PhantomMSK Hip

K210136 · OrthoGrid Systems, Inc. · LLZ · Feb 18, 2021 · Radiology

Device Facts

Record IDK210136
Device NamePhantomMSK Hip
ApplicantOrthoGrid Systems, Inc.
Product CodeLLZ · Radiology
Decision DateFeb 18, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

PhantomMSK Hip is an image-processing software indicated to assist in the positioning of Total Hip Replacement and Hip Preservation components. It is intended to assist in precisely positioning Total Hip Preservation components intraoperatively by measuring their positions relative to the bone structures of interest provided that the points of interest can be identified from radiology images. Clinical judgement and experience are required to properly use the device. The device is not for primary image interpretation. The software is not for use on mobile phones.

Device Story

PhantomMSK Hip is non-invasive image-processing software for intraoperative fluoroscopic assessment during Total Hip Replacement and Hip Preservation. Input: fluoroscopic images from C-arm. Operation: software uses radiopaque calibration array attached to C-arm image intensifier to calculate and correct for fluoroscopic distortion caused by electromagnetic interference or phosphor curvature. Output: templating, measurement, and distortion-corrected images displayed outside sterile field. Used by surgeons in OR to assist in implant/anatomic alignment. Surgeon uses output to guide component positioning; clinical judgment required. Benefits: improved precision in component placement via distortion-corrected visualization.

Clinical Evidence

Bench testing only. Verification and validation testing performed at code and system levels. Testing confirmed performance of templating, measurement, and distortion correction tools using Philips Zenition 50 Digital Video Signal. Calibration array compatibility verified for C-arm image intensifier diameters 34.5-38.5 cm and 28.6-33 cm. All requirements met pass/fail criteria.

Technological Characteristics

Software-based image processing system; operates on commercial off-the-shelf (COTS) hardware (PC, monitor, mouse). Includes radiopaque calibration array for C-arm attachment. Distortion correction algorithm based on calibration array geometry. Connectivity: PACS export. Non-vendor specific image processing.

Indications for Use

Indicated for patients undergoing Total Hip Replacement or Hip Preservation procedures. Assists surgeons in intraoperative component positioning and alignment via fluoroscopic image analysis.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ February 18, 2021 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. OrthoGrid Systems, Inc. % Mr. Trevor Hafer Product Development Engineer 3865 Wasatch Boulevard, Suite 301 SALT LAKE CITY UT 84109 Re: K210136 Trade/Device Name: PhantomMSK Hip Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: January 15, 2021 Received: January 19, 2021 Dear Mr. Hafer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {1}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K210136 Device Name PhantomMSK Hip #### Indications for Use (Describe) PhantomMSK Hip is an image-processing software indicated to assist in the positioning of Total Hip Replacement and Hip Preservation components. It is intended to assist in precisely positioning Total Hip Preservation components intraoperatively by measuring their positions relative to the bone structures of interest provided that the points of interest can be identified from radiology images. Clinical judgement and experience are required to properly use the device. The device is not for primary image interpretation. The software is not for use on mobile phones. Type of Use (Select one or both, as applicable) | <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |----------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) SUMMARY ### OrthoGrid Systems, Inc.'s PhantomMSK Hip | Date Prepared: | 1/15/2021 | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | OrthoGrid Systems Inc.<br>3865 Wasatch Blvd. Suite 301Salt Lake City, UT 84109<br>Phone: +1 (801) 330 3113 or +1 (801) 703 5866 Ext 107<br>Facsimile: +1 (801) 801 206 2756 | | Contact Person: | Trevor Hafer<br>Product Development Engineer<br>Thafer@orthogrid.com<br>M: +1 (801) 793 7015 | | Name of Device: | PhantomMSK Hip | | Classification Name: | Picture Archiving and Communications System (21 CFR 892.2050) | | Regulatory Class: | Class II | | Product Code: | LLZ | | Predicate Devices: | PhantomMSK (K182332) | ## Device Description: The PhantomMSK Hip is a non-invasive software system that provides image analysis tools for Total Hip Replacement and Hip Preservation procedures that use fluoroscopic imaging to assist with implant and anatomic alignment. PhantomMSK Hip provides templating, measurement, and distortion adaptation (correction) tools for intraoperative fluoroscopic image assessment. PhantomMSK Hip does not include any custom computer hardware and is a software-based device that can be run on a "commercial off-the-shelf" system (i.e. PC, keyboard, mouse, touchscreen monitor etc.) that meet minimum performance requirements. Furthermore, PhantomMSK Hip operates on image principles that are not vendor specific. To operate PhantomMSK Hip, a fluoroscopic image is acquired from a C-arm and displayed outside the sterile field, where the image analysis tools can be used at the surgeon's discretion. Fluoroscopic distortion is attributed to external electromagnetic interference and the mapping of the planar image on a curved input phosphor. The PhantomMSK Hip uses software features in conjunction with its radiopaque calibration array, which attaches to the C-arm image intensifier, to calculate and adapt for fluoroscopic distortion. ## Indications for Use: PhantomMSK Hip is an image-processing software indicated to assist in the positioning of Total Hip Replacement and Hip Preservation components. It is intended to assist in precisely positioning Total Hip Replacement and Hip Preservation components intraoperatively by measuring their positions relative to the bone structures of interest provided that the points of interest can be identified from {4}------------------------------------------------ radiology images. Clinical judgement and experience are required to properly use the device. The device is not for primary image interpretation. The software is not for use on mobile phones. # Substantial Equivalence: The PhantomMSK Hip is shown to be substantially equivalent to its predicate, PhantomMSK. The PhantomMSK Hip has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended surgical use of the device and do not raise new question of safety or effectiveness when used as labeled. | Company | OrthoGrid Systems, Inc. | OrthoGrid Systems, Inc. | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Name | PhantomMSK Hip | PhantomMSK | | FDA 510(k) # | - | K182332 | | Product Code | LLZ | LLZ | | Principles of<br>Operation | The subject device is an<br>image-processing software. A<br>fluoroscopic image is acquired<br>from a C-arm by the software<br>and displayed outside the<br>sterile field, where the image<br>analysis tools can be used at<br>the surgeon's discretion. | The predicate device is an image-<br>processing software. A<br>fluoroscopic image is acquired<br>from a C-arm by the software and<br>displayed outside the sterile field,<br>where the image analysis tools<br>can be used at the surgeon's<br>discretion. | | Image Data | C-arm, X-ray | C-arm, X-ray | | Features | Templating, distortion<br>correction, angle<br>measurements, relative &<br>calibrated measurements,<br>drawing, compare, and<br>overlay tool, picture in picture,<br>PACS export, tutorial mode. | Templating, distortion correction,<br>angle measurements, relative<br>measurements | # Testing: Verification and validation testing were done at the code and system level according to written test protocols established before testing was conducted to ensure that all templating overlay, measurement, and distortion correction tools performed as expected. Specifically, testing was performed with the Philips Zenition 50 Digital Video Signal. The calibration array is intended for Carms having an outer diameter of the image intensifier ranging between 34.5 to 38.5 cm and 28.6 to 33 cm. {5}------------------------------------------------ The test results were reviewed by designated technical professionals to ensure acceptability criteria were satisfied prior to the release of the software. The results are summarized in test summary reports. The conclusion states that: - . Verification strategies and test procedures used are appropriate - System test procedures trace to requirements - . All requirements are tested or otherwise verified - o Test results meet the required pass/fail criteria ## Conclusions The PhantomMSK Hip is shown to be substantially equivalent to its predicate, PhantomMSK. The PhantomMSK Hip has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended surgical use of the device and do not raise new question of safety or effectiveness when used as labeled. In addition, performance data demonstrate that the PhantomMSK Hip does not raise different questions of safety or effectiveness. Thus, the PhantomMSK Hip is substantially equivalent.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%