ZBPro Diagnostic

K243252 · Zbeats, Inc. · DPS · Jul 10, 2025 · Cardiovascular

Device Facts

Record IDK243252
Device NameZBPro Diagnostic
ApplicantZbeats, Inc.
Product CodeDPS · Cardiovascular
Decision DateJul 10, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2340
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

ZBPro Diagnostic is a cloud-based medical device intended for use by qualified healthcare professionals in the detection and analysis of common cardiac arrhythmias in Holter ECG data in the adult, non-paced population. The product supports downloading and analyzing Lead II, CM5 (Ch 1), or Modified-MLII (Ch 2+) on retrospective 3-lead and 5-lead 24/48-hour Holter ECG recordings collected using standard Ag/AgCl wet electrodes in adult, non-paced patients. ZBPro is not intended for use with multi-lead analysis, wearable patches, or pediatric/paced recordings. ZBPro Diagnostic can also be electronically interfaced and perform analysis with data transferred from other computer-based ECG systems, such as an ECG management system. ZBPro Diagnostic provides ECG signal processing and analysis, QRS and Ventricular Ectopic Beat detection, QRS feature extraction, R-R interval measurement, heart rate measurement, and rhythm analysis. ZBPro Diagnostic is not for use in life-supporting or sustaining systems or ECG monitor and Alarm devices. The product can be integrated into computerized ECG monitoring devices. In this case, the medical device manufacturer will identify the indication for use depending on the application of their device. ZBPro Diagnostic interpretation results are not intended to be the sole means of diagnosis. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information.

Device Story

Cloud-based SaMD for retrospective analysis of ambulatory Holter ECG recordings; inputs include 24-48 hour ECG data (Lead II, CM5, Modified-MLII) uploaded via web browser/API. Proprietary algorithm performs signal processing, QRS detection, ventricular ectopic beat detection, R-R interval/heart rate measurement, and rhythm analysis. Used by qualified healthcare professionals (physicians/clinicians) in clinical settings to review and annotate ECGs. Output consists of annotated ECG reports and rhythm analysis provided as advisory information. Clinicians use output to support diagnostic decision-making; device does not provide standalone diagnosis. Benefits include automated, efficient analysis of Holter data to assist in arrhythmia detection.

Clinical Evidence

Bench testing only. Performance validated using adjudicated databases per ANSI/AAMI EC57 and IEC 60601-2-47 reporting conventions. Robustness evaluated via ZBPro Noise Stress Testing (ZNST) database. Software verification and validation confirmed design inputs and specifications met. Usability testing conducted with Certified Cardiovascular Technicians.

Technological Characteristics

Cloud-hosted (AWS) SaMD; web-browser interface. Supports 3-lead/5-lead Holter ECG (Lead II, CM5, Modified-MLII). Proprietary automated ECG interpretation algorithm. Standards: AAMI ANSI IEC 62304 (software lifecycle), IEC 62366-1 (usability), AAMI ANSI EC57 (rhythm/ST measurement), AAMI ANSI IEC 60601-2-47 (ambulatory ECG safety/performance).

Indications for Use

Indicated for adult, non-paced patients requiring detection and analysis of common cardiac arrhythmias from retrospective 3-lead or 5-lead 24/48-hour Holter ECG recordings. Not for pediatric or paced populations, multi-lead analysis, wearable patches, life-support, or real-time monitoring.

Regulatory Classification

Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION July 10, 2025 ZBeats Inc. % Christopher Ford Regulatory Affairs Consultant Zaliance Medical Solutions 5153 West Wooley Rd Oxnard, California 93035 Re: K243252 Trade/Device Name: ZBPro Diagnostic Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS, DQK Dated: June 13, 2025 Received: June 13, 2025 Dear Christopher Ford: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K243252 - Christopher Ford Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K243252 - Christopher Ford Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jennifer W. Shih -S Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K243252 Device Name ZBPro Diagnostic Indications for Use (Describe) ZBPro Diagnostic is a cloud-based medical device intended for use by qualified healthcare professionals in the detection and analysis of common cardiac arrhythmias in Holter ECG data in the adult, non-paced population. The product supports downloading and analyzing Lead II, CM5 (Ch 1), or Modified-MLII (Ch 2+) on retrospective 3-lead and 5-lead 24/48-hour Holter ECG recordings collected using standard Ag/AgCl wet electrodes in adult, non-paced patients. ZBPro is not intended for use with multi-lead analysis, wearable patches, or pediatric/paced recordings. ZBPro Diagnostic can also be electronically interfaced and perform analysis with data transferred from other computer-based ECG systems, such as an ECG management system. ZBPro Diagnostic provides ECG signal processing and analysis, QRS and Ventricular Ectopic Beat detection, QRS feature extraction, R-R interval measurement, heart rate measurement, and rhythm analysis. ZBPro Diagnostic is not for use in life-supporting or sustaining systems or ECG monitor and Alarm devices. The product can be integrated into computerized ECG monitoring devices. In this case, the medical device manufacturer will identify the indication for use depending on the application of their device. ZBPro Diagnostic interpretation results are not intended to be the sole means of diagnosis. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information. Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 {4} ZBeats, Inc. ZBPro Diagnostic Traditional 510(K) Submission # 510(K) Summary ## 5.1 Applicant/Submitter Company Name : ZBeats, Inc Street Address : 25 Health Sciences Drive City : Stony Brook NY State : NY Zip Code : 11790 Phone Number : (631) 220-2463 ## 5.2 Contact Person Full Name : Bin Fang Phone : (631) 220-2463 ## 5.3 Correspondent Information Full Name : Christopher Ford Phone : (916) 612-9954 ## 5.4 Date of Preparation Date of Preparation : 07/10/2025 ## 5.5 Device Information | Device Name | ZBPro Diagnostic | | --- | --- | | Model Number | | K243252 {5} ZBeats, Inc. ZBPro Diagnostic Traditional 510(K) Submission | Common Name | ECG Analysis System | | --- | --- | | Regulation | 21 CFR 870.2340, 21 CFR 870.1425 | | Product Code | DPS;DQK | | Reviewing Panel/Branch | Cardiovascular | | Submission Number | K243252 | # 5.6 Predicate Device(s) Table - 5.1 Predicate Device(s) | Predicate Type | 510(k) Number | Device Name | Manufacturer | Product Code | | --- | --- | --- | --- | --- | | Primary | K212112 | Cardiologs Holter Platform | Cardiologs Technologies | DPS, DQK | # 5.7 Indications for Use ZBPro Diagnostic is a cloud-based medical device intended for use by qualified healthcare professionals in the detection and analysis of common cardiac arrhythmias in Holter ECG data in the adult, non-paced population. The product supports downloading and analyzing Lead II, CM5 (Ch 1), or Modified-MLII (Ch 2+) on retrospective 3-lead and 5-lead 24/48-hour Holter ECG recordings collected using standard Ag/AgCl wet electrodes in adult, non-paced patients. ZBPro is not intended for use with multi-lead analysis, wearable patches, or pediatric/paced recordings. ZBPro Diagnostic can also be electronically interfaced and perform analysis with data transferred from other computer-based ECG systems, such as an ECG management system. ZBPro Diagnostic provides ECG signal processing and analysis, QRS and Ventricular Ectopic Beat detection, QRS feature extraction, R-R interval measurement, heart rate measurement, and rhythm analysis. ZBPro Diagnostic is not for use in life-supporting or sustaining systems or ECG monitor and Alarm devices. The product can be integrated into computerized ECG monitoring devices. In this case, the medical device manufacturer will identify the indication for use depending on the application of their device. ZBPro Diagnostic interpretation results are not intended to be the sole means of diagnosis. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information. # 5.8 Device Description ZBPro is cloud-based Software as a Medical Device which aids healthcare professionals in interpreting ambulatory ECG recordings. The software comprises a secure web interface and a backend server hosted on Amazon Web Server (AWS). Authenticated users upload compatible 24-48 hour Holter ECG recordings via a web browser through an Application Programming Interface (API). ZBPro's proprietary ECG interpretation algorithm analyzes and annotates ECGs to provide supportive information for ECG and arrhythmia analysis. ZBPro provides beat-by-beat ECG signal processing and analysis, QRS detection, Ventricular Ectopic Beat detection, R-R interval measurement, heart rate and Heart Rate Variability measurement, and rhythm analysis. ZBPro consists of: K243252 Page 2 of 7 {6} ZBeats, Inc. ZBPro Diagnostic Traditional 510(K) Submission 1. A web interface which provides tools to upload data, measure, analyze and review numerous ECGs and patient diary logs, make manual annotation and generate ECG reports. 2. An automated proprietary ECG interpretation algorithm which measures and analyzes ECGs to provide adjunct information for ECG diagnosis. The backend application is established in Amazon Web Services (AWS) and accessed through an Internet connection and a web browser to perform ECG analysis and generate reports. # 5.9 Intended Use/Indications for Use Comparison ZBPro Diagnostic (K243252) and the Cardiologs Holter Platform (K212112) are both cloud-based ECG analysis platforms intended for retrospective review of ambulatory electrocardiogram (ECG) recordings to assist qualified healthcare professionals in detecting and analyzing cardiac arrhythmias. Both systems support integration with external ECG acquisition devices and provide core analytical functions including QRS detection, rhythm classification, ectopic beat detection, and heart rate trend analysis. The technological differences between ZBPro Diagnostic and its predicate are limited to algorithmic design and scope of use. These differences are mitigated through validated performance, labeling restrictions, and risk controls. ZBPro Diagnostic is substantially equivalent to the Cardiologs Holter Platform and raises no new questions of safety or effectiveness as compared to the predicate. Both devices are intended for use by qualified healthcare professionals to detect and analyze common cardiac arrhythmias using retrospective ambulatory ECG data. The primary indication (arrhythmia assessment in Holter ECG recordings in adult patients) is identical in both devices, and both utilize well-established principles of signal processing, QRS detection, rhythm classification, and clinical data presentation. The primary differences relate to patient population and recording source compatibility. ZBPro Diagnostic is limited to the adult, non-paced population, whereas Cardiologs includes pediatric and paced patient populations and accepts data from additional sources (e.g., event recorders, 12-lead ambulatory ECG). These differences reflect scope of validation rather than underlying technological capabilities and do not introduce new questions of safety or effectiveness. ZBPro's intended use is clearly delineated in its labeling and supported by clinical validation. Both devices support electronic interfacing with other ECG systems and offer nearly identical functional capabilities, including ECG signal processing, ventricular ectopy detection, interval and heart rate analysis, and comprehensive rhythm interpretation. Neither system is intended for use in life-sustaining applications or real-time monitoring, and both clearly state that their results are to be used in conjunction with clinician oversight, not as a standalone diagnostic decision-maker. While Cardiologs has broader input compatibility and population coverage, ZBPro maintains equivalence in intended use and analytical functionality. The observed differences are managed through labeling and intended use limitations. Therefore, these distinctions do not impact the substantial equivalence of the two systems in their shared core function to provide automated ECG interpretation support for adult Holter monitoring data. # 5.10 Comparison of Technological Characteristics with Predicate The subject device has the same intended use and substantially equivalent technological characteristics that do not raise questions of safety or effectiveness as the predicate. Table - 5.2 Feature Comparison: ZBPro Diagnostic to the predicate device illustrates a comparison of detection features between ZBPro Diagnostic and the cited predicate. Table - 5.2 Feature Comparison: ZBPro Diagnostic to the predicate device | Feature | ZBPro Diagnostic | Cardiologs Holter Platform | | --- | --- | --- | | Heart rate determination for non-paced adult | Yes | Yes | | QRS Detection | Yes | Yes | | Non-paced arrhythmia interpretation for adult patients | Yes | Yes | | Non-paced ventricular arrhythmia calls | Yes | Yes | K243252 {7} ZBeats, Inc. ZBPro Diagnostic Traditional S10(K) Submission Table - 5.3 Comparison of technological characteristics: ZBPro Diagnostic to the predicate device provides a comparison of the technological characteristics of the ZBPro Diagnostic with the cited predicate. Table - 5.3 Comparison of technological characteristics: ZBPro Diagnostic to the predicate device | | ZBPro Diagnostic (this submission) | Cardiologs Holter Platform | Comparison to predicate device | | --- | --- | --- | --- | | Device Trade Name | ZBPro Diagnostic | CardioLogs ECG Analysis Platform | N/A | | Manufacturer | ZBeats, Inc. | CardioLogs Technologies | N/A | | 510(K) No. | (this submission) | K212112 | N/A | | Regulation Number | 21 CFR 870.2340 21 CFR 870.1425 | 21 CFR 870.2340 21 CFR 870.1425 | Same | | Product Code | DQK, DPS | DQK, DPS | Same | | Class | 2 | 2 | Same | | Device Class/Name | Electrocardiograph | Electrocardiograph | Same | | Indications for Use | ZBPro Diagnostic is a cloud-based medical device intended for use by qualified healthcare professionals in the detection and analysis of common cardiac arrhythmias in Holter ECG data in the adult, non-paced population. The product supports downloading and analyzing Lead II, CM5 (Ch 1), or Modified-MLII (Ch 2+) on retrospective 3-lead and 5-lead 24/48-hour Holter ECG recordings collected using standard Ag/AgCl wet electrodes in adult, non-paced patients. ZBPro is not intended for use with multi-lead analysis, wearable patches, or pediatric/paced recordings. ZBPro Diagnostic can also be electronically interfaced and perform analysis with data transferred from other computer-based ECG systems, such as an ECG management system. ZBPro Diagnostic provides ECG signal processing and analysis, QRS and Ventricular Ectopic Beat detection, QRS feature extraction, R-R interval measurement, heart rate measurement, and rhythm analysis. ZBPro Diagnostic is not for use in life-supporting or sustaining | Cardiologs Holter Platform is intended for use by qualified healthcare professionals for the assessment of arrhythmias using ECG data in the adult and pediatric population. The product supports downloading and analyzing data recorded in compatible formats from any device used for arrhythmia diagnostics such as Holter, event recorder, 12-lead ambulatory ECG devices, or similar devices when assessment of the rhythm is necessary. The Cardiologs Holter Platform can also be electronically interfaced and perform analysis with data transferred from other computer-based ECG systems, such as an ECG management system. Cardiologs provides ECG signal processing and analysis, QRS and Ventricular Ectopic Beat detection, QRS feature extraction, interval measurement, heart rate measurement, and rhythm analysis. Cardiologs is not for use in life-supporting systems or ECG monitor and Alarm devices. The product can be integrated into computerized ECG monitoring | Both are cloud-based medical devices intended for use by qualified healthcare professionals in the detection and analysis of common cardiac arrhythmias in Holter ECG data in the adult, non-paced population. However, Cardiologs is also intended for use on pediatric patients and paced patients. This increased compatibility does not introduce new questions of safety or efficacy as compared to the predicate as both systems use well-established principles of ECG signal detection and analysis and are validated for their intended uses. Both products support downloading and analyzing equivalent vectors on retrospective 3-lead and 5-lead 24/48-hour Holter ECG recordings taken with standard Ag/AgCl wet electrodes, however Cardiologs supports event recorders, 12-lead ambulatory ECG devices, and similar recording devices. This increased compatibility does not introduce new questions of safety or efficacy compared to the predicate as both systems use well-established principles of ECG signal detection and analysis and are validated for their intended uses. | K243252 {8} ZBeats, Inc. ZBPro Diagnostic Traditional 510(K) Submission | | ZBPro Diagnostic (this submission) | Cardiologs Holter Platform | Comparison to predicate device | | --- | --- | --- | --- | | | systems or ECG monitor and Alarm devices. The product can be integrated into computerized ECG monitoring devices. In this case, the medical device manufacturer will identify the indication for use depending on the application of their device. ZBPro Diagnostic interpretation results are not intended to be the sole means of diagnosis. It is offered to physicians and clinicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient background, clinical history, symptoms, and other diagnostic information. | devices. In this case, the medical device manufacturer will identify the indication for use depending on the application of their device. Cardiology interpretation results are not intended to be the sole means of diagnosis and are offered to physicians on an advisory basis only in conjunction with the physician's knowledge of ECG patterns, patient history, and other diagnostic information. | | | **Fundamental scientific technology** | ZBPro consists of: A web interface which provides tools to upload data, measure, analyze and review numerous ECGs and patient diary logs, make manual annotation and generate ECG reports. An automated proprietary ECG interpretation algorithm which measures and analyzes ECGs to provide adjunct information for ECG diagnosis. The backend application is established in Amazon Web Services (AWS) and accessed through an Internet connection and a web browser to perform ECG analysis and generate reports. | The Cardiologs Holter Platform consists of: An interface which provides tools to measure, analyze and review numerous ECGs coded in java language under the Angular and D3.js frameworks; An automated proprietary ECG interpretation support algorithm which measures and analyzes ECGs to provide supportive information for ECG diagnosis, written in Python language. This application can be accessed through an Internet connection and a web browser, or is directly connected to the Cardiologs’ Application Programming Interface (API). | ZBPro Diagnostic does not present any major technological differences compared to the predicate device. Substantially equivalent: Both devices provide a web interface which provides tools to upload, analyze, review ECG data. Both consist of a proprietary ECG interpretation support algorithm and are not intended to be the sole means of diagnosis. The interpretations are offered to physicians on an advisory or assistive basis. While the two platforms differ in input generalization and architectural approach, these differences do not alter the shared fundamental technology purpose of ECG rhythm analysis for Holter monitoring. ZBPro Diagnostic achieves equivalent clinical functionality within its labeled scope and is therefore substantially equivalent in its technological characteristics to the predicate device. Both are accessed through an internet connection and a web browser, but the predicate device can also be accessed through a direct connection to the API. This connection difference should not have any effect on intended use or performance and does not introduce new issues of safety or effectiveness as compared to the predicate. | ZBPro Diagnostic and the Cardiologs Holter Platform are both cloud-based ECG analysis systems designed to receive, process, and support clinician review of retrospective ambulatory ECG recordings. Both platforms provide secure, browser-accessible environments, RESTful API integration, and clinical tools for ECG visualization, annotation, and cardiologist confirmation. K243252 Page 5 of 7 {9} ZBeats, Inc. ZBPro Diagnostic Traditional 510(K) Submission The core technologies used by each platform are similar in functional intent but differ in algorithmic framework implementation. ZBPro Diagnostic processes standard-format Holter ECG files (MIT-BIH, EDF, ISHNE) acquired with validated lead configurations (Lead II, CM5 (Ch 1), or Modified-MLII (Ch 2+)). From a technological architecture perspective, both systems: a. Operate via secure, cloud-hosted infrastructure; b. Offer browser-based access for clinical and technical users; c. Use RESTful APIs for interoperability with EHRs, RPM systems, and diagnostic services; d. Provide ECG editing tools, rhythm markup capabilities, and cardiologist review workflows. ZBPro Diagnostic is currently validated for use in the Google Chrome browser, while Cardiology offers broader cross-platform compatibility, including support for Safari and iOS environments. This distinction affects user workflow preferences but not the core diagnostic capabilities. All observed differences are mitigated through ZBPro's labeling, risk controls, and software safeguards. The system's limited scope, focused on adult Holter patients using defined input vectors, does not rely on expanded device compatibility and is substantially equivalent. In summary, while the two platforms differ in input generalization and architectural approach, these differences do not alter the shared fundamental technology purpose: ECG rhythm analysis for Holter monitoring. ZBPro Diagnostic achieves equivalent clinical functionality within its labeled scope and is therefore substantially equivalent in its technological characteristics to the predicate device. ## 5.11 Summary of Performance Data Performance tests have been performed in compliance with the following recognized consensus standards: - AAMI ANSI IEC 62304 2006 - Medical device software - Software life-cycle processes - IEC 62366-1 Edition 1.0 2015-02 - Medical devices - Application of usability engineering to medical devices. - AAMI ANSI EC57:2012 - Testing and Reporting Performance Results of Cardiac Rhythm And ST-Segment Measurement Algorithms - AAMI ANSI IEC60601-2-47:2012 - Medical Electrical Equipment - Part 2-47: Particular Requirements For The Basic Safety And Essential Performance Of Ambulatory Electrocardiographic Systems ## 5.12 Software verification and validation Testing described in this 510(k) consisted of verification of all design input requirements and product specifications. All clinical input requirements were validated against a gold standard. No residual anomalies appeared during verification and software validation tests. General usability tests, analyzing the users' ability to import, display, store, analysis, distribute, and manage ECG data, were performed by certified cardiovascular technicians and met all requirements. All software validation testing was completed successfully and met all requirements. Based upon the Intended Use, Indications for Use, product technical information, performance evaluation, and standards compliance provided in this premarket notification, ZBPro Diagnostic has been shown to be substantially equivalent to the cited predicate. Usability of ZBPro Diagnostic was tested throughout development, validating the effectiveness of risk control measures associated with the user interface. ZBeats conducted usability tests of the ZBPro user interface with Certified Cardiovascular Technicians to evaluate the usability of the design, labeling and risk. The protocols cover the testable technical requirements relevant to the user interface as described in the User Interface Specification. Performance validation testing included comprehensive rhythm classification analyses on an adjudicated database in accordance with ANSI/AAMI EC57 and IEC 60601-2-47 reporting conventions, as detailed in the ZBPro Diagnostic Performance Validation Report. Additional noise stress testing using the ZBPro Noise Stress Testing (ZNST) database was conducted to evaluate algorithm robustness under degraded signal conditions. K243252 {10} ZBeats, Inc. ZBPro Diagnostic Traditional S10(K) Submission ## 5.13 Conclusion Based upon the Intended Use, Indications for Use, product technical information, performance evaluation, and standards compliance provided in this premarket notification, ZBPro Diagnostic has been shown to be substantially equivalent to the cited predicate. K243252 Page 7 of 7
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...