic sub-groups, and acquisition systems; and
- (v) The test dataset must include a minimum of 3 geographically diverse sites, separate from sites used in training of the model.
- (2) Software verification, validation, and hazard analysis must be performed. Software documentation must include:
- A description of the model/algorithm, algorithm inputs/outputs, and supported (i) patient population:
- (ii) Integration testing in the intended software system or software environment; and
- A description of the expected impact of all applicable sensor acquisition (iii) hardware characteristics on performance and any associated hardware specifications, including:
- (A) A description of input signal / data quality control measures; and
{10}------------------------------------------------
- A description of all mitigations for user error or failure of any subsystem (B) components (including signal detection, signal analysis, data display, and storage) on output accuracy.
- (3) Human factors assessment of the intended users in the intended use environment must evaluate the risk of misinterpretation of device output.
- (4) Labeling must include:
- A summary of the performance testing methods, tested hardware, (i) tested/supported patient population, results of the performance testing for tested performance measures/metrics, summary-level descriptions of patient demographics and associated subgroup analyses for training and test datasets, and the expected minimum performance of the device:
- (ii) Device limitations or subpopulations for which the device may not perform as expected:
- (iii) Warning that the user should not rely on the lack of a suspected finding to ruleout follow-up;
- A statement that the device output should not replace a full clinical evaluation of (iv) the patient and that the output may not be sufficient as the sole basis for further testing;
- Warnings identifying sensor acquisition factors that may impact measurement (v) results:
- (vi) Guidance for interpretation of the measurements and typical follow-up testing; and
- (vii) The type(s) of hardware sensor data used, including specification of compatible sensors for data acquisition.
# BENEFIT-RISK DETERMINATION
The risks of the device are based on nonclinical testing as well as data collected in a clinical study described above.
Primary risk is false negatives where a subject with HCM would be said to not likely have the disease. The original clinical testing and additional analyses provided adequate information to assess the likely performance of the device in the intended use population. The specificity is sufficiently high to address the risk of false negatives and there is low risk associated with routine follow-up testing with false positives. Due to the low prevalence of the disease, the positive predictive value is low but HCM is hard to diagnose, so the risk of false positives is acceptable given the benefit of identifying HCM in otherwise asymptomatic patients. Usability testing was performed to ensure that the output is not misinterpreted as a definitive diaenosis. This testing was performed in cardiologists rather than non-specialty clinicians; however, this is likely adequate as non-cardiologists would likely be more prone to referring to further follow-up. which is the intended use of the device.
The probable benefits of the device are also based on nonclinical laboratory studies as well as data collected in a clinical study as described above.
{11}------------------------------------------------
The ability to provide additional detection of and insight into a disease that is relatively uncommon and difficult to diagnose in subjects represents a benefit to the intended patient population.
# Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
# Benefit/Risk Conclusion
In conclusion, given the available information above, for the following indication statement:
Viz HCM is intended to be used in parallel to the standard of care to analyze recordings of 12lead ECG made on compatible ECG devices. Viz HCM is capable of analyzing the ECG. detecting signs associated with hypertrophic cardiomyopathy (HCM), and allowing the user to view the ECG and analysis results. Viz HCM is indicated for use on 12-lead ECG recordings collected from patients 18 years of age or older. Viz HCM is not intended for use on patients with implanted pacemakers. Viz HCM is limited to analysis of ECG data and should not be used in-lieu of full patient evaluation or relied upon to make or confirm diagnosis. Viz HCM identifies patients for further HCM follow-up and does not replace the current standard of care methods for diagnosis of HCM. The results of the device are not intended to rule-out HCM follow-up. The probable benefits outweigh the probable risks for the Viz HCM. The device provides benefits, and the risks can be mitigated by the use of general controls and the identified special controls.
# CONCLUSION
The De Novo for the Viz HCM is granted and the device is classified as follows:
Product Code: QXO Device Type: Cardiovascular machine learning-based notification software Regulation Number: 21 CFR 870.2380 Class: II
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.