K260104 · Orthosoft Inc. (d/b/a) Zimmer CAS · QHE · Feb 4, 2026 · Orthopedic
Device Facts
Record ID
K260104
Device Name
Signature™ ONE System
Applicant
Orthosoft Inc. (d/b/a) Zimmer CAS
Product Code
QHE · Orthopedic
Decision Date
Feb 4, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Automatic segmentation of scapula and humerus
—
—
—
Production grade scapula and humerus CT segmentations
—
—
—
Indications for Use
The Signature™ ONE System is indicated, based on patient-specific radiological images with identifiable placement anatomical landmarks, to assist in pre-operative planning and/or intra-operative guiding of surgical instruments for shoulder replacement surgical procedures on patients not precluded from being radiologically scanned. The Signature™ ONE System is designed for use on a skeletally mature patient population. The targeted population has the same characteristics as the population that is suitable for the implants compatible with the Signature™ ONE System. The Signature™ ONE System is to be used with the glenoid components of the following shoulder implant systems in accordance with their indications and contraindications: Zimmer® Trabecular Metal Reverse Plus® Shoulder, Comprehensive® Total Shoulder System, Comprehensive® Reverse Shoulder System, Comprehensive® Reverse Augmented Baseplates and Alliance® Glenoid System. The Signature™ ONE System pre-operative planning is also compatible with the humeral components of the following shoulder implant systems in accordance with their indications and contraindications: Comprehensive® Total Shoulder System, Comprehensive® Reverse Shoulder System, and Identity™ Shoulder System. The Signature™ ONE System Guides and bone models are intended for single use only.
Device Story
System assists pre-operative planning and intra-operative guidance for shoulder arthroplasty. Inputs: patient-specific CT images. Process: internal software uses AI/ML-based automatic segmentation and manual landmarking to create 3D bone models; surgeon reviews/approves plan via portal; additive manufacturing (SLS) produces polyamide (nylon) surgical guides and bone models. Output: patient-specific guides/models used by surgeons in OR to transfer pre-operative plan to anatomy. Benefits: improved surgical accuracy and planning. Workflow: CT images processed by Zimmer Biomet operators; surgeon interacts via Zimmer Biomet Portal; guides/models shipped to hospital for sterilization and use.
Clinical Evidence
Bench testing only. Performance data included verification and validation testing, engineering analysis, and usability engineering. Software V&V conducted per IEC 62304 and FDA guidance for enhanced documentation. No clinical data presented.
Technological Characteristics
Materials: Polyamide (nylon). Manufacturing: Additive manufacturing (Selective Laser Sintering - SLS). Software: Internal CT reconstruction/segmentation (AI/ML-based) and planning applications. Connectivity: Zimmer Biomet Portal for external user interaction. Sterilization: Non-sterile, sterilized at hospital. Shelf life: 6 months.
Indications for Use
Indicated for skeletally mature patients undergoing shoulder replacement surgery (anatomic or reverse) who are eligible for radiological scanning. Used with specific Zimmer shoulder implant systems.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{1}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
February 4, 2026
Orthosoft Inc. (d/b/a) Zimmer CAS
Nilam Dave
Regulatory Affairs Specialist
75 Queen St. Suite 3300
Montreal, QC H3C 2N6
Canada
Re: K260104
Trade/Device Name: Signature™ ONE System
Regulation Number: 21 CFR 888.3660
Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis
Regulatory Class: Class II
Product Code: QHE, KWT, KWS, PHX, MBF, HSD
Dated: January 6, 2026
Received: January 13, 2026
Dear Nilam Dave:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}
K260104 - Nilam Dave
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K260104 - Nilam Dave
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Farzana Sharmin -S
Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K240106 | ? |
| Please provide the device trade name(s). | | ? |
| Signature™ ONE System | | |
| Please provide your Indications for Use below. | | ? |
| The Signature™ ONE System is indicated, based on patient-specific radiological images with identifiable placement anatomical landmarks, to assist in pre-operative planning and/or intra-operative guiding of surgical instruments for shoulder replacement surgical procedures on patients not precluded from being radiologically scanned.
The Signature™ ONE System is designed for use on a skeletally mature patient population. The targeted population has the same characteristics as the population that is suitable for the implants compatible with the Signature™ ONE System.
The Signature™ ONE System is to be used with the glenoid components of the following shoulder implant systems in accordance with their indications and contraindications: Zimmer® Trabecular Metal Reverse Plus® Shoulder, Comprehensive® Total Shoulder System, Comprehensive® Reverse Shoulder System, Comprehensive® Reverse Augmented Baseplates and Alliance® Glenoid System.
The Signature™ ONE System pre-operative planning is also compatible with the humeral components of the following shoulder implant systems in accordance with their indications and contraindications: Comprehensive® Total Shoulder System, Comprehensive® Reverse Shoulder System, and Identity™ Shoulder System.
The Signature™ ONE System Guides and bone models are intended for single use only. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{5}
K260104 - Page 1 of 5
# 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Signature™ ONE System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance documents, 'The Special 510k Program Guidance for Industry and Food and Drug Administration Staff', issued on September 13, 2019, 'Deciding When to Submit a 510(k) for a Change to an Existing Device Guidance for Industry and Food and Drug Administration Staff', issued on October 25, 2017, and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device Guidance for Industry and Food and Drug Administration Staff', issued on October 25, 2017.
**Sponsor:**
Orthosoft Inc (d/b/a. Zimmer CAS)
75 Queen St., Suite 3300
Montreal, QC, CANADA H3C 2N6
Establishment Registration Number: 9617840
**Contact Person:**
Nilam Dave
Regulatory Affairs Specialist
Telephone: +1 (206) 638-8947
**Date:**
06 Jan 2026
**Subject Device:**
Trade Name: Signature™ ONE System
Common Name: Signature™ ONE System, ONE Planner Shoulder
**Classification Name:**
- QHE– Shoulder joint metal/polymer semi-constrained cemented prosthesis (21 CFR 888.3660)
**Additional Product Codes:**
- KWT- Shoulder joint metal/polymer non-constrained cemented prosthesis (21 CFR 888.3650)
- KWS- Shoulder joint metal/polymer semi-constrained cemented prosthesis (21 CFR 888.3660)
- PHX- Shoulder joint metal/polymer semi-constrained cemented prosthesis (21 CFR 888.3660)
- MBF- Shoulder joint metal/polymer/metal nonconstrained or semi-constrained porous-coated uncemented prosthesis (21 CFR 888.3670)
- HSD- Shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis (21 CFR 888.3690)
{6}
K260104 - Page 2 of 5
## Predicate Device:
| 510(k) Number | Device Name | Manufacturer |
| --- | --- | --- |
| K232425 | Signature™
ONE System | Zimmer CAS |
## Purpose and Device Description:
The Signature™ ONE System is developed to assist in pre-operative planning (using the Signature™ ONE Planner) and to accurately transfer a pre-operative plan to orthopedic surgical procedures (using the Signature™ ONE Guides and bone model) if desired in skeletally mature individuals for Total Shoulder Arthroplasty. Both anatomic and reverse (TSA and RSA respectively) approaches are supported.
The Signature ONE Guides and Bone Models are designed and manufactured of polyamide (nylon) using additive manufacturing selective laser sintering (SLS), based on the approved/finalized pre-surgical plan and shipped prior to surgery. The guides and bone models are provided non-sterile and sterilized at the hospital. They are used intraoperatively to assist the surgeon in reproducing the plan on the scapula. The Signature ONE System surgical technique remains close to the conventional shoulder arthroplasty workflow.
The Signature™ ONE System uses a Non-Device Medical Device Data System (MDDS) called the Zimmer Biomet Portal for the interaction with external users (i.e. imaging technician and the surgeon). The internal users (i.e. the Zimmer Biomet operators) use manufacturing software applications to prepare the patient cases for the surgeon.
The purpose of the submission is that the CT Reconstruction internal software application was updated with an improved AI/ML locked model for automatic segmentation and also retraining the model with new production grade scapula and humerus CT segmentations. The AI/ML model is used in the Segmentation step only within the CT Reconstruction internal software application prior to manual segmentation performed by the Zimmer Biomet operator. The use of AI has not changed since predicate K232425. In addition, the Landmarking step within the internal software application was updated with option to select additional configurable landmarks in the Planning application (used to determine the reference coordinate systems to provide native bone information).
{7}
K260104 - Page 3 of 5
The overall Intended Use, Indications for Use, manufacturing process, materials, sterilization methods, principle of operation, and software workflow have not changed from the predicate K232425.
## Indications for Use:
The Signature™ ONE System is indicated, based on patient-specific radiological images with identifiable placement anatomical landmarks, to assist in pre-operative planning and/or intra-operative guiding of surgical instruments for shoulder replacement surgical procedures on patients not precluded from being radiologically scanned.
The Signature™ ONE System is designed for use on a skeletally mature patient population. The targeted population has the same characteristics as the population that is suitable for the implants compatible with the Signature™ ONE System.
The Signature™ ONE System is to be used with the glenoid components of the following shoulder implant systems in accordance with their indications and contraindications: Zimmer® Trabecular Metal Reverse Plus® Shoulder, Comprehensive® Total Shoulder System, Comprehensive® Reverse Shoulder System, Comprehensive® Reverse Augmented Baseplates and Alliance® Glenoid System.
The Signature™ ONE System pre-operative planning is also compatible with the humeral components of the following shoulder implant systems in accordance with their indications and contraindications: Comprehensive® Total Shoulder System, Comprehensive® Reverse Shoulder System, and Identity™ Shoulder System.
The Signature™ ONE System Guides and bone models are intended for single use only.
## Summary of Technological Characteristics:
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use/ Indications for Use: Same as predicate device
- Principle of Operation: The subject and predicate devices both utilize pre-operative images, and
{8}
K260104 - Page 4 of 5
intraoperative guidance of instruments. The Principle of Operation remains unchanged since device clearance in K232425.
- Guides/Bone Models: The subject and predicate device utilize 3D printing (SLS) to manufacture the guides. The guides are non-sterile single use and have a shelf life of 6 months. The existing guides and bone models remain unchanged since predicate clearance K232425.
- Technological Characteristics: The subject and predicate device have similar technological characteristics (e.g. image modality, anatomical site, hardware components, and input file type) from the predicate device cleared in K232425.
- Software Workflow: The subject and predicate device utilize internal manufacturing software applications and an external software planning application that the surgeon interacts with to review, modify and approve the plan. The overall fundamental software workflow is identical between the subject and predicate device in that CT DICOM images are triaged, segmentation of the image occurs, landmarks are placed on the 3D reconstructed bone model, planning/approval by the surgeon and creation of the guide and bone model. The workflow remains unchanged from the predicate device K232425.
- Software Output: The AI/ML model used in the Segmentation step within the CT Reconstruction internal software application was updated by retraining the model with production grade scapula and humerus CT segmentations where as in the predicate the AI/ML model was trained with production grade scapula scans and the humerus scans were exploratory quality used to develop the model. The Landmarking step within the internal software application was updated with option to select additional configurable landmarks, and like the predicate uses existing landmarks to determine the reference coordinate systems to provide native bone information.
## Summary of Performance Data (Nonclinical and/or Clinical):
The following performance data was provided in support of the substantial equivalence determination:
## Device Performance Testing:
{9}
K260104 - Page 5 of 5
Verification and Validation Testing for Signature™ ONE System was conducted with the following aspects:
- Performance Tests: Performance tests documented to ensure the performance of the implemented features and verify related design inputs.
- Engineering Analysis: Analysis documented to ensure the performance of the implemented features and verify related design inputs.
- Usability Engineering: Performance of the system in regards to human factors engineering. This remains unchanged and applicable from the predicate K232425.
- Validation: Validation performed to validate related user needs, intended use and safety and effectiveness. This remains unchanged and applicable from the predicate K232425.
## Software Verification and Validation Testing
Software tests were conducted to satisfy the Enhanced Level of Documentation per requirements of Content of Premarket Submissions for Device Software Functions Guidance for Industry and Food and Drug Administration Staff issued on June 14, 2023 and IEC 62304 (Medical Device Software- Software Life Cycle Processes). The testing demonstrates that the Signature™ ONE System does not raise any different questions of safety and effectiveness as compared to the predicate devices.
## Substantial Equivalence Conclusion
The proposed and predicate device have the same Intended Use and Indications for Use, principle of operation, Guides/Bone Models, similar technological characteristics and same software workflow. The new AI/ML model for automatic segmentation is used in the same way as the predicate device. The additional configurable landmarks in the Landmarking step does not change the existing landmarking application and the overall software workflow remains the same as the predicate device. In summary, any differences between the devices does not raise different questions of safety and effectiveness and the proposed device is at least as safe and effective as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.