The sponsor utilized post-market surveillance data from the EU market to confirm the safety, performance, and usability conclusions of the device, supporting the clinical evaluation.
Post-Market Surveillance; EU Market Data; Performance Monitoring
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
EU Post-Market Surveillance (PMS); Post-market surveillance
Patients in the EU market using the ARTHUR device
Not applicable for this study
Performance, safety, and usability data
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Musculoskeletal ultrasound image capture
—
Scan at least 85% of MCP 1-5, PIP 1-5, wrist RC-IC and DIP 1-4 per hand
Successfully acquires diagnostically usable ultrasound images for at least 85% of targeted joints
—
—
Performance and usability study conducted at a clinic in USA
—
Indications for Use
ARTHUR by ROPCA is intended to hold an external ultrasound transducer of cleared ultrasound equipment for automatic image capturing of finger and wrist joints. ARTHUR is indicated for adult patients, that are under investigation or monitoring of musculoskeletal diseases. The system shall be used in a professional healthcare facility environment. ARTHURs mode of operation is to use B-mode (2D), Color Doppler and/or Power Doppler mode on connected ultrasound machines. The clinical application is musculoskeletal ultrasound of the hand and wrist joints.
Device Story
ARTHUR is a robotic accessory for FDA-cleared ultrasound machines; automates image capture of finger, hand, and wrist joints. System uses camera technology to detect patient hand anatomy; robotic arm positions ultrasound probe for scanning. Operates in B-mode (2D), Color Doppler, and Power Doppler. Used in professional healthcare facilities; patient guided by touchscreen interface; supervised by healthcare professional. Device captures grayscale images and Doppler videos; stored for clinician review to assist in diagnosis and monitoring of musculoskeletal diseases. Includes safety features: collision detection, kinematic force limits (max 50N, 10N normal use), emergency stop, and mechanical probe release. Does not provide diagnostic decisions; output used by clinicians for assessment.
Clinical Evidence
Bench and usability testing performed. Performance study demonstrated ARTHUR successfully captures diagnostically usable ultrasound images for at least 85% of targeted joints (MCP 1-5, PIP 1-5, wrist RC-IC, DIP 1-4). Usability study confirmed clinical personnel and patients could complete examinations safely. No clinical diagnostic performance metrics (sensitivity/specificity) reported as device is an image-capture accessory; clinical assessment remains with the physician.
Technological Characteristics
Robotic arm accessory for third-party ultrasound probes. Materials/specs: kinematic control system, collision detection, mechanical release. Energy: electrical. Connectivity: interfaces with cleared ultrasound machines. Software: Class C (IEC 62304). Safety: force limited to 50N max, 10N normal use. Sterilization: N/A (accessory). Algorithm: fixed, non-adaptive AI models for joint detection and image evaluation.
Indications for Use
Indicated for adult patients under investigation or monitoring for musculoskeletal diseases of the hand and wrist joints.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
MELODY, Remote Control System for Ultrasound Probe (K161354)
Sofia Automated Tomographic Ultrasound (ATUS) (K080555)
Submission Summary (Full Text)
{1}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
ROPCA ApS
Trine Winther
Head of Quality
Cortex Park 26 E
Odense M, 5230
Denmark
June 18, 2026
Re: K260190
Trade/Device Name: Arthur
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX, QIH
Dated: May 15, 2026
Received: May 15, 2026
Dear Trine Winther:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260190 - Trine Winther
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K260190 - Trine Winther
Page 3
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2026.06.18 13:00:20 -04'00' For
Yanna Kang, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260190 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| ARTHUR | | |
| Please provide your Indications for Use below. | | ? |
| ARTHUR by ROPCA is intended to hold an external ultrasound transducer of cleared ultrasound equipment for automatic image capturing of finger and wrist joints. ARTHUR is indicated for adult patients, that are under investigation or monitoring of musculoskeletal diseases. The system shall be used in a professional healthcare facility environment. ARTHURs mode of operation is to use B-mode (2D), Color Doppler and/or Power Doppler mode on connected ultrasound machines. The clinical application is musculoskeletal ultrasound of the hand and wrist joints. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
| 510(k) #: K260190 | 510(k) Summary | Prepared on: 2026-06-18 |
| --- | --- | --- |
| Contact Details | K260190 | 21 CFR 807.92(a)(1) |
| Applicant Name | ROPCA ApS | |
| Applicant Address | Cortex Park 26 E Odense M 5230 Denmark | |
| Applicant Contact Telephone | +45 22 53 22 07 | |
| Applicant Contact | Ms. Trine Winther | |
| Applicant Contact Email | tw@ropca.com | |
| Device Name | 21 CFR 807.92(a)(2) | |
| Device Trade Name | ARTHUR | |
| Common Name | Ultrasonic pulsed echo imaging system | |
| Classification Name | System, Imaging, Pulsed Echo, Ultrasonic | |
| Regulation Number | 892.1560, 892.1570, 892.1550, 892.2050 | |
| Product Code(s) | IYO, ITX, IYN, QIH | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K161354 | MELODY, Remote Control System for Ultrasound Probe | IYO |
| K080555 | Sofia Automated Tomographic Ultrasound (ATUS) | ITX |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| ARTHUR stands for ARTHritis Ultrasound Robot and is an accessory to already verified and FDA-cleared off the shelf ultrasound machines. | | |
| Upon connection with the ultrasound unit, ARTHUR can automatic capture ultrasound images of various joints of the fingers, hands, and wrists that can be utilized for medical review. The patient is supervised by a healthcare professional to follow the instructions from ARTHUR. | | |
| The main interface to the platform is an intuitive touchscreen that will guide the patient user through the stages of the examination process, supervised by a healthcare professional(e.g., hand placement, gel application). Through advanced camera technology, ARTHUR automatically detects the patient's hands - identifying their individual hand joints. Once proper hand placement is identified, the robotic arm moves an attached ultrasound probe to each individually selected joint, scanning them thoroughly while ensuring safe and proper contact between the patient and probe. During the scanning procedure, ARTHUR communicates with the ultrasound machine, simultaneously changing modes between greyscale and Doppler images. One grayscale image and one doppler video for each joint is saved on the platform for the clinician to review and use as assistance input for diagnosis and monitoring of musculoskeletal diseases. | | |
| ARTHUR is intended for use on adult patients, that are under clinical investigation for, or monitoring of, musculoskeletal diseases and operated/supervised by trained healthcare professionals. The intended environment of the device are healthcare facilities, as a healthcare professional is required. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
{6}
ARTHUR by ROPCA is intended to hold an external ultrasound transducer of cleared ultrasound equipment for automatic image capturing of finger and wrist joints.
ARTHUR is indicated for adult patients, that are under investigation or monitoring of musculoskeletal diseases.
The system shall be used in a professional healthcare facility environment.
ARTHURs mode of operation is to use B-mode (2D), Color Doppler and/or Power Doppler mode on connected ultrasound machines.
The clinical application is musculoskeletal ultrasound of the hand and wrist joints.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The indications for use of the subject device are similar to the predicate device's indications for use as both devices are indicated to hold an already cleared third-party ultrasound probe by a software controlled robotic arm to perform an ultrasound scan.
The difference lies in how the software controlled robotic arm is operated. The predicate device is operated remotely by the sonographer with support of an on-site assistant, and the subject device's ultrasound image collection operated by software algorithms including artificial intelligence and is also supported by an on-site assistant. However, this difference has no influence on safety and performance, which is also supported by the reference device, as it also performs a fully automatic scan. Thus, the indications can be considered equivalent to the predicate as the artificial intelligence memes human behavior. Another difference lies in anatomical location, that also does not raise any new safety or performance issues, since both the subject and predicate device are accessories to already cleared ultrasound devices. Therefore, the intended anatomical sites, or any limitations thereon, are determined by the ultrasound device and its probes.
## Technological Comparison
21 CFR 807.92(a)(6)
The general technological characteristics of the subject device and the predicate device are substantial equivalent, as both devices use software-controlled robotic arms to hold cleared third-party ultrasound probes for ultrasound scans. The primary technological difference between the two devices lies in the method of operation of the software-controlled robotic arm. The predicate device is operated remotely by a sonographer with assistance from an on-site operator, whereas the subject device operates fully automated by software algorithms that control probe positioning and ultrasound image acquisition supported by an on-site assistant. Hereby, the subject device's software uses algorithms to detect joint positions and control the robotic movements of the ultrasound probe, followed by a real-time analysis of the acquired ultrasound images, which enables the system to automatically identify and store the optimal ultrasound image of each joint. Both the joint position detection and the ultrasound image evaluation algorithms are based on artificially developed models that are fixed and non-adaptive during production. As the models are static, performance verification and validation are feasible and controlled. Therefore automatic movement of the ultrasound probe is comparable to the reference device. The ultrasound images will in the following treatment be review and used for clinical assessment.
Despite the difference in automation level, the technological characteristics do not raise new or different questions of safety or effectiveness, as the subject device is designed with Class C software in accordance with IEC 62304 and incorporates safety features including collision detection and a kinematic control system that limits the maximum applied force to 50 N and a force of 10N in normal use. The normal force is similar to the predicate device of 20 N. The differences are due to different anatomical structures and required pressure due to image quality. Despite this, there are no safety concerns, since this force limit is significantly below the thresholds recommended in ISO 15066 for human-robot interaction.
In addition, the device is equipped with different safety controls as an emergency button to stop the robotic arm movement abruptly or a mechanical release mechanism for the ultrasound probe to allow immediate disengagement from the patient in worst-case scenarios. The mechanical design, together with the software safety configuration, minimizes the risk of injury, trapping, or unintended contact during operation.
Based on the same intended use and the controlled differences in technological characteristics, the subject device is substantially equivalent to the predicate device. Performance data in support of the substantial equivalence determination e.g., biocompatibility testing in accordance to ISO 10993-1, electrical safety and electromagnetic compatibility according to IEC 60601-1 and IEC 60601-1-2 and other verification and validation testing is provided.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
ARTHUR by ROPCA is an accessory to a verified ultrasound machine(s) and provides an automatic way to capture ultrasound images of wrist and finger joints. According to the market analysis included in this evaluation, ARTHUR is a new solution for a growing market since it automates an already used and proven clinical method. The state of the art investigation also showed a clear indication of ultrasound being the preferred and dominant imaging modality currently available for early diagnosis and disease monitoring in possible musculoskeletal diseases.
{7}
Based on the risk management process, the residual risk and risk benefit analysis finds that ARTHUR does not expose the patient to any unacceptable risk. It showed an overall low residual risk for ARTHUR to capture images of insufficient quality or no images at all. However, this is of no hazard related consequence to the patient, thus this risk is mitigated by the doctor. The risk for the doctor is to revert to manually scanning the patient as it is done currently or use joint palpation. There is no risk related to the outcome of the treatment for the patients by introducing ARTHUR in the process.
The performance study and the usability study showed that ARTHUR was safe to use. Both the clinical personnel and the patients were able to understand the instructions and to complete the tests with few non-hazardous situations.
The performance studies showed conformity with the clinical benefit. Furthermore, ARTHUR by ROPCA conforms with the criteria for performance and benefit risk, and the literature investigated did not contradict the essential performance requirement. Based on the statistical analysis we have provided evidence that 'ARTHUR shall succeed in scanning at least 85% on average of MCP 1-5, PIP 1-5, wrist RC-IC and DIP 1-4 per hand.' thereby it can be claimed 'ARTHUR by ROPCA automatically captures standardised ultrasound images of the hands for clinical use.'
The Post-Market Surveillance (PMS) activities from EU market has not shown any changes to the conclusions from the initial release of this clinical evaluation and will be continued as defined in the PMS plan and ROPCA will continuously collect performance, safety and usability data which will be processed and used to ensure continued service and performance monitoring.
An additional Usability and performance study conducted at a clinic in USA was performed and has showed good results whereas all criteria and the clinical claim are fulfilled.
ARTHUR is a low-risk medical device incorporating multiple safety features, including software-based safety controls and a mechanical release mechanism for the third-party, FDA-cleared ultrasound probe. These safety features are designed to minimize the risk of injury during operation.
ARTHUR automatically acquires ultrasound images intended to support the diagnosis and monitoring of musculoskeletal diseases. The device does not provide diagnostic decisions; all acquired ultrasound images are intended to be reviewed and interpreted by qualified clinical professionals.
Performance testing demonstrates that ARTHUR successfully acquires diagnostically usable ultrasound images for at least 85% of targeted joints. Based on the intended use, technological characteristics, safety controls, and performance data provided in this submission, ARTHUR does not raise new or different questions of safety or effectiveness compared to the predicate device and is therefore substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.