MDx-Chex for BCY is intended for use as an external positive and negative assayed control to monitor the performance of the qualitative detection of yeast by the Diasorin LIAISON PLEX Yeast Blood Culture assay on the LIAISON PLEX System. The MDx-Chex for BCY Positive and Negative Controls are composed of a buffered solution with stabilized erythrocytes and leukocytes in a matrix of blood culture media components. Positive Control: Yeast: Candida albicans, Candida dubliniensis, Candida glabrata, Candida krusei, Candida parapsilosis, Candida tropicalis, Candida auris, Candida kefyr, Candida lipolytica, Candida lusitaniae, Candida guilliermondii, Cryptococcus neoformans/gattii, Candida famata, Candida haemulonii/duobushamulonii. Negative Control: buffered solution only. This product is not intended to replace manufacturer controls provided with the device.
Device Story
MDx-Chex for BCY is a ready-to-use liquid quality control kit; contains positive and negative controls for the Diasorin LIAISON PLEX Yeast Blood Culture (BCY) assay. Positive control contains stabilized yeast pathogens (14 species) in a matrix of blood culture media, erythrocytes, and leukocytes; negative control contains buffered solution only. Used in clinical laboratories to monitor assay performance, including lysis, nucleic acid isolation/purification, PCR inhibitor removal, amplification, and detection. Processed like a patient sample. Provides healthcare providers with verification that the LIAISON PLEX system is functioning correctly; ensures reliability of qualitative yeast detection results. Also configured as a verification kit for equipment installation, workflow development, and operator proficiency.
Clinical Evidence
No clinical data. Performance was established via bench testing, including multi-site reproducibility (100% PPA/NPA, n=180), repeatability (100% PPA/NPA, n=120), lot-to-lot reproducibility, and closed-vial stability (75 days at 2-25°C). Shipping stability was validated under simulated winter and summer temperature extremes.
Technological Characteristics
Ready-to-use liquid control; buffered solution with stabilized erythrocytes and leukocytes in blood culture media matrix. Monitors lysis, nucleic acid isolation/purification, PCR inhibitor removal, amplification, and detection. Compatible with LIAISON PLEX System. No specific energy source or connectivity; standalone control material.
Indications for Use
Indicated for use as an external positive and negative assayed control to monitor the performance of the qualitative detection of yeast by the Diasorin LIAISON PLEX Yeast Blood Culture assay on the LIAISON PLEX System. Intended for clinical laboratory use.
Regulatory Classification
Identification
An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.
Special Controls
An assayed quality control material for clinical microbiology assays must comply with the following special controls: (1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate: (i) Analyte concentration; Expected values: (ii) Analyte source: (iii) (iv) Base matrix; (v) Added components; (vi) Safety and handling information; and, (vii) Detailed instructions for use. (2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including: (i) Description of the process for value assignment and validation. (ii) Description of the protocol(s) used to establish stability. (iii) Line data establishing precision/reproducibility. (iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance. (v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method. (vi) Where applicable, detailed documentation related to studies for surrogate controls. (3) Premarket notification submissions must include an adequate mitigation (e.g., realtime stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling. (4) Your 21 CFR 809.10 compliant labeling must include the following: (i) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following: (A) Assayed control material analyte(s); (B) Whether the material is intended for quantitative or qualitative assays: (C) Stating if the material is a surrogate control; (D)The system(s), instrument(s), or test(s) for which the quality control material is intended. (ii) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following statement: "This product is not intended to replace manufacturer controls provided with the device." (iii)A limiting statement that reads "Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements."
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate:
(i) Analyte concentration;
(ii) Expected values;
(iii) Analyte source;
(iv) Base matrix;
(v) Added components;
(vi) Safety and handling information; and
(vii) Detailed instructions for use.
(2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including:
(i) Description of the process for value assignment and validation.
(ii) Description of the protocol(s) used to establish stability.
(iii) Line data establishing precision/reproducibility.
(iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance.
(v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method.
(vi) Where applicable, detailed documentation related to studies for surrogate controls.
(3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling.
(4) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following:
(A) Assayed control material analyte(s);
(B) Whether the material is intended for quantitative or qualitative assays;
(C) Stating if the material is a surrogate control; and
(D) The system(s), instrument(s), or test(s) for which the quality control material is intended.
(ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.”
(iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K254167
B Applicant
Streck, LLC
C Proprietary and Established Names
MDx-Chex for BCY
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| PMN | Class II | 21 CFR 866.3920-Assayed quality control material for clinical microbiology assays | MI-Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a substantial equivalence determination for the MDx-Chex for BCY.
B Measurand:
Multi-analyte quality control materials
C Type of Test:
MDx-Chex for BCY is intended for in vitro diagnostic use as an external assayed quality control to monitor the performance of the qualitative detection of yeast by the Diasorin LIAISON PLEX Yeast Blood Culture assay on the LIAISON PLEX System. The MDx-Chex for BCY Positive and Negative Controls are composed of a buffered solution with stabilized erythrocytes and leukocytes in a matrix of blood culture media components. Positive Control: Yeast: Candida
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
albicans, Candida dubliniensis, Candida glabrata, Candida krusei, Candida parapsilosis, Candida tropicalis, Candida auris, Candida kefyr, Candida lipolytica, Candida lusitaniae, Candida guilliermondii, Cryptococcus neoformans/gattii, Candida famata, Candida haemulonii/duobushamulonii. Negative Control: buffered solution only.
## Intended Use/Indications for Use:
### A Intended Use(s):
See Indications for Use below.
### B Indication(s) for Use:
MDx-Chex for BCY is intended for use as an external positive and negative assayed control to monitor the performance of the qualitative detection of yeast by the Diasorin LIAISON PLEX Yeast Blood Culture assay on the LIAISON PLEX System. The MDx-Chex for BCY Positive and Negative Controls are composed of a buffered solution with stabilized erythrocytes and leukocytes in a matrix of blood culture media components. Positive Control: Yeast: Candida albicans, Candida dubliniensis, Candida glabrata, Candida krusei, Candida parapsilosis, Candida tropicalis, Candida auris, Candida kefyr, Candida lipolytica, Candida lusitaniae, Candida guilliermondii, Cryptococcus neoformans/gattii, Candida famata, Candida haemulonii/duobushamulonii. Negative Control: buffered solution only. This product is not intended to replace manufacturer controls provided with the device.
### C Special Conditions for Use Statement(s):
Rx-For Prescription Use Only
For in vitro diagnostic use only
### D Special Instrument Requirements:
MDx-Chex for BCY is to be used on the Diasorin LIAISON PLEX system only.
## IV Device/System Characteristics:
### A Device Description:
MDx-Chex for BCY is a quality control kit consisting of positive and negative controls for the Diasorin LIAISON PLEX Yeast Blood Culture assay (BCY) on the LIAISON PLEX System. The MDx-Chex for BCY Positive Control is positive for 14 different yeast species (Table 1) that can cause bloodstream infections. The MDx-Chex for BCY Negative Control is negative for all yeast targets in the LIAISON PLEX BCY test.
Each control mix contains stabilized human blood components (erythrocytes and leukocytes) and blood culture media components that have been identified as inhibitors to nucleic acid purification and hybridization assays (e.g., hemoglobin, leukocyte DNA, and sodium polyanetholesulfonate anticoagulant). The MDx-Chex for BCY quality control kit contains stabilized blood components, simulated blood culture media components, and inactivated, intact yeast microorganisms resulting in a full-process, cellular-based control for the Diasorin
K254167- Page 2 of 12
{3}
LIAISON PLEX Yeast Blood Culture panel. Full-process controls evaluate the entire analytical workflow, including sample lysis, nucleic acid isolation and purification, hybridization, detection, and analysis, as well as the impact of inhibitors present in blood culture samples and pre-analytical variables.
Table 1: Yeast in MDx-Chex for BCY and detected by the Diasorin LIAISON PLEX Yeast Blood Culture assay on the LIAISON PLEX System
| Candida albicans | Candida krusei |
| --- | --- |
| Candida auris | Candida lipolytica |
| Candida dubliniensis | Candida lusitaniae |
| Candida famata | Candida parapsilosis |
| Candida glabrata | Candida tropicalis |
| Candida guilliermondii | Candida haemulonii/duobushaemulonii |
| Candida kefyr | Cryptococcus neoformans/gattii |
B Principle of Operation:
Not Applicable
V Substantial Equivalence Information:
A Predicate Device Name(s):
MDx-Chex for BCID2
B Predicate 510(k) Number(s):
K212576
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K254167 (candidate device) | K212576 (predicate) |
| --- | --- | --- |
| Device Trade Name | MDx-Chex for BCY | MDx-Chex for BCID2 |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | MDx-Chex for BCY is intended for use as an external positive and negative assayed control to monitor the performance of the qualitative detection of yeast by the Diasorin LIAISON PLEX Yeast Blood Culture assay on the LIAISON PLEX System. The MDx-Chex for BCY Positive and Negative Controls are composed of a buffered solution with | MDx-Chex for BCID2 is intended for use as an external positive and negative assayed control to monitor the performance of the qualitative detection of yeast, Gram positive and Gram negative bacteria, as well as associated antimicrobial resistance genes, by the BioFire |
K254167- Page 3 of 12
{4}
K254167- Page 4 of 12
| Device & Predicate Device(s): | K254167
(candidate device) | K212576
(predicate) |
| --- | --- | --- |
| | stabilized erythrocytes and leukocytes in a matrix of blood culture media components.
Positive Control: Yeast: *Candida albicans*, *Candida dubliniensis*, *Candida glabrata*, *Candida krusei*, *Candida parapsilosis*, *Candida tropicalis*, *Candida auris*, *Candida kefyr*, *Candida lipolytica*, *Candida lusitaniae*, *Candida guilliermondii*, *Cryptococcus neoformans/gattii*, *Candida famata*, *Candida haemulonii/duobushamulonii*.
Negative Control: buffered solution only. This product is not intended to replace manufacturer controls provided with the device. | FilmArray Blood Culture Identification 2 (BCID2)
Panel on FilmArray systems.
Control 1 - GN: Gram negative bacteria: *Acinetobacter colcoaceticus-baumannii* complex, *Bacteroides fragilis*, *Enterobacter cloacae* complex, *Escherichia coli*, *Klebsiella aerogenes*, *Klebsiella oxytoca*, *Klebsiella pneumoniae* group, *Proteus spp.*, *Salmonella spp.*, *Serratia marcescens*, *Haemophilus influenza*, *Neisseria meningitides*, *Pseudomonas aeruginosa*, *Stenotrophomonas maltophilia*; antimicrobial resistance genes: KPC, CTX-M, IMP, NDM, OXA-48-like, VIM, *mcr-1*.
Control 2 - GPY: Gram positive bacteria: *Enterococcus faecalis*, *Enterococcus faecium*, *Listeria monocytogenes*, *Staphylococcus aureus*, *Staphylococcus epidermidis*, *Staphylococcus lugdunesis*, *Streptococcus agalactiae*, *Streptococcus pneumonia*, *Streptococcus pyogenes*; yeast: *Candida albicans*, *Candida auris*, *Candida glabrata*, *Candida krusei*, *Candida parapsilosis*, *Candida tropicalis*, *Cryptococcus neoformans/gatti*; antimicrobial resistance genes: *mecA/C* and MREJ, |
{5}
| Device & Predicate Device(s): | K254167
(candidate device) | K212576
(predicate) |
| --- | --- | --- |
| | | vanA/B. This product is not intended to replace manufacturer controls provided with the device. |
| Physical Format | Ready-to-Use Liquid | Same |
| Direction for Use | Process like patient sample | Same |
| Assay Steps Monitored | Lysis, nucleic acid isolation/purification/PCR inhibitor removal, amplification, detection, identification/data reporting | Same |
| General Device Characteristic Differences | | |
| Number of targets monitored in one assay | Multiple - 14 yeast organisms | Multiple, >30 targets |
| Composition | Intact inactivated yeast, human erythrocytes and leukocytes, and relevant components of blood culture media | Intact inactivated bacteria, antimicrobial resistance genes, human erythrocytes and leukocytes, and relevant components of blood culture media |
VI Standards/Guidance Documents Referenced:
- ISO 14971,2019: Medical devices - Application of risk management to medical devices
- ISO 15223,2021: Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements.
- CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
Note: The results for all analytical studies described below represent percent agreement based on comparing the expected call to the obtained test result (positive or negative). The terms, PPA (positive percent agreement) and NPA (negative percent agreement) were used to describe this agreement.
K254167- Page 5 of 12
{6}
1. Precision/Reproducibility:
Reproducibility
A multi-site reproducibility study was conducted with the MDx-Chex for BCY where testing was completed at three sites with seven operators and consisted of ten positive control samples and ten negative control samples for each of three lots of MDx-Chex for BCY for a total of 30 samples per control type and 60 samples per lot. Both positive and negative controls were tested on different days (2 tubes x 1 lot x 1 day, for 10 days and 3 different lots) for a total of 180 runs (90 runs per MDx-Chex for BCY control type) from all testing sites and all MDx-Chex for BCY lots.
All samples were prepared and analyzed on the DiaSorin LIAISON PLEX System per the Instructions for Use. This resulted in a positive percent agreement (PPA) of 100% (90/90) for the MDx-Chex for BCY Positive Control (Table 2) and a negative percent agreement (NPA) of 100% (90/90) for the MDx-Chex for BCY Negative Control (Table 3).
Table 2: MDx-Chex for BCY Positive Control Multi-Site Reproducibility Summary
| Category | Site #1 | | Site #2 | | Site #3 | | PPA for All Sites Combined | 95% Confidence Interval |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | # Observed/# Expected Results¹ | PPA | # Observed/# Expected Results¹ | PPA | # Observed/# Expected Results¹ | PPA | | |
| MDx-Chex for BCY Positive Control | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 100% (90/90) | 96%-100% |
¹ Expected result for the Positive Control is positive.
Table 3: MDx-Chex for BCY Negative Control Multi-Site Reproducibility Summary
| Category | Site #1 | | Site #2 | | Site #3 | | NPA for All Sites Combined | 95% Confidence Interval |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | # Observed/# Expected Results¹ | NPA | # Observed/# Expected Results¹ | NPA | # Observed/# Expected Results¹ | NPA | | |
| MDx-Chex for BCY Negative Control | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 100% (90/90) | 96%-100% |
¹ Expected result for the Negative Control is negative.
The results support the reproducibility of the MDx-Chex for BCY across three separately manufactured control lots between sites, days, and operators when used with the DiaSorin LIAISON PLEX BCY test on different DiaSorin LIAISON PLEX systems.
Repeatability
An internal repeatability study was conducted to assess performance of MDx-Chex for Blood Culture Yeast (BCY) using at least two DiaSorin LIAISON PLEX systems. The testing schedule consisted of evaluating two MDx-Chex for BCY control types (1 control type per day) from three lots over 20 days for a total of 120 runs. Samples were analyzed on the
K254167- Page 6 of 12
{7}
DiaSorin LIAISON PLEX System per the Instructions for Use. Three operators tested each control type using three LIAISON PLEX BCY test cartridge lots and two LIAISON PLEX systems. This resulted in a PPA of 100% (60/60) for the MDx-Chex for BCY Positive Control (Table 4) and a NPA of 100% (60/60) for the MDx-Chex for BCY Negative Control (Table 5).
Table 4: MDx-Chex for BCY Positive Control Repeatability Study Summary
| Category | # Observed/#Expected Results^{1} | PPA | 95% Confidence Interval |
| --- | --- | --- | --- |
| MDx-Chex for BCY Positive Control | 60/60 | 100% | 94%-100% |
1 Expected result for the Positive Control is positive.
Table 5: MDx-Chex for BCY Negative Control Repeatability Summary
| Category | # Observed/#Expected Results^{1} | NPA | 95% Confidence Interval |
| --- | --- | --- | --- |
| MDx-Chex for BCY Negative Control | 60/60 | 100% | 94%-100% |
1 Expected result for the Negative Control is negative.
The results support the conclusion that MDx-Chex for Blood Culture Yeast shows repeatability across three separately manufactured control lots when used with the LIAISON PLEX BCY Test on the DiaSorin LIAISON PLEX system.
## Lot-to-Lot Reproducibility
Lot-to-lot reproducibility was demonstrated by testing three lots of MDx-Chex for BCY. Samples were prepared per the control Instructions for Use and tested using the same DiaSorin LIAISON PLEX BCY test cartridge lot tested on one DiaSorin LIAISON PLEX system over ten days. Ten positive and ten negative MDx-Chex for BCY control tubes were tested on the same DiaSorin LIAISON PLEX system for 60 data points total (30 per lot per positive and negative control). All MDx-Chex for BCY positive and negative control lots passed with ≥ 90% positive or negative agreement with expected results, respectively. Results are presented in Tables 6 and 7.
Table 6: Lot-to-Lot Reproducibility for MDx-Chex for BCY Positive Controls
| Category | Lot # | # Observed/# Expected Results^{1} | PPA | 95% Confidence Interval |
| --- | --- | --- | --- | --- |
| MDx-Chex for BCY Positive Control | 5202 | 10/10 | 100% | 69-100% |
| | 5209 | 10/10 | 100% | 69-100% |
| | 5223 | 10/10 | 100% | 69-100% |
1 Expected result for the Positive Control is positive.
Table 7: Lot-to-Lot Reproducibility for MDx-Chex for BCY Negative Controls
| Category | Lot # | # Observed/# Expected Results^{1} | PPA | 95% Confidence Interval |
| --- | --- | --- | --- | --- |
| MDx-Chex for BCY Negative Control | 5202 | 10/10 | 100% | 69-100% |
| | 5209 | 10/10 | 100% | 69-100% |
| | 5223 | 10/10 | 100% | 69-100% |
1 Expected result for the Negative Control is negative.
K254167- Page 7 of 12
{8}
The results support that MDx-Chex for BCY is reproducible across three separately manufactured lots when used with the DiaSorin LIAISON PLEX BCY Test.
## Within-run Precision
A within-run precision study was conducted to assess performance of one MDx-Chex for BCY lot, using the same DiaSorin LIAISON PLEX BCY test cartridge lot tested on the same day with one DiaSorin LIAISON PLEX System. The study consisted of 10 tests for each positive and negative control tube generated from one MDx-Chex for BCY lot (total of 20 tests per control kit). All MDx-Chex for BCY positive and negative control lots passed with $\geq 90\%$ positive or negative agreement with expected results, respectively. Results are presented in Tables 8 and 9.
Table 8: Within-run Precision Summary for MDx-Chex for BCY Positive Controls
| Category | Lot # | # Observed/# Expected Results¹ | PPA | 95% Confidence Interval |
| --- | --- | --- | --- | --- |
| MDx-Chex for BCY Positive Control | 5223 | 10/10 | 100% | 69-100% |
¹ Expected result for the Positive Control is positive.
Table 9: Within-run Precision Summary for MDx-Chex for BCY Negative Controls
| Category | Lot # | # Observed/# Expected Results¹ | PPA | 95% Confidence Interval |
| --- | --- | --- | --- | --- |
| MDx-Chex for BCY Negative Control | 5223 | 10/10 | 100% | 69-100% |
¹ Expected result for the Negative Control is negative.
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
Not applicable.
4. Detection Limit and Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
### Closed-vial Stability
A real-time closed-vial stability study is ongoing using three lots of MDx-Chex for BCY with currently available data summarized below. Twenty positive and 20 negative control samples from each MDx-Chex for BCY lot were stored at room temperature (20-25 °C) and refrigerated temperature (2-8 °C) and collected at different timepoints for testing. The first
K254167- Page 8 of 12
{9}
data collection timepoint was on Day 0 (ship date), followed by Day 75 for each lot and storage condition. Samples were prepared and analyzed on the DiaSorin LIAISON PLEX System per the MDx-Chex for BCY Instructions for Use. Data collected on Days 0 and at a minimum of 75 days are presented in Tables 10 and 11.
Table 10: Closed-vial Stability of MDx-Chex for BCY Positive Control- PPA
| Shelf-Life | Storage Temperature | # Observed/# Expected Results^{1} | PPA | 95% Confidence Interval | PPA ≥ 90% Acceptance |
| --- | --- | --- | --- | --- | --- |
| Day 0 | N/A | 60/60 | 100% | 94-100% | Pass |
| Day 75+ | 2-8°C | 60/60 | 100% | 94-100% | Pass |
| | 20-25°C | 60/60 | 100% | 94-100% | Pass |
Expected result for the Positive Control is positive.
Indicates that lots stored at 2-8°C were tested for at least 75 days; Lot 5202 (80 days), Lot 5209 (83 days), Lot 5223 (78 days). Lots stored at 20-25°C were tested for at least 75 days; Lot 5202 (80 days), Lot 5209 (83 days), Lot 5223 (79 days).
Table 11: Closed-vial Stability of MDx-Chex for BCY Negative Control- NPA
| Shelf-Life | Storage Temperature | # Observed/# Expected Results^{1} | NPA | 95% Confidence Interval | NPA ≥ 90% Acceptance |
| --- | --- | --- | --- | --- | --- |
| Day 0 | N/A | 60/60 | 100% | 94-100% | Pass |
| Day 75+ | 2-8°C | 60/60 | 100% | 94-100% | Pass |
| | 20-25°C | 60/60 | 100% | 94-100% | Pass |
Expected result for the Negative Control is negative.
Indicates that lots stored at 2-8°C were tested for at least 75 days; Lot 5202 (80 days), Lot 5209 (83 days), Lot 5223 (79 days). Lots stored at 20-25°C were tested for at least 75 days; Lot 5202 (83 days), Lot 5209 (84 days), Lot 5223 (80 days).
The data for the ongoing closed-vial stability study support the conclusion that MDx-Chex for BCY has a closed-vial stability for 75 days when stored at 2-25°C. The study is ongoing to support the use of the MDx-Chex for BC-GN control kit throughout the labeled shelf life.
## Shipping Stability
A shipping stability study was performed to validate the stability of MDx-Chex for BCY during shipment. One MDx-Chex for BCY lot was subjected to simulated winter and summer shipping temperature profiles over 5 days. Temperature stress conditions for summer include a 120-hour exposure period with the following temperature profile: 5h at 22°C, 14h at 26°C, 5.5h at 22°C, 5.5h at 25°C, 10h at 22°C, 2h at 40°C, 15h at 29°C, 5h at 40°C, 17.5h at 26°C, 11.5h at 29°C, 10h at 22°C, 7.5h at 30°C, and 11.5h at 24°C. Temperature stress conditions for winter include a 120-hour exposure period with the following temperature profile: 5h at 22°C, 14h at 15°C, 5.5h at 22°C, 5.5h at 17°C, 10h at 22°C, 2h at -10°C, 15h at 10°C, 5h at -10°C, 17.5h at 15°C, 11.5h at 10°C, 10h at 22°C, 7.5h at 8°C, and 11.5h at 18°C.
Samples at each storage temperature (2°C and 25°C) were exposed to winter and summer temperature extremes and then were stored back at each respective storage temperature (2°C and 25°C) for a week prior to being tested using LIAISON PLEX BCY Test. Data were collected from 20 samples per positive and negative control for each simulated shipping profile within the 75-day closed-vial stability testing period. An additional 20 samples per control type will be tested at the end of shelf-life.
K254167- Page 9 of 12
{10}
All summer and winter shipping conditions passed with $\geq 90\%$ positive and negative agreement with expected results and presented in Tables 12 and 13.
Table 12: Shipping Stability Study Results for the MDx-Chex for BCY Positive Control
| Category | Storage Temperature^{1} | # Observed/# Expected Results^{2} | PPA | 95% Confidence Interval |
| --- | --- | --- | --- | --- |
| Summer * | 2-8°C | 20/20 | 100% | 83-100% |
| | 20-25°C | 20/20 | 100% | 83-100% |
| Winter * | 2-8°C | 20/20 | 100% | 83-100% |
| | 20-25°C | 20/20 | 100% | 83-100% |
1. Samples were stored at each respective temperature prior to exposure to simulated summer or winter conditions and incubated back at each respective storage temperature prior to testing on the DiaSorin LIAISON PLEX system.
2. Expected result for the positive Control is positive.
*Indicates initial dataset for shipping study which was collected within the 75-day CVS testing period.
Table 13: Shipping Stability Study Results for the MDx-Chex for BCY Negative Control
| Category | Storage Temperature^{1} | # Observed/# Expected Results^{2} | NPA | 95% Confidence Interval |
| --- | --- | --- | --- | --- |
| Summer * | 2-8°C | 20/20 | 100% | 83-100% |
| | 20-25°C | 20/20 | 100% | 83-100% |
| Winter * | 2-8°C | 20/20 | 100% | 83-100% |
| | 20-25°C | 20/20 | 100% | 83-100% |
1. Samples were stored at each respective temperature prior to exposure to simulated summer or winter conditions and incubated back at each respective storage temperature prior to testing on the DiaSorin LIAISON PLEX system.
2. Expected result for the negative Control is negative.
*Indicates initial dataset for shipping study which was collected within the 75-day CVS testing period.
Results support that the MDx-Chex for BCY control kit is stable and functional for 75 days after exposure to extreme summer and winter shipping temperature conditions. The study is ongoing to support the shipping stability of the MDx-Chex for BC-GN control kit throughout the labeled shelf life.
6. Assay Cut-Off:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable.
2. Matrix Comparison:
The matrix of MDx-Chex for BCY is made of stabilized blood components (erythrocytes and leukocytes) in a simulated blood culture media. Since the matrix is not identical to that of the routine DiaSorin LIAISON PLEX BCY assay sample, blood culture media, a test was performed to investigate the effect of the matrix on the assay.
To confirm that the simulated blood culture matrix does not impact performance of the DiaSorin LIAISON PLEX BCY assay, Candida tropicalis (1×10⁶ cells/mL final
K254167- Page 10 of 12
{11}
concentration) was spiked into MDx-Chex for BCY matrix as well as into BD BACTEC Plus Aerobic/F culture vial with negative whole blood to simulate a clinical sample. These samples were then tested in triplicate using DiaSorin LIAISON PLEX BCY assay. Three replicates of each simulated matrix with no spike-in organism were also tested to serve as negative controls. The results presented in Tables 14 and 15 showed PPA and NPA of 100%, respectively.
Table 14: Matrix Equivalency Study Results for Control and Clinical Matrix Spiked with Candida tropicalis, tested on DiaSorin LIAISON PLEX BCY Assay
| Matrix Type | # Observed/# Expected Results^{1} | PPA | 95% Confidence Interval |
| --- | --- | --- | --- |
| MDx-Chex for BCY Matrix + Candida tropicalis | 3/3 | 100% | 29-100% |
| Clinical Matrix + Candida tropicalis | 3/3 | 100% | 29-100% |
1 Expected result for the Spiked-in matrices are positive for Candida tropicalis.
Table 15: Matrix Equivalency Study Results for Unspiked Control and Clinical Matrix, tested on DiaSorin LIAISON PLEX BCY Assay
| Matrix Type | # Observed/# Expected Results^{1} | NPA | 95% Confidence Interval |
| --- | --- | --- | --- |
| MDx-Chex for BCY Matrix | 3/3 | 100% | 29-100% |
| Clinical Matrix | 3/3 | 100% | 29-100% |
1 Expected result for non-spiked matrices are negative.
The results demonstrate that MDx-Chex for BCY matrix has no effect on target detection (i.e., no inhibition and/or false negative results) when tested with the LIAISON PLEX BCY Test. Data for MDx-Chex for BCY was identical to the results for the simulated clinical blood culture sample.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Clinical Cut-Off:
Not applicable.
4. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
K254167- Page 11 of 12
{12}
# D Expected Values/Reference Range:
MDx-Chex for BCY is a qualitative control and the expected results are listed in Table 16 below.
Table 16: Pathogens detected by MDx-Chex for BCY Control
| Pathogen | Positive Control | Negative control |
| --- | --- | --- |
| Candida albicans | Detected | Not detected |
| Candida auris | Detected | Not detected |
| Candida dubliniensis | Detected | Not detected |
| Candida famata | Detected | Not detected |
| Candida glabrata | Detected | Not detected |
| Candida guilliermondii | Detected | Not detected |
| Candida kefyr | Detected | Not detected |
| Candida krusei | Detected | Not detected |
| Candida lipolytica | Detected | Not detected |
| Candida lusitaniae | Detected | Not detected |
| Candida parapsilosis | Detected | Not detected |
| Candida tropicalis | Detected | Not detected |
| Candida haemulonii/duobushaemulonii | Detected | Not detected |
| Cryptococcus neoformans/gattii | Detected | Not detected |
# VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
# IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K254167- Page 12 of 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.