K260145 · Zhenjiang Assure Medical Equipment Co., Ltd. · IOR · Sep 15, 2026 · Physical Medicine
Device Facts
Record ID
K260145
Device Name
YJ-K1K2 Wheelchair (YJ-013F)
Applicant
Zhenjiang Assure Medical Equipment Co., Ltd.
Product Code
IOR · Physical Medicine
Decision Date
Sep 15, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.3850
Device Class
Class 1
Indications for Use
The YJ-K1K2 Wheelchair(YJ-013F) is to provide mobility to persons limited to a sitting position.
Device Story
Manual, foldable mechanical wheelchair; provides mobility to users restricted to sitting position. Features main frame, seat, adjustable footrests, four wheels, hand rims for manual propulsion, and brakes. Operated by user or assistant via push handles. Used in indoor/outdoor environments. Provides mobility; facilitates transport via folding mechanism. Benefits patient by enabling independent or assisted movement.
Clinical Evidence
No clinical data. Bench testing only. Device performance verified against ISO 7176-1, -3, -5, -7, -8, -11, -13, -15, -22, and ISO 16840-10.
Technological Characteristics
Mechanical wheelchair; steel frame; nylon upholstery; PU armrest padding; aluminum brake/fork; PP/steel handrims. Manual operation; foldable cross-brace design. Dimensions: 990x580x820mm. Weight capacity: 150 lbs. Complies with ISO 7176 series and ISO 16840-10. No electronics or software.
Indications for Use
Indicated for persons limited to a sitting position requiring mobility assistance.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{1}
**U.S. FOOD & DRUG**
ADMINISTRATION
September 15, 2026
Zhenjiang Assure Medical Equipment Co., Ltd.
% Eric Shi
Deputy R&D Manager
Shanghai Jiushun Enterprise Management Technology Service Co
15 Floor, 25 Floor, Bao An Tower #800, Dongfang Rd.
Shanghai, 200122
China
Re: K260145
Trade/Device Name: YJ-K1K2 Wheelchair (YJ-013F)
Regulation Number: 21 CFR 890.3850
Regulation Name: Mechanical Wheelchair
Regulatory Class: Class I, reserved
Product Code: IOR
Dated: January 8, 2026
Received: January 20, 2026
Dear Eric Shi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260145 - Eric Shi
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K260145 - Eric Shi
Page 3
Sincerely,
# Tushar Bansal -S
for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260145
Device Name
YJ-K1K2 Wheelchair(YJ-013F)
Indications for Use (Describe)
The YJ-K1K2 Wheelchair(YJ-013F) is to provide mobility to persons limited to a sitting position.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
# **CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# **\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}
Sponsor: Zhenjiang Assure Medical Equipment Co., Ltd.
Subject Device: YJ-K1K2 Wheelchair(YJ-013F)
# 510(k) Summary
# K260145
## 1. Applicant Information
Company Name: Zhenjiang Assure Medical device Co., Ltd.
Establishment Registration Number: 3007309323
Address: No.297, Chuqiao Road, Zhenjiang City, Jiangsu Province, China.
Phone:+86-13852941204
Contact Person (including title): Mei Ying (Deputy R&D Manager)
E-mail: shelly@assuremedical.com.cn
## 2. Primary Correspondent/Consultant Information
Company Name: Shanghai Jiushun Enterprise Management Technology Service Co., Ltd.
Address: 15 floor, 25 floor, Bao An Tower, No.800, Dongfang Road, Shanghai, China.
Zip code: 200122
Contact Person (including title): Eric Shi (Deputy R&D Manager)
Phone:+86-18621771627
E-mail: eric-shi@isosh.com
Prepared Date: 3 April 2026
## 3. Subject Device Information
Type of 510(k) submission: Special
Trade/Device Name: YJ-K1K2 Wheelchair(YJ-013F)
Regulation Name: Mechanical Wheelchair
Product Code: IOR
Regulation Number: 890.3850
Regulation Class: 1
## 4. Predicate Device Information
Sponsor: Zhenjiang Assure Medical Equipment Co., Ltd.
1
{6}
Sponsor: Zhenjiang Assure Medical Equipment Co., Ltd.
Subject Device: YJ-K1K2 Wheelchair(YJ-013F)
Name of Device: YJ-K1/K2 Wheelchair
Model: YJ-K1F 16" NDE; YJ-K1F 16" NDS; YJ-K1F 18" NDE; YJ-K1F 18" NDS; YJ-K1F 20"
NDE;YJ-K1F 20" NDS;YJ-K1F 18" NDEE;YJK1F 16" VDE;YJ-K1F 16" VDS;YJ-K1F 18"
VDE;YJ-K1F 18" VDS;YJ-K1F 20" VDE;YJ-K1F 20" VDS;YJ-K2 16" VDFE;YJ-K2 16" VDFS;YJ-K2
18" VDFE;YJ-K2 18" VDFS;YJ-K2 20" VDFE;YJ-K2 20" VDFS;YJ-K1G 18" NFFE;YJ-K1G 18"
NFFS;YJ-K2TC 16" NAFE;YJ-K2TC 18" NAFE;YJK2TC 20" NAFE.
Regulation Name: Mechanical Wheelchair
Product Code: IOR
Regulation Number: 890.3850
Regulation Class: 1
510(k) number:K232230
## 5. General Description
The YJ-K1K2 Wheelchair(YJ-013F) are mechanical wheelchairs which are a manually operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position. It can be folded for transport by bring the two sides together.
The manual wheelchair incorporates a main frame, a seat, two adjustable footrests and four wheels.
The larger rear wheels have hand rims of slightly smaller diameter projecting just beyond the tire.
These allows the user to manoeuvre the chair by pushing them on without requiring them to grasp the tires. The manual wheelchairs have brakes that bear on the tires of the rear wheels and two push handles at the upper rear of the frame to allow for manual propulsion by an assistant.
### 5.1 Model(s)
YJ-013F
### 5.2 Intended Use/Indications For Use
The YJ-K1K2 Wheelchair(YJ-013F) is to provide mobility to persons limited to a sitting position.
### 5.3 Product Composition
Main frame, back upholstery, seat upholstery, handgrip, armrest, front wheel, rear wheel, hand rim, crossbar, footplates, brake, leg rest, Anti-tipper (selected models), brake extender (selected models).
### 5.4 Human body Contacting Materials
2
{7}
Sponsor: Zhenjiang Assure Medical Equipment Co., Ltd.
Subject Device: YJ-K1K2 Wheelchair(YJ-013F)
| Patient Contacting Component | Material Name |
| --- | --- |
| Frame | Steel |
| Seat+Back upholstery | Nylon |
| Armrest padded material | PU |
| Handrim | Polypropylene (PP)/Steel |
| Brake | Aluminum |
| Front Fork | Aluminum |
| Footplate | Plastic/Aluminum |
| Tire | PU |
5.5 Specification
| | YJ-013F |
| --- | --- |
| Overall length*width*height | 990mm*580mm*820mm |
| Folded dimension | 990mm*330mm*610mm |
| Seat width | 12" 305mm14" 360mm |
| Weight (Total) | 19.15kg |
| Weight Capacity | 150LBS |
| Seat depth | 340mm |
| Backrest angle | 5.5° (fixed, non-adjustable) |
| Backrest height | 450mm |
| Backrest width | 12" 305mm14" 360mm |
| Footrest to seat distance | 350-430mm |
| Armrest length | 370mm |
| Armrest width | 50mm |
| Hand rim diameter | 415mm |
| Static stability uphill | 10.00° |
| Static stability downhill | 10.00° |
| Static stability sideways | 10.20° |
| Diameter of front wheels | Φ6X1.25"PU |
| Diameter of rear wheels | Φ22" PU |
| Uphill holding angle | 7.25° |
| Downhill holding angle | 7.20° |
## 6. Summary of Comparison and Technological Characteristics
Table 1-General Comparison
3
{8}
Sponsor: Zhenjiang Assure Medical Equipment Co., Ltd.
Subject Device: YJ-K1K2 Wheelchair(YJ-013F)
| Elements of Comparison | Predicate Device | Subject Device | Comment |
| --- | --- | --- | --- |
| K number | K232230 | N/A | N/A |
| Manufacturer | Zhenjiang Assure Medical device Co., Ltd | Zhenjiang Assure Medical device Co., Ltd | Same |
| Proprietary Name | Reclining wheelchair | Reclining wheelchair | Same |
| Classification | 1 | 1 | Same |
| Regulation Name | Mechanical Wheelchair | Mechanical Wheelchair | Same |
| Regulation Number | 890.3850 | 890.3850 | Same |
| Product Code | IOR | IOR | Same |
| Indications For Use | The YJ-K1/K2 Manual Wheelchair is to provide mobility to persons limited to a sitting position. | The YJ-K1K2 Wheelchair(YJ-013F) is to provide mobility to persons limited to a sitting position | Same |
Table 2 Performance And Safety Comparison
| Elements of Comparison | Predicate Device | Subject Device | Comment |
| --- | --- | --- | --- |
| Model | YJ-K1 | YJ-013F | N/A |
| Design Characteristic | Manual Operation, Four-Wheels, Foldable, Cross-Brace, Pull-to Lock, Armrest, Backrest, Foot rest, skirt guard, Anti-tipper (selected models), brake extender(selected models) | Manual Operation, Four-Wheels, Foldable, Cross-Brace, Pull-to Lock, Armrest, Backrest, Foot rest, skirt guard, Anti-tipper (selected models), brake extender(selected models) | Same |
| Brake control | occupant-operated brake only | occupant-operated brake only | Same |
4
{9}
Sponsor: Zhenjiang Assure Medical Equipment Co., Ltd.
Subject Device: YJ-K1K2 Wheelchair(YJ-013F)
| Operation Environment | For indoor/outdoor use | For indoor/outdoor use | Same |
| --- | --- | --- | --- |
| Control Mode | Mechanical | Mechanical | Same |
| Size(unfold) | 1220 (L) *660/640/700 (W) *857/908mm (H) | 990mm*580mm*820mm | Note 1 |
| Stowage length/width/height | 1220(L)*280/285/310(W)*857/908mm(H) | 990mm*330mm*610mm | Note 1 |
| Weight (Total) | 25.5kg | 19.15kg | Note 1 |
| Weight Capacity | 300LBS | 150LBS | Note 1 |
| Seat width | 410-510mm | 305-330mm | Note 1 |
| Backrest width | 410-510mm | 305-330mm | Note 1 |
| Seat depth | 460mm | 340mm | Note 1 |
| Front wheel | Φ8X1"PVC | Φ6X1.25"PU | Note 1 |
| Rear wheels | Φ24" PU | Φ22" PU | Note 1 |
| Armrest | Fixed/flip up | Fixed/flip up | Same |
| Foot rest | swing away footrest/ elevating leg rest | swing away footrest/ elevating leg rest | Same |
| Rear Axle Position | Single | Single | Same |
| Safety Feature | Manual Wheel Lock | Manual Wheel Lock | Same |
| Braking time | The manual wheelchair brake is mechanical structure and manual braking, and the braking time can be ignored when it is used immediately. | The manual wheelchair brake is mechanical structure and manual braking, and the braking time can be ignored when it is used immediately. | Same |
| Braking distance | brake is mechanical structure and manual braking, and the braking distance can be ignored. | brake is mechanical structure and manual braking, and the braking distance can be ignored. | Same |
| Frame material | Steel | Steel | Same |
5
{10}
Sponsor: Zhenjiang Assure Medical Equipment Co., Ltd.
Subject Device: YJ-K1K2 Wheelchair(YJ-013F)
| Seat+Back upholstery | PVC | Nylon | Note 2 |
| --- | --- | --- | --- |
| Handrim | Plastic/Steel | Polypropylene (PP)/Steel | Note 2 |
| Armrest padded material | PVC | PU | Note 2 |
| Brake | Steel | Aluminum | Note 2 |
| Front Fork | Steel | Aluminum | Note 2 |
| Footplate | Plastic/Aluminum | Plastic/Aluminum | Same |
| Performance | Complywith:ISO7176-1ISO7176-3ISO7176-5ISO7176-7ISO7176-8ISO7176-11ISO7176-13ISO7176-15ISO7176-16 | Complywith:ISO7176-1ISO7176-3ISO7176-5ISO7176-7ISO7176-8ISO7176-11ISO7176-13ISO7176-15ISO7176-22ISO16840-10 | Similar 1 |
| Biocompatibility | Comply with:ISO10993-1ISO10993-5ISO10993-10 | Comply with:ISO10993-1ISO10993-5ISO10993-10 | Same |
#### Note 1
Compared to the predicate device, the subject device has different values on the unfold size, stowage size, device weight, capacity, seat width, backrest width and tire size. However, the subject device has passed ISO 7176-5:2021 (Dimensions, mass and maneuvering space) and ISO 7176-7:1998 (Measurement of seating and wheel dimensions), so the above differences do not raise any new risk of safety or effectiveness.
#### Note 2
6
{11}
Sponsor: Zhenjiang Assure Medical Equipment Co., Ltd.
Subject Device: YJ-K1K2 Wheelchair(YJ-013F)
Compared to the predicate device, the subject device has implemented material changes including Seat+back upholstery (Nylon), Armrest padded material (Polyurethane (PU)), Brake material change (Aluminum), Front fork material change (Aluminum), and handrim material (Polypropylene (PP)). However, these differences does not impact safety and effectiveness of the subject device.
The Nylon and Aluminum materials have been previously incorporated in FDA-cleared mechanical wheelchairs approved under the same regulation (21 CFR 890.3850) and product code IOR, with similar intended use and technology (e.g., K252370 and K112948). The contact classification of these changed components remains intact skin contact without any change in the nature, duration and degree of human body contact.
The Armrest padded material (Polyurethane (PU)) and the handrim material (Polypropylene (PP)) of the subject device are listed in Section B of Attachment G of FDA's guidance on the "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", and these materials has a documented history of safe use in legally marketed medical devices in contact with intact skin (e.g., through Medical Device Reporting (MDR) analysis, literature search). This does not impact safety and effectiveness.
# Similar 1
ISO 7176-16:2012 was withdrawn and replaced by ISO 16840-10:2021, so ISO 7176-16:2012 is therefore no longer valid, nor recognized by FDA. Therefore, to establish adequate fire resistance (and thus, overall safety) of your device in support of substantial equivalence, we have provided additional information and/or rationale to demonstrate adequate resistance to ignition of the postural support components of your device, in accordance with ISO 16840-10:2021 or alternative scientifically valid means.
# Similar 2
The subject device and the predicate device utilize the same patient-contacting materials and contact classification (intact skin). Both devices comply with ISO 10993-1, ISO 10993-5, and ISO 10993-10 for biological evaluation.
7
{12}
Sponsor: Zhenjiang Assure Medical Equipment Co., Ltd.
Subject Device: YJ-K1K2 Wheelchair(YJ-013F)
## 7. Summary of Non-Clinic Performance Testing
### 7.1 Bench Performance
Non-clinical tests were conducted to verify that the subject device met all design specifications and was substantially equivalent to the predicate device.
The test results demonstrated that the proposed device complies with the following standards:
16-195 ISO 7176-1 Third edition 2014-10-01
Wheelchairs-Part 1: Determination of static stability
16-192 ISO 7176-3 Third edition 2012-12-15
Wheelchairs-Part 3: Determination of effectiveness of brakes
16-163 ISO 7176-5 Second edition 2008-06-01
Wheelchairs-Part 5: Determination of overall dimensions, mass and manoeuvring space
16-196 7176-7 First Edition 1998-05-15
Wheelchairs-Part 7: Measurement of seating and wheel dimensions
16-197 7176-8 Second editon 2014-12-15
Wheelchairs-Part 8: Requirements and test methods for static, impact and fatigue strengths
16-190 ISO 7176-11 Second edition 2012-12-01
Wheelchairs-Part 11: Test dummies
16-25 ISO 7176-13 First edition 1989-08-01
Wheelchairs-Part 13: Determination of coefficient of friction of test surfaces
16-27 7176-15 First edition 1996-11-15
Wheelchairs-Part 15: Requirements for information disclosure, documentation and labeling
16-198 ISO 7176-22 Second edition 2014-09-01
Wheelchairs-Part 22: Set-up procedures
16-233 ISO 16840-10 Second edition 2021-06 Corrected version 2022-01
Resistance to Ignition of Postural Support Devices
### 7.2 Biocompatibility
8
{13}
Sponsor: Zhenjiang Assure Medical Equipment Co., Ltd.
Subject Device: YJ-K1K2 Wheelchair(YJ-013F)
No new biocompatibility testing was conducted. The subject device is submitted via the FDA Special 510(k) pathway, which aims to expedite review when existing cleared test data can support substantial equivalence without additional testing.
All material changes in the subject device (nylon for seat+back upholstery, polyurethane (PU) for armrest padded material, polypropylene (PP) for the handrim and aluminum for the brake and front fork) have either been previously used in FDA-cleared mechanical wheelchairs (e.g., K252370 and K112948).) under the same regulation and product code, with identical intended use and intact skin contact classification, or are listed (i.e. polypropylene and polyurethane) in section B of Attachment G of FDA's guidance on the "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"". No new biocompatibility risks are introduced by these material changes, as the contact conditions (contact site, duration, frequency) remain consistent with the previously cleared devices.
## 8. Clinical Test Conclusion
No clinical study is included in this submission.
## 9. Summary Of Verification And Validation
The manufacturer has performed non-clinical performance testing based on its risk assessment utilizing Failure Mode Effect Analysis (FMEA).
Following Quality System processes, required testing was conducted to validate the cumulative modifications made to the subject devices.
Based on the risk analysis, there is an identification of the verification and/or validation activities required to comply with 21 CFR 820.30.
The manufacturer use outsourced performance testing as verification and/or validation activities.
## 10. Conclusion
The subject device and the predicate device have the same intended use and technological characteristics. All differences have not raised any new safety or effectiveness issues, and the device is substantially equivalent to the cleared predicate device (K232230) in terms of core performance, safety standards, and intended use, meeting the core requirement of FDA Special 510(k) that changes do not affect device safety and effectiveness.
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.