K260077 · OrthoGrid Systems, Inc. · QIH · May 29, 2026 · Radiology
Device Facts
Record ID
K260077
Device Name
OrthoGrid Hip AI® 4.0
Applicant
OrthoGrid Systems, Inc.
Product Code
QIH · Radiology
Decision Date
May 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K260077 · May 29, 2026
OrthoGrid Hip AI® 4.0
OrthoGrid Systems, Inc.
De-identified fluoroscopic images from five clinical sites; Retrospective THA surgical case data
Real-world fluoroscopic images were used to train and test the AI models for landmark detection and anatomy/implant segmentation. The data reflects actual intraoperative imaging and surgical conditions.
AI model training; Retrospective clinical images; Fluoroscopic data; THA procedures
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
AI Models Development Dataset; Retrospective analysis of clinical fluoroscopic images
Patients undergoing Total Hip Arthroplasty (THA); Sample Size: 502 patients (total across models); Number of Sites: 5
Not applicable for this study
Landmark detection, implant segmentation, and femur alignment accuracy
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Anatomic landmark detection (Teardrops)
Transformer-based models
Mean error below 3.70 mm (Right) and 3.22 mm (Left)
0.95 ± 0.73 mm (Right); 0.64 ± 0.49 mm (Left)
Training: 84% of 301 patients; Tuning: 6% of 301 patients; Tuning Evaluation: 10% of 301 patients
—
84 THA surgical cases, 342 images
>1 (orthopedic surgeons)
Anatomic landmark detection (Lesser Trochanters)
Transformer-based models
Mean error below 4.84 mm (Right) and 3.77 mm (Left)
2.02 ± 1.35 mm (Right); 1.54 ± 1.16 mm (Left)
Training: 84% of 301 patients; Tuning: 6% of 301 patients; Tuning Evaluation: 10% of 301 patients
—
84 THA surgical cases, 342 images
>1 (orthopedic surgeons)
Cup positioning (Version and Inclination)
CNN and Transformer-based models
Mean error below 3.48° (Version) and 2.42° (Inclination)
Training: 82% of 389 patients; Tuning: 6% of 389 patients; Tuning Evaluation: 12% of 389 patients
—
84 THA surgical cases, 342 images
>1 (orthopedic surgeons)
Femur alignment
CNN and Transformer-based models
Mean error below 2.85°
1.27 ± 0.88°
Training: 76% of 502 patients; Tuning: 12% of 502 patients; Tuning Evaluation: 12% of 502 patients
—
51 THA surgical cases, 182 images
>1 (orthopedic surgeons)
Indications for Use
OrthoGrid Hip AI® 4.0 is an image-processing software indicated to assist in the positioning of Total Hip Replacement components. The software uses Artificial Intelligence (AI) to allow real-time initialization of grid placement relative to the bone structures of interest in C-arm fluoroscopic images to assist in precisely positioning Total Hip Replacement components intraoperatively by measuring their positions. Artificial Intelligence will only offer automated initialization of grid placement if the points of interest can be identified from radiology images. Clinical judgement and experience are required to properly use the device. The device is not for primary image interpretation. The software is not for use on mobile phones.
Device Story
SaMD using non-generative AI to analyze intraoperative 2D fluoroscopic images (C-arm) for THA procedures; provides digital annotations/visual guidance for pelvic plane tracking, cup positioning, leg length, and offset. Input: 2D fluoroscopic images; Processing: CNN and Transformer-based models identify anatomic landmarks (Teardrops, Lesser Trochanters) and implants; Output: automated initialization of measurement tools and visual indicators. Used in OR by surgeons; displayed outside sterile field. Physician retains control to review, edit, or override all AI-identified landmarks and measurements. Benefits: assists precise component positioning via automated landmark detection, reducing manual setup time.
Clinical Evidence
Bench testing and clinical validation performed. AI models validated using 84 THA surgical cases (342 images) for landmarks/implants and 51 cases (182 images) for anatomy. Performance compared against ground truth from three orthopedic surgeons. Mean errors for landmarks (0.64–2.02 mm) and angles (0.44–1.38°) fell within established clinical variability thresholds. Sensitivity 100% under study conditions. No clinical data on human patients provided; validation used cadaveric specimens and retrospective de-identified clinical images.
Technological Characteristics
Software-only device (SaMD) running on commercial off-the-shelf (COTS) hardware. Uses deep learning (CNN and Transformer architectures). Compatible with C-arm image intensifiers and flat-panel detectors. Includes optional radiopaque calibration arrays (9 and 12-inch) for distortion correction. Cybersecurity controls per FDA guidance. Locked/frozen ML models; no adaptive learning.
Indications for Use
Indicated for patients undergoing direct anterior total hip replacement (THA) procedures. Contraindicated when clear fluoroscopic images cannot be obtained or key landmarks (e.g., Lesser Trochanters, Teardrops) are not identifiable due to patient anatomy, surgical technique (e.g., excessive reaming), or OR equipment limitations.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 29, 2026
OrthoGrid Systems, Inc.
Shweta Conner
Regulatory Affairs, Sr. Specialist
7026 S. Commerce Park Dr.
Suite 200
Midvale, Utah 84047
Re: K260077
Trade/Device Name: OrthoGrid Hip AI® 4.0
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: April 29, 2026
Received: April 29, 2026
Dear Shweta Conner:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260077 - Shweta Conner
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K260077 - Shweta Conner
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260077 | ? |
| Please provide the device trade name(s). | | ? |
| OrthoGrid Hip AI® 4.0 | | |
| Please provide your Indications for Use below. | | ? |
| OrthoGrid Hip AI® 4.0 is an image-processing software indicated to assist in the positioning of Total Hip Replacement components. The software uses Artificial Intelligence (AI) to allow real-time initialization of grid placement relative to the bone structures of interest in C-arm fluoroscopic images to assist in precisely positioning Total Hip Replacement components intraoperatively by measuring their positions. Artificial Intelligence will only offer automated initialization of grid placement if the points of interest can be identified from radiology images. Clinical judgement and experience are required to properly use the device. The device is not for primary image interpretation. The software is not for use on mobile phones. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K260077
# 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the OrthoGrid Hip AI® 4.0 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance documents, 'The Special 510k Program Guidance for Industry and Food and Drug Administration Staff', issued on September 13, 2019, 'Deciding When to Submit a 510(k) for a Change to an Existing Device Guidance for Industry and Food and Drug Administration Staff', issued on October 25, 2017, and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device Guidance for Industry and Food and Drug Administration Staff', issued on October 25, 2017.
Sponsor: OrthoGrid Systems, Inc.
7026 South Commerce Park Drive
Suite 200
Midvale, UT, 84047
Establishment Registration Number: 9617840
Contact Person: Shweta Conner
Regulatory Affairs Sr, Specialist
Telephone: 901-569-3004
Date: May 29, 2026
Subject Device: Trade Name: OrthoGrid Hip AI® 4.0
Common Name: Automated Radiological Image Processing Software
Classification Name:
QIH– Medical Image Management And Processing System
(21 CFR §892.2050)
Predicate Device:
| Manufacturer | Device Name | 510(k) Number |
| --- | --- | --- |
| OrthoGrid Systems, Inc. | PhantomMSK Hip | K210136 |
# Device Description and Purpose:
The OrthoGrid Hip AI® 4.0 is a non-invasive Software as a Medical Device (SaMD) that uses non-generative Artificial Intelligence (AI) models to analyze fluoroscopic images to aid physicians with intra-operative implant and anatomic alignment and radiographic alignment for direct anterior total hip replacement procedures.
Alignment considerations during total hip replacement include pelvic tilt and obliquity, cup inclination and version, Leg Length and global offset. To assist users with these alignments, OrthoGrid Hip AI® 4.0 is an Artificial Intelligence-enabled product that provides digital annotations and visual guidance points on the fluoroscopic image using anatomic landmarks for functional pelvic plane tracking, cup positioning, leg length and offset. To operate OrthoGrid Hip AI® 4.0, a fluoroscopic image is acquired from a C-arm and displayed outside the sterile field, where the image analysis tools can be used at the user's discretion.
{5}
OrthoGrid Hip AI® 4.0 operates on image principles that are compatible with both C-arm Image Intensifiers and Flat Panel C-arms. OrthoGrid Hip AI® 4.0 offers radiopaque Calibration Arrays in 9 and 12-inch sizes as an accessory for correction of fluoroscopic distortion for C-arm Image Intensifiers. Fluoroscopic distortion is attributed to external electromagnetic interference and the mapping of the planar image on a curved input phosphor.
OrthoGrid Hip AI® 4.0 does not include any custom computer hardware, and the software can be run on a "commercial off-the-shelf" system (i.e. PC, keyboard, mouse, touchscreen monitor etc.) that meets minimum performance requirements. As part of the System, all required computer accessories are provided to the customer.
The basis of this submission is the addition of optional non-generative artificial intelligence (AI) element for detection in fluoroscopic image of anatomic landmarks and objects to provide automated selection and initialization of measuring tools and pelvic position visual indicator. Addition of anatomical cup inclination and version angles measurement option per clinical preference. Introduction of optional ability to save and/or retrieve case specific data via remote or manual transfer.
# Indications for Use:
OrthoGrid Hip AI® 4.0 is an image-processing software indicated to assist in the positioning of Total Hip Replacement components. The software uses Artificial Intelligence (AI) to allow real-time initialization of grid placement relative to the bone structures of interest in C-arm fluoroscopic images to assist in precisely positioning Total Hip Replacement components intraoperatively by measuring their positions. Artificial Intelligence will only offer automated initialization of grid placement if the points of interest can be identified from radiology images.
Clinical judgement and experience are required to properly use the device. The device is not for primary image interpretation. The software is not for use on mobile phones.
# Contraindications:
OrthoGrid Hip AI® 4.0 may not be suitable for use in cases where clear fluoroscopic images cannot be obtained, and the clinician is unable to identify key landmarks, such as the Lesser Trochanters or Teardrops:
- This may be due to patient anatomic abnormalities.
- This may be due to surgical technique, for example, excessive acetabular reaming resulting in the inability to visualize the radiographic Teardrop.
- This may be due to operating room equipment in use, such as the size of the operating table limiting the tilt of the C-arm, and therefore the detection of Teardrops.
Substantial Equivalence:
| Category | Subject Device OrthoGrid Hip AI 4.0 (K260077) | Predicate Device PhantomMSK Hip (K210136) | Relationship | Notes |
| --- | --- | --- | --- | --- |
| Device | Automated radiological image processing software | System, image processing, radiological | Similar | The device is updated as the subject device uses Artificial Intelligence to |
| | | technology | | increase in the accuracy of the image processing |
| Method | Automated radiology | System, radiology | Similar | The device is used to identify the radiology images |
| Method | Image processing | System, image processing, radiology | Similar | The device is used to identify the radiology images |
{6}
| Category | Subject Device | Predicate Device | Relationship | Notes |
| --- | --- | --- | --- | --- |
| | OrthoGrid Hip AI 4.0 (K260077) | PhantomMSK Hip (K210136) | | automatically identify the anatomical landmarks, whereas the predicate device requires the user to manually identify and place points of interest on fluoroscopic images. Differences between the devices do not raise new questions of safety and effectiveness and verification and validation activities demonstrate that the subject device is at least as safe and effective as the legally marketed predicate device. |
| Product Code | QIH | LLZ | Similar | The product code is updated to reflect the subject device’s use of Artificial Intelligence to automatically identify the anatomical landmarks, whereas the predicate device requires the user to manually identify and place points of interest on fluoroscopic images. Differences between the devices do not raise new questions of safety and effectiveness and verification and validation activities demonstrate that the subject device is at least as safe and effective as the legally marketed predicate device. |
| Regulation | 21 CFR 892.2050 | 21 CFR 892.2050 | Identical | |
| Regulation Description | Medical image management and processing system | Medical image management and processing system | Identical | |
| Device Class | II | II | Identical | |
| Classification Panel | Radiology | Radiology | Identical | |
{7}
| Category | Subject Device OrthoGrid Hip AI 4.0 (K260077) | Predicate Device PhantomMSK Hip (K210136) | Relationship | Notes |
| --- | --- | --- | --- | --- |
| Intended Use | Image-processing software intended to assist in the intra-operative positioning of total hip replacement components during direct anterior total hip arthroplasty (THA) procedures. | Image-processing software intended to assist in the intra-operative positioning of total hip replacement components during direct anterior total hip arthroplasty (THA) procedures. | Identical | |
| Indications for Use | OrthoGrid Hip AI® 4.0 is an image-processing software indicated to assist in the positioning of Total Hip Replacement components. The software uses Artificial Intelligence (AI) to allow real-time initialization of grid placement relative to the bone structures of interest in C-arm fluoroscopic images to assist in precisely positioning Total Hip Replacement components intraoperatively by measuring their positions. Artificial Intelligence will only offer automated initialization of grid placement if the points of interest can be identified from radiology images.Clinical judgement and experience are required to properly use the device. The device is not for primary image interpretation. The software is not for use on mobile phones. | PhantomMSK Hip is an image-processing software indicated to assist in the positioning of Total Hip Replacement and Hip Preservation components. It is intended to assist in precisely positioning Total Hip Replacement and Hip Preservation components intraoperatively by measuring their positions relative to the bone structures of interest provided that the points of interest can be identified from radiology images.Clinical judgement and experience are required to properly use the device. The device is not for primary image interpretation. The software is not for use on mobile phones. | Similar | The Indications for Use statement has been updated to reflect that the subject device utilizes AI to automatically identify the anatomical landmarks.Differences between the devices do not raise new questions of safety and effectiveness and verification and validation activities demonstrate that the subject device is at least as safe and effective as the legally marketed predicate device. |
| Principle of Operation | A fluoroscopic image is acquired from a C-arm by the software and displayed outside the sterile field, where the | A fluoroscopic image is acquired from a C-arm by the software and displayed outside the sterile field, where the | Identical | |
{8}
| Category | Subject Device OrthoGrid Hip AI 4.0 (K260077) | Predicate Device PhantomMSK Hip (K210136) | Relationship | Notes |
| --- | --- | --- | --- | --- |
| | image analysis tools can be used at the surgeon’s discretion. | image analysis tools can be used at the surgeon’s discretion. | | |
| Image Modality | Uses intraoperative fluoroscopic images acquired from a C-arm. | Uses intraoperative fluoroscopic images acquired from a C-arm. | Identical | |
## Summary of Nonclinical Performance Data:
Verification and Validation Testing for OrthoGrid Hip AI® 4.0 was conducted with the following aspects:
- Performance Tests: Performance tests documented to ensure the performance of the implemented features and verify related design inputs.
- Usability Engineering: Performance of the system in regard to human factors engineering.
- Validation: Validation performed to validate related user needs, intended use and substantial equivalence.
- Software Testing: Software tests were conducted to satisfy requirements of the FDA Guidance for the Content Premarket Submissions for Software Contained in Medical Devices and IEC 62304 (Medical Device Software- Life Cycle Process). The software was considered a “Enhanced” level of concern. The testing demonstrates the subject device OrthoGrid Hip AI® 4.0 does not raise any new issues of safety and effectiveness as compared to the predicate device.
- Cybersecurity Testing: Cybersecurity tests were conducted and successfully met requirements per FDA Guidance “Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions”.
## Description of AI Functionality and its Integration in the System:
The OrthoGrid Hip AI® 4.0 system incorporates Machine Learning (ML) models that operate on 2D fluoroscopic images to support intraoperative AP Pelvis and AP Hip workflows. The AI models identify landmarks, anatomy, and Total Hip Arthroplasty (THA) implants, which are used to initialize software functions within the system.
These AI outputs are used to automatically initialize software tools that assist with:
- Leg length and offset measurements.
- Cup version and Cup inclination measurements using the trans-teardrop reference line.
- Scene identification to assist workflow progression.
- Anatomic orientation/alignment and positioning within an image.
- Femoral shaft alignment for the AP Hip workflow Overlay feature.
## User Control and Override:
The detected scene type can be manually overridden by the user at any time. All AI-identified landmarks that impact system measurements are visible and fully editable, allowing the user to
{9}
manually adjust landmark positions in real time at any stage of the procedure. The system automatically updates measurements and calculations immediately based on the user's input.
The OrthoGrid Hip AI® 4.0 outputs are intended to assist, not replace, professional clinical judgment. The operating physician is responsible for reviewing, verifying, and as needed, adjusting all AI-identified landmarks and AI-driven measurements, and retains full responsibility for final clinical decisions.
# ML Models Input/Output:
The ML models' input is a 2D fluoroscopic intra-operative anteroposterior (AP) or near AP image showing the pelvis and/or proximal femur, captured from compatible C-Arm systems.
The models generate either segmentation of relevant anatomy (femur) or implant (cup, stem) or landmark positions (e.g., Teardrops, Lesser Trochanters).
ML model outputs are transformed into user-controllable elements, such as landmark positions (e.g., AP landmarks and cup landmarks), femur axis alignment, and inputs used for initial scene selection. These elements are presented to the user at the appropriate workflow stage. Users can interact with these elements, which can be reviewed and manually adjusted as needed to ensure clinical accuracy.
# Model Architecture:
The OrthoGrid Hip AI® 4.0 ML models use deep learning architectures, including Convolutional Neural Networks (CNNs) and Transformer-based models. CNN- and Transformer-based models are used for anatomy and implant segmentation, while Transformer-based models are used for landmark detection. All models are optimized for analysis of 2D fluoroscopic images to support landmark identification and anatomy and implant segmentation.
# Model Development Summary:
The OrthoGrid Hip AI® 4.0 system incorporates Machine Learning (ML) models trained, and tested using real-world, de-identified fluoroscopic images collected from five geographically distinct clinical sites within the United States. The data reflect THA procedures and capture the complexity of intraoperative imaging and realistic surgical conditions consistent with the device's intended use.
The ML models were developed using real THA fluoroscopic images. No synthetic images were used. Table 1 summarizes the datasets used for each OrthoGrid Hip AI® 4.0 models, including the number of patients, number of images, and the approximate percentage of data used for training, tuning, and tuning evaluation (held out for technical performance testing) to ensure objective evaluation and reliable model performance.
Table 1: AI Models Development Dataset. Tuning Evaluation sets were held out from training.
| ML Model | # Patients | # Images | Training, Tuning, Tuning Evaluation (% of Patients) |
| --- | --- | --- | --- |
| Landmark | 301 | 15990 | 84% – 6% – 10% |
| Implant | 389 | 16476 | 82% – 6% – 12% |
| Femur | 502 | 23484 | 76% – 12% – 12% |
{10}
In addition to technical performance testing on tuning evaluation datasets, the AI models were clinically validated using cadaveric specimens with known demographic characteristics. These images were captured under controlled simulated THA surgical conditions to reflect the intended clinical environment.
# ML Model Performance on Tuning Evaluation Dataset:
The AI models were evaluated on the tuning evaluation dataset, and the results for the tested sample within the study conditions are summarized in Table 2 in terms of Accuracy, Sensitivity and Specificity.
Table 2 : Reference VS AI Predictions - ML Models: Accuracy, Sensitivity, and Specificity
| ML Model | Accuracy ± STD | Sensitivity | Specificity | Sample size N (Patients, Images) |
| --- | --- | --- | --- | --- |
| Left & Right Teardrop (Pixel) | 1.6 ± 2.1 | 100% | 99% | N= (29, 1542), study conditions |
| Left & Right Lesser Trochanter (Pixel) | 2.8 ± 4.5 | 98% | 96% | N= (29, 1542), study conditions |
| Implant | 95 ± 5% | 100% | 99% | N= (45, 1923), study conditions |
| Femur | 98 ± 3% | 100% | 100% | N= (60, 2393), study conditions |
# Notes:
- Accuracy: Indicates how closely AI predictions match the reference. Evaluated using distance differences for landmarks or the spatial overlap for implant and anatomy models.
- Sensitivity (Recall): Reflects the ability of the AI to correctly identify true landmarks or anatomy or implant.
- Specificity: Reflects the ability of the AI to correctly ignore areas where landmarks or anatomy or implant is absent.
- Sample size $N$ : Values are presented as (number of patients, number of images)
# ML Model Performance on Clinical Validation Dataset:
# ML Models Outputs
The performance of the three integrated ML models' outputs was evaluated against clinical Ground Truth (GT) generated by three qualified orthopedic surgeons, whose inter-surgeon error variability was quantified to establish the validation threshold. This threshold is based on typical differences observed between surgeons' measurements, defined as the value below which $97.5\%$ of these differences fall, representing the natural range of clinical measurement variability. AI models were considered validated when their mean error, calculated with respect to the GT, remained at or below this value. The tuning and evaluation datasets included: Landmark and Implant Models: 84 THA surgical cases, 342 images; Anatomy Model: 51 THA surgical cases, 182 images.
All measures met the validation criteria for the tested sample size and study conditions, demonstrating that the AI outputs are consistent, reliable, and fall within the variability range of
{11}
the observed errors among surgeons in clinical settings as reported in Table 3. The detection rate (i.e. Sensitivity) of all AI models was $100\%$ under study conditions.
Table 3: Manual vs AI Predictions - Global Average Results Compared to Surgeon Ground Truth for Evaluated Measures (Teardrops, Lesser Trochanters, Cup Positioning, and Femur Alignment Angles)
| Measure | Mean Error (Accuracy) ± STD | Validation Threshold | Agreement Rate AI vs Manual | Sample size N (cases, images) |
| --- | --- | --- | --- | --- |
| Teardrop (mm) | 0.95 ± 0.73 (Right) | 3.70 | 100% | N= (84, 253), study conditions |
| | 0.64 ± 0.49 (Left) | 3.22 | 100% | N= (84, 245), study conditions |
| Lesser Trochanter (mm) | 2.02 ± 1.35 (Right) | 4.84 | 100% | N= (84, 189), study conditions |
| | 1.54 ± 1.16 (Left) | 3.77 | 100% | N= (84, 199), study conditions |
| Cup Version (°) | 1.38 ± 0.95 | 3.48 | 100% | N= (84, 255), study conditions |
| Cup Inclination (°) | 0.44 ± 0.43 | 2.42 | 100% | N= (84, 255), study conditions |
| Femur Alignment Angle (°) | 1.27 ± 0.88 | 2.85 | 98.04% | N= (51, 182), study conditions |
Notes:
- Mean Error: Difference between AI predictions and manual ground truth, where the ground truth is defined as the average measurement from the three surgeons.
The AI models were clinically validated in a statistical study using data that may not fully represent all patient populations. Key demographic representation included:
Gender: $57\%$ female, $43\%$ male
Age: Concentrated in 60-79 (61%); limited representation under 50 (7%)
- C-Arm detector size: Most commonly 12-inch (40.5%), with smaller proportions of 9-inch (26.2%) and flat-panel detectors (33.3%)
The results showed that the AI models met the predefined performance criteria, with a statistical strength of $95\%$ confidence and $90\%$ reliability, and no systematic deviations were observed across the evaluated factors.
Performance results along with other factors were also collected and reported, including:
- BMI: Concentrated under 25 (61%), with limited representation over 35 (4.5%).
- Ethnicity: Predominantly White (90.5%), with limited representation of Asian (2.4%) and Hispanic (7.1%) participants.
- Bone disease: Most common condition is arthritis (29.8%), with smaller representation of osteoporosis (14.3%).
{12}
While some of these additional subgroups had limited sample sizes and could not support statistically significant conclusions, the data was carefully reviewed and ML model performance remained consistent, as no systematic bias or performance degradation was observed across these categories, supporting confidence in the overall AI performance results. However, performance in underrepresented populations has not been independently validated. Physicians should exercise due judgment when interpreting AI outputs for patients whose characteristics differ substantially from the validation dataset.
## AI Integrated with OrthoGrid Hip AI® 4.0:
A total of 102 paired system output measurements were compared between fully automatic AI outputs and manually adjusted measurements all using different workflows, AP Pelvis or AP Hip workflows (Ipsilateral or Contralateral).
Of these, 89.2% were concordant, while 10.8% showed some degree of discrepancy. The maximum observed differences were ≤ 3 mm for Leg Length and Offset measurements and ≤ 1° for Cup inclination and anteversion angles. This demonstrates a high degree of surgeon agreement with AI predictions, supporting confidence in model performance under studied conditions.
## ML Models Limitations:
AI performance characteristics were established under controlled validation study conditions and the intended use environment. Actual clinical performance may vary depending on factors such as patient anatomy, surgical technique, and intraoperative conditions. Atypical imaging conditions/positions, including landmarks near image boundaries or partially outside the field of view, unusual occlusions, or poor image quality (e.g., blurred, contrast) may result in inaccurate or failed landmarks detection.
## AI Performance Monitoring:
The ML Models are locked/frozen and do not engage in continuous or adaptive learning once deployed. The locked model design ensures consistent, validated performance characteristics throughout the product lifecycle.
OrthoGrid Hip AI® 4.0 ML performance is monitored after release. Updates to the ML models may be applied through a controlled software change, including a full software version update, when supported by new data or validated findings.
## Conclusion:
The conclusions drawn from the nonclinical testing demonstrate that the device is as safe, as effective, and performs as well or better than the legally marketed predicate device, PhantomMSK Hip (K210136).
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.